Carrier composition for bone substitute materials
11324859 · 2022-05-10
Assignee
Inventors
Cpc classification
A61L2300/112
HUMAN NECESSITIES
A61L27/3821
HUMAN NECESSITIES
A61L27/446
HUMAN NECESSITIES
A61L27/18
HUMAN NECESSITIES
A61L2400/12
HUMAN NECESSITIES
A61L27/58
HUMAN NECESSITIES
A61L27/18
HUMAN NECESSITIES
A61L2430/02
HUMAN NECESSITIES
A61L27/025
HUMAN NECESSITIES
A61L27/427
HUMAN NECESSITIES
International classification
A61L27/18
HUMAN NECESSITIES
Abstract
The present invention relates to a carrier composition for particulate and granular bone substitute materials which is a hydrogel comprising a mixture of ethylene oxide (EO)-propylene oxide (PO) block copolymers and silica nanoparticles embedded therein. The present invention further relates to a bone substitute material containing osteoconductive and/or osteoinductive particles or granules in addition to the novel carrier composition. Processes for producing the novel carrier composition and the novel bone substitute material are likewise provided in the context of the invention.
Claims
1. A carrier composition for particulate and granular bone substitute materials, wherein the carrier composition is a hydrogel comprising: (a) an ethylene oxide (EO)-propylene oxide (PO) block copolymer or a mixture of ethylene oxide (EO)-propylene oxide (PO) block copolymers; and (b) silica nanoparticles with a size of between about 0.5 nm and about 10 nm.
2. The carrier composition according to claim 1, wherein the proportion of water in the hydrogel ranges from 60% to 90%.
3. The carrier composition according to claim 1, wherein the proportion of ethylene oxide (EO)-propylene oxide (PO) block copolymers in the hydrogel is between about 10% and 40% (w/w).
4. A carrier composition according to claim 1, wherein the proportion of silica nanoparticles is between about 2% and 12% (w/w).
5. A carrier composition according to claim 1, wherein the silica nanoparticles have a size between about 0.5 nm and 1.5 nm.
6. A carrier composition according to claim 1, wherein the silica nanoparticles form fractal aggregation clusters having an average size of less than 200 nm.
7. A carrier composition according to claim 1, wherein the ethylene oxide (EO)-propylene oxide (PO) block copolymers in the carrier composition have a molecular weight distribution between about 1,000 g/mol and 70,000 g/mol.
8. A carrier composition according to claim 1, wherein at least 30% (w/w) of the ethylene oxide (EO)-propylene oxide (PO) block copolymers in the carrier composition consist of a poloxamer.
9. A bone substitute material comprising: (a) a carrier composition according to claim 1; and (b) osteoconductive and/or osteoinductive particles or osteoconductive and/or osteoinductive granules.
10. A bone substitute material according to claim 9, wherein the osteoconductive or osteoinductive particles have a size between about 5 μm and 100 μm.
11. A bone substitute material according to claim 9, wherein the osteoconductive or osteoinductive particles are hollow spheres having an opening.
12. A bone substitute material according to claim 11, wherein the hollow spheres form clusters of a size between about 100 μm and 3,000 μm.
13. A bone substitute material according to claim 9, wherein the osteoconductive or osteoinductive particles or the osteoconductive and/or osteoinductive granules consist of hydroxyapatite crystallites which have the morphology of the biological hydroxyapatite of the bone and are coated with a matrix of silica xerogel.
14. A bone substitute material according to claim 9, wherein the osteoconductive and/or osteoinductive particles or the osteoconductive and/or osteoinductive granules are coated with a silica gel.
15. A process for the preparation of a bone substitute material comprising the steps of (a) providing a carrier composition according to claim 1; (b) optionally treating the carrier composition with gamma radiation; and (c) mixing the carrier composition with osteoconductive and/or osteoinductive particles or with osteoconductive and/or osteoinductive granules.
16. A carrier composition according to claim 8, wherein said poloxamer is poloxamer 407.
17. A carrier composition according to claim 8, wherein said poloxamer has an average molecular weight in the range of 9,800 to 14,600 g/mol.
