MAGNETIC RESONANCE IMAGING METHOD TO NON-INVASIVELY MEASURE BLOOD OXYGEN SATURATION
20170224259 · 2017-08-10
Inventors
- Orlando P. Simonetti (Columbus, OH)
- Rizwan Ahmad (Hilliard, OH, US)
- Lee C. Potter (Riverlea, OH, US)
- Juliet Jaison Varghese (Columbus, OH, US)
Cpc classification
G01R33/5602
PHYSICS
G01R33/50
PHYSICS
International classification
A61B5/145
HUMAN NECESSITIES
A61B5/055
HUMAN NECESSITIES
G01R33/50
PHYSICS
G01R33/56
PHYSICS
Abstract
Methods for data acquisition and processing of magnetic resonance (MR) imaging to obtain the oxygen saturation (O2sat) of blood using a relationship between transverse relaxation time (T2) of blood and oxygen saturation. The method includes obtaining multiple images at various T2 preparation times. Next, non-linear curve fitting may be used to solve for arterial or venous O2sat. The disclosure provides a calibration-free method for accurate quantitative assessment of blood in the heart and deep vessels, even in locations having limited accessibility with other diagnostic techniques.
Claims
1. A method for determining oxygen saturation (O2sat) of blood using magnetic resonance (MR) image data, comprising: acquiring multiple transverse relaxation time (T2) prepared source images, each of the source images having distinct pulse timing; and deriving multiple T2-maps from the T2 prepared source images; and using a chemical exchange model and applying a non-linear curve fitting to the multiple T2-maps to determine arterial or venous O2sat.
2. The method of claim 1, wherein the T2 prepared source images are obtained by varying a combination of a number of refocusing pulses and τ.sub.180 values.
3. The method of claim 2, wherein the refocusing pulses range from 2 to 12 for each T2 map, and where the τ.sub.180 values range from 7.5 ms to 25 ms.
4. The method of claim 1, further comprising: receiving electrocardiogram data from a patient under examination; and triggering the acquiring of the T2 prepared source images using the electrocardiogram data.
5. The method of claim 4, further comprising measuring arterial O2sat using non-invasive pulse oximetry during the acquisition of MR data.
6. The method of claim 1, wherein the chemical exchange model is a Luz-Meiboom (L-M) model.
7. The method of claim 6, wherein a subset of the quantities in the L-M model are measured independently.
8. The method of claim 6, wherein ranges or probabilistic priors are enforced, during curve fitting, on unknown quantities in the L-M model.
9. The method of claim 8, further wherein the unknown quantities are at least one of % SbO.sub.2, T.sub.2o, τ.sub.ex, and α, wherein % SbO.sub.2 is blood O2sat, T.sub.2o is a constant, and τ.sub.ex and α are biophysical parameters.
10. The method of claim 8, where measurements from arteria and venous blood are processed jointly.
11. The method of claim 1, wherein the multiple T2-maps are single slice maps in which a pixel value reflects T2 of blood in chambers of a heart and blood vessels in a chest of a patient from which the T2 prepared source images were acquired.
12. A method for determining oxygen saturation (O2sat) of blood using magnetic resonance (MR) image data, comprising: obtaining multiple transverse relaxation time (T2) prepared source images, each of the T2 source images having distinct pulse timing; and directly using a measured signal, S, from the multiple T2 prepared source images in a chemical exchange model and applying a non-linear curve fitting to determine venous O2sat.
13. The method of claim 12, wherein the T2 prepared source images are obtained by varying a combination of a number of refocusing pulses and τ.sub.180 values.
14. The method of claim 13, wherein the refocusing pulses range from 2 to 12 for each T2 map, and where the τ.sub.180 values range from 7.5 ms to 25 ms.
15. The method of claim 12, further comprising: receiving electrocardiogram data from a patient under examination; and triggering the acquiring of the T2 prepared source images using the electrocardiogram data.
16. The method of claim 15, further comprising measuring arterial O2sat using non-invasive pulse oximetry during the acquisition of MR data.
17. The method of claim 12, wherein the chemical exchange model is a Luz-Meiboom (L-M) model
18. The method of claim 17, wherein ranges or probabilistic priors are enforced, during curve fitting, on unknown quantities in the L-M model.
19. The method of claim 12, wherein a relationship of the measured signal S to T2 is determined according to:
S=S0*exp(−TE/T2), wherein S0 is an initial signal, and TE a product of the number of refocusing pulses and τ.sub.180.
20. The method of claim 19, wherein S0 is determined as the measured signal S of a T2 prepared image with 0 refocusing pulses.
Description
BRIEF DESCRIPTION OF THE DRAWINGS
[0016] The above and/or other aspects will become more apparent by describing certain exemplary embodiments with reference to the accompanying drawings.
