Separation of Radiometals
20220118379 · 2022-04-21
Inventors
- Fedor ZHURAVLEV (Roskilde, DK)
- Kristina Søborg PEDERSEN (Roskilde, DK)
- Jesper FONSLET (Birkerød, DK)
- Joseph Michael IMBROGNO (North Stonington, CT, US)
- Andrea Adamo (Cambridge, MA, US)
- Klavs F. Jensen (Lexington, MA)
Cpc classification
C01G9/003
CHEMISTRY; METALLURGY
C22B34/1259
CHEMISTRY; METALLURGY
G21G1/00
PHYSICS
B01D2311/04
PERFORMING OPERATIONS; TRANSPORTING
C01G53/003
CHEMISTRY; METALLURGY
A61K51/0478
HUMAN NECESSITIES
International classification
B01D61/14
PERFORMING OPERATIONS; TRANSPORTING
C22B3/26
CHEMISTRY; METALLURGY
C22B34/12
CHEMISTRY; METALLURGY
Abstract
Method of separation of a radiometal ion from a target metal ion, comprising a first liquid-liquid extraction step in which an organic phase comprising an extractant and an interfacial tension modifier is mixed with an aqueous phase comprising the radiometal ion and the target metal ion in order that the radiometal ion is at least partially transferred to the organic phase, followed by a first phase separation step, wherein the phase separation is carried out in flow comprising the use of a microfiltration membrane to separate the phases based on the interfacial tension between the phases such that a permeate phase passes through the membrane and a retentate phase does not.
Claims
1. A method of separation of a radiometal ion from a target metal ion, comprising a first liquid-liquid extraction step in which an organic phase comprising an extractant and an interfacial tension modifier is mixed with an aqueous phase comprising the radiometal ion and the target metal ion in order that the radiometal ion is at least partially transferred to the organic phase, followed by a first phase separation step, wherein the phase separation is carried out in flow comprising the use of a microfiltration membrane to separate the phases based on the interfacial tension between the phases such that a permeate phase passes through the membrane and a retentate phase does not, wherein: a. the radiometal ion is a .sup.68Ga ion, the target metal ion is a .sup.68Zn ion, the extractant is selected from one or more dialkyl ethers R.sup.1OR.sup.2, wherein the two alkyl groups R.sup.1 and R.sup.2 can be the same or different, or can together form a cyclic ether, and can optionally be substituted, and the interfacial tension modifier is selected from one or more aromatic hydrocarbons, which may optionally be halogenated, and/or one or more C2-C9 alkanes, which may optionally be halogenated; or b. the radiometal ion is a .sup.89Zr ion, the target metal ion is a .sup.natY ion, the extractant is a solvent able to function as a bidentate ligand for .sup.89Zr via two oxygen atoms, and the interfacial tension modifier is a solvent having similar properties to the extractant, but that are not able to function as a bidentate ligand for the .sup.89Zr ion, such that it does not interfere with the ability of the extractant to interact with the .sup.89Zr ions; or c. the radiometal ion is a .sup.45Ti ion, the target metal ion is a .sup.natSc ion, the extractant is a solvent able to function as a bidentate ligand for .sup.45Ti via two oxygen atoms, and the interfacial tension modifier is a solvent having similar properties to the extractant, but that are not able to function as a bidentate ligand for the .sup.45Ti ion, such that it does not interfere with the ability of the extractant to interact with the .sup.45Ti ions; or d. the radiometal ion is a .sup.64Cu ion, the target metal ion is a .sup.64Ni ion, the extractant is selected from: one or more trialkyl phosphine oxides; one or more alkylphosphoric acid monoalkyl esters; one or more diketones having the structure R.sup.3—C(═O)CH.sub.2C(═O)—R.sup.4, in which R.sup.3 and R.sup.4 are each independently an alkyl or an aryl group; and one or more aldoximes or ketoximes in which the substituent(s) of the oxime group are aromatic groups; and the interfacial tension modifier is a solvent comprising one or more straight or branched chain cyclic or acyclic aliphatic alkanes having from five to sixteen carbon atoms, which may optionally be substituted, and/or a solvent comprising one or more aromatic hydrocarbons, which may optionally be substituted.
2. (canceled)
3. (canceled)
4. The method according to claim 1, wherein the first liquid-liquid extraction step is conducted in flow, and wherein the first liquid-liquid extraction step comprises mixing the aqueous phase and the organic phase such that stable liquid-liquid segmented flow of the mixture is established.
5. (canceled)
6. The method according to claim 1, wherein the aqueous phase comprises a concentration of aqueous hydrochloric acid or nitric acid of greater than or equal to 3M.
7. (canceled)
8. (canceled)
9. The method according to claim 1, wherein the radiometal ion and the target metal ion are defined as follows: a. the radiometal ion is a .sup.68Ga(III) ion and the target metal ion is a .sup.68Zn(II) ion; or b. the radiometal ion is a .sup.89Zr(IV) ion and the target metal ion is a .sup.natY(III) ion; or c. the radiometal ion is a .sup.45Ti(IV) ion and the target metal ion is a .sup.natSc(III) ion; or d. the radiometal ion is a .sup.64Cu(II) ion and the target metal ion is a .sup.64Ni(TI) ion.
10. The method according to claim 1, wherein the radiometal ion is a Ti ion and the target metal ion is a Sc ion, and: the aqueous phase is a solution in 12M HCl; the extractant is selected from the group consisting of maltol, vanillin, eugenol, and guaiacol (o-methoxyphenol); and the interfacial tension modifiers are selected from the group consisting of fluorobenzene, trifluorotoluene, thiophene and anisole.
11. The method according to claim 10, wherein the extractant is guaiacol and the interfacial tension modifier is anisole.
12. (canceled)
13. (canceled)
14. The method according to claim 1, wherein the radiometal ion is a Ga ion and the target metal ion is a Zn ion, the extractant is selected from the group consisting of diethylether, butylmethyl ether, diisopropyl ether, tetrahydropyran, methyl hexyl ether, dibutyl ether and diamyl ether, and the interfacial tension modifier is selected from the group consisting of: a fluorinated aromatic hydrocarbon; an aromatic hydrocarbon; an alkoxybenzene; a halogenated alkane; and an alkane.
15. The method according to claim 14, wherein the aqueous phase is a solution in 6M HCl, and the extractant is selected from diethyl ether, diisopropyl ether, dibutyl ether, butyl methyl ether and hexyl methyl ether.
16-18. (canceled)
19. The method according to claim 14, wherein the extractant is selected from butyl methyl ether, diisopropyl ether, dibutyl ether and diethyl ether, and the interfacial tension modifier is selected from the group consisting of toluene, anisole, 1,2-dichloroethane, trifluorotoluene and heptane.
20-22. (canceled)
23. The method according to claim 14, further comprising a back extraction procedure comprising, following the first phase separation step, a first back-extraction step in which an organic phase comprising the radiometal ion is mixed with an aqueous solution of a protic acid in order that the radiometal ion is at least partially transferred to the aqueous solution, followed by a back-extraction phase separation step, in which the phase separation is carried out in flow comprising the use of a microfiltration membrane to separate the phases based on the interfacial tension between the phases such that a permeate phase passes through the membrane and a retentate phase does not, in order to obtain an aqueous solution comprising the radiometal ion.
24. The method according to claim 23, wherein the aqueous solution of a protic acid is an aqueous solution of less than 6 M HCl.
25-31. (canceled)
32. The method according to claim 1, wherein the radiometal ion is a Zr ion and the target metal ion is a Y ion, the extractant is selected from the group consisting of maltol, vanillin, eugenol, and guaiacol (0-methoxyphenol), and the interfacial tension modifier is selected from the group consisting of fluorobenzene, trifluorotoluene, thiophene and anisole.
33. The method according to claim 32, wherein the extractant is guaiacol (o-methoxyphenol) and the interfacial tension modifier is anisole.
34-37. (canceled)
38. The method according to claim 1, wherein the radiometal ion is a Zr ion and the target metal ion is a Y ion, the extractant is 0.1 M trioctylphosphine oxide (TOPO), the interfacial tension modifier is hexane, and the aqueous phase is a solution in 6 M HCl.
39. The method according to claim 1, wherein the radiometal ion is a Cu ion and the target metal ion is a Ni ion, the extractant is selected from: one or more trialkylphosphine oxides in which the alkyl groups are selected from: straight chain or branched hydrocarbons having from six to ten carbon atoms; one or more alkylphosphoric acid monoalkyl esters having the structure R.sup.5—P(═O)(OH)—OR.sup.6, where R.sup.5 and R.sup.6 are each independently a branched or unbranched C.sub.6 to C.sub.10 alkyl group; one or more diketones having the structure R.sup.3—C(═O)CH.sub.2C(═O)—R.sup.4, in which R.sup.3 and R.sup.4 are each independently an optionally halogenated branched or unbranched C.sub.1 to C.sub.10 alkyl group or a substituted or unsubstituted phenyl group; one or more aldoximes or ketoximes having an aromatic substituent wherein the benzene ring is substituted with both an oxygen and an alkyl group.
40. The method according to claim 39, in which the extractant is selected from the group consisting of: Cyanex 923 (TRPO), trioctylphosphine oxide, 2-ethylhexylphosphoric acid mono-2-ethylhexyl ester (PC-88A), 1-phenyldecane-1,3-dione, heptadecane-8,10-dione, 1,3-diphenylpropane-1,3-dione, 5-nonylsalicylaldoxime, 5-dodecylsalicylaldoxime, Acorga® P50, or 2-hydroxy-5-nonylacetophenone oxime.
41. (canceled)
42. The method according to claim 39, in which the interfacial tension modifier is selected from n-pentane, n-hexane, n-heptane, n-octane, n-nonane, n-decane, n-undecane, i-hexane, neo-hexane, i-heptane, neo-heptane, cyclohexane, cycloheptane, cyclooctane, kerosene, light petroleum, benzene, naphthalene, toluene, ethylbenzene, dimethylbenzene, iso-octane and mixtures thereof.
43-52. (canceled)
53. The method according to claim 1, wherein the microfiltration membrane is a PTFE membrane having a pore size of 0.2 Om, and the PFA diaphragm has a thickness of 0.002″ (0.0508 mm).
54. (canceled)
55. A method of generation of radiometal ions from a target metal, comprising: a. providing a solid target metal or an aqueous solution of ions of the target metal; b. irradiation of the solid target metal or the target metal ion solution with a particle beam to produce a solid mixture of radiometal and target metal, or a mixture of radiometal ions and target metal ions in aqueous solution c. separation of the radiometal ions from the target metal ions according to the method of claim 1.
56-59. (canceled)
60. The method of generation of radiometal ions from a target metal according to claim 55, wherein step a. comprises providing a solid target metal, and, after step b. and before step c., the method further comprises a step of dissolution of the solid mixture of radiometal and target metal to produce an aqueous solution comprising radiometal ions and target metal ions.
61-63. (canceled)
64. A method to make a radiolabelled pharmaceutical, comprising the steps of: a. using the method of claim 55 to provide separated radiometal ions; and b. reacting the separated metal ions with a reactive precursor of the radiolabelled pharmaceutical in a manner to obtain the radiolabelled pharmaceutical.
65. The method of claim 64, wherein the radiometal ions comprise .sup.68Ga, and wherein step b. comprises using a back extraction procedure to obtain an aqueous solution comprising the separated radiometal ions, wherein the aqueous solution comprising the separated radiometal ions is reacted with the precursor of the radiolabelled pharmaceutical in a manner in step b. to obtain the radiolabelled pharmaceutical.
66. (canceled)
67. Apparatus for conducting separation of a radiometal ion from a target metal ion by means of a liquid-liquid extraction and phase separation carried out in continuous flow, comprising: a first inlet for an aqueous phase comprising the radiometal ion and the target metal ion; a second inlet for an organic phase comprising an extractant and an interfacial tension modifier; one or more mixers for mixing the organic phase and the aqueous phase; tubing to convey the mixture of the organic phase and the aqueous phase; a phase separation apparatus comprising a microfiltration membrane to separate the organic phase from the aqueous phase based on the interfacial tension between the phases such that a permeate phase passes through the membrane and a retentate phase does not; a first outlet for the aqueous phase exiting the phase separation apparatus; a second outlet for the organic phase exiting the phase separation apparatus.
