APPARATUS FOR ULTRASOUND FLOW VECTOR IMAGING AND METHODS THEREOF
20170215836 · 2017-08-03
Inventors
Cpc classification
G01S15/58
PHYSICS
A61B8/12
HUMAN NECESSITIES
A61B8/5223
HUMAN NECESSITIES
G16H50/20
PHYSICS
G01S7/52085
PHYSICS
A61B8/4494
HUMAN NECESSITIES
G01S7/52073
PHYSICS
G16H50/30
PHYSICS
A61B8/5215
HUMAN NECESSITIES
A61B8/5207
HUMAN NECESSITIES
International classification
Abstract
Apparatus and methods of use are provided for complex flow imaging and analysis that is non-invasive, accurate, and time-resolved. It is particularly useful in imaging of vascular flow with spatiotemporal fluctuations. This apparatus is an ultrasound-based framework called vector projectile imaging (VPI) that can dynamically render complex flow patterns over an imaging view at millisecond time resolution. The VPI apparatus and methods comprise: (i) high-frame-rate broad-view data acquisition (based on steered plane wave firings); (ii) flow vector estimation derived from multi-angle Doppler analysis (coupled with data regularization and least-squares fitting); and (iii) dynamic visualization of color-encoded vector projectiles (with flow speckles displayed as adjunct).
Claims
1-10. (canceled)
11. An ultrasound imaging system comprising: a data acquisition unit, which transmits, via an ultrasonic array transducer, a series of plane waves into a tissue and receives waves reflected from the tissue to obtain beam-formed data frames; a flow vector estimation unit which estimates flow vectors at pixels of the data frames using data of the data frames; and a display rendering circuit which displays, on a display, the flow vectors estimated as moving projectiles, wherein positions of the projectiles is dynamically updated between data frames to depict trajectories of the projectiles.
12. The system of claim 11, wherein, the projectiles are color coded, and color code of the projectiles is related to velocity magnitude at the pixels.
13. The system of claim 11, wherein projectile length of the projectiles is related to velocity magnitude at the pixels.
14. The system of claim 11, wherein direction of the projectiles shows flow direction at the pixels.
15. The system of claim 11, wherein the display rendering circuit displaying the flow vectors estimated as moving projectiles comprises: selecting a set of pixels in a data frame as projectile launch points; displaying the flow vectors at the set of pixels selected as projectiles; calculating incremental displacement of the next frame based on corresponding axial-lateral velocities of the projectiles and inter-frame period; updating positions of projectiles in the next frame based on the displacement calculated; and displaying projectiles of the next frame at the positions updated.
16. The system of claim 1, wherein a projectile is displayed only if its instantaneous pixel position falls within a flow region.
17. The system of claim 11, wherein, when an instantaneous pixel position of a projectile does not fall within a flow region, a new projectile is regenerated at an original projectile launch point of this projectile.
18. The system of claim 11, wherein the display rendering circuit further displays grayscale flow speckles together with the projectiles.
19. An ultrasound imaging method, comprising: transmitting, via an ultrasonic array transducer, a series of plane waves into a tissue and receiving waves reflected from the tissue to obtain beam-formed data frames; estimating flow vectors at pixels of the data frames using data of the data frames; and displaying, on a display, the flow vectors estimated as moving projectiles, wherein position of the projectiles is dynamically updated between data frames to depict trajectories of the projectiles.
20. The method of claim 19, wherein, the projectiles are color coded, and color code of the projectiles is related to velocity magnitude at the pixels.
21. The method of claim 19, wherein projectile length of the projectiles is related to velocity magnitude at the pixels.
22. The method of claim 19, wherein direction of the projectiles shows flow direction at the pixels.
23. The method of claim 19, wherein displaying the flow vectors estimated as moving projectiles comprises: selecting a set of pixels in a data frame as projectile launch points; displaying the flow vectors at the set of pixels selected as projectiles; calculating incremental displacement of the next frame based on corresponding axial-lateral velocities of the projectiles and inter-frame period; updating positions of projectiles in the next frame based on the displacement calculated; and displaying projectiles of the next frame at the positions updated.
24. The method of claim 19, wherein a projectile is displayed only if its instantaneous pixel position falls within a flow region.
25. The method of claim 19, wherein, when an instantaneous pixel position of a projectile does not fall within a flow region, a new projectile is regenerated at an original projectile launch point of this projectile.