[0017]
[0018]
[0019]
[0020]
[0021]
[0022]
[0023]
[0024]
[0025]
[0026]
[0027]
[0028]
[0029]
[0030]
[0031]
[0032]
[0033]
DETAILED DESCRIPTION
[0034] Reference will now be made to exemplary embodiments described with reference to the accompanying drawings, wherein like reference numerals refer to like elements throughout. In this regard, the exemplary embodiments may have different forms and should not be construed as being limited to the descriptions set forth herein. Accordingly, the exemplary embodiments are merely described below, by referring to the figures, to explain aspects of the present description.
[0035] Overview of Method for Clinical Utilization of MRI to Determine Blood Oximetry
[0036] In the model described in Eq. 1, T.sub.2b can be measured using MRI, and Hct can be measured from a small blood sample. Other parameters such as ω.sub.0 and τ.sub.180 can be controlled based on the choice of magnetic field strength and T2-preparation pulse design, respectively. This leaves the desired parameter, blood oxygen saturation, % SbO.sub.2 and three other nuisance parameters (T.sub.2o, τ.sub.ex and α) as unknowns.
[0037] While a single measurement of the apparent blood T2 would not be sufficient to estimate all of the unknown patient-specific parameters in the L-M model, the present disclosure provides for a method that allows these unknown parameters to be estimated on a patient-specific basis. Noting that τ.sub.180 is a controllable parameter, acquiring multiple T2 measurements, each at a different τ.sub.180, provides the diversity of data that is needed to characterize the patient-specific relationship between blood T2 and O2sat. Once a sufficient number of effective T2 weighted images have been acquired (from which quantitative T2 maps can be created as an intermediate step in the analysis), each using a different τ.sub.180, the four unknown model parameters (% SbO.sub.2, T.sub.2o, τ.sub.ex, and α) can be stably estimated using a constrained non-linear least squares curve fit. Although the solution remains viable when the nuisance parameters are unknown a priori, a constrained model is used in order to avoid the possibility of convergence to any local minima, and thereby improve accuracy. This approach provides a framework for patient-specific, calibration-free T2 oximetry (shown in
[0038] Using a technique to perform rapid, quantitative characterization of the myocardial T2 to identify inflammation and edema, τ.sub.180, was adjusted among four refocusing pulses in the T2 preparation train in order to vary the echo times, T2p, which is defined as T2p=number of refocusing pulses×T.sub.180. For the estimation of blood O2sat, T2p is extended by increasing the number of refocusing pulses based on a segmented Malcolm Levitt phase cycling pattern (0, 2, 4, 8 and 12 refocusing pulses). Thus, four T2 maps may be generated for a given blood pool, using τ.sub.180 values of 12, 15, 20, and 25 ms. Therefore, the T2p times corresponding to each T2 map used in the study were 0, 24, 48, 96 and 144 ms (for τ.sub.180=12 ms); 0, 30, 60, 90 and 180 ms (for τ.sub.180=15 ms); 0, 40, 80 and 160 ms (for τ.sub.180=20 ms); and 0, 50, 100 and 200 ms (for τ.sub.180=25 ms), respectively.
[0039] As the MR imaging planes of the heart usually include both arterial and venous blood pools, each T2 map can provide a T2 measurement of both arterial and venous blood. It is also possible to measure the O2sat of arterial blood by non-invasive pulse oximetry. Therefore, joint processing of venous and arterial blood T2 measurements, together with a known value of arterial O2sat, can provide additional information that aids in accurate parameter estimation.
[0040]
[0041] The graph illustrated in
[0042] Implementations of Image Acquisition and Data Processing Technique to Determine Blood O2Sat
[0043] With reference to
[0044] With reference to
[0045] Referring again to
[0046] In a second implementation (202, 208), the method may be performed without the intermediate step of generating a T2 map. Thus, in the second implementation, 202 is followed by 208, where the Luz-Meiboom (L-M) exchange model may be directly applied to the MR signal in the measured T2 prepared source images and a non-linear curve fitting may be used to solve for venous O2sat.
[0047] Thus, in each implementation, by addressing the flow sensitivity of the T2 preparation pulse and the inaccuracies introduced by oversimplification of the model relating T2 to O2sat, the level of accuracy and reproducibility for this technique will be raised to a level required for clinical application in cardiac patients. Further, the methodology of the present disclosure can be easily incorporated into current MRI scanning protocols for acquired and inherited heart disease to provide additional useful and practical diagnostic information.
[0048] Further details of each of steps 204, 206 and 208 are provided below.
[0049] Optimization of the T2-Preparation Pulse Timing (204)
[0050] With the above overview as an introduction to the present disclosure, optimization of the T2-preparation pulse timing will now be described. The standard T2 preparation utilized for myocardial and blood T2 mapping uses four refocusing pulses and τ.sub.180 up to 50 ms to achieve the longer echo times needed for blood T2 quantification. Such wide pulse spacing increases sensitivity to higher orders of motion (e.g., acceleration) and diffusion, and leads to heterogeneity and signal loss in the blood pool. While there are no gradients applied during this non-selective preparation, dephasing results from flow in the presence of local susceptibility gradients.