68-94. (canceled)
95. The method of claim 64, wherein the radiometal ions comprise .sup.45Ti and/or .sup.89Zr.
Description
BRIEF DESCRIPTION OF THE DRAWINGS
[0108]
[0109]
[0110]
[0111]
[0112]
[0113]
[0114]
[0115]
[0116]
[0117]
[0118]
[0119]
[0120]
[0121]
[0122]
DETAILED DESCRIPTION
[0123] Apparatus for Separation of Radiometal Ions
[0124] As noted above, liquid-liquid extraction (LLE) is a widely used means of separation of components of a solution by partitioning the components between two different solvents. Traditionally, this has been conducted between immiscible solvents which separate under the influence of gravity in such apparatus as a separatory funnel, or which are forced to separate by use of apparatus such as a centrifuge. The extraction is based on the relative solubilities of the components of the solution in the chosen immiscible liquids, usually an aqueous phase and an organic phase. In some cases, the system of organic and aqueous phases plus one or more components may form an emulsion or “third phase” which can prevent or make less effective the partitioning of the components between easily separable organic and aqueous phases.
[0125] “Extraction” is used to describe the transfer of a component from the aqueous phase to the organic phase, whereas “stripping” or “back-extraction” describes the transfer of a component of interest from the organic phase to the aqueous phase. Removal of an unwanted component from the organic phase is described as “scrubbing”.
[0126] In order that an extraction or back-extraction takes place efficiently, it is necessary for the organic and aqueous phases containing the components to be extracted to be thoroughly mixed, in order to permit partitioning of the components between the phases according to their relative solubilities in the phases, followed by a means of separating the two phases from one another. Traditionally, in a separatory funnel, this would be carried out by shaking the separatory funnel containing the phases and components, followed by allowing the phases to separate under gravity, and running off the phases in turn from the bottom of the funnel. More than one extraction step can be carried out to ensure the maximum extraction of the desired components.
[0127] It has been recognized by the present inventors that it would be desirable to avoid the handling of solutions containing radioactive materials by an operator, as would be required by these traditional manual methods of liquid-liquid extraction. Nonetheless, such separations are attractive as they are simple and inexpensive to carry out, compared with other methods such as chromatographic methods that may require expensive media, and may result in the need to dispose of radioactively-contaminated media following use. In addition, liquid-liquid extraction can be conducted on a wide range of scales.
[0128] Recently, new methods of, and apparatus for, conducting mixing of fluids and separation of fluids have been developed, which methods may be conducted in a continuous manner. The present inventors have recognized that such apparatus and methods could potentially have applicability in the separation of metal ions, in particular the separation of radiometal ions from target metal ions. As far as the present inventors are aware, no such use has been made of these new methods and apparatus in the field of the present invention.
[0129] Continuous flow synthesis apparatus has been developed recently that allows reactions to be carried out in a continuous manner. A review.sup.52 of transformations that have been carried out in continuous flow systems lists transformations such as hydrogenations and reductions, oxidation, acid or base catalyzed bond-forming reactions, transition metal catalyzed bond-forming reactions, esterification reactions, protection and deprotection reactions, photocatalysis and enzymatic reactions. Particular attention has been paid to the use of reactive and/or toxic, particularly gaseous, reagents, as they can be generated and used in a closed system: for example, cyanogen bromide.sup.53, chlorine azide.sup.54, ethylene gas.sup.55, (meth)acryloyl chloride.sup.56, chlorine gas.sup.57 and diazomethane.sup.58. The use of such systems in the automated production of drugs has been suggested.sup.59. Some continuous purifications involving metals have been reported, such as removal of excess ligand from nanoparticles.sup.60 and extraction of leached copper from a target compound.sup.61. However, so far as the present inventors are aware, such technologies have not been applied to the separation of metal ions from one another.
[0130] WO2004/087283 describes systems which may be used for liquid-liquid separations, amongst other uses, and in which the separation is carried out by means of differential wetting of arrays of capillary tubes. For example, a hydrophilic and a hydrophobic liquid may be mixed, and the mixture brought in contact with one or more capillary tubes having a hydrophobic coating. The hydrophobic liquid thus wets the capillary tube and rises up it, whereas the hydrophilic liquid does not wet the capillary tube and does not enter it. In this way, the hydrophobic liquid passes through the array of capillary tubes and is separated from the hydrophilic liquid.
[0131] WO2014/026098 describes a membrane separation apparatus suitable for the separation of a first fluid (permeate) from a second fluid (retentate) based on the interfacial tension of the two fluids. In particular, it relates to a system in which a pressure controller is included in the apparatus to apply pressure across the microfiltration membrane that is independent of the pressure downstream of the device, and which can control the selectivity of the membrane for the passage of the fluids, such that one fluid can be allowed to pass selectively through the membrane thus separating it from the other fluid. Such a membrane separation unit is available from Zaiput Flow Technologies. A schematic of such a separator 10 is depicted on the right hand side of
[0132] Without wishing to be bound by theory.sup.48, it is believed that, in order that the separation is complete, the capillary pressure P.sub.cap associated with the fluid passageways in the membrane and the mixture of fluids to which the membrane is exposed, must not be exceeded, or both fluids will be forced through the membrane. Thus, ΔP.sub.mem, the pressure difference across the membrane, may not exceed P.sub.cap. P.sub.cap is quantified as:
[0133] where θ is the contact angle formed between the solid material of the membrane, the first fluid to be separated and the second fluid to be separated, r is the radius of the membrane pores, and y is the interfacial tension with respect to the first fluid to be separated and the second fluid to be separated.
[0134] Further, in order to ensure that the whole of the first fluid passes through the membrane, ΔP.sub.mem must exceed the pressure P.sub.per needed to cause the permeate liquid to pass through the membrane. P.sub.per is quantified as:
[0135] where μ is the viscosity of the permeate phase, Q is the entering permeate fluid volumetric flow rate, L is the membrane thickness, n is the number of pores, and R is the pore radius; this assumes that the membrane acts as an array of cylindrical pores.
[0136] In addition, the separator must be operated at a flow rate which is suited to the available membrane area; if the flow rate is excessive, both phases may exit both outlets.
[0137] Where the pressure drop along the length of the membrane is negligible compared to P.sub.cap−P.sub.per, then ΔP.sub.mem can be assumed to be constant along the length of the membrane, and the conditions for successful separation are
P.sub.cap>ΔP.sub.mem>P.sub.per
[0138] The first inequality is satisfied by selection of the microfiltration membrane material and pore size in a range appropriate for the separation, and the second by ensuring that the pressure on the retentate side of the membrane is greater than that on the permeate side of the membrane; this additional pressure is provided by the pressure controller. In practice, the actual operating range of pressures is often narrower than the theoretical range given above.
[0139]
[0140] While the apparatus for mixing of the two phases is described here as a mixing tee followed by static mixers, it will be appreciated that other combinations of mixing apparatus (either passive or active) can be used depending on the degree of mixing required, the nature of the fluids to be mixed, and the volume of fluid to be mixed. For example: fewer or more static mixers may be employed; the mixing tee may be a Y-junction mixer or a T-junction mixer; a packed bed reactor housing sand, stainless steel or glass beads may replace one or more of the mixers depicted in
[0141] The variable length mixing loop, along with the other tubing used in the apparatus, is made from a material chosen with reference to its chemical compatibility with the solvent system, and the pressures that it will need to withstand in operation; the present inventors have found that PFA tubing is suitable for use with the solvent systems used herein. The length of the mixing loop, and the other mixers used, are selected in order to ensure an adequate degree of mixing of the phases, and a residence time in the apparatus sufficient to ensure efficient partitioning of the metal ions between the phases, for the chosen solvent system and the metal ions to be separated.
[0142] For the separation of radiometals, the total production time is a critical parameter in the system since the radiometal is continuously undergoing decay back to the target metal. Therefore, the shortest possible residence time is desirable, and the mixing of the phases must be optimized to ensure efficient extraction in as short a time as possible.
[0143] It has been found by the present inventors that achieving liquid-liquid segmented flow (sometimes referred to as “slug flow”, though this term more usually refers to gas-liquid mixtures) in the mixed fluid stream passing through the tubing 80, mixing loop 130 and on to the membrane separator 10 is of importance in the present invention. Liquid-liquid segmented flow describes a flow pattern through a tube or pipe in which a first fluid is dispersed in a second fluid in the form of segments of varying length. During stable liquid-liquid segmented flow, the first fluid is shed from the back of the segment at the same rate as it is picked up at the front of the segment, and so the segment length remains constant as it travels along the tube. The present inventors have found that the high mass transfer in liquid-liquid segmented flow systems is particularly beneficial in allowing the efficient partition of components between two phases for the purposes of liquid-liquid extraction. Accordingly, the mixers provided in the apparatus are selected such that liquid-liquid segmented flow is provided in the mixing loop 130 for the combination and volume of fluids used. The present inventors have found that, in the solvent systems used herein, liquid-liquid segmented flow is achieved by mixing of the phases through mixing tee 90. When static mixers are used also, the performance of the extraction was further improved.
[0144] Liquid-liquid segmented flow may be determined by visual inspection of the mixture as it flows through the tubing, or may be detected for example by a phototransistor device which clips on to the outside of the tubing and detects a phase interface by alteration in current flow depending on the amount of light received. These devices can detect phase interfaces even in mixtures of colourless liquids. Such devices are available from Optek Technology (OPB350 and OCB350 series). These devices can also be used at the outlets of the separator to detect whether retention or breakthrough of a phase has occurred.
[0145] As discussed above, the membrane separator 10 comprises two main components: a polymer microfiltration membrane 30 and a thin diaphragm 60 (
[0146] The diaphragm 60 acts to modulate the pressure between the aqueous and organic sides of the membrane 30. The aqueous phase is retained by the membrane 30, while the organic phase permeates through the membrane 30. The physical properties and geometry of the membrane 30 as well as the chemical nature of the aqueous and organic phases and their interactions with the membrane surface determine the capillary and permeation pressures. The interactions between the aqueous and the organic phases determine the interfacial tension. The interplay between these parameters determines whether the conditions are within the operating range of the system. If they are not, incomplete phase separation will occur.
[0147] Pressure control may be provided by controlling the pressure at each of the outlets of the separation apparatus; however, to do so makes it difficult to integrate the apparatus with other downstream components. Accordingly, it is preferable to use a pressure controller, as described in WO2014/026098 and shown in
[0148] The diaphragm may be made from a polymer selected from the group consisting of perfluoroalkoxyalkane (PFA), latex, polytetrafluoroethylene (PTFE), fluorinated ethylene propylene (FEP), fluoroelastomers (FMK), perfluoroelastomers (FFKM), tetrafluoro ethylene/polypropylene rubbers (FEPM), neoprene, nitrile rubber, and polyethylene. The diaphragm material should be selected primarily with regard to its resistance to the solvent system to be used in the separation: for example, its acid resistance and/or resistance to organic solvents. The elasticity of the diaphragm following long periods of deformation or physical degradation will also be affected by the choice of diaphragm material. As perfluoroalkoxyalkane (PFA) is a very robust material and is mechanically strong, it is the preferred choice of diaphragm material.
[0149] The choice of diaphragm thickness is important as this directly affects the pressure exerted on across the membrane; this must be selected in combination with the membrane properties and solvent system to arrive at a functional apparatus for a given separation. Preferably, the diaphragm thickness is 0.002″ (0.0508 mm).