26. The method of claim 19, further comprising displaying grayscale flow speckles together with the projectiles.
Description
BRIEF DESCRIPTION OF THE DRAWINGS
[0014] The foregoing and other objects and advantage of the present invention will become more apparent upon reference to the following specification and annexed drawings in which:
[0015]
[0016]
[0017]
[0018]
[0019]
[0020]
[0021]
[0022]
[0023]
[0024]
[0025]
[0026]
[0027]
DETAILED DESCRIPTION OF THE INVENTION
[0028] An arrangement of apparatus for carrying out the vector projectile imaging (VPI) of the present invention is illustrated in
[0029] To achieve data acquisition an ultrasonic array transducer 10 is positioned on the outside of the patient's body adjacent the vasculature in which the fluid (e.g., blood) is to be imaged. The array 10 transmits a series of ultrasonic unfocused, steered plane waves 11 into the tissue at a high rate. The transmission angle with respect to the transducer surface 13 is changed after each wave. The transducer 10 also receives the waves reflected from the tissue at different receive steering angles and stores it in Pre-beam-Formed Data Acquisition device 12. This data is in the form of frames for each angle.
[0030] Each received frame of data is applied to a separate Beam Former 30 (enclosed in dotted lines), which includes Beam-Forming circuits 14.sub.1 to 14.sub.N. The output of each beam forming circuit is sub-sampled in Sub-Sampling circuits 16.sub.1 to 16.sub.K for levels 1 through K. In turn, the output of each sub-sampling circuit is processed in a Regularized Flow Estimation circuit 18. Then the outputs of the flow estimation circuits for a frame are averaged in circuits 20.
[0031] A Least Squares Vector Estimation circuit 22 performs two major processing stages, i.e., (i) regularizing of the frequency shift estimation on each frame and; (ii) axial-lateral vector component estimation based on least-squares fitting. This circuit uses a customized estimation algorithm with post-hoc data regularization. In Display Rendering circuit 24 a duplex visualization method is used in which the primary channel shows color-encoded particle projectiles (little arrows) whose color code and length are both related to the velocity magnitude at a particular location in the imaged vasculature. The direction of the projectiles shows flow direction. The position of these projectiles is dynamically updated between frames to quantitatively highlight flow paths together with changes in magnitude and orientation. The secondary visualization channel depicts grayscale flow speckles that are derived based on the slow-time filter power. This supplementary flow information serves as an adjunct depiction of flow trajectories. The output of circuit 24 is the VPI image 26
[0032] Data Acquisition
[0033] Data acquisition in VPI is based upon the use of plane wave transmissions and parallel beam-forming (both performed from multiple angles). As shown in
[0034] Using a VPI framework, acquisition of flow vector information at high frame rates is facilitated by performing steered plane wave transmissions whose operating principles have recently matured for ultrasound imaging applications. See, Montaldo G, Tanter M, Bercoff J, Benech N, Fink M., “Coherent plane-wave compounding for very high frame rate ultrasonography and transient elastography,” IEEE Trans. Ultrason. Ferroelec. Freq. Contr., (2009), 56: 489-506, which is incorporated herein in its entirety. As shown in
[0035] For each plane wave transmission at a given Tx angle, a set of N beam-formed data frames is generated in parallel based on the corresponding array of channel-domain pulse echoes (waves reflected from the tissue) which are acquired from the transducer 10 at the Pre-Beam Forming Data Acquisition Device 12. As illustrated in
[0036] In order to monitor temporal changes in flow dynamics, plane wave transmission and receive beam-forming are carried out multiple times for all MN Tx-Rx angle pairs (as indicated in
[0037] Flow Vector Estimation
[0038] At each slow-time instant, the VPI method of the present invention performs flow vector estimation independently at all pixel positions based on their corresponding set of MN slow-time ensembles.
[0039] Flow vector estimation in VPI works by processing, for every pixel position, its corresponding set of slow-time signals from all Tx-Rx angle pairs.
[0040] As illustrated in
[0041] To obtain consistent slow-time frequency shift estimates as necessary for accurate vector computation, a customized estimation algorithm with post-hoc data regularization is devised to individually process every filtered slow-time ensemble. As shown in
[0042] Following the estimation steps, a two-stage post-hoc processing strategy is performed to regularize entries of the slow-time frequency estimate ensembles. First, to remove spurious estimates at time instants where flow is not detected by a Tx-Rx angle pair, entries in the slow-time frequency ensembles are set to zero if their respective slow-time power estimate at that instant is below a predefined threshold (i.e. a flow classification mask 38 is applied similar to the color gain mask in color flow imaging). Second, akin to previous efforts in color flow signal processing, phase unwrapping is applied to the sifted slow-time frequency estimates to account for possible aliasing artifacts that may well occur when performing the lag-one autocorrelation algorithm. See, Lai X, Torp H, Kristoffersen K, “An extended autocorrelation method for estimation of blood velocity,” IEEE Trans. Ultrason. Ferroelec. Freq. Contr., (1997), 44: 1332-1342, which is incorporated herein in its entirety. This step effectively extends the dynamic range of the slow-time frequency estimates beyond the Nyquist sampling limit. Note that both regularization steps are performed individually on all MN ensembles of slow-time frequency estimates.