[0051] In accordance with the present disclosure, τ.sub.180 is reduced to a relatively short spacing (e.g., 25 ms or less) in order to reduce sensitivity to flow. In an example embodiment, multiple T2 maps are formed, each with a different τ.sub.180. The source images that are used to generate these maps use distinct numbers of refocusing pulses, including, but not limited to, 2, 4, 8, 12 pulses. In one non-limiting embodiment, the number of refocusing pulses in the source images of a single T2 map is doubled from the standard four to eight, thereby reducing the maximum τ.sub.180 from 50 ms to 25 ms. As such, signal loss due to flow and diffusion is reduced.
[0052] The accuracy of T2 measured using the reduced τ.sub.180 prep was verified against the gold standard spin echo technique in a static phantom to have less than 10% error. Preliminary studies in a cohort of eight healthy volunteers showed the effectiveness of shortening τ.sub.180 in reducing blood pool variability caused by flow, as shown in
[0053] Shortening the average τ.sub.180 from 30 ms to 15 ms reduced the coefficient of variation in T2 measured in the right ventricle by 34%, and in the left ventricle by 60% in a cohort of healthy volunteers. Additional evidence that the modified T2 prep produces equivalent results with or without flow is shown in
[0054] Non-Linear Parameter Estimation (206)
[0055] The Luz-Meiboom (L-M) model is widely accepted as a valid description of the mechanism of T2 relaxation facilitated by proton exchange between sites at two different resonant frequencies. The L-M model equates the reciprocal of the T2 relaxation to a sum of two terms: the reciprocal of the relaxation time of blood at 100% oxygen saturation and a term proportional to (1-O2sat).sup.2. The proportionality constant for the second term depends on five physical parameters: the hematocrit (Hct), or volume ratio of red blood cells (RBC) to the total volume of blood; the proton exchange time (τ.sub.ex); a dimensionless parameter, alpha (α), related to the magnetic susceptibility of deoxyhemoglobin; the proton resonance frequency (ω.sub.0); and the 180° pulse interval of a spin echo or T2 preparation train (τ.sub.180). The pulse interval and resonance frequency are under experimental control. Other unknown parameters are considered nuisance parameters, and importantly, may vary from individual to individual. At a given field strength (where ω.sub.0 is fixed), by manipulating the pulse interval, multiple T2 maps are generated in both venous and arterial blood pools; in this manner, the quantity of interest, blood O2sat, and all unknown nuisance parameters may be jointly estimated, bypassing the existing need for technique or patient-specific in vitro calibration.
[0056] With regard to the L-M model specified in Eq. 1, it has a number of parameters that are either controlled or calibrated. A common practice is to lump the calibration parameters into two composite calibration parameters, T.sub.2o and a constant, K, defined by Wright et al. as:
[0057] which has units ms.sup.−1 or s.sup.−1. K lumps together parameters that depend on the scanning conditions (τ.sub.180 and ω.sub.0) with other parameters that depend on patient-specific characteristics of the blood (Hct, τ.sub.ex, and α). Substituting K into the L-M model results in the simplified equation proposed by Wright et al., which was introduced above:
[0058] While computationally the simplest, this model does not take into account the effect of individual hematocrit, nor the inter-individual variation in the other biophysical parameters, τ.sub.ex, and α, and could lead to inaccurate estimation of O2sat in patients with abnormal hematocrit levels. Wright et al. also observed a dependence of Hct on the constant term, T.sub.2o. Since then, a more comprehensive theoretical model, incorporating patient specific hematocrit, has been proposed by Golay et al.
[0059] The additional parameters from Eq. 1 being the relaxation time of plasma (T.sub.2p), the relaxation rates of diamagnetic blood components, oxyhemoglobin and deoxyhemoglobin (R2.sub.dia, R2.sub.oxy and R2.sub.deoxy), and the frequencies of diamagnetic blood components, oxyhemoglobin and deoxyhemoglobin (ω.sub.dia, ω.sub.deoxy and ω.sub.oxy) under the influence of an external magnetic field.
[0060] These different versions of the model have been extensively used for in vitro characterization of the magnetic properties and dependence of blood T2 in both human and animal blood, and across different field strengths. The conventional approach is to generate empirically determined calibration factors for the unknown parameters in the model, which are then applied to the general population to aid in the estimation of O2sat from T2. Studies involving patient specific calibration factors have also been performed to estimate O2sat more accurately, but this still involves drawing blood samples from each individual to perform an in vitro calibration, making the whole process cumbersome and impractical.
[0061] The method for patient-specific, calibration free T2 based MR oximetry of the present disclosure has been evaluated by a controlled graded hypoxemia experiment in a porcine model across a range of oxygen saturation levels (detailed below). The performance of the method in the different versions of the L-M model with four (Eq. 1), three (Eq. 2) and six (Eq. 6) unknown parameters respectively, as defined in Eqns. 1-6, are also examined below. Examining Eqns. 1-6, the parameters of the L-M model can be either directly measured, controlled or estimated. The measurable parameters are blood T2 in the regions of interest (arterial and venous), and hematocrit, which can be measured from a small blood sample. The controllable parameters are ω.sub.0, and τ.sub.180, the inter-echo spacing of the T2 preparation pulses in the imaging sequence. This leaves the remaining unknown/nuisance parameters to be estimated along with % SbO.sub.2—the blood oxygen saturation. The technique described in this disclosure samples the curve that defines the T2-O2 relationship at different values of τ.sub.180.