[0150] The microfiltration membrane may be made from a polymer selected from the group consisting of polytetrafluoroethylene (PTFE), polyvinylidene fluoride (PVDF), cellulose acetate, polysulfane, polysulfone, polyethersulfone, polypropylene, polyethylene, and polyvinyl chloride (PVC). As well as chemical compatibility with the solvent system to be used in the separation, the membrane material should be selected having regard to the wettability of the material by the organic phase to be used (for hydrophobic materials such as listed above). It has been found by the present inventors that polytetrafluoroethylene (PTFE) fulfils these requirements for the solvent systems used herein.
[0151] The choice of membrane pore size affects the upper and lower boundaries of the pressure of the system, as explained in detail above. Each individual membrane will have a range of pore sizes, and so the pore size specified herein is the manufacturer's specification of pore size, which will represent an average value. The pore size must be selected in combination with the membrane material, the diaphragm thickness and the solvent system to arrive at a functional apparatus for a given separation. Suitably, the microfiltration membrane has a pore size selected in the range 0.1 to 1.0 μm, such as 0.1 to 0.5 μm, or 0.2 to 0.5 μm. Preferably, the microfiltration membrane has a pore size of 0.2 μm.
[0152] It will be appreciated from the foregoing discussion that it may be possible for a number of different combinations of membrane material, pore size and diaphragm thickness to provide suitable separation conditions for a given solvent system. Similarly, a number of different solvent systems may be separable in an apparatus having a given combination of membrane material, pore size and diaphragm thickness.
[0153] Turning now to
[0154] Turning now to
[0155] In the scrubbing stage 272 shown in
[0156] Turning now to
[0157] Separation Methods
[0158] The selection of an appropriate extractant to conduct a liquid-liquid extraction in which the phase separation is conducted in flow comprising the use of a microfiltration membrane to separate the phases based on interfacial tension is crucial: the system must provide selective extraction of the metal ion of interest, as little extraction as possible of the target metal ion, must be stable in the presence of the strongly acidic solutions often used in the generation of radiometals (to dissolve irradiated metal foils or irradiated electroplated layers, and/or to avoid the hydrolysis of susceptible metals such as Ti or Zr), and must have a sufficiently high interfacial tension with the aqueous phase that complete separation can be achieved using the microfiltration membrane. This is a much more demanding set of criteria than need be applied to standard batch liquid-liquid separations carried out on the basis of density.
[0159] While the present inventors have made reference to literature reports of extraction of particular individual metal ions in batch processes, and in some cases to separation of metal ions from one another in batch processes, it was not expected that these literature conditions would be directly applicable in the processes of the present invention. Indeed, a number of the literature conditions simply did not work at all under the necessary conditions. For example, it had been reported that liquid-liquid batch separation of .sup.68Ga ions from .sup.68Zn ions in an aqueous solution of protic acid could be carried out using isopropyl ether.sup.28. However, under the conditions required for .sup.68Ga production (in particular, the presence of around 7 m ZnCl.sub.2 concentration), the phase equilibrium simply did not allow efficient separation; the present inventors found that 65% of the ZnCl.sub.2 present migrated into the organic phase, thus heavily contaminating the .sup.68Ga solution with Zn. Other conditions were found to be too inefficient for application in the process of the present invention, for example the use of 1-octanol in the separation of .sup.45Ti from Sc, which allowed only 50% extraction efficiency. Yet other extractants attempted by the present inventors did not provide clean phase separation when the microfiltration membrane was applied, but instead led to breakthrough, retention or the formation of emulsions.
[0160] It was surprisingly discovered by the present inventors that it is possible to adjust the properties of an extractant with respect to the aqueous phase such that an extractant that did not provide clean phase separation when used alone could do so with the addition of a carefully selected second solvent, here referred to as an interfacial tension modifier. The interfacial tension modifier must not interfere with the interactions between the radiometal ion and the extractant, must not extract the target metal ion to any significant degree, must not dissolve water to any significant extent, and must be able to adjust the properties of the interfacial tension of the overall solvent system (extractant, aqueous phase and interfacial tension modifier (if present)) with respect to the microfiltration membrane such that complete separation of the phases by the microfiltration membrane was possible. The amount of the interfacial tension modifier must be selected carefully to provide optimum separation, as must the relative flow rates of the organic phase (extractant and interfacial tension modifier) and the aqueous phase. The interfacial migration (ie the tendency of one phase to contaminate the other) is a critical parameter which must be minimized to prevent the contamination of aqueous phase with the organic phase, which would make the process incompatible with recycling the target metal solution for a further irradiation step, due to stringent organic-free requirements for aqueous cyclotron solution targets, particularly ZnCl.sub.2.
[0161] Thus, for the separation of .sup.45Ti from Sc, and for the separation of .sup.89Zr from Y, the extractant chosen is a solvent having the ability to function as a bidentate ligand for the radiometal via two oxygen atoms, preferably thus forming a five membered ring, as well as having a suitable interfacial tension with 12 M (37%) HCl. Suitable extractants may be maltol, vanillin, eugenol, and guaiacol (o-methoxyphenol), with guaiacol being particularly preferred. The interfacial tension modifier is a solvent having similar properties to the extractant, though not having the ability to function as a bidentate ligand for the radiometal ion, such that it does not interfere with the ability of the extractant to interact with the radiometal ions, as well as the ability to modify the interfacial tension of the overall system to allow complete separation. Suitable interfacial tension modifiers may be fluorobenzene, trifluorotoluene, thiophene and anisole, with anisole being particularly preferred. For the preferred system of guaiacol and anisole, an amount of anisole of at least 10% v/v is found to perform particularly well, and the optimum flow ratio for the organic phase to the aqueous phase to be greater than 3 to 1, and in some cases 5 to 1. Alternatively, where the radiometal ion is a Zr ion and the target metal ion is a Y ion, it is preferable that the extractant is 0.1 M trioctylphosphine oxide (TOPO), the interfacial tension modifier is hexane, and the aqueous phase is a solution in 6 M HCl.
[0162] For the separation of .sup.68Ga from Zn, the use of ether extractants was found to work well on combination with an interfacial tension modifier selected from the group consisting of: a fluorinated aromatic hydrocarbon; an aromatic hydrocarbon; an alkoxybenzene; a halogenated alkane, for example selected from the group consisting of 1,2-dichloroethane, 1,1,2-trichloroethane, 1,1,1-trichloroethane, hexachloroethane and bromoethane; and an alkane; particularly, selected from the group consisting of toluene, anisole, 1,2-dichloroethane, trifluorotoluene and heptane, with trifluorotoluene being the most preferred. Preferably, the ratio of the extractant to the interfacial tension modifier is 1:2 by volume, and the optimum flow ratio for the organic phase to the aqueous phase is greater than 3 to 1.
[0163] For the separation of .sup.64Cu from .sup.64Ni, the use of a phosphine oxide extractant was found to work well with an interfacial modifier selected from an aromatic hydrocarbon and an aliphatic alkane which may be cyclic or acyclic; for example selected from the group consisting of n-pentane, n-hexane, n-heptane, n-octane, n-nonane, n-decane, n-undecane, i-hexane, neo-hexane, i-heptane, neo-heptane, cyclohexane, cycloheptane, cyclooctane, kerosene, light petroleum, benzene, naphthalene, toluene, ethylbenzene, dimethylbenzene and iso-octane and mixtures thereof; particularly selected from the group consisting of toluene, heptane and hexane. Preferably, the concentration of the extractant in the interfacial tension modifier is at least 0.1 M, such as from 0.1 M to 0.4 M, and the optimum flow ratio for the organic phase to the aqueous phase is from 5:1 to 3:1.
EXAMPLES
[0164] Separation of Ti and Sc
[0165] Materials
[0166] Guaiacol (99%), anisole (99%), 1-octanol (99%), titanium (IV) chloride (neat), titanium (IV) chloride solution (0.09 M in 20% HCl), hydrochloric acid (37%), sulfuric acid (95.0-98.0%) and pyridine-2,6-dicarboxylic acid (dipic) (98%) were purchased from Sigma Aldrich and used without further purification. TLC plates (Silica gel on TLC Al foil) were purchased from Sigma Aldrich. Scandium (III) chloride (anhydrous, 99.9%) and scandium foil (250 μm, 99.9% pure, rare earth analysis) were purchased from Alfa Aesar. Custom Ti and Sc ICP standards were purchased from Inorganic Ventures (100 ppm of each metal in a 5% HCl solution). Salan.sup.62 and (salan)Ti(dipic).sup.3 were synthesized according to the literature procedures.
[0167] The membrane separator module was similar to those manufactured by Zaiput Flow Technologies. The aqueous and the organic phases were combined through a tee and mixed with two static mixers and mixing tubing. The aqueous phase was retained by the membrane, while the organic phase permeated through the membrane. Under the conditions of direct extraction, the radionuclide .sup.45Ti was selectively extracted from scandium into the organic phase. Pall PTFE membranes were used for all experiments (47 mm diameter, 0.1/0.2/0.5 μm pore size, polypropylene support for the 0.1/0.2 μm pore sizes). Perfluoroalkoxy alkane (PFA) diaphragms (0.001″/0.002″/0.005″ (0.0254/0.0508/0.1270 mm)) were purchased from McMaster Carr. All PFA tubing ( 1/16″ (1.5875 mm) OD, 0.03″ (0.762 mm) ID) was purchased from Idex Health and Science. PTFE static mixers were purchased from Stamixco. The 15 mL plastic centrifuge tubes with screw caps were purchased from VWR.
[0168] Radionuclide Production and Separation
[0169] For all experiments, the cyclotron target material, scandium, was used at its natural abundance level.
[0170] .sup.45Ti was produced by 10-20 μA proton irradiation of 30-60 mg scandium foil, for 5-15 min using a GE PETtrace cyclotron. To minimize coproduction of the .sup.44Ti (half-life=60.0 years), a 500 μm thick aluminium foil was used to degrade the incidental 16 MeV beam to approximately 13 MeV. The irradiated foil was digested in 30-37% M HCl. The mixture was filtered and centrifuged if necessary. The solution was diluted with concentrated hydrochloric acid to make the final dilution ca. 12 M in HCl. These dilutions were used as the aqueous phase for the LLEF.
[0171] Instrumentation and Methods
[0172] The solutions for the continuous membrane-based separation were pumped using either the KDS 100 Legacy Syringe (radioactive experiments) or the Harvard Apparatus PHD 2000 Programmable and Infusion syringe pumps (non-radioactive experiments). The NMR spectra were taken on Agilent 400 MR operating at 400.445 MHz (.sup.1H). Radio-TLC was performed with a Raytest MiniGita TLC scanner using chloroform/ethyl acetate (1/1, (v/v)) as a mobile phase. The HPLC and radio-HPLC analyses were performed on a Hitachi Chromaster equipped with a Carrol&Ramsey 105-S radio-detector and a Hitachi 5430 double diode array detector. Column: Phenomenex Luna 3μ C18(2) (100 Å, 100 mm×2.00 mm). Flow: 0.5 mL/min. Eluents: (A) 0.1% (v/v) CF.sub.3COOH in Milli-Q water, (B) 0.1% (v/v) CF.sub.3COOH in CH.sub.3CN. The radiochemical identity of [.sup.45Ti](salan)Ti(dipic) was established by comparing its retention time with that of its natural abundance isotopomer. The radiochemical conversion (RCC) was determined by radio-HPLC or radio-TLC and calculated as:
RCP=(Area.sub.product/Total Area)*100%.
[0173] The extent of extraction (extraction %) was determined from the radioactivity measurements and using inductively coupled plasma atomic emission spectroscopy (ICP-AES, Agilent 5100 Dual View) of the aqueous phase. Samples of the aqueous phase were collected before the LLE and after 5, 15, 30, and 45 minutes of LLE. 0.35 mL of each sample was digested in 5 mL with 10% (v/v) H.sub.2SO.sub.4 for 6 hours at 160° C. 2.7 mL of the digested sample was diluted up to 10 mL with Milli-Q water to reach a total acid concentration of 5% (v/v). Calibration standards (Inorganic Ventures) were prepared to match the sample matrix (Sc) with concentrations of 22.2, 18, 15, 10, and 5 ppm Ti and Sc and run prior to every set of samples. Samples were analyzed in radial view at a viewing height of 8 mm. The extraction to the organic phase was calculated from the concentration of Ti and Sc in the aqueous phase before and after the LLE by E=((c.sub.before LLE−c.sub.after LLE)/c.sub.before LLE)−100%.