[0043] Least-Squares Vector Computation Algorithm
[0044] Once the MN slow-time frequency shifts are estimated from all Tx-Rx angle pairs, the axial and lateral components of the flow vector are derived using multi-angle Doppler analysis principles. See, Dunmire et al. 2000 cited above, which is incorporated in its entirety. This computational task is equivalent to solving an over-determined system of equations whereby MN data values (i.e. frequency shift estimates from all the Tx-Rx angle pairs at one slow-time instant) are used as inputs to solve for two unknowns (i.e. axial velocity and lateral velocity). Specifically, the flow vector v=(v.sub.z, v.sub.x) can be estimated through a least-squares fitting approach. The advantage of adopting this algebraic framework is that each resulting flow vector estimate would be optimized in the sense that its mean squared error is minimized for a given input. In turn, in cases with noisy data input, consistent estimation performance can still be maintained compared with the two-equation, two-unknown formulation that corresponds to the conventional two-angle vector Doppler method (Ekroll et al. 2013; Kripfgans et al. 2006 cited above).
[0045] In the least-squares vector computation method, v can be calculated by carrying out a matrix operation with each MN×1 measurement vector u (consisting of individual frequency shift values).
[0046] The least-squares flow vector estimator can be considered a generalized form of the cross-beam Doppler estimation method in which multiple Tx-Rx angle pairs are used in lieu of the two-angle approach that has been reported previously. See, Kripfgans et al. 2006; Tortoli et al. 2006, 2010; and Ekroll et al. 2013, all cited above and incorporated herein in their entirety. In the case with MN combinations of Tx-Rx angle pairs, the m.sup.th Tx angle can be denoted as θ.sub.m and the n.sup.th Rx angle can be denoted as φ.sub.n. Note that, since plane wave transmissions and dynamic receive focusing are used, θ.sub.m and φ.sub.n would remain the same for all pixels within the imaging view. It is well-known from the Doppler equation that, with this angle pair configuration, the slow-time frequency shift φ.sub.mn for an object moving at velocity magnitude v and angle α is equal to:
where c.sub.o is the acoustic speed and f.sub.o is the ultrasound center frequency. See, Dunmire et al. 2000, cited above and incorporated herein in its entirety. Following similar derivations, the mathematical form expressed in (A1) can be modified by noting two points: (i) there exists a trigonometry relation [cos(A−B)=cos(A)cos(B)+sin(A)sin(B)]; and (ii) the axial and lateral velocity components are respectively equal to v.sub.z=v cos(α) and v.sub.x=v sin(α). See, Tsang I K H, Yiu B Y S, Yu A C H, “A least-squares vector flow estimator for synthetic aperture imaging,” Proc. IEEE Ultrason. Symp., (2009), 1387-1390, which is incorporated herein in its entirety. Substituting these relations into (A1), the following revised form of the Doppler equation can be obtained:
[0047] The flow vector estimator seeks to solve for v.sub.x and v.sub.z, the two unknowns in (A2), by forming an over-determined system of equations from MN realizations of (A2) as made available through the use of different Tx-Rx angle pairs. In matrix notation, this system of equations can be expressed in the following form for a given flow vector v=(v.sub.x, v.sub.z):
where A is the angle-pair matrix (MN×2 in size) and u as the measurement vector (MN×1 in size). Note that u.sub.mn is essentially equal to the right hand side of (A2) for a given Tx-Rx angle pair (i.e. u.sub.mn=c.sub.oφ.sub.mn/f.sub.o).