[0062] As shown in the example embodiment in
[0063] Non-Linear Parameter Estimation (208)
[0064] In the second implementation, the signal from the source images is directly employed in the L-M model instead of calculating an intermediate T2 measurement. Here, determining T2 maps is not necessary as O2sat estimates can be derived directly from a series of T2-prepared images without the intermediate step of T2 estimation. Referring to the discussion above, the equation relating the signal to T2 is expressed as S=S0*exp(−TE/T2), where S is the measured signal in each source image, S0 is the initial signal, and the echo time or T2 preparation time, TE, is the product of the number of refocusing pulses and τ.sub.180. S0 becomes an additional nuisance parameter to be solved for. Alternatively, S0 can be treated as the signal measured in the T2 prepared image with 0 refocusing pulses.
[0065] Preliminary Implementation and Testing—Volunteer Study
[0066] Example Equipment
[0067] In this example embodiment, all MR imaging experiments are performed using a 1.5 T MRI system (MAGNETOM, Avanto, Siemens) with maximum gradient amplitude of 40 mT/m and slew rate of 200 mT/m/msec. A cohort of volunteers were additionally examined at a 3 T MRI system (Magnetom Tim Trio, Siemens).
[0068] T2 Preparation Schemes
[0069] Based on preliminary evaluation of the compromise between flow sensitivity and T2 sensitivity to oxygen saturation, in this example, the preparation pulses will utilize τ.sub.180 ranging from 7.5 ms up to 25 ms, with the number of refocusing pulses ranging from 2 to 12 for each T2 map. Although additional pulses increase the overall specific absorption rate (SAR) of the pulse sequence, (for example, by a factor of approximately 1.8 from 4-pulse to 8-pulse), this has not proven to be a limiting factor at 1.5 T, the field strength at which most clinical cardiac MRI is performed, for the combination of pulses and τ.sub.180 times mentioned above. Additionally, evaluating the sequences (T2 prepared MR images with 8 refocusing pulses and T2 maps with constant τ.sub.180) in seven volunteers at 3.0 T found it to be well within SAR limits (for example, 1.35±0.24 W/kg at τ.sub.180 of 7.5 ms for variable number of pulses up to 12).
[0070] Volunteer Study at 1.5 T
[0071] The study aimed to evaluate if the technique described in this disclosure provided estimates of O2sat within normal physiological range. The study was conducted in eleven healthy volunteers (two females, mean age 27.7±6.7 years). After obtaining written informed consent to participate in the study, a venous blood sample was drawn from the subject's arm for a measurement of the hematocrit fraction (to estimate O2sat). A three lead electrocardiogram was placed on the subject's chest in order to monitor heart rate and for cardiac triggering of MR images. A six-element phased array body coil was placed over the thorax and combined with six elements of a spine array coil for signal reception. In addition, a pulse oximeter probe was placed on the finger for monitoring arterial oxygen saturation during the acquisition of MR images.
[0072] To estimate venous O2sat, four T2 maps were acquired in a short axis or horizontal long axis view of the right and left ventricles were utilized. Some of the volunteers had multiple data acquisitions (range, two to five data sets). The images were cardiac triggered and acquired free breathing during late diastole. Each T2 map was acquired using a different inter-echo spacing of 12, 15, 20 and 25 ms respectively. The other imaging parameters of the T2 maps were TR: 4000 to 5000 ms (four to six R-R intervals), one signal average, flip angle=70°, parallel acceleration=2, Bandwidth=1182 Hz/pixel, spatial resolution=2.4×2.4×10 mm. The acquisition time of each T2 map was approximately 20 to 30 seconds.
[0073] The T2 value of arterial and venous blood were measured by manually drawing contours around the lumen of the left and right ventricles in each of the T2 maps. For each volunteer, the multiple T2 measurements of the arterial and venous blood pool, along with hematocrit and non-invasive arterial O2sat were processed jointly to estimate venous oxygen saturation along with other nuisance parameters (T.sub.2o, τ.sub.ex and α). The initial estimates and bounds for the unknown parameters were: 0.8 [0, 1] for venous O2sat, 300 [0, 400] ms for 7′.sub.2o, 3 [2, 7] ms for τ.sub.ex and 0.5 [0.2, 0.6] ppm for α. These initial estimates and constraints were kept constant for all the measurements and analysis in this volunteer study.