[0174] Batch LLE and Separation
[0175] For the non-radioactive work, 0.5 mL 0.01 M TiCl.sub.4 and 0.01 M ScCl.sub.3 in 37% HCl was mixed with extractant(s) and shaken for 2 min in a centrifuge tube. The phases were allowed to separate by gravity. The concentration of Ti and Sc in the aqueous phase before and after the LLE was measured by ICP-AES.
[0176] For the batch LLE of .sup.45Ti, a centrifuge tube was charged with 2 mL of the solution of .sup.45Ti (10-50 MBq) in 37% HCl and 2 mL of the organic phase. The mixture was shaken vigorously, spun for 15 minutes, and centrifuged at 4000 rpm to separate the phases.
[0177] An Eppendorf 5702 centrifuge was used to assist in phase separation. For batch experiments, the phase mixing was performed using a IKA ROCKER 3D digital shaker.
[0178] Continuous Membrane-Based LLE and Separation
[0179] The continuous liquid-liquid extraction and phase separation in flow was performed using a membrane-based separator with a PFA diaphragm for integrated pressure control. A flow schematic of the apparatus is depicted in
Example 1—Investigation of Extractants for Batch Liquid-Liquid Extraction of Ti Ions from a Solution Also Containing Sc Ions
[0180] The preliminary screening experiments were performed in batch using gravity separation. It had been reported.sup.64 that .sup.45Ti can be extracted from aqueous HCl into 1-octanol, presumably as [.sup.45Ti]Ti n-octyloxide 1, Structure 1.
##STR00001##
TABLE-US-00001 TABLE 1 Liquid-liquid batch extraction of .sup.45Ti from cyclotron-irradiated Sc foil digested in 37% HCl, except entry 1a. Entry Extraction system (organic phase) D EE (%) 1 1-octanol, neat.sup.a 0.04 4 2 1-octanol, neat 0.88 47 3 1,2-Decanediol, 0.1M in 1-octanol 1.2 54 4 2,3-naphthalene diol, 0.1M in 1-octanol 1.1 52 5 C.sub.10F.sub.21CH.sub.2CH(OH)CH.sub.2OH.sup.b <0.001 <0.1 6 Guaiacol, neat 3 75 D is the distribution coefficient, D = [.sup.45Ti(org)]/[.sup.45Ti(aq)] and EE is the extraction efficiency, EE = 100% * [.sup.45Ti(org)]/[.sup.45Ti(total)] as measured by a radiation detector. .sup.a20% HCl; .sup.btrifluorotoluene/hexafluoropropanol (1/1, (v/v))
We observed little extraction when a solution of .sup.45Ti in 20% HCl was used (Table 1, entry 1). Using 37% HCl (12 M) significantly improved the extraction. 1,2-Decanediol used as co-extractant as a 0.1 M solution in 1-octanol gave only a slight improvement over neat 1-octanol. 2,3-naphthalene diol, reported, in the context of the extractive spectrophotometric determination of Ti content in rocks, to extract Ti at pH=4,.sup.65 showed similarly modest performance. (Table 1, entries 3 and 4). During these batch extractions, we noticed a significant increase in the volume of the 1-octanol phase suggestive that conc. HCl was migrating into the organic phase. In an attempt to improve the phase separation we turned to perfluorinated extractants..sup.66 Disappointingly, the fluorous analog of 1-octanol, CF.sub.3(CF.sub.2).sub.5CH.sub.2CH.sub.2OH, formed an emulsion. A 0.05 M solution of C.sub.10F.sub.21CH.sub.2CH(OH)CH.sub.2OH in trifluorotoluene/hexafluoropropanol (1/1, (v/v)) failed to extract any .sup.45Ti. (Table 1, entry 5). Previously, it was reported, in the context of investigating catalytic complexes of Ti, that guaiacol (o-methoxyphenol) and maltol were able easily to form moisture-sensitive but isolable complexes with titanium tetrachloride..sup.67 Gratifyingly, using neat guaiacol as an extractant, we were able to extract 75% of activity into the organic phase, presumably as Structure 2.
##STR00002##
Example 2—Investigation of Extractants and Conditions for Liquid-Liquid Extraction and Phase Separation in Flow of Ti Ions from a Solution Also Containing Sc Ions
[0181] The LLE and phase separation in flow were performed using a membrane-based separator with a PFA diaphragm for integrated pressure control (
[0182] A flow schematic of the experimental setup is analogous to what has been depicted in
[0183] To find the optimum extraction conditions, the diaphragm thickness, membrane pore size, organic phase composition, flow rate ratios, and residence times were varied. Once the optimal materials and conditions were determined, a study was conducted to determine the shortest residence time, thereby minimizing the dead volume and overall processing time, while still maintaining high extraction. The residence time was varied by varying the length of the mixing tubing after the static mixers while maintaining a constant flow rate. All systems were operated for 60 min each and samples were collected every 15 min.
[0184] Investigation of Membrane Pore Size, Diaphragm Thickness and Organic Phase Composition
[0185] Due to the harsh nature of both the aqueous and organic solvents used for this extraction, both the membrane and diaphragm had to be extremely stable. Therefore, PTFE membranes were used for all experiments described herein.
[0186] Complete separation requires both the diaphragm thickness and membrane pore size be chosen so that P.sub.dia lies between the P.sub.cap and P.sub.perm pressures. In general, low interfacial tension mixtures often require smaller pore size membranes and thinner diaphragms. In these experiments, polytetrafluoroethylene (PTFE) membranes were tested using the following pore sizes: 0.1, 0.2, and 0.5 μm. Three different diaphragm film thicknesses were also tested: 0.001″, 0.002″, and 0.005″ (0.025 mm, 0.051 mm, and 0.127 mm).
[0187] Since guaiacol had shown the most selective and highest extraction efficiency for Ti over Sc in batch, it was chosen as a candidate for translation into flow, with 1-octanol also being investigated.
TABLE-US-00002 TABLE 2 Membrane Diaphragm Pore Size Thickness [μm] [in (mm)] Performance 37% HCl and 1-octanol 0.5 0.005 (0.127) Breakthrough 0.002 (0.051) Breakthrough 0.001 (0.025) Breakthrough 0.2 0.001 (0.025) Breakthrough 0.1 0.001 (0.025) Retention/Breakthrough Separation performance for various membrane pore sizes and diaphragm thicknesses for 37% HCl mixed with 1-octanol
[0188] It can be seen from Table 2 that no satisfactory conditions could be found for the use of 1-octanol as extractant. As the extraction efficiency was only around 50%, it was decided not to investigate this extractant further.
[0189] A solution of 0.01 M TiCl.sub.4 and 0.01 M ScCl.sub.3 in 37% HCl was extracted into guaiacol using the membrane separator.
[0190] Occasional retention and/or breakthrough of the aqueous into the organic phase was observed with a 0.2 μm PTFE/PP membrane, 0.002″ (0.051 mm) diaphragm thickness, and 0.2 mL/min total flow rate. The situation was remedied by adding various amounts of anisole as an interfacial tension modifier, which is structurally similar to guaiacol but acted to increase the interfacial tension, as shown in Table 3. Although the guaiacol/anisole mixture performed much better than 1-octanol, its extraction efficiency was not high enough. Therefore, the ratio of guaiacol to anisole was varied as well, as summarized in Table 5.
TABLE-US-00003 TABLE 3 Membrane Diaphragm Pore Size Thickness [μm] [in (mm)] Performance 37% HCl and 1:1 Guiacol/Anisole 0.5 0.005 (0.127) Breakthrough 0.002 (0.051) Breakthrough 0.001 (0.025) Retention 0.2 0.002 (0.051) Complete separation Separation performance for various membrane pore sizes and diaphragm thicknesses for 37% HCl mixed with 1:1 guaiacol/anisole.
[0191] Investigation of Flow Rate Ratios and Organic Phase Compositions Since a 0.2 μm membrane and a 0.002″ (0.051 mm) diaphragm was the only combination that led to complete phase separation, it was used for all of the optimization experiments.
[0192] The organic mixtures of guaiacol and anisole were used to extract Ti, (0.01 M) from 37% HCl at a total flow rate of 0.20 mL/min and aqueous to organic ratios of 1/5, 1/3, 1/1, 3/1, and 5/1 (v/v). Corresponding flow rates are shown in Table 4.
TABLE-US-00004 TABLE 4 Aq/Org Aq. Flow Org. Flow Flow Ratio Rate Ratio Rate Ratio [—] [mL/min] [mL/min] 1:5 0.03 0.17 1:3 0.05 0.15 1:1 0.10 0.10 3:1 0.15 0.05 5:1 0.17 0.03 Aqueous to organic flow rate ratios with corresponding volumetric flow rates (total flow rate = 0.20 mL/min).
[0193] The composition of the organic phase needed to both selectively extract only Ti and have a high enough interfacial tension with the HCl phase that complete separation could be achieved. It was determined that extraction was directly correlated with guaiacol concentration, that is a higher guaiacol concentration led to higher extraction up to a maximum Ti extraction of 90% with 90% guaiacol. Guaiacol concentrations above 90% led to incomplete phase separation. A summary of the phase separation performance using various organic phase compositions is shown in Table 5.
[0194] In addition to the composition of the organic phase, the relative ratios of aqueous to organic flow rates were also varied. When comparing relative flow rate ratios of 1/1, 1/3, and 1/5 (v/v) (aq. to org.) it was determined that 1/1 gave the lowest extraction. A ratio of 1/3 gave a higher extraction, but 1/5 did not yield a further increase in performance. All ratios where the aqueous flow rate was higher led to lower extraction efficiency. Therefore, a flow rate ratio of 1/3 was chosen as to avoid using excess solvent (Table 5).
TABLE-US-00005 TABLE 5 Org. Phase Aq. Org. Guaiacol/Anisole Flow Rate Flow Rate Performance (v/v) [mL/min] [mL/min] [—] Aqueous Phase: 37% HCl (12M) 0.2 μm PTFE membrane and 0.002” (0.051 mm) Diaphragm thickness 10/90 0.10 0.10 Complete Sep. 75/25 0.05 0.15 Complete Sep. 0.10 0.10 0.10 0.30 0.25 0.75 90/10 0.03 0.17 Complete Sep. 0.05 0.15 0.25 0.75 95/05 0.10 0.10 Retention/Breakthrough Phase separation performance for different guaiacol to anisole ratios in the organic phase using different aqueous to organic flow rate ratios.
[0195] Investigation of Scalability and Stability of the Extraction
[0196] In order to determine the scalability and stability of the extraction, the total flow rates were scaled five-fold to a total flow rate of 1.00 mL/min, while maintaining the same flow rate ratios and residence times. The extraction performance was identical to the original scale, and the maximum extraction of 90% was still achieved at 90% guaiacol and a flow rate ratio of 1/3 (
[0197] Investigation of Residence Time
[0198] After an optimal system was developed, the residence time of mixing was varied to minimize the dead volume and decrease the total amount of time spent in the system. This was achieved by increasing or decreasing the length of the PFA tubing used for mixing. The following lengths were tested with their corresponding residence times at 0.20 mL/min: 10 cm (13.7 s), 25 cm (34.2 s), 54 cm (73.9 s), 108 cm (147.8 s), 216 cm (295.6 s).
[0199] The total production time is a critical parameter in this system since the radioactive Ti is continuously undergoing decay back to Sc (t½=3 hours). Therefore, the shortest possible residence time is desirable. Residence times of the mixing tubing were varied from 13.7 s up to −5 min. The extraction efficiency was the same for all residence times. Therefore the shortest residence time was the most optimal, and resulted in a total system residence time of less than 1 min (
[0200] Conclusion—Optimal Conditions
[0201] Overall, an organic phase consisting of 90% guaiacol and 10% anisole, total flow rates of 0.20 or 1.00 mL/min, an aqueous to organic flow rate ratio of 1/3, and a residence time of 13.7 s led to highest and most efficient extraction resulting in 90.3±1.1% extraction of Ti and 0% extraction of Sc.
