Cartesian sampling for dynamic magnetic resonance imaging (MRI)
11294009 · 2022-04-05
Assignee
Inventors
- Rizwan Ahmad (Columbs, OH, US)
- Ning Jin (Powell, OH, US)
- Orlando Simonetti (Columbus, OH)
- Yingmin Liu (Columbus, OH, US)
- Adam Rich (Columbus, OH, US)
Cpc classification
G01R33/543
PHYSICS
G01R33/5608
PHYSICS
G01R33/5619
PHYSICS
International classification
G01V3/00
PHYSICS
G01R33/54
PHYSICS
G01R33/561
PHYSICS
G01R33/56
PHYSICS
Abstract
A variable density Cartesian sampling method that allows retrospective adjustment of temporal resolution, providing added flexibility for real-time applications where optimal temporal resolution may not be known in advance. The methods provide for a computationally efficient sampling methods where a first step includes producing a uniformly random sampling pattern using a golden ratio on a grid, and the second step is applying a nonlinear stretching operation to create a variable density sampling pattern. Diagnostic quality images may be recovered at different temporal resolutions.
Claims
1. A method for CArtesian sampling with Variable density and Adjustable temporal resolution (CAVA), comprising: creating a uniformly random sampling pattern on a first grid by creating the first grid as a k-t Cartesian grid with N.sub.s phase-encoding (PE) lines; starting, from a randomly selected first PE index, p.sub.s (l), advancing an index of a subsequent PE lines gN.sub.s, yielding:
p.sub.s(i+1)=p.sub.s(i)+gN.sub.s
.sub.N.sub.
•
.sub.N.sub.
2. An MM apparatus, comprising: a scanner that generates magnetic fields used for the MR examination; a measurement space having a patient table; and a controller having an evaluation module, wherein the evaluation module executes instructions for performing a method for CArtesian sampling with Variable density and Adjustable temporal resolution (CAVA) that includes: creating a uniformly random sampling pattern on a first grid; for each frame, circularly rotating the uniformly random sampling pattern for a previous frame; repeating the creating and circularly rotating for all frames; projecting samples from the first grid to a second grid using a nonlinear stretching function; reconstructing diagnostic quality images from the second grid; and visually presenting the diagnostic quality images, wherein the evaluation module creates the uniformly random sampling pattern further comprising creating the first grid as a k-t Cartesian grid with N.sub.s phase-encoding (PE) lines, and wherein evaluation module further starts, from a randomly selected first PE index, p.sub.s (l), advancing an index of a subsequent PE lines gN.sub.s, yielding:
p.sub.s(i+1)−p.sub.s(i)+gN.sub.s
.sub.N.sub.
•
.sub.N.sub.
3. A method for Golden-shift Ordered Cartesian (GOC) sampling, comprising: creating a uniformly random sampling pattern on a k-t Cartesian grid with Ns phase-encoding (PE) lines; circularly rotating the sampling pattern of a previous frame; and determining if all frames have been completed, and if so, applying a nonlinear stretching operation to create a variable density sampling pattern with N PE lines.
4. A method for CArtesian sampling with Variable density and Adjustable temporal resolution (CAVA), comprising: creating a uniformly random sampling pattern on a first grid; projecting samples from the first grid to a second grid using a nonlinear stretching function; reconstructing diagnostic quality images from the second grid; and visually presenting the diagnostic quality images, wherein the projecting yields:
5. An MM apparatus, comprising: a scanner that generates magnetic fields used for the MR examination; a measurement space having a patient table; and a controller having an evaluation module, wherein the evaluation module executes instructions for performing a method for CArtesian sampling with Variable density and Adjustable temporal resolution (CAVA) that includes: creating a uniformly random sampling pattern on a first grid; for each frame, circularly rotating the uniformly random sampling pattern for a previous frame; repeating the creating and circularly rotating for all frames; projecting samples from the first grid to a second grid using a nonlinear stretching function; reconstructing diagnostic quality images from the second grid; and visually presenting the diagnostic quality images, wherein the projecting yields:
6. The method of claim 1, wherein:
7. The method of claim 1, further comprising creating a variable density sampling pattern with N PE lines using the nonlinear stretching operation.
8. The method of claim 7, wherein the variable density sampling pattern has adjustable temporal resolution.
9. The method of claim 1, wherein the projecting yields:
10. The method of claim 1, further comprising averaging k-space data sampled using CAVA over time to produce a fully sampled k-space.
11. The apparatus of claim 2, wherein:
12. The MM apparatus of claim 2, wherein the evaluation module further creates a variable density sampling pattern with N PE lines using the nonlinear stretching operation.