[0074] T2 maps acquired in a volunteer in a in a mid-short axis view of the ventricles at 1.5 T for six different τ.sub.180 are shown in
TABLE-US-00001 TABLE 1 Estimated parameters Mean ± SD Range Venous O2sat (%) 73.04 ± 7.55% .sup. 61-89% .sup. T.sub.2O (ms) 245.13 ± 13.98 ms 220.95-275.45 ms τ.sub.ex (ms) 5.85 ± 1.44 ms 3.01-6.99 ms α (ppm) .sup. 0.31 ± 0.07 ppm .sup. 0.23-0.45 ppm
[0075] In this proof of concept study, the proposed calibration-free non-invasive MR oximetry method was implemented in a preliminary cohort of healthy volunteers at 1.5 T, and venous O2sat was estimated in the right ventricle. The estimated venous O2sat in the majority of the volunteers remained within the normal range of 60% to 80%. The accuracy of the venous O2sat measurement could not be verified against any invasive reference standard in this cohort, as it is impractical to conduct a research study wherein an invasive procedure has to be performed in healthy volunteers.
[0076] Volunteer Study at 3 T
[0077] A preliminary evaluation was conducted in a cohort of healthy volunteers at 3 T magnetic field strength. Subject consent and preparation procedures were the same as the volunteer study at 1.5 T. Following the acquisition of breath held localizers to identify the mid-short axis imaging plane. Four T2 prepared SSFP quantitative T2 maps (τ.sub.180=10, 12, 15 and 20 ms, TR>3000 to 4000 ms, FA=40°, 2.8×2.8×10 mm.sup.3, NEX=2, free breathing) were acquired in seven volunteers (mean age: 32.6±12.2 years, four females) on a 3 T MRI system (Tim Trio, Siemens Healthineers, Erlangen, Germany). The T2p times for the maps were 0, 20, 40, 80 and 120 ms for τ.sub.180=10 ms, 0, 24, 48, 96 and 144 ms for τ.sub.180=12 ms, 0, 30, 60 and 120 ms for τ.sub.180=15 ms and 0, 40, 80 and 160 ms for τ.sub.180=20 ms. The average acquisition time for the T2 maps was approximately 40 to 60 seconds.
[0078] The T2 of venous and arterial blood were measured in each map in an ROI in the right and left ventricle. For each volunteer, the multiple T2 measurements of the arterial and venous blood were processed jointly to estimate venous O2sat along with other nuisance parameters. The values of hematocrit (Hct) and arterial O2sat were assumed at 0.41 and 0.97 respectively for all volunteers. The initial constraints and bounds for the unknown parameters were the same as used in the volunteer study at 1.5 T.
[0079]
TABLE-US-00002 TABLE 2 Estimated parameters Mean ± SD Range Venous O2sat (%) 73.12 ± 4.26 63.80-76.26 T.sub.2O (ms) 149.65 ± 9.82 140.91-167.63 τ.sub.ex (ms) 3.99 ± 1.49 2.0-6.51 α (ppm) 0.39 ± 0.07 0.30-0.50
[0080] The average venous O2sat was similar for volunteers at both field strengths. The estimated venous O2sat for all volunteers was within the normal physiological range at 3 T.
[0081] The feasibility of the proposed oximetry technique was demonstrated in a cohort of healthy volunteers at 3 T in this study. In conclusion, the volunteer studies at 1.5 T and 3 T served to implement and establish proof of the proposed concept at clinically used field strength.
[0082] Testing and Validation—Animal Hypdxemia Study
[0083] The above methodology was tested by measuring T2 at different τ.sub.180 in a porcine model of graded hypoxia and in healthy volunteers. With respect to the animal hypoxemia study,
[0084] The study was conducted with the approval of the Institutional Animal Care and Use Committee (IACUC). Seven pigs were anesthetized with isoflurane and ventilated on 100% O2. Balloon catheters were inserted into the right atrium and proximal aorta for sampling venous and arterial blood respectively. After placement in the MR scanner, the animals were subjected to controlled graded hypoxemia by varying the ratio of oxygen to nitrogen gas inhaled. It was sought to achieve arterial O2sat levels ranging from 100% down to 70% in each animal.
[0085] Each inspired gas mixture was maintained for at least 10 minutes to allow oxygen saturation levels to stabilize before blood sampling and imaging. Arterial and venous blood samples (roughly 0.1 mL) were drawn from the aortic and right atrial catheters before and after imaging at each hypoxemic stage; the samples were immediately analyzed with a Vetscan I-stat 1 handheld blood gas analyzer (Abaxis Inc., Union City, Calif., USA). The arterial and venous O2sat for each hypoxemic stage was determined by averaging the saturation levels measured before and after imaging (approximately ten minutes apart). The hematocrit was measured in each blood sample and averaged across all measurements to determine the value for each animal.
[0086] After stepping through stages from highest to lowest level of inspired oxygen, the animal was allowed to recover for approximately 15-20 minutes by breathing 100% O2. The animal was then euthanized after a second set of measurements were made at 100% arterial O2sat.