[0048] From linear algebra principles, it is well known that v in Equation (3) can be found by multiplying the pseudo-inverse of A with v: a solution that is often referred to as the least-squares fitting solution. See, Moon T K, Stirling W C, “Mathematical Methods and Algorithms for Signal Processing,” Upper Saddle River: Prentice-Hall, (2000), which is incorporated herein in its entirety. Thus, with each MN×1 measurement vector u (consisting of individual frequency shift values), v can be calculated by carrying out the following matrix operation:
where the T superscript denotes a matrix transpose operation and entity (A.sup.TA).sup.−1A.sup.T is well-known in linear algebra as the pseudo-inverse of matrix A. See Moon T K, Stirling W C, “Mathematical Methods and Algorithms for Signal Processing,” Upper Saddle River: Prentice-Hall, (2000), which is incorporated herein in its entirety. Note that the pseudo-inverse (A.sup.TA).sup.−1A.sup.T is essentially a 2×MN matrix of constant values (as long as the Tx-Rx angle pairs remain the same). Thus, the same pseudo-inverse is applicable to different pixel positions. It is also worth pointing out that, for the least-squares estimator given in Equation (4), the resulting flow vector estimate can be considered as an optimal solution in the sense that its mean-squared error is minimized for a given input. It is worth emphasizing that Equation (4) is carried out individually at every pixel position and at each slow-time instant.
[0049] Dynamic Visualization Procedure
[0050] Using the computation protocol according to the present invention, frames of flow vector information can be generated at a rate of f.sub.VPI, which equals to f.sub.DAQ/K [i.e. f.sub.PRF/(MK)] for a step size of K slow-time samples when executing the sliding window implementation. To facilitate dynamic rendering of these flow vector estimates, a novel duplex visualization method is used. Its primary visualization channel shows color-encoded particle projectiles (small arrows) whose color code and projectile length are both related to the velocity magnitude (on a scale from zero to a tunable maximum value). The direction of the projectiles shows flow direction. The position of these projectiles is dynamically updated between frames to quantitatively highlight flow paths together with changes in magnitude and orientation. The secondary visualization channel depicts grayscale flow speckles that are derived based on the slow-time filter power. This supplementary flow information serves as an adjunct depiction of flow trajectories in ways similar to that offered by the B-flow imaging technique. See, Chiao R Y, Mo L Y, Hall A L, Miller S C, Thomenius K E, “B-mode blood flow (B-flow) imaging,” In: Proceedings, IEEE Ultrasonics Symposium, San Juan, Puerto Rico, 22-25 October. New York: IEEE; (2000), p. 1469-1472; and Lovstakken L, Bjaerum S, Martens D, Torp H, “Blood flow imaging—A new real-time, 2-D flow imaging technique,” IEEE Trans Ultrason Ferroelectr Freq Control (2006), 53:289-299. Note that the graphical representation of multidirectional flow dynamics rendered by our duplex visualization approach is essentially different from the dot-based particle visualization algorithm that has been reported recently in ultrasound flow imaging (Flynn et al. 2011).
[0051] VPI provides quantitative flow visualization through dynamic rendering of color-encoded particle projectiles. In
[0052] The dynamic projectile method of the VPI visualization protocol works as follows: First, as shown in
[0053] Hardware and Parameters
[0054] The VPI invention has been implemented using a research-purpose, channel-domain imaging platform that allows the transmission and reception operations of each array element to be configured individually. This platform is a composite system in which the front-end of a SonixTouch research scanner (Ultrasonix, Richmond, BC, Canada) was coupled to a pre-beam-formed data acquisition tool, and data was streamed to a back-end computing workstation through a universal serial bus link (specifications listed in Table 1a). An L14-5 linear array (Ultrasonix) was used as the operating transducer. See, Cheung C C P, Yu A C H, Salimi N, Yiu B Y S, Tsang I K H, Kirby B, Azar R Z, Dickie K, “Multi-channel pre-beamform data acquisition system for research on advanced ultrasound imaging methods,” IEEE Trans. Ultrason. Ferroelec. Freq. Contr., (2012) 59: 243-253, which is incorporated herein in its entirety.
[0055] Pressure field recordings of this ultrasound transmission hardware were taken using a membrane hydrophone (HMB-0500, Onda, Sunnyvale, Calif., USA) that was mounted on a three-axis micro-positioner (ASTS-01, Onda). When operating in plane wave excitation mode, our scanner hardware generated a derated peak negative pressure of 0.72 MPa (located at 2-cm depth). This pressure value, for our given pulsing parameters, corresponded to a mechanical index of 0.32; spatial-peak, temporal-average intensity of 0.16 W/cm2; and spatial-peak, temporal-peak intensity of 27 W/cm2 (assuming operation in 37_C degassed water). These numbers were well within the safety limits defined by the U.S. Food and Drugs Administration. See Duck F A, “Medical and non-medical protection standards for ultrasound and infrasound,” Prog Biophys Mol Biol (2007) 93:176-191.