Example 3—Optimisation of Conditions for Liquid-Liquid Extraction and Phase Separation in Flow of .SUP.45.Ti Ions from a Solution Also Containing Sc Ions
[0202] With the extraction conditions optimized for .sup.natTi, we turned to the radioactive isotopomer, .sup.45Ti. While the concentration of Sc was comparable in both non-radioactive and radioactive cases (0.01 M vs. 0.03-0.13 M correspondingly) the concentration of the radiometal in the Sc-containing matrix solution was lower than that of its natural abundance isotopomer by 10 orders of magnitude, ranging from 1 to 10 picomoles. At these concentrations, even trace levels of impurities or water could potentially lead to side-reaction or hydrolysis and, as a consequence, change the extraction efficiencies of .sup.45Ti. To our delight, however, the LLE of .sup.45Ti in flow using a guaiacol-anisole 9/1 (v/v) mixture and a flow rate ratio of 1/3 (aq. to org.), with a residence time of 13.7 s showed that the extraction efficiency of .sup.45Ti was consistent with that of .sup.natTi (84.8±2.4% and 90.3±1.1% correspondingly), (
Example 4—Synthesis of .SUP.45.Ti-Containing Radiotracer
[0203] Finally, to examine if the extracted solution of .sup.45Ti can be directly used for radiolabelling, we attempted a synthesis of [.sup.45Ti](salan)Ti(dipic) 3, a Ti-antineoplastic, previously used for animal .sup.45Ti-PET and ex vivo radiotracing (Scheme 1)..sup.18
##STR00003##
[0204] To that end, the organic phase after the continuous LLE of .sup.45Ti was collected and reacted with an equimolar solution of salan and 2,6-pyridine dicarboxylic acid (dipic) in pyridine at 60° C. An essentially complete (98.7%) conversion to the desired product 3 was observed within 15 min as evidenced by radio-TLC (red peak for the product 3 and only traces of unreacted 2, green peak), proving the high quality and reactivity of extracted .sup.45Ti (
[0205] Separation of Zr and Y
[0206] Materials
[0207] Guaiacol (99%, natural), anisole (99%, ReagentPlus), hydrochloric acid (37%, ACS reagent), sulfuric acid (95.0-98.0%), and trioctylphosphine oxide (TOPO, ≥98.5%) were purchased from Sigma Aldrich. High purity hydrochloric acid (37%, “Ultrapur”) was purchased from Merck. All purchased chemicals were used without further purification. Yttrium foil (99.9%) was purchased from Alfa Aesar. ZrCl.sub.4 and YCl.sub.3 were purchased from Sigma Aldrich.
[0208] Pall PTFE membranes were used for all experiments (47 mm diameter, 0.1/0.2/0.5 μm pore size, polypropylene support). Perfluoroalkoxy alkane (PFA) diaphragms (0.001″/0.002″/0.005″ (0.0254/0.0508/0.1270 mm)) were purchased from McMaster Carr. All PFA tubing ( 1/16″ (1.5875 mm) OD, 0.03″ (0.762 mm) ID) was purchased from Idex Health and Science. PTFE static mixers were purchased from Stamixco. Two syringe pumps (KDS 100 Legacy Syringe Pump) and a dose calibrator (CRC-55tR, CII Capintec, Inc.) were used for the experiments.
[0209] Radionuclide Production and Separation
[0210] For all experiments, the cyclotron target material (yttrium) was used at its natural abundance level.
[0211] .sup.89Zr was produced by proton bombardment of yttrium foils on a PETTrace PT800 cyclotron. The 640 μm thick, 5 mm×5 mm foils were cut and sandwiched between a silver disc and a 500 μm Al degrader and placed in the target holder, providing direct water cooling on the rear face of the silver. Based on SRIM calculations the Al foil degrades the incident proton energy from the nominal 16.5 to approx. 13.1 MeV, bringing the energy below the threshold for co-production (<100 μb) of both .sup.88Y and .sup.88Zr. The irradiated foil was digested in 30-37% HCl. The mixture was filtered and centrifuged if necessary. If needed, the solution was diluted with water to make the final dilution ca. 6 M in HCl. These dilutions were used as the aqueous phase for the LLEF.
[0212] Instrumentation and Methods
[0213] For .sup.89Zr work, the extent of extraction (extraction %) was determined from the radioactivity measurements and using inductively coupled plasma atomic emission spectroscopy (ICP-AES, Agilent 5100 Dual View) of the aqueous phase. Samples of the aqueous phase were collected before the LLE and after 5, 15, 30, and 45 minutes of LLE. 0.35 mL of each sample was digested in 5 mL with 10% (v/v) H.sub.2SO.sub.4 for 6 hours at 160° C. 2.7 mL of the digested sample was diluted up to 10 mL with Milli-Q water to reach a total acid concentration of 5% (v/v).
[0214] An Eppendorf 5702 centrifuge was used to assist in phase separation. The membrane separator module was similar to those manufactured by Zaiput Flow Technologies. The solutions for the continuous membrane-based separation were pumped using either the KDS 100 Legacy Syringe (radioactive experiments) or the Harvard Apparatus PHD 2000 Programmable and Infusion syringe pumps (non-radioactive experiments). For batch experiments, the phase mixing was performed using an IKA ROCKER 3D digital shaker.
[0215] The continuous liquid-liquid extraction in flow (LLEF) was performed using the apparatus depicted in
Example 5—Investigation of Extractants and Conditions for Liquid-Liquid Extraction and Phase Separation in Flow of Zr Ions from a Solution Also Containing Y Ions
[0216] Earlier reports indicated that zirconium can be extracted from its acidic solutions into an organic phase containing trioctylphosphine oxide (TOPO)..sup.35 However, the preliminary batch experiments containing the equimolar solutions of 0.01 M ZrCl.sub.4 and YCl.sub.3 in 37% HCl produced a 3-phase mixture.
[0217] We began by testing whether the phase separation in this extraction system can be improved under the LLEF conditions.
[0218] Starting with the conditions optimized for Ti/Sc extraction (0.2 μm membrane, 0.002″ (0.051 mm) diaphragm, and 0.05/0.15 mL/min aq/org flow rate) we found that extensive breakthrough of the third phase occurred at both 0.2 μm and 0.1 μm membrane pore size (Table 6, entry 1). Lowering the concentrations of ZrCl.sub.4 and YCl.sub.3 to 0.001 M and then to 0.0005 M did not result in any improvement in the phase separation either (Table 6, entries 3-4). Only by lowering the concentration of HCl from 12 M to 6 M can a complete phase separation be achieved (Table 6, entry 4).
[0219] Unable to use the literature conditions for LLEF of Zr, we turned to 9/1, (v/v) guaiacol/anisole mixture, which performed extremely well for the phase separation and LLEF of Ti. Complete phase separation occurred in 37% HCl at both 0.001 M and 0.01 M ZrCl.sub.4 and YCl.sub.3 in a wide range of flow rates (Table 6, entries 5-6).
TABLE-US-00006 TABLE 6 Flow rate Flow rate Aqueous (aqueous) (organic) Separation Entry Organic phase phase (mL/min) (mL/min) performance 1 0.1M TOPO 37% HCl 0.05 0.15 Breakthrough in hexane (third phase) 2 0.1M TOPO 0.001M ZrCl.sub.4 and 0.05 0.15 Breakthrough in hexane YCl.sub.3 in 37% HCl (third phase) 3 0.1M TOPO 0.0005M ZrCl.sub.4 and 0.05 0.15 Breakthrough in hexane YCl.sub.3 in 37% HCl (third phase) 4 0.1M TOPO 0.001M ZrCl.sub.4 and 0.05 0.15 Complete in hexane YCl.sub.3 in 6M HCl phase separation 5 90% guaiacol 0.001M ZrCl.sub.4 and 0.033 0.167 Complete 10% anisole YCl.sub.3 in 37% HCl phase separation 6 90% guaiacol 0.01M ZrCl.sub.4 and 0.05 0.15 Complete 10% anisole YCl.sub.3 in 37% HCl 0.25 0.75 phase 0.033 0.167 separation Phase separation performance for different extractant/interfacial tension modifier mixtures in the organic phase, different starting concentrations in the aqueous phase, and using different aqueous to organic flow rate ratios.
[0220] The extraction efficiencies of Zr from 0.01 M ZrCl.sub.4 solution in 37% HCl, also containing 0.01 M YCl.sub.3 were investigated at different flow rates using the guaiacol/anisole, 9/1 (v/v) mixtures (
Example 6—Investigation of Extractants and Conditions for Liquid-Liquid Extraction and Phase Separation in Flow of .SUP.89.Zr Ions from a Solution Also Containing Y Ions
[0221] The extraction of .sup.89Zr from its solution in 37% HCl, also containing 0.01 M YCl.sub.3 was explored at low flow rates (0.033/0.166 mL/min, aq/org) using the guaiacol/anisole, 9/1 (v/v) mixtures. (
[0222] Separation of Gallium and Zinc
[0223] Materials
[0224] Anisole (99%, ReagentPlus), hydrochloric acid (37%, ACS reagent), zinc chloride (≥98%), sulfuric acid (95.0-98.0%), dibutyl ether, butyl methyl ether, tetrahydropyran, hexyl methyl ether, α,α,α-trifluorotoluene, and toluene, were purchased from Sigma Aldrich. Diethyl ether, diisopropyl ether (≥99%), and high purity hydrochloric acid (37%, “Ultrapur”) were purchased from Merck. Heptane (99.7%) and 1,2-dichloroethane were purchased from VWR Chemicals. All purchased chemicals were used without further purification. Zinc foil (99.9%) was purchased from Alfa Aesar.
[0225] Pall PTFE membranes were used for all experiments (47 mm diameter, 0.1/0.2/0.5 μm pore size, polypropylene support). Perfluoroalkoxy alkane (PFA) diaphragms (0.001″/0.002″/0.005″ (0.0254/0.0508/0.1270 mm)) were purchased from McMaster Carr. All PFA tubing ( 1/16″ (1.5875 mm) OD, 0.03″ (0.762 mm) ID) was purchased from Idex Health and Science. PTFE static mixers were purchased from Stamixco. The 15 mL plastic centrifuge tubes with screw caps were purchased from VWR.
[0226] Combined Radiogallium (.sup.66Ga, .sup.67Ga, .sup.68Ga) and .sup.65Zn Production and Purification
[0227] For all experiments, the cyclotron target material (zinc) was used at its natural abundance level.
[0228] Production: These radionuclides were produced simultaneously, by proton bombardment of stacked Zn and Cu foils. The incident 16.5 MeV proton beam would first encounter a 250 μm thick, 831 mg Zn foil before entering a 500 μm thick, 327 mg Cu foil. Incident energy on the Cu foil was calculated to appx. 12.8 MeV, making the 500 μm foil a thick target (range in Cu only 370 μm). The foils were irradiated for 160 minutes at 10 μA resulting in an integrated current of 26.2 μAh. The irradiated Zn foil, containing gallium radioisotopes was dissolved in a small amount of 3 M or 6 M hydrochloric acid and then added to either the 7 molal (m) or 1 M stock solution of ZnCl.sub.2 also prepared in 3 M or 6 M hydrochloric acid.
[0229] .sup.65Zn Purification: The irradiated Cu foil (327 mg, containing 5.6 MBq of .sup.65Zn) was dissolved in 1.7 mL of concentrated HNO.sub.3 at 60° C. The deep blue solution was evaporated to dryness at 150° C. using vigorous Ar flow. The blue solid was re-dissolved in 2.5 mL 1 M HCl, and loaded onto TK200 resin (3 g). The resin was first eluted with 1 M HCl, which removed all the copper (a total of 14 mL), and then with water, which eluted the zinc (a total of 25 mL). The fractions containing the highest amount of .sup.65Zn were collected, the solution was evaporated to dryness, and added to either the 7 molal (m) or 1 M stock solution of ZnCl.sub.2 prepared as described above. The resulting solution, containing 100-300 kBq of .sup.65Zn and radiogallium (present mostly as .sup.67Ga) and simulating a cyclotron-irradiated liquid target mixture was used as the aqueous phase for the LLE.