[0087] MRI Protocol
[0088] All imaging was performed on a 1.5 T magnet (MAGNETOM Avanto, Siemens Healthineers, Erlangen, Germany) with a maximum gradient amplitude of 45 mT/m and slew rate of 200 mT/m/ms. A flexible six-element phased array body coil was placed on the thorax over the heart and combined with elements of a spine array coil for signal reception. MRI at each stage of hypoxemia included the acquisition of four T2 maps, each using a different τ.sub.180, in a single short axis view including both right and left ventricles. At each stage of hypoxemia, four T2 maps, each with T.sub.180 of 12, 15, 20 and 25 ms were acquired in a randomized order to avoid any bias that may be caused by the drifting of the O2sat levels. The images were cardiac triggered and acquired free breathing in late systole to avoid rapid, disrupted flow during diastolic filling. The imaging parameters were TR: 4000 to 5000 ms (seven to fourteen cardiac cycles), two signal averages (NEX), flip angle=70°, parallel acceleration=2, bandwidth=1182 Hz/pixel, spatial resolution=2.8×2.8×10 mm. The acquisition time of each map was approximately 40 to 50 seconds. Cardiac output was measured at the aortic outflow using a real-time velocity sequence at each hypoxemia stage (TR/TE=96.4/5.1 ms, TA=5 sec, spatial resolution=3.8×3.1×10 mm). Heart rate was monitored using a 3-lead wireless electrocardiogram, and arterial oxygen saturation was monitored by placing a pulse oximeter probe on the lower lip of the animal.
[0089] Image Analysis
[0090] All image analysis was performed on a Leonardo workstation (Siemens Healthineers, Erlangen, Germany). The T2 value of arterial and venous blood were measured by manually drawing contours around the lumen of the right ventricular (venous) and left ventricular (arterial) blood pools in each of the four T2 maps acquired at each hypoxemia stage. These measurements (four arterial and four venous), along with the measured hematocrit and reference arterial O2sat at each stage were processed jointly to estimate venous O2sat along with other nuisance parameters (T.sub.2o, τ.sub.ex, and α) in the L-M model.
[0091] Parameter estimation from the L-M model was performed with a constrained non-linear least squares method using an interior point algorithm in Matlab R2016a (The Mathworks, Natick, Mass., USA). Blood T2, hematocrit, and arterial O2sat (by blood gas analysis) were all measured, and ω.sub.0 (4×10.sup.8 rad/s at 1.5 T) and τ.sub.180 (12, 15, 20 and 25 ms) were known.
[0092] The performance of the proposed MR technique was compared against the previously proposed solution of applying a global predetermined calibration factor (K=25 s.sup.−) to a single T2 measurement (τ.sub.180=12 ms) using the simplified model in Eq. 2. There are two unknown parameters in this simplified model, T.sub.2o and SbO.sub.2. As previously proposed by Wright et al, T.sub.2o was first calculated for each hypoxemia stage using the reference measurement of arterial O2sat measured in samples drawn by invasive catheterization. The calculated T.sub.2o, predetermined K and measured venous blood T2 were then used in Eq. 2 to solve for venous O2sat.
[0093] Statistical Analysis
[0094] All variables are reported in mean±standard deviation. Statistical analysis was performed in SPSS (Version 23, IBM Corp, Armonk, N.Y., USA). Linear regression was performed to compare the relationship between the venous O2sat estimated using MRI against the reference venous O2sat measured by blood gas analysis. The bias and limits of agreement between the two methods were evaluated by the Bland Altman method. Statistical significance was inferred for p<0.05.
[0095] Results
[0096] Thirty three paired measurements of arterial and venous blood at different oxygen saturation levels were obtained from the seven animals. Two of the animals died during the experiment. However, there were three usable data sets (pertaining to inhalation conditions/hypoxemia stages) in one of these animals and one usable data set in the other; these were used in the analysis. Six data sets, where the venous O2sat fell below 40% were excluded from the analysis since rapid breathing, along with high and variable heart rates at these severe hypoxemia stages significantly degraded image quality. Therefore, the data used in the final analysis included twenty seven measurements—from one hypoxemia stage in one animal, three hypoxemia stages in two animals, and five hypoxemia stages in the other four animals.
[0097] Using exhaustive search, a set of bounds that minimized average absolute error in estimated venous O2sat was chosen. These bounds (listed in Table 3) acted as constraints for non-linear least squares curve fitting of the remaining 19 data sets. These bounds limit the solution space and thus avoid local minima or physiologically improbable values.
TABLE-US-00003 TABLE 3 Estimated Parameters Initial Value Constraints % SbO.sub.2 0.8 [0-1] T.sub.2O (ms) 300 [200-400] τ.sub.ex (ms) 3 [0-6] α (ppm) 0.545 [0.52-0.57]
[0098] The average baseline characteristics for all animals are listed in Table 4. The mean hematocrit fraction in all animals was 0.25±0.03 (range, 0.20 to 0.29). For the training set, the venous O2sat levels measured by catheter sampling ranged from 45% to 81%; for the testing set, the venous O2sat ranged from 47% to 87%.