13. The MRI apparatus of claim 12, wherein the variable density sampling pattern has adjustable temporal resolution.
14. The MRI apparatus of claim 2, wherein the projecting yields:
15. The MRI apparatus of claim 2, wherein the evaluation module further averages k-space data sampled using CAVA over time to produce a fully sampled k-space.
16. The method of GOC sampling of claim 14, wherein creating the uniformly random sampling pattern comprises sampling on a smaller grid with N.sub.s PE lines
17. The method of GOC sampling of claim 14, circularly rotating the sampling pattern of the previous frame comprises starting from a randomly selected first PE index, p.sub.s (l), an index of the subsequent PE lines is sequentially advanced by gN.sub.s, with g=(1+√{square root over (5)})/2.
18. The method of GOC sampling of claim 14, further comprising applying nonlinear stretching to create a sampling pattern that has variable density.
19. The method of claim 4, creating the uniformly random sampling pattern further comprising creating the first grid as a k-t Cartesian grid with N.sub.s phase-encoding (PE) lines.
20. The method of claim 19, wherein:
21. The method of claim 4, further comprising creating a variable density sampling pattern with N PE lines using the nonlinear stretching operation.
22. The method of claim 21, wherein the variable density sampling pattern has adjustable temporal resolution.
23. The method of claim 4, further comprising averaging k-space data sampled using CAVA over time to produce a fully sampled k-space.
24. The MRI apparatus of claim 5, wherein the evaluation module creates the uniformly random sampling pattern further comprising creating the first grid as a k-t Cartesian grid with N.sub.s phase-encoding (PE) lines.
25. The apparatus of claim 5, wherein:
26. The MRI apparatus of claim 5, wherein the evaluation module further creates a variable density sampling pattern with N PE lines using the nonlinear stretching operation.
27. The MRI apparatus of claim 26, wherein the variable density sampling pattern has adjustable temporal resolution.
28. The MRI apparatus of claim 5, wherein the evaluation module further averages k-space data sampled using CAVA over time to produce a fully sampled k-space.
Description
BRIEF DESCRIPTION OF THE DRAWINGS
(1) A detailed description of certain aspects of the present disclosure in accordance with various example implementations will now be provided with reference to the accompanying drawings. The drawings form a part hereof and show, by way of illustration, specific implementations and examples. In referring to the drawings, like numerals represent like elements throughout the several figures.
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
(11)
(12)
(13)
(14)
(15)
(16)
(17)
(18)
(19)
DETAILED DESCRIPTION
(20) The present disclosure is directed to a new sampling method called herein Golden-shift Ordered Cartesian sampling (GOC). GOC provides the benefits of VISTA; however, uses a novel computationally efficient methodology. The present disclosure is also directed to a new Cartesian data sampling scheme, called CArtesian sampling with Variable density and Adjustable temporal resolution (CAVA). Distinguishing features of CAVA include (i) temporal resolution that can be adjusted retrospectively, (ii) variable density to preferentially sample the central k-space at higher rate, (iii) maintaining incoherence, and (iv) ensuring a fully sampled time average to facilitate sensitive map estimation. For validation, CAVA was used to collect real-time, free-breathing, phase-contrast MRI (PC-MRI) from ten healthy volunteers. Peak velocity (PV) and stroke volume (SV) in the aorta, recovered from the highly accelerated CAVA data are compared to fully sampled segmented imaging. A t-test found the PV and SV using the real-time CAVA data were not significantly different from fully sampled segmented imaging.
(21) Example Environment
(22)
(23) The MRI apparatus 100 includes a scanner 103 that generates magnetic fields used for the MR examination within a measurement space 104 having a patient table 102. In accordance with the present disclosure, the scanner 103 may include a wide bore 70 cm superconducting magnet having a field strength of approximately 0.5-3.0 Tesla (T).
(24) A controller 106 includes an activation unit 111, a receiver device 112 and an evaluation module 113. During a phase-sensitive flow measurement, MR data are recorded by the receiver device 112, such that MR data are acquired in, e.g., a measurement volume or region 115 that is located inside the body of a patient 105. The MRI apparatus 100 may include an 18-coil array (e.g., arranged as two 3×3 grids); support parallel imaging using SPIRIT, GRAPPA, SENSE, VISTA, AMP, FISTA, SCoRE, and/or Bayesian Inference; and perform analog-to-digital conversion (ADC) at a gantry of the MRI apparatus 100.
(25) An evaluation module 113 prepares the MR data such that they can be graphically presented on a monitor 108 of a computing device 107 and such that images can be displayed. In addition to the graphical presentation of the MR data, a three-dimensional volume segment to be measured can be identified by a user using the computing device 107. The computing device may include a keyboard 109 and a mouse 110. The computing device may include a Xeon central processing unit (CPU) or better; 16 GB of random access memory (RAM); Multi-GPU, K20 or Titan Z reconstruction hardware; support DiCOM 3