18. A bone substitute material according to claim 14, wherein the silica concentration in the silica gel is between about 3% and 10%.
19. The process of claim 15, wherein the osteoconductive and/or osteoinductive particles or the osteoconductive and/or osteoinductive granules are coated with a silica hydrogel.
Description
DESCRIPTION OF THE FIGURES
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
(11)
(12)
(13)
(14)
(15)
EXAMPLES
(16) The following examples illustrate the effectiveness as well as the advantages of the carrier composition according to the invention and the bone substitute material formulated from it.
Example 1: Production of Hydrogels with and without SiO.SUB.2
(17) For comparison purposes, SiO.sub.2-free and SiO2-containing hydrogels were produced on the basis of Kolliphor P 407. For the production of the SiO.sub.2-free hydrogels, 23.5 g Kolliphor P 407 from BASF were mixed with 76.5 g water. For the hydrogels containing SiO.sub.2, a sol was prepared by ion exchange with a SiO.sub.2 concentration of 4% and 6%, respectively. Concentrated sodium water glass solution from Merk (specification: Na.sub.2O: 7.5-8.5%; SiO.sub.2: 25.5-28.5%) was used and diluted with ultrapure water. A Lewatit MonoPlus SP 112Na+ column was used as an ion exchanger. The soles had a pH value of 2.7 and were cooled down to 5° C. In each 76.5 g of the sol 23.5 g of Kolliphor P 407 were stirred in. The resulting hydrogels contain polymers with the molecular mass distribution shown in
(18) Peak 1: Position: 5,550 g/mol, proportion: 17.8%
(19) Peak 2: Position: 11,000 g/mol, proportion: 8.2%
(20) Peak 3: Position: 13,470 g/mol, proportion: 73.1%
(21) Peak 4: Position: 25,500 g/mol, proportion: 0.8%
(22) The peaks at 5,550 g/mol and 11,000 g/mol represent fragments of Kolliphor 407. The peak at 25,500 g/mol results from the cross-linking of two chains.
(23) Some of the prepared samples were treated with gamma radiation (17.5 to 30 kGray, radiation source: cobalt 60, maximum activity 111 PBq). The gamma radiation leads to a cross-linking of the polymer chains. At the same time, chains are also broken. The result is a polymer with a broad molecular mass distribution.
(24) These hydrogels contain polymers with the molecular mass distribution shown in
(25) Peak 1: Position: 5,400 g/mol, proportion: 8.1%
(26) Peak 2: Position: 11,000 g/mol, proportion: 4.0%
(27) Peak 3: Position: 13,400 g/mol, proportion: 37.0%
(28) Peak 4: Position: 17,000 g/mol, proportion: 2.7%
(29) Peak 5: Position: 25,500 g/mol, proportion: 4.2%
(30) Peak 6: Position: 35,000 g/mol, proportion: 0.2%
(31) After irradiation, the proportion of the continuous mass distribution was 43.8%. The original Kolliphor 407 only has a proportion of 37%. Molecules with a continuous size distribution of up to approx. 70,000 g/mol have the largest proportion of 43.8%.
(32) Another part of the prepared samples was stored at elevated temperature for a longer period of time. After storage for 55 days at 60° C., the hydrogels contained polymers with the molecular mass distribution shown in
(33) Peak 1: Position: 5,350 g/mol, proportion: 10.6%
(34) Peak 2: Position: 8,200 g/mol, proportion: 13.7%
(35) Peak 3: Position: 13,470 g/mol, proportion: 23.6%
(36) Peak 4: Position: 17,700 g/mol, proportion: 1.9733%
(37) Peak 5: Position: 24,700 g/mol, proportion: 1.5018%.
(38) Peak 6: Position: 35,000 g/mol, proportion: 0.0%
(39) It can be seen that also in this case molecules with a size distribution ranging from about 1,000 g/mol to about 70,000 g/mol show the largest proportion of 48.7%.