[0230] Instrumentation and Methods
[0231] Gallium and zinc were quantified by measuring radioactivities from .sup.67Ga, .sup.68Ga, and .sup.65Zn radioisotopes using the CRC-55tR, CII (Capintec, Inc) dose calibrator and Princeton Gammatech LGC 5 and Ortec GMX 35195-P gamma spectrometers.
[0232] An Eppendorf 5702 centrifuge was used to assist in phase separation. The membrane separator module was similar to those manufactured by Zaiput Flow Technologies. The solutions for the continuous membrane-based separation were pumped using either the KDS 100 Legacy Syringe (radioactive experiments) or the Harvard Apparatus PHD 2000 Programmable and Infusion syringe pumps (non-radioactive experiments). For batch experiments, the phase mixing was performed using an IKA ROCKER 3D digital shaker.
[0233] For the phase separation studies, a centrifuge tube was charged with 1.3 mL of the 7 m ZnCl.sub.2-3 M HCl, or 7 m ZnCl.sub.2-6 M HCl solution and various amounts of organic phase were added. The mixture was shaken for 30 minutes and centrifuged at 4000 rpm to separate the phases.
[0234] For the batch LLE of gallium, a centrifuge tube was charged with 1.3 mL of the 7 m ZnCl.sub.2-3 M HCl, or 7 m ZnCl.sub.2-6 M HCl solution, also containing .sup.67Ga, .sup.68Ga, and .sup.65Zn radioisotopes, and various amounts of organic phase were added. The mixture was shaken for 30 minutes and centrifuged at 4000 rpm to separate the phases.
[0235] The continuous liquid-liquid extraction and phase separation in flow was performed using a membrane-based separator with a PFA diaphragm for integrated pressure control. A flow schematic of the apparatus is depicted in
Example 7—Phase Separation Studies Using Several Dialkyl Ethers and Hydrochloric Acid, Also Containing Concentrated Zinc Chloride
[0236] The earlier work established that dialkyl ethers, and in particular diethyl ether, efficiently and selectively extracted gallium from 5-6 M hydrochloric acid solutions in batch..sup.68-71 Since dialkyl ethers are generally non-toxic, readily available low boiling point liquids, we decided to evaluate this class of compounds for further development in LLE and membrane-based separation of gallium from zinc. Tetrahydropyran (THP) diethyl (Et.sub.2O), diisopropyl (.sup.iPr.sub.2O), dibutyl (Bu.sub.2O), butyl methyl (BuOMe), and hexyl methyl (HexOMe) ethers were chosen as the extractants. The preliminary experiments showed that the presence of concentrated (7 m) ZnCl.sub.2 dramatically influenced the phase equilibrium. A single phase was observed by mixing equal volumes of diethyl ether, and a concentrated solution of ZnCl.sub.2 prepared in 6 M HCl. Lowering the concentration of HCl to 5, and then to 4 M still produced a single phase. At 3.5 M HCl two phases finally separated but extensive migration of aqueous into the organic phase was observed (Table 7, entry 8, Et.sub.2O/aq=6.22, (v/v)). Lowering the concentration of HCl further led to a gradual decrease in the ratio Et.sub.2O/aq, (v/v), ie a decrease in the migration of aqueous into the organic phase (Table 7, entries 2-4). Unexpectedly, this trend was opposite to what one observed when ZnCl.sub.2 was not present..sup.69 The extraction efficiency of Ga and Ga/Zn selectivity were also disappointing.
TABLE-US-00007 TABLE 7 Extraction in Et.sub.2O (%) Entry HCl (M) Et.sub.2O/aq, (v/v) .sup.68Zn .sup.68Ga 1 0 1.02 15.29 8.23 2 0.5 1.21 61.51 22.09 3 1 1.51 66.47 28.9 4 1.5 1.74 5 2 2.02 73.19 47.37 6 2.5 2.58 7 3 3.82 80.24 64.91 8 3.5 6.22 Batch extraction of zinc and gallium into diethyl ether from a solution of ZnCl.sub.2 prepared by dissolving 1 g of salt in 1 mL of aqueous HCl of a given strength.
Example 8—the Effect of Adding an Interfacial Tension Modifier on Phase Separation Using Several Dialkyl Ethers and Hydrochloric Acid, Also Containing 7 m Zinc Chloride
[0237] The interfacial migration is a critical parameter which had to be minimized to prevent the contamination of aqueous phase with the organic phase, which would make the process incompatible with the implementation of continuous LLEF due to stringent organic-free requirements for the ZnCl.sub.2-based aqueous cyclotron solution targets. The significant interfacial migration would also lead to low interfacial tension, which might cause a phase breakthrough during membrane separation. Our strategy was to find a suitable interfacial tension modifier which provided for reliable phase separation with no or little interfacial migration while keeping good Ga extraction efficiency and high Ga/Zn selectivity. Given its low capacity to dissolve water.sup.72, toluene was initially chosen as an interfacial tension modifier for screening the phase separation in the series R.sub.1OR.sub.2/ZnCl.sub.2—HCl.
[0238]
Example 9—the Batch Extraction of Ga and Zn Using Several Dialkyl Ethers, TFT, and Hydrochloric Acid, Also Containing 7 m Zinc Chloride
[0239] Having established a preferred interfacial tension modifier for phase separation, we investigated its performance for Ga/Zn extraction selectivity in the series of extractants R.sub.1OR.sub.2/ZnCl.sub.2—HCl at 6 M and 3 M HCl. Table 8 shows that TFT in combination with any of the six ethers in the screening set allowed for excellent gallium extraction efficiencies (entries 1-6). At 6 M HCl, the .sup.iPr.sub.2O and BuOMe were found to be the best performers extracting up to 95% Ga in batch. As expected, the Ga extraction efficiency decreased substantially in 3 M HCl. Nevertheless, a 2/1 (v/v) mixture of TFT and .sup.iPr.sub.2O was able to extract 77% of Ga and only 1% of Zn (entry 3). THP co-extracted the highest amount of Zn from 6 M and 3 M HCl.
TABLE-US-00008 TABLE 8 ZnCl.sub.2/ 6M HCl ZnCl.sub.2/ 3M HCl Entry Ether Ga extraction (%) Zn extraction (%) Ga extraction (%) Zn extraction (%) 1 THP 92 (5) 13 (1) 79 (1) 15 (3) 2 Et.sub.2O 94 (1) 3 (2) 73 (16) 4 (2) 3 .sup.iPr.sub.2O 97 (1) 5 (2) 78 (1) 1 (1) 4 Bu.sub.2O 89 (1) 0.5 (1) 20 (9) 1 (1) 5 BuOMe 97 (1) 5 (2) 48 (1) 0.3 (1) 6 HexOMe 93 (3) 0.9 (1) 26 (4) 0.2 (1) 7 Am.sub.2O 79 (6) 0.3 (1) 11 (2) 2 (1) The batch extraction of Ga and Zn for each of the dialkyl ethers in a 1:2 ratio with TFT, and hydrochloric acid, also containing 7 m zinc chloride. The figures in parentheses following the percentages are the standard deviations obtained over three runs of the extractions.
Example 10—the Liquid Liquid Extraction in Flow of Ga and Zn Using Several Dialkyl Ethers, TFT, and Hydrochloric Acid, Also Containing 7 m Zinc Chloride, Followed by Back-Extraction of Ga into 0.1 M HCl
[0240] Next, we translated the batch experiments into fully continuous flow experiments using the apparatus depicted in
[0241] The aqueous phase was formed by a 7 m ZnCl.sub.2/3 M HCl mixture and the organic phase consisted of a 2/1, (v/v) mixture of TFT used as an interfacial tension modifier and the series of ethers were used as the extractant. The aqueous and the organic phases were combined through a tee and mixed with two static mixers and mixing tubing. The aqueous phase was retained by the membrane, while the organic phase permeated through the membrane. For the membrane-based separator, we used optimized conditions established in the previously described work on .sup.45Ti separation: flowrate org/aq, (mL/h)=45/15; 0.2 μm membrane pore size, 0.002″ (0.051 mm) diaphragm, and 108 cm mixing tube. The .sup.68Ga was selectively extracted into the organic phase. The organic phase can then be either collected or directly re-routed into the second separation module, where 0.1 M HCl was used as the aqueous phase. After the second stage LLEF, .sup.68Ga was selectively back-extracted into the aqueous phase together with the residual Zn.
[0242]
[0243] Table 9 shows that gallium stripping was uniformly high (>90%) across the series. On the other hand, little selectivity was observed for Zn stripping, so that a single-stage LLEF/back-extraction sequence delivered the desired gallium solution in 0.1 M HCl containing as much as 10 mg/mL of zinc (the presence of greater than 10 mg/mL of Zn is indicated in
TABLE-US-00009 TABLE 9 Ether Ga stripping % Zn stripping % Et.sub.2O 99 91 Pr.sub.2O Run 1: 93 Run 1: 58 Run 2: 98 Run 2: 93 BuOMe Run 1: 98 Run 1: 47 Run 2: 98 Run 2: 76 THP 99 98 HexOMe 97 72 Stripping of gallium and zinc from R.sub.1OR.sub.2/TFT, ½ (v/v) using 0.1M HCl.
Example 11—the Two-Stage Liquid-Liquid Extraction in Flow of Ga and Zn Using Diisopropyl Ether, TFT, and Hydrochloric Acid, Also Containing 7 m Zinc Chloride Followed by Back-Extraction of Ga into 0.1 M HCl
[0244] To decrease the amount of co-extracted Zn, two consecutive (two-stage) liquid-liquid extractions/back-extractions in flow were performed. In this experiment, two extraction/back-extraction modules can be combined. After the first stage of extraction/back-extraction, the mixture Ga containing residual zinc was acidified to 6 M HCl and subjected to the second stage of extraction/back-extraction under analogous conditions.
[0245] After the second stage, the 71% of original gallium was recovered in the final solution for radiolabeling, which also contained 100 μg/mL of Zn.
TABLE-US-00010 TABLE 10 The two-stage LLEF of Ga and Zn using the mixture of dialkyl ethers, TFT, and hydrochloric acid, also containing 7 m zinc chloride 1.sup.st stage extraction/ 2.sup.nd stage extraction/ 1.sup.st and 2.sup.nd stages back-extraction back-extraction combined Extraction Stripping Extraction Stripping Extraction Stripping % % % % % % Ga 74 98 99 99 73% 97% Zn 1 96 3 49 0.03 47%
Example 12—the Two-Stage Liquid-Liquid Extraction in Flow of Ga and Zn Using Several Dialkyl Ethers, TFT, and Hydrochloric Acid, Also Containing 7 m Zinc Chloride Also Including Scrubbing of Residual Zn with 0.1 M HCl and Back-Extraction of Ga into 0.1 M HCl
[0246] Even higher purity gallium solution can be obtained, if an extra step of liquid-liquid extraction of Zn using 8 M HCl is included (
Example 13—the Single-Stage Liquid-Liquid Extraction in Flow of Ga and Zn Using Diisopropyl Ether, TFT, and Hydrochloric Acid, Also Containing 7 m Zinc Chloride from a Cyclotron Target Solution
[0247] Radioqallium production: Approximately 3.5 ml of target solution (5 M ZnCl.sub.2 in 3 M aq. HCl) was loaded in a GE PETtrace liquid target. The target chamber was made of niobium to limit corrosion. The target front foil was 250 μm niobium foil, bringing the proton energy down to 12.5 MeV from the nominal 16.5 MeV. The target was not pressurized, but left open to ensure no pressure buildup in the chamber. Bombardment was performed at a current of 5 μA for 6 minutes. After irradiation the produced Ga-68 was quantified by gamma spectroscopy on a 10% GeLi detector, calibrated using certified Eu-152 and Ba-133 sources. The produced activity at saturation was calculated to 204 MBq/μA.