[0056] A vector estimation configuration with three Tx angles (−10°, 0°, +10°) and three Rx angles (−10°, 0°, +10°) was implemented (i.e. M=3, N=3, MN=9). To realize such a configuration, a steered plane wave pulsing sequence was programmed on the platform by executing relevant functions in the TEXO software development kit (Ultrasonix) to define array channel delays that only generate angle steering without focusing. Typical pulse-echo imaging parameters were used as summarized in Table 1b, and pre-beam-formed channel-domain data was acquired on reception.
TABLE-US-00001 TABLE 1 Parameters Used for Experimental Realization of VPI Parameter Value (a) Imaging Platform Number of Tx/Rx Channels 128 Array Pitch 0.3048 mm Pre-Beam formed Data Sampling Rate 40 MHz Pre-Beam formed Data Bit Resolution 12 (b) Data Acquisition Imaging Frequency 5 MHz Tx Pulse Duration 3 cycles (0.6 μs) Pulse Repetition Frequency 10 kHz Effective Data Acquisition Rate 3.33 kHz Maximum Imaging Depth 2 cm Data Acquisition Duration 1 s (c) Beam Forming Pre-Beam-formed Data Filter Pass band 3-7 MHz Filter Design Method Equiripple Data Frame Size 200 × 380 pixels Pixel Dimension 0.1 × 0.1 mm (d) Slow-time data processing Normalized clutter filter cutoff 0.05 Sliding Window for flow estimation 128 samples Sliding Window Step Size 8 samples (d) VPI Visualization Nominal Frame Rate 416 fps Launch Point Density 4% Mean Projectile Lifetime 30 frames Flow Speckle Dynamic Range 40 dB
TABLE-US-00002 TABLE 2 Alternative Experimental Parameters Parameter Value Imaging Specifications Imaging frequency 5 MHz # of array channels 128 Pulse duration 3 cycles (0.429 μs) Pulse repetition frequency 10 kHz Tx-Rx steering angles −10°, 0°, +10° Flow Pattern Specifications Peak inlet flow rate 6 ml/s Pulse cycle frequency 1.2 Hz
[0057] A 3-Tx, 3-Rx VPI configuration was implemented. Also, for each Tx-Rx angle pair, a 3-level sub-sampling was imposed during flow estimation. VPI was tested on anatomically realistic flow models that resembled healthy and stenosed carotid bifurcation. These are suitable geometries because flow dynamics within them are known to be multi-directional and significantly time-varying. The phantoms are wall-less designs based on lost-core casting with polyvinyl alcohol gel. Pulsatile flow is supplied through the use of a gear pump with programmable flow rates. The resulting images are shown in
[0058] After streaming the acquired data offline to the back-end processor, various image formation and visualization operations were carried out as required for VPI. First, to improve the channel-domain signal-to-noise ratio, a finite-impulse-response band pass filter (minimum order; parameters listed in Table 1c) was applied to the pre-beam-formed data of each channel using Matlab (R2012a; Mathworks, Natick, Mass., USA). After that, delay-and-sum beam-forming from the three Rx angles were executed using a graphical processing unit (GPU) based parallel computing approach like that disclosed in Yiu et al. 2011, which was previously cited and is incorporated by reference in its entirety. Note that an array of two GTX-590 GPUs (NVidia, Santa Clara, Calif., USA) was used for this operation to facilitate processing at real-time throughput. Subsequently, an implementation for speckle imaging was used as disclosed in Yiu B Y S, Tsang I K H, Yu A C H, “GPU-based beam former: fast realization of synthetic aperture imaging and plane wave compounding,” IEEE Trans. Ultrason. Ferroelec. Freq. Contr., (2011), 58: 1698-1705. In addition, clutter filtering (in the form of a minimum order high-pass finite-impulse-response filter; see Table 1d for parameters) and lag-one autocorrelation were performed individually on the nine Tx-Rx angle pairs at each pixel position. Other downstream operations related to vector estimation (post-hoc regularization and least-squares fitting) were then conducted in Matlab. Ad hoc persistence and median filtering were also performed. At last, VPI image sets were obtained and rendered using the duplex dynamic visualization algorithm of the present invention and the parameters listed in Table 1d. Note that the flow information in our VPI image sets was overlaid on top of a background B-mode image frame that was formed from spatial compounding of the beam-formed data frames formed from the nine Tx-Rx angle pairs.