TABLE-US-00004 TABLE 4 Baseline Characteristics Mean ± SD Heart Rate (bpm) 97 ± 18 Cardiac output (L/min) 5.42 ± 0.97 Arterial blood pH 7.45 ± 0.05 Venous blood pH 7.39 ± 0.05 Arterial pO2 (mmHg) 511.86 ± 17.53 Venous pO2 (mmHg) 44.64 ± 5.23
[0099] The heart rate and cardiac output increased with progressive hypoxemia in all animals. Arterial and venous blood T2 values decreased with lower levels of inspired O2. T2 measurements of arterial and venous blood generally decreased with increasing τ.sub.180. The T2 maps acquired at four different stages of hypoxemia in one animal are shown in
[0100] Estimation of the unknown parameters within the specified bounds revealed that the bound for T.sub.2o was active for six of the nineteen data sets, eleven sets for a and nine data sets for τ.sub.ex. The O2sat values estimated by MRI ranged from 49% to 90%. The mean and standard deviation and range of the three nuisance parameters estimated from curve fitting of the L-M model were: 225.40±27.19 ms, (200-278.88 ms) for T.sub.2o; 4.98±1.13 ms. (3.12-6 ms) for τ.sub.ex; and 0.56±0.01 ppm, (0.52-0.57 ppm) for a.
[0101] Linear regression and Bland Altman plots comparing the proposed technique and predetermined calibration method against catheter based venous O2sat measurements are shown in
[0102] One of the data sets, corresponding to an arterial O2sat of 100% and venous O2sat of 87% could not be used to determine O2sat from the predetermined calibration factors. In this data, the venous T2 (283.9 ms) was measured to be higher than the arterial T2 (274.1 ms). Of the remaining eighteen data sets analyzed, a significant linear relationship was seen for the calibration factor K=25 s.sup.−1 (y=1.09x, R=0.996, p<0.001, 95% Cl=1.05 to 1.14). The 95% Cl indicates that the slope was significantly different from 1 for K=25 s.sup.−1, but not for the method disclosed in this invention. The Bland Altman plots also revealed a significant bias for K=25 s.sup.−1 (−6.8±5.7%, p=0.0001, 95% Cl=−9.6 to −3.9%). The limits of agreement for the predetermined calibration method were larger than the technique in this disclosure (−18.0% to 4.4% for K=25 s.sup.−1).
[0103] Conclusions from Animal Hypoxemia Experiment
[0104] A method to non-invasively determine blood oxygen saturation using quantitative T2 maps has been described and validated against invasive blood gas analysis across a range of oxygen saturation levels in a porcine model of progressive hypoxemia. The range of venous O2sat examined in this study covered the normal and abnormal levels seen with cardiovascular disease. O2sat estimated by the proposed MRI method demonstrated a very good agreement with invasive blood gas analysis. This novel approach obviates the need for patient-specific in vitro calibration, and provides auto-calibrated estimation of venous oxygen saturation by MRI.
[0105] The dependence of blood T2 on the inter-echo spacing in the CPMG pulse train was exploited to acquire a range of effective T2 values as a function of τ.sub.180 and thereby generated sufficient data to perform a reliable fit of the four unknown parameters of the L-M model. In this example embodiment, these multiple estimates of blood T2 have been acquired using single-shot, T2-prepared, SSFP quantitative T2 maps of the ventricular blood pools. The T2 maps were acquired in a free breathing state, with the intention of designing and implementing a protocol that would be feasible in cardiovascular patients, many of whom have difficulty holding their breath.
[0106] The range of τ.sub.180 times were chosen such that a reasonable tradeoff between sensitivity to oxygen saturation and insensitivity to flow induced dephasing was achieved. As seen in
[0107] In the interest of overall image acquisition time and practical application in the clinic, the example embodiment in the present disclosure utilizes the relatively simple model described in Eq. 1 to demonstrate the proposed solution for MR oximetry. The same methodology presented here could be used to estimate the parameters of more complex models.
[0108] Besides providing non-invasive access to virtually any location in the cardiovascular system, this MRI-based method also offers the ability to assess average O2sat over large regions of interest. Cather-based blood sampling in chambers where sufficient mixing has not occurred may result in inaccurate O2sat measurements as a small, localized sample is obtained. Non-invasive measurement of O2sat by MRI, on the other hand, allows the spatial averaging of O2sat within a large region of interest within cardiac chambers or blood vessels. This provides effective “mixing” of the blood within the image plane and thus may overcome one limitation of diagnostic invasive catheterization. In the future, three-dimensional imaging techniques may be developed to provide full spatial coverage of cardiac chambers and large vessels.
[0109] Fitting the L-M model by a constrained non-linear least squares in the animal experiment described here showed the activation of bounds for the nuisance parameters, which implies that the solution, up to some degree, was influenced by the selection of the bounds and not just the data. However, the selection of bounds was not subjective and was completely data driven; it is conjectured that these bounds will generalize to a broader population at least within a species. Even in cases where these bounds may become suboptimal, employing bounds provides more flexibility and reliance on measured data over using fixed, generic values for the unknown parameters.