(40) For all samples the viscosity was measured as a function of the shear rate (StrainSweep Test, oscillation rheometer ARES—T.A. Instruments). The results are shown in
(41) In addition, the shear modulus was measured as a function of frequency. This measurement provides information about the vibration behaviour of viscoelastic materials under oscillating shear stress and allows conclusions to be drawn about the interaction of the molecules in the system.
(42)
(43) This effect is also documented by the measurements of the complex shear modulus as a function of shear, which are shown in
(44)
(45)
Example 2: Embedding of Porous Bone Substitute Materials
(46) Osteoinductive granules of hydroxyapatite (HA) in the form of fir cones were used (Nanobone, Artoss GmbH, Rostock, Germany). These were on average 3 mm long and had a diameter between 0.5 and 1.0 mm. The HA showed a crystallographic morphology similar to that of biological HA. This HA was embedded in a highly porous matrix of silica xerogel. The porosity of the granules was about 50%, the specific surface area was about 200 m.sup.2/g, and the pore size distribution showed a maximum at 4 nm.
(47) The granules were impregnated in a mass ratio of 1:1 with a pure silica sol with a SiO.sub.2 concentration of 6% and a pH value of 7.0. In contact with the solid, the silica sol gels. Granules are produced which are filled with a silica gel and are coated with same.
(48) To produce the poloxamer-silica hydrogel, 35 g Kolliphor P 407 (BASF) were stirred in 65 g silica sol with a SiO.sub.2 content of 6%. The sol was previously cooled to 1° C. Cross-linking is achieved by gamma irradiation in the range of 17.5 to 30 kGrey. This polymer-silica hydrogel was mixed with the coated granules in a mass ratio of 1:1. The resulting pasty bone substitute material is very easy to shape and can be inserted into bone defects with an applicator.
(49) The stability of the coating of the granules with pure silica hydrogel was controlled by subjecting the material to accelerated ageing for 1 year according to ASTM F 1980-07. After removing the poloxamer-silica hydrogel by rinsing with water, granules coated with pure silica hydrogel could be seen under the microscope.
Example 3: Functionality in Animal Experiments
(50) The experiments were carried out with female rabbits (New Zealand White, 3-4 kg, Charles River, Sulzfeld, Germany). The bone substitute material produced according to Example 2 was implanted bilaterally into the hind legs. The cut through the cutis and subcutis has a length of approx. 2.5 cm. The musculature was also severed in a small area in order to keep the injuries as small as possible, and then the periosteum was carefully detached from the bone at the defect site to be placed. A cylindrical defect (5 mm in diameter and 10 mm in length) was then inserted into each of the lateral condyles of the femora. A standard drill (Ø 4.5 mm) was used for this purpose. During defect settlement, the area was rinsed with 0.9% NaCl solution to prevent necrosis of bone tissue due to heat exposure.
(51) Anaesthesia was administered subcutaneously to the neck fold by injection of 10% ketamine (30-60 mg/kg body weight) and 2% xylazine (5 mg/kg body weight). After 10 min, 0.3 ml atropine (0.5 mg/ml) was administered. In addition, novamine sulfone (500 mg/ml) was injected as an analgesic and penicillin G (intramuscular 150,000 i.U.) as an antibiotic. Local anaesthesia was performed with 2 ml xylocitin-loc (2%/ml). After implantation, the wound area was rinsed with gentamicin (80 mg/2 ml, 1:5 dilution with NaCl). The wound closure (point seam) was made with vicryl suture material.
(52) After trial periods of 4, 8 and 12 weeks, the corresponding trial groups were removed from the trial. The euthanasia was performed on the anaesthetised animal (10% ketamine and 2% xylazine, subcutaneously) using Release® (300 mg/ml corresponding to: 1 ml/kg body weight) intravenously. Histological sections were made for the evaluation. The defect regions were explanted, decalcified and embedded in paraffin. A hematoxylin and eosin stain was applied.
(53) Result: After 4 weeks neither the polymer-silica hydrogel nor the pure silica hydrogel was detectable. A complete resorption occurred. Changes in the temporal sequence to granules embedded in the patient's blood were not detectable during defect healing.