[0059] In order to evaluate the accuracy of the flow vector estimation algorithm used in our VPI technique, steady-flow calibration experiments were first conducted using a multi-vessel flow phantom that we fabricated in-house. The phantom comprised three wall-less straight tubes whose long axes were aligned along the same plane; each tube had a different diameter (2, 4 and 6 mm) and flow angle (−10°, 0° and +10°), and they were positioned at different depths (1.5, 4 and 6 cm). By use of an investment casting protocol similar to that described in our previous work (Yiu and Yu 2013, previously cited), the phantom was fabricated with polyvinyl alcohol cryogel as the tissue-mimicking material, whose acoustic attenuation coefficient and acoustic speed were respectively measured to be 0.24 dB/cm, MHz and 1518 m/s.
[0060] The phantom was connected to a gear pump (AccuFlow-Q, Shelley Medical Imaging, London, ON, Canada) that supplied continuous circulation of blood-mimicking fluid (Shelley; 1037 kg/m3 density, 3.95 3 106 m2/s viscosity) at a flow rate of 2.5 mL/s. The transducer scan plane was aligned to the long axis of the three vessels (which were expanded to diameters of 2.2, 4.4 and 6.3, respectively, because of flow-mediated dilation). Raw data were then acquired for the three-Tx, three-Rx configuration based on the parameters described above, and the lumen velocity profiles were estimated using VPI's flow vector computation algorithm. Results were correlated with the theoretical parabolic profiles, whose centerline velocities were 131, 31 and 16 cm/s, respectively, for the three dilated vessels
[0061] To assess the practical efficacy of the VPI technique, this framework was used to image complex flow dynamics inside anatomically realistic carotid bifurcation phantoms. In particular, efficacy of VPI was evaluated through an anthropomorphic carotid bifurcation phantom study.
[0062]
[0063] Note that the carotid bifurcation vasculature is rather suitable for this investigation because it possessed curved vessel geometries in the vicinity of the junction between three branches: the common carotid artery (CCA), the internal carotid artery (ICA), and the external carotid artery (ECA). In other words, it effectively allowed testing of the ability of the three-Tx, three-Rx VPI configuration to track flow patterns with significant multi-directional and spatiotemporal variations. Another point of merit in using these geometries is that their flow dynamics have already been extensively characterized by others using optical particle image velocimetry (Kefayati S, Poepping T L “Transitional flow analysis in the carotid artery bifurcation by proper orthogonal decomposition and particle image velocimetry.” Med. Eng. Phys., 2013; 35: 898-909.; Poepping et al. (2010)) and computational fluid dynamics (Steinman D A, Poepping T L, Tambasco M, Rankin R N, Holdsworth D W, “Flow patterns at the stenosed carotid bifurcation: effect of concentric versus eccentric stenosis,” Ann. Biomed. Eng., 2000; 28: 415-423. This information effectively provides an established reference for comparison with the flow patterns rendered by VPI.
[0064] Two different carotid bifurcation phantom models were used for experimentation (i) healthy co-planar geometry (
[0065] A steady-flow calibration experiment was first conducted by connecting the healthy bifurcation phantom to a gear pump (AccuFlow-Q; Shelley Medical Imaging, London, ON, Canada) that supplied continuous circulation of blood mimicking fluid (Shelley; 1037 kg/m.sup.3 density, 3.95×10.sup.6 m.sup.2/s viscosity) at 5 ml/s flow rate. The transducer scan plane was aligned to the CCA long axis (expanded to 6.4 mm diameter due to flow-mediated dilation), and raw data was acquired for the three-Tx, three-Rx configuration based on the parameters described earlier. The lumen velocity profile was then estimated using VPI's flow vector computation algorithm. Results were compared to the theoretical parabolic profile (with 31 cm/s centreline velocity).
[0066] Next, pulsatile flow experiments were performed using the bifurcation phantoms. The flow pulse, with a 72 bpm pulse rate (i.e. 1.2 Hz) and a 5 ml/s systolic flow rate, resembled a carotid pulse pattern that featured a primary systolic upstroke and a secondary dicrotic wave (we-defined in the pump system). VPI cineloops were then generated by processing raw data acquired under such flow settings to determine the ability of VPI in visualizing complex flow features. To facilitate comparison, Doppler spectrograms were computed at representative pixel positions in the imaging view by reprocessing the raw slow-time ensembles at those places.