[0248] Liquid-liquid extraction of radioqallium followed by irradiated target solution purification: 2.5 mL of the irradiated target solution was used as the aqueous phase for LLE. iPr.sub.2O/TFT (1/2) was used as the organic phase. The phases were separated using the membrane separator with a 0.2 μm PTFE/PP membrane and a 2 mil (0.0508 mm) diaphragm. The aqueous flow rate was 0.25 mL/min and the organic flow rate was 0.75 mL/min. Samples of the aqueous and organic phase after the LLE were collected and the activity of .sup.66,67,68Ga (radiogallium) was measured with gamma spectroscopy. 57% of radiogallium was extracted into the organic phase. The aqueous phase was then passed through a C18 cartridge after the LLE, in order to remove trace organic solvent, and used directly for a second irradiation.
Example 14—the Single-Stage Liquid-Liquid Extraction in Flow of Ga and Zn Using Diisopropyl Ether, TFT, and Hydrochloric Acid, Also Containing 7 m Zinc Chloride from a Re-Used Cyclotron Target Solution
[0249] Radioqallium production from a re-used cyclotron target solution: Approximately 3.5 ml of a 1:1 target solution (5 M ZnCl.sub.2 in 3 M aq. HCl) and LLE-purified target solution from the first bombardment (Example 13) was loaded in a GE PETtrace liquid target. The target chamber was made of niobium to limit corrosion. The target front foil was 250 μm niobium foil, bringing the proton energy down to 12.5 MeV from the nominal 16.5 MeV. The target was not pressurized, but left open to ensure no pressure buildup in the chamber. Bombardment was performed at a current of 5 μA for 5 minutes. After ended irradiation the produced Ga-68 was quantified by gamma spectroscopy on a 10% GeLi detector, calibrated using certified Eu-152 and Ba-133 sources. The produced activity at saturation was calculated to 258 MBq/μA.
[0250] Liquid-liquid extraction of radiogallium from the re-used irradiated target solution: The LLE procedure described in Example 13 was used to extract Ga from 2.5 ml of the target solution from the second bombardment, which led to a radiogallium extraction of 62%.
[0251] Separation of Cu and Ni
[0252] Materials
[0253] Hydrochloric acid (37%, ACS reagent), zinc chloride (≥98%), trioctylphosphine oxide (≥98.5%), cobalt chloride, iron chloride, silver chloride, copper (II) chloride and toluene were purchased from Sigma Aldrich. High purity hydrochloric acid (37%, “Ultrapur”) was purchased from Merck. Heptane (99.7%) and hexane were purchased from VWR Chemicals. Nickel-64 (99.6% isotope-enriched) was purchased from Campro Scientific. All purchased chemicals were used without further purification.
[0254] Pall PTFE membranes were used for all experiments (47 mm diameter, 0.1/0.2/0.5 μm pore size, polypropylene support). Perfluoroalkoxy alkane (PFA) diaphragms (0.001″/0.002″/0.005″ (0.0254/0.0508/0.1270 mm)) were purchased from McMaster Carr. All PFA tubing ( 1/16″ (1.5875 mm) OD, 0.03″ (0.762 mm) ID) was purchased from Idex Health and Science. PTFE static mixers were purchased from Stamixco. The 15 mL plastic centrifuge tubes with screw caps were purchased from VWR.
[0255] Radionuclide Production and Separation
[0256] For all experiments, the cyclotron target material was Nickel-64 (99.6% isotope-enriched).
[0257] Copper-64 Production and Purification
[0258] Production: .sup.64Cu was produced via the .sup.64Ni(p,n).sup.64Cu reaction using a water-cooled solid target mounted on the beam line of a PETtrace (GE Healthcare) cyclotron. The target consisted of approximately 80 mg of .sup.64Ni metal (enriched to 99%) electroplated on a silver disk backing. The target was irradiated with a proton beam with an incident energy of 16.5 MeV and a beam current of 20 μA. After irradiation, the silver disk backing was transferred into a hot cell where it was treated with 30% HCl for 30 min. at 60° C., and then for 5 min. at 80° C., resulting in a clear green solution containing a mixture of .sup.64CuCl.sub.2 and .sup.64NiCl.sub.2.
[0259] Purification: The solution was decanted, diluted to 6M HCl and loaded onto a Dowex 1×8 (chloride form 200-400 mesh) column. The column was washed with 21 mL of 6 M HCl, and then with 33 mL of 5 M HCl. Finally, the column was eluted with 10 mL of 1 M HCl, which elutes .sup.64Cu. Final evaporation from aqueous HCl yielded 2-6 GBq of .sup.64Cu as .sup.64CuCl.sub.2 with specific activity, 300-3000 TBq/mmol and radionuclidic purity of 99%.
[0260] Instrumentation and Methods
[0261] .sup.64Cu was quantified by measuring radioactivities from .sup.64Cu radioisotopes using the CRC-55tR, CII (Capintec, Inc) dose calibrator and Princeton Gammatech LGC 5 and Ortec GMX 35195-P gamma spectrometers. Cu, Ni, Ag, Fe, Co and Zn were quantified by ICP.
[0262] An Eppendorf 5702 centrifuge was used to assist in phase separation. The membrane separator module was similar to those manufactured by Zaiput Flow Technologies. The solutions for the continuous membrane-based separation were pumped using either the KDS 100 Legacy Syringe (radioactive experiments) or the Harvard Apparatus PHD 2000 Programmable and Infusion syringe pumps (non-radioactive experiments). For batch experiments, the phase mixing was performed using an IKA ROCKER 3D digital shaker.
[0263] The continuous liquid-liquid extraction and phase separation in flow was performed using a membrane-based separator with a PFA diaphragm for integrated pressure control. A flow schematic of the apparatus is depicted in
Example 15—Phase Separation Performance and Liquid-Liquid Extraction in Flow of Copper and Copper-64 Using Trioctylphosphine Oxide (TOPO) in Toluene, Hexane, and Heptane, Also Containing Various Amounts of Cu, Ni, Co, Zn, Fe, and Ag in 6M Hydrochloric Acid
[0264] A single stage liquid-liquid extraction in flow of copper and copper-64 was performed using the setup depicted in
TABLE-US-00011 TABLE 11 Phase separation performance and LLE in flow of copper and copper-64 from 6M HCl containing various amounts of metal impurities. Flow rate Flow rate Diaphragm Organic (aqueous) (organic) Separation Entry Membrane (mm) Aqueous phase phase (mL/min) (mL/min) performance 1 0.2 μm 0.0508 6M HCl 0.4M 0.05 0.15 Complete PTFE/PP (2 mil) 0.001M CuCl.sub.2 TOPO in 0.10 0.10 phase 0.01M NiCl.sub.2 toluene 0.0333 0.1667 separation 2 0.2 μm 0.0508 6M HCl 0.1M 0.05 0.15 Complete PTFE/PP 0.001M CuCl.sub.2 TOPO in 0.25 0.75 phase 0.01M NiCl.sub.2 hexane separation 3 0.2 μm 0.0508 60 ppm 0.1M 0.05 0.15 Complete PTFE/PP Cu, Ni, Co, TOPO in phase Zn, Fe and heptane separation Ag in 6M HCl 4 0.2 μm 0.0508 .sup.64Cu in 6M 0.1M 0.05 0.15 Complete PTFE/PP HCl TOPO in 0.25 0.75 phase heptane separation
[0265] With 0.4 M TOPO in toluene up to 93% copper extraction was achieved with ratios of 1:3 and 1:5, while the extraction was 85% with a 1:1 ratio (
[0266] Extraction with 0.4 M TOPO is not sensitive to a change in interfacial tension modifier (toluene vs. hexane) and remains efficient (93%) when the concentration of TOPO is decreased to 0.1 M. No nickel is extracted to the organic phase.
[0267]
[0268] Whilst the invention has been described with reference to preferred embodiments, it will be appreciated that various modifications are possible within the scope of the invention.
[0269] In this specification, unless expressly otherwise indicated, the word ‘or’ is used in the sense of an operator that returns a true value when either or both of the stated conditions is met, as opposed to the operator ‘exclusive or’ which requires that only one of the conditions is met. The word ‘comprising’ is used in the sense of ‘including’ rather than in to mean ‘consisting of’. All prior teachings acknowledged herein are hereby incorporated by reference. No acknowledgement of any prior published document herein should be taken to be an admission or representation that the teaching thereof was common general knowledge in Australia or elsewhere at the date hereof.