[0110] Alternative Methodologies
[0111] Exchange Model/Imaging Sequence
[0112] MR oximetry techniques have been previously described to estimate blood oxygen saturation in the pediatric and adult population. However, these methods have not gained clinical acceptance and this may be due to the need for technique-specific and patient-specific in vitro calibration procedures. While other studies have performed T2 measurements across a range of τ.sub.180, the goal of these experiments was primarily to derive predetermined calibration parameters for the estimation of O2sat from a single T2 measurement. The technique presented in this disclosure is universal and would be applicable for any mathematical representation of the exchange/diffusion model that describes the dependence of blood T2 to its corresponding O2sat as a function of the inter-echo spacing (τ.sub.180). The data could therefore, potentially be acquired with any imaging sequence that achieves T2 weighting with a train of refocusing pulses. This technique would be especially valuable in patients, who may present under different hemodynamic and metabolic states and altered blood magnetic properties than the normal population.
[0113] Exploration with Four Equations
[0114] In Eq. (1) examined in the present disclosure, there are four unknowns.
[0115] Therefore, a minimum of four equations are required to estimate the four unknown parameters from the L-M model. In an example embodiment, the animal hypoxemia data sets were evaluated to estimate venous O2 sat using only T2 maps acquired at two different τ.sub.180 times of 12 and 20 ms. This provided two equations of venous and arterial blood each from these two T2 maps. The correlation graph shown in
[0116] Estimation of Venous O2 Sat from T2 Prepared Source Images
[0117] To illustrate the example embodiment described in (202, 208), an alternative form of the T2 map was acquired during the animal hypoxemia experiment, in which a constant number of 8 refocusing pulses was implemented across the different TE times in the T2 prepared source images. The resulting T2 prepared images, therefore, had varying τ.sub.180 times (TE=T2p×τ.sub.180). The signal measured in the blood from the source images were then used to estimate venous O2 sat using the method described in (202, 208). The correlation graph shown in
[0118] Six-Parameter Model
[0119] As another example embodiment, the performance of the calibration-free approach was evaluated with the more comprehensive model defined in Eq. (5). The terms in Eq. (5) were regrouped as:
The initial constraints are estimate and bounds constraints for the six unknown parameters in this model were: 0.70 [0,1] for venous O2sat, 0.004 [0.002,0.005] ms.sup.−1 for R.sub.2p′, 3[1,8] ms for τ.sub.ex, 0.2 [0.1, 0.4] ppm for α, 0.06[0.02, 0.1] ppm for β, and 0.003[0.001,0.02] ms.sup.−1 for Δk.
[0120] Exploration without Using a Reference Measurement
[0121] In another example embodiment, the technique in the present disclosure was also evaluated without the use of an arterial T2 and O2sat reference measurement to estimate venous O2sat. The four venous T2 measurements at τ.sub.180 times of 12, 15, 20 and 25 ms were evaluated together with the hematocrit.
[0122] The venous O2sat estimated using the four equations of venous T2 alone ranged from 26% to 74%.
[0123] Testing in a Cohort of Patients
[0124] The technique described in this disclosure was further evaluated in a cohort of nine patients with cardiovascular disease. The patients (age 51.44±19.66 years, four females) were clinically indicated for invasive right heart catheterization and cardiac MRI. Four short-axis T2 maps including both arterial (left ventricle, LV) and venous (right ventricle, RV) blood were acquired at τ.sub.180=12, 15, 20 and 25 ms respectively at 1.5 T. The resulting eight blood T2 measurements (four venous and four arterial), non-invasive arterial O2sat obtained with a pulse oximeter (SpO2), and hematocrit on the day of MRI, were jointly processed to fit the L-M model using the approach described above. Venous O2sat was then compared against the invasive reference measurement.
[0125] Results
[0126] Example T2 maps acquired in one patient are shown in
[0127] The result of this preliminary evaluation in patients by direct comparison against invasive O2sat measurements is promising and demonstrates good agreement with the gold standard measurement of invasive catheterization.
CONCLUSION
[0128] Thus, in view of the disclosure above, a novel method to non-invasively determine oxygen saturation has been developed. The proposed solution for non-invasive estimation of blood O2sat in the heart was implemented and validated across a range of physiological and pathological O2sat levels in healthy volunteers, animals and cardiovascular patients. In this example embodiment, effective T2 measurements of arterial and venous blood using quantitative T2 prepared SSFP quantitative T2 maps were acquired at distinct inter-echo spacings and fit to the L-M model to non-invasively estimate O2sat and other nuisance parameters. The estimation of venous O2sat from these effective T2 measurements of the blood pool were in good agreement with the reference measurement obtained by invasive catheterization. An evaluation of the previously proposed method of applying a predetermined calibration factor to a simplified L-M model was also used to estimate venous O2sat; the results show greater bias and larger variability compared to the proposed calibration-free oximetry method.