[0067] The vector computation algorithm of VPI was found to be capable of deriving flow vector estimates at high accuracy. Corresponding results obtained from the steady-flow calibration experiment are shown in
[0068] The plots are the flow vector profiles in the CCA of a healthy bifurcation phantom (with 5 ml/s constant flow rate) with the transducer placed in parallel with the CCA vessel. As can be observed in
[0069] The estimated flow speed magnitude across the lumen of the three vessels was generally found to resemble a parabolic shape that matched well with the theoretical prediction. As illustrated in
[0070] Calibration results showed that, using three transmit angles and three receive angles (−10°, 0°, +10° for both), VPI can accurately compute flow vectors even when the transducer was placed in parallel to the vessel (6.4 mm dilated diameter; 5 ml/s steady flow rate). The practical merit of VPI was further demonstrated through an anthropomorphic flow phantom investigation that considered both healthy and stenosed carotid bifurcation geometries. For the healthy bifurcation with 1.2 Hz carotid flow pulses, VPI was able to render multi-directional and spatiotemporally varying flow patterns (using 416 fps nominal frame rate, or 2.4 ms time resolution). In the case of stenosed bifurcation (50% eccentric narrowing), VPI enabled dynamic visualization of high-speed flow jet and recirculation zones.
[0071] Using the VPI technique, time-resolved quantitative visualization of multi-directional and spatiotemporally varying flow patterns that emerge within curvy vasculature under pulsatile flow conditions were achieved for a healthy carotid bifurcation with 72 bpm pulse rate. As an illustration,
[0072] The nominal VPI frame rate (f.sub.VPI) was 416 fps, and it was played back at 50 fps (f.sub.DAQ was 3,333 Hz). The rendered flow dynamics were found to be consistent with well-established findings obtained from computational predictions See, Berger S A, Jou L D, “Flows in stenotic vessels,” Annu. Rev. Fluid Mech., (2000) 32: 347-382). In particular, it can be readily observed that the temporal evolution of flow speed and flow direction rendered by VPI in different parts of the vasculature are, as expected, synchronized with the stroke of the flow pulse. Also, in the ECA branch (lower branch), streamlined forward flow (without reversal) along the vasculature was evident throughout the pulse cycle. This latter observation effectively demonstrates that VPI's flow vector estimation procedure is robust against flow angle variations, which do arise in the ECA as its inlet segment is inherently tortuous.
[0073] The technical merit of VPI is perhaps more notably demonstrated by its rendering of flow disturbances in the ICA branch of the healthy carotid bifurcation. This branch corresponds to the upper branch in
[0074] The millisecond time resolution (2.4 ms for 416 fps nominal frame rate) of VPI effectively enabled tracking of when a flow disturbance emerged in the carotid bulb of the healthy bifurcation vasculature.
[0075] To further demonstrate VPI's efficacy in accurately visualizing highly complex flow dynamics,
[0076] One striking observation to be noted is that, during systolic upstroke, the formation of a high-velocity flow jet (red arrows) can be dynamically visualized at the site of stenosis. Indeed, the flow jet continued to propagate along the inner wall side of the ICA sinus (i.e. the unstenosed side) and spurted across the ICA lumen. It then collided against the outer ICA wall near the distal end of the carotid bulb where the ICA vessel started to become straightened. Upon hitting the outer wall, the jet direction was reoriented tangentially against the wall and eventually ramped off before it dissipated further downstream.
[0077] In
[0078] As illustrated in
[0079] Using ultrasound to visualize complex flow dynamics is inherently not a straightforward task. In developing a prospective solution, two practical vascular flow conditions must be taken into account: (i) at a given time instant, flow speed and direction (i.e. the flow vector) may vary spatially because of the tortuous nature of vascular geometry; and (ii) over a cardiac pulse cycle, flow components would deviate temporally due to pulsatile behaviour. VPI has been designed to capture and render these spatiotemporal dynamics in blood flow.
[0080] From a technical standpoint, VPI is equipped with three key features that enable time-resolved visualization of flow vectors over an imaging view. First, it performs high-frame-rate broad-view data acquisition via multi-angle plane wave imaging principles, so as to achieve the high time resolution required to monitor flow pulsations and their spatial variations over an imaging view (
[0081] The practical merit of VPI in visualizing complex flow dynamics is demonstrated through a carotid bifurcation phantom study with controlled flow conditions that are otherwise not possible in-vivo (
[0082] As an integrative insight into VPI's application potential in delineating specific details of complex flow patterns,
[0083] In the case of healthy bifurcation, it should be noted that the recirculation zone in the carotid bulb region is larger at the end of post-systolic down stroke (
[0084]
[0085] For the stenosed carotid bifurcation, the spatial extent of its two flow recirculation zones shows substantial differences. While VPI did not detect significant changes in the size of the carotid bulb recirculation zone over the pulse cycle (
[0086] As mentioned above, the adjunct display of grayscale flow speckles can enhance the flow visualization performance of the algorithm because inter-frame flow speckle displacements can serve to highlight the flow trajectory path.
[0087]
[0088] Being a newly developed technique with fine temporal resolution and flow vector estimation capabilities, VPI can be leveraged to investigate various forms of complex flow dynamics. For instance, besides using VPI to study flow patterns in the carotid bifurcation as demonstrated here, this technique can be used to examine multi-directional flow dynamics inside diseased vascular features such as aneurysms. Also, VPI can be applied to visualize flow turbulence with fluttering features that require fine temporal resolution to render coherently. Realizing these applications would effectively substantiate the diagnostic value of VPI in complex flow analysis.
[0089] As the fine temporal resolution offered by VPI hinges on the use of broad-view data acquisition sequences in which the ultrasound firings are unfocused in nature, the flow signals returned from deeper vasculatures would inevitably be weaker as a consequence. This issue, which would be physically worsened by depth dependent attenuation, may pose a challenge when diagnosing certain patients whose vasculature tends to be positioned farther away from the skin surface. Hence, flow signal enhancement techniques may be used to reinforce the efficacy of the VPI framework when used in different in-vivo scan settings. One particular strategy that can be used is the incorporation of coded excitation principles into the transmission pulse sequence design. See Zhao H, Mo L Y L, Gao S, “Barker-coded ultrasound color flow imaging: Theoretical and practical design considerations,” IEEE Trans Ultrason Ferroelectr Freq Control (2007), 54:319-331, which is incorporated herein in its entirety. Alternatively, microbubble contrast agents may be introduced to boost the flow signal level when performing VPI. See, Tremblay-Darveau C, Williams R, Milot L, Bruce M, Burns P N, “Ultrafast Doppler imaging of microbubbles,” In Proceedings 2012 IEEE Ultrasonics Symposium, Dresden, Germany, 7-10 October. New York: IEEE; (2012), p. 1315-1318.
[0090] Another aspect of VPI to be further refined is its engineering considerations regarding the technique's real time realization. In the arrangement of
[0091] Alternatively, VPI can be further used for echocardiography investigations where vector visualization of intracardiac flow fields currently relies on either post-processing of color flow imaging data or the use of microbubble contrast agents to perform echo particle image velocimetry. To realize VPI for echocardiography applications, the flow vector estimation framework would need to be further refined to account for the non-stationary tissue clutter that arises due to myocardial contraction. For instance, when deriving the frequency shift estimates of each Tx-Rx angle pair, advanced signal processing solutions that are resilient against tissue motion biases could be adapted for this purpose, such as maximum likelihood estimation and adaptive-rank eigen-estimation. See, Lovstakken L, Bjaerum S, Torp H, “Optimal velocity estimation in ultrasound color flow imaging in presence of clutter,” IEEE Trans. Ultrason. Ferroelec. Freq. Contr., (2007), 54: 539-549; and Yu A C H, Cobbold R S C, “Single-ensemble-based eigen-processing methods for color flow imaging—Part II. The matrix pencil estimator,” IEEE Trans. Ultrason. Ferroelec. Freq. Contr., (2008), 55: 573-587, both of which are incorporated herein in their entirety.
[0092] VPI can be considered a new approach in leveraging ultrasound for flow estimation purposes. In particular, it represents a drastic transformation in the way that flow information is acquired, estimated, and rendered in comparison to conventional color flow imaging. Since VPI is essentially non-invasive, this technique should hold promise in being introduced as a routine diagnostic tool to investigate complex flow dynamics in the human vasculature. For instance, in the context of carotid diagnostics, VPI can potentially be adopted as a more instinctive way to assess the severity of carotid stenosis compared with the conventional Doppler spectrogram mode that is routinely performed as part of the clinical practice for carotid disease management. If such clinical translation effort can be realized, the present role of ultrasound in vascular diagnostics can undoubtedly be expanded.
[0093] The invention is not to be limited in scope by the specific embodiments described herein. Indeed, various modifications of the invention in addition to those described will become apparent to those skilled in the art from the foregoing description and accompanying figures. Such modifications are intended to fall within the scope of the appended claims.