REFERENCES
[0270] 1. Huang, R.; Wang, M.; Zhu, Y.; Conti, P.; Chen, K. Current Topics in Medicinal Chemistry 2015, 15, 795. [0271] 2. Brasse, D., Nonat, A. Dalton Trans. 2015, 44 (11), 4845. [0272] 3. Ramogida, C. F.; Orvig, C. Chemical Communications 2013, 49 (42), 4720. [0273] 4. Zeglis, B. M.; Houghton, J. L.; Evans, M. J.; Viola-Villegas, N.; Lewis, S. J. Inorg. Chem. 2014, 53 (4), 1880. [0274] 5. Martiniova, L.; Palatis, L. D.; Etchebehere, E.; Ravizzini, G. Current Radiopharmaceuticals 2016, 9 (3), 187. [0275] 6. Witowska-Patena, E.; Mazurek, A.; Dziuk, M. Central European Journal of Urology 2017, 70 (1), 37. [0276] 7. Jailian, A. R. Iranian Journal of Nuclear Medicine 2016, 24 (1), 1. [0277] 8. Eder, M.; Neels, O.; Müller, M.; Bauder-Wüst, U.; Remde, Y.; Schsfer, M.; Hennrich, U.; Eisenhut, M.; Afshar-Oromieh, A.; Haberkorn, U.; et al. Novel Preclinical and Radiopharmaceutical Aspects of [68Ga]Ga-PSMA-HBED-CC: A New PET Tracer for Imaging of Prostate Cancer. Pharmaceuticals 2014, 7 (7), 779. [0278] 9. Velikyan, I. Molecules 2015, 20, 12913-12943. [0279] 10. Chakravarty, R.; Chakraborty, S.; Radhakrishnan, E.; Kamaleshwaran, K.; Shinto, A.; Dash, A. Nuclear Medicine and Biology 2017, 46, 1. [0280] 11. Filosofov, D. V.; Garland, M.; John, K. D.; Knapp Jr., F. F.; Kunetsoz, R.; Mausner, L.; Mirzadeh, S.; Pillai, M. R. A.; Ponsard, B.; Rösch, F. IAEA Radioisotopes and Radiopharmaceuticals Series No 2, 1. [0281] 12. Chakravarty, R.; Chakraborty, S.; Ram, R.; Vatsa, R.; Bhusari, P.; Shukla, J.; Mittal, B. R.; Dash, A Journal of Labelled Compounds and Radiopharmaceuticals 2016, 59 (3), 87. [0282] 13. Engle, J. W.; Lopez-Rodriguez, V.; Gaspar-Carcamo, R. E.; Valdovinos, H. F.; Valle-Gonzalez, M.; Trejo-Ballado, F.; Severin, G. W.; Barnhart, T. E.; Nickles, R. J.; Avila-Rodriguez, M. A. Applied Radiation and Isotopes 2012, 70 (8), 1792. [0283] 14. Jensen, M.; Clark, J. In Proceedings of 13.sup.th International Workshop on Targetry and Target Chemistry; Danmarks Techniske Universitet, Risø Nationallaboratioriet for Bæredygtig Energi, 2011, pp 288-292. [0284] 15. Pandey, M. K.; Byrne, J. F.; Jiang, H.; Packard, A. B.; DeGrado, T. R. American Journal of Nuclear Medicine and Molecular Imaging 2014, 4 (4), 303. [0285] 16. WO2015/175972 [0286] 17. Merrill, J. C.; Lambrecht, R. M.; Wolf, A. P. International Journal of Applied Radiation and Isotopes 1978, 29, 115. [0287] 18. Sadheghi, M.; Enferadi, M.; Nadi, H. Radiochemistry 2011, 53, 411. [0288] 19. Fazaeli, Y.; Abouzadeh, M.; Aardaneh, K.; Kakavand, T.; Bayat, F.; Yousafi, K. Nuclear Technology and Radiation Protection 2014, 29, 28. [0289] 20. Severin, G. W.; Nielsen, C. H.; Jensen, A. I.; Fonslet, J.; Kjor, A.; Zhuravlev, F. J. Med. Chem 2015, 58, 7591. [0290] 21. Ishiwata, K.; Ido, T.; Monma, M.; Murakami, M.; Fukuda, H.; Kameyama, M.; Yamada, K.; Endo, S.; Yoshioka, S.; Sato. T.; Matsuzawa, T. International Journal of Radiation Applications and Instrumentation Part A: Applied Radiation and Isotopes 1991, 42, 707. [0291] 22. Vāvere, A.; Laforest, R.; Welch, M. J. Nuclear Medicine and Biology 2005, 32, 117. [0292] 23. Vāvere, A.; Welch, M. J. Journal of Nuclear Medicine 2005, 46, 683. [0293] 24. Gagnon, K.; Severin, G. W.; Barnhart, T.; Engle, J. W.; Valdovinos, H. F.; Nickles, R. J. in AIP Conference Proceedings, AIP, 2012, vol 1509, pp 211-214. [0294] 25. Severin, G. W.; Engle, J. W.; Barnhart, T. E.; Nickles, R. J. .sup.89Zr Radiochemistry for Positron Emission Tomography. Medicinal Chemistry 2011, 7 (5). [0295] 26. Dilworth, J. R.; Pascu, S. I. The Chemistry of PET Imaging with Zirconium-89. Chemical Society Reviews 2018, 47 (8), 2554-2571. [0296] 27. Wright, B. D.; Whittenberg, J.; Desai, A.; DiFelice, C.; Kenis, P. J. A.; Lapi, S. E.; Reichert, D. E. Microfluidic Preparation of a .sup.89Zr-Labeled Trastuzumab Single-Patient Dose. Journal of Nuclear Medicine 2016, 57 (5), 747-752. [0297] 28. Meijs, W. E.; Herscheid, J. D. M.; Haisma, H. J.; Wijbrandts, R.; van Langevelde, F.; Van Leuffen, P. J.; Mooy, R.; Pinedo, H. M. Production of Highly Pure No-Carrier Added .sup.89Zr for the Labelling of Antibodies with a Positron Emitter. Applied Radiation and Isotopes 1994, 45 (45), 1143-1147. [0298] 29. Sadeghi, M.; Kakavand, T.; Rajabifar, S.; Mokhtari, L.; Rahimi-Nezhad, A. Nukleonica 2009, 54 (1), 25. [0299] 30. Lahiri, S.; Banerjee, S.; Das, N. R. Journal of Radioanalytical and Nuclear Chemistry 1997, 218 (2), 215. [0300] 31. Nakamura, T.; Sakai, A.; Nishihama, S.; Yoshizuka, K. Solvent Extraction and Ion Exchange 2009, 27 (4), 501. [0301] 32. Mane, C. P.; Anuse, M. A. Journal of the Chinese Chemical Society (Taipei, Taiwan) 2008, 55 (4), 807. [0302] 33. S. A. Kandil, B. Scholten, Z. A. Saleh, A. M. Youssef, S. M. Qaim, H. H. Coenen, Journal of Radioanalytical and Nuclear Chemistry 2007, 274, 1, 45-52. [0303] 34. Dejesus, O. T.; Nickles, R. J. Production and Purification of .sup.89Zr, a Potential PET Antibody Label. International Journal of Radiation Applications and Instrumentation. Part A, Applied Radiation and Isotopes 1990, 41 (8), 789-790. [0304] 35. SATO, T. THE EXTRACTION OF ZIRCONIUM(IV) FROM AQUEOUS ACID SOLUTIONS BY TRIOCTYLPHOSPHINE OXIDE. Solvent Extraction and Ion Exchange 1983, 1 (2), 251-261. [0305] 36. X. Fan, D. J. Parker, M. D. Smith, A. Ingram, Z. Yang, J. P. K. Seville, Nuclear Medicine and Biology 2006, 33, 7, 929-944. [0306] 37. K. S. Bhatki, A. T. Rane, M. B. Kabadi, Journal of Radioanalytical Chemistry, 1969, 2, 1-2, 73-77. [0307] 38. S. Banerjee, S. Lahiri, S. B. Manohar, A. Ramaswami, Applied Radiation and Isotopes, 2002, 56, 4, 571-575. [0308] 39. M. Sadeghi and B. Shirazi, Applied Radiation and Isotopes, 2008, 66, 12, 1810-1813. [0309] 40. Y. Wen, Z. Qin, W. Liu, Lanzhou Daxue Xuebao, Ziran Kexueban 2002, 38, 1, 58-61. [0310] 41. A. Gopalakrishna, H. Naik, S. V. Suryanarayana, Y. Naik, V. T. Nimje, B. k. Nayak, S. K. Sarkar, S. Padmanabhan [0311] 42. D. Nayak and S. Lahiri Applied Radiation and Isotopes 2001, 54, 2, 189-193 [0312] 43. A. Alian Nat. Bur. Stand. (US) Spec. Publ. 1969, 312, 1, 473-474 [0313] 44. M. Saeki, Y. Sasaki, A. Yokoyama, JAEA-Conf 2010. 2009-007 (Ronbunshu—Nippon Genshiryoku Kenkyu Kaihatsu Kiko Ryoshi Bimu Oyo Kenkyu Bumon Dai 9-kai Hikari Ryoshi Kagaku Kenkyu Shinpojumu, 2008) 121-124. [0314] 45. K. Aardaneh, T. N. van der Walt, Journal of Radioanalytical and Nuclear Chemistry 2006, 268, 1, 25-32. [0315] 46. M. K. Das, Madhusmita, S. Chattopadhyay, S. S. Das, L. Barua, M. A. Neyar, U. Kumar, A. De Journal of Radioanalytical and Nuclear Chemistry 2016, 310, 1, 423-432. [0316] 47. J. Britton and T. F. Jamieson, Nature Protocols 2017 12, 11, 2423. [0317] 48. A. Adamo, P. L. Heider, N. Weeranoppanant, K. F. Jensen Industrial and Engineering Chemistry Research 2013, 52, 10802-10808. [0318] 49. A. G. Gaikwad and A. M. Rajput Journal of Rare Earths 2010 28, 1, 1-6. [0319] 50. R. C. Luckay, X. Sheng, C. E. Strasser, H. G. Raubenheimer, D. A. Safin, M. G. Babashkina, A. Klein Dalton Transactions 2009 39, 8227-8236. [0320] 51. J. Konczyk, A. Nowik-Zajac, C. A. Kozlowski Separation Science and Technology 2016 51, 14, 2394-2410. [0321] 52. K. Masuda, T. Ichitsuka, N. Koumura, K. Sato, S. Kobayashi, Tetrahedron, 2018, 74, 1705-1730. [0322] 53. G. Glotz, R. Lebel, D. Dallinger and C. O. Kappe, Angewandte Chemie Intl Edn 2017, 56, 13786-13789. [0323] 54. B. Leforestier, M. Vögtle Synlett 2016, 27, 1957-1962. [0324] 55. C. Battilocchio, G. Iannucci, S. Wang, E. Godineau, A. Kolleth, A. De Mesmaeker and S. V. Ley React. Chem. Eng. 2017, 2, 295-298. [0325] 56. J. Salaklang, V. Maes, M. Conradi, R. Dams, T. Junkers React. Chem. Eng. 2018, 3, 41-47. [0326] 57. F. J. Strauss, D. Cantillo, J. Guerra, C. O. Kappe React. Chem. Eng. 2016, 1, 472-476. [0327] 58. H. Yang, B. Martin and B. Schenkel Organic Process Research and Development 2018, 22, 446-456. [0328] 59. P. Zhang, N. Weeranoppanant, D. A. Thomas, K. Tahara, T. Stelzer, M. G. Russell, M. O'Mahony, A. S. Myerson, H. Lin, L. P. Kelly, K. F. Jansen, T. F. Jamieson, C. Dai, Y. Cui, N. Briggs, R. L. Beingessner and A. Adamo, Chem. Eur. J. 2018, 24, 2776-2784. [0329] 60. Y. Shen, N. Weeranoppanant, L. Xie, Y. Chen, M. R. Lusardi, J. Imbrogno, M. G. Bawendi and K. F. Jansen, Nanoscale 2017, 9, 7703-7707. [0330] 61. H. P. L. Gemoets, G. Laudadio, K. Verstraete, V. Hessel and T. Nosl, Angewandte Chemie Intl Edn 2017, 56, 7161-7165. [0331] 62. E. Y. Tshuva, I. Goldberg and M. Kol, J. Am. Chem. Soc. 2000, 122, 10706-10707. [0332] 63. T. A. Immel, M. Grutzke, A.-K. Spate, U. Groth, P. Ohlschlager and T. Huhn, Chem. Commun. 2012, 48, 5790-5792. [0333] 64. J. Siikanen, The Journal of Nuclear Medicine, 54, 1095. [0334] 65. R. Mondal and P. Tarafder, Microchimica Acta, Mikrochimica Acta, Microchim. Acta, Microch Act, Microchim Acta, Mikrochim Acta, Mikroch Act, Mikrochimica Acta Micro and Trace Analysis, Microchimica Acta, 2004, 148, 327-333. [0335] 66. E. de Wolf, G. van Koten and B.-J. Deelman, Chemical Society Reviews, 1999, 28, 37-41. [0336] 67. P. Sobota, K. Przybylak, J. Utko, L. Jerzykiewicz, A. Pombeiro, M. da Silva and K. Szczegot, Chemistry—a European Journal, 2001, 7, 951-958. [0337] 68. Swift, E. H. A NEW METHOD FOR THE SEPARATION OF GALLIUM FROM OTHER ELEMENTS. J. Am. Chem. Soc. 1924, 46 (11), 2375-2381. [0338] 69. Irving, H. M.; Rossotti, F. J. C. The Solvent Extraction of Group IIIB Metal Halides. Analyst 1952, 77 (920), 801-812. [0339] 70. Nachtrieb, N. H.; Fryxell, R. E. The Extraction of Gallium Chloride by Isopropyl Ether. J. Am. Chem. Soc. 1949, 71 (12), 4035-4039. [0340] 71. BROOKS, R. R.; LLOYD, P. J. Influence of Molecular Structure on the Liquid/Liquid Extraction of the Chloro-Complexes of Gallium and Indium with Aliphatic Ethers. Nature 1961, 189, 375. [0341] 72. Kirchnerová, J.; Cave, G. C. B. The Solubility of Water in Low-Dielectric Solvents. Canadian Journal of Chemistry 1976, 54 (24), 3909-3916. [0342] 73. EP3101660. [0343] 74. Tolmachev, V; Lundqvist, H; Einarsson, L. Production of .sup.61Cu from a natural nickel target Appl. Radiat. Isot. 1998, 49(1-2), 79-81. [0344] 75. Vāvere, A. L.; Lewis, J. S. Dalton Transactions, 2007, 4893-4902. [0345] 76. Devi, N. B.; Nayak, B. The Journal of the Southern African Institute of Mining and Technology, 2014, 114, 937-943. [0346] 77. Devi, N. B.; Mishra, S. The Journal of the Southern African Institute of Mining and Technology, 2012, 112, 859-864. [0347] 78. Sepúlveda, R.; Castille, J.; Plaza, A.; Senchez, J.; Torres, A.; Romero, J. The Journal of Supercritical Fluids 2017, 128, 26-31. [0348] 79. Djebabra, S.; Barkat, D. J Mater. Environ. Sci. 2015, 6 (11), 3382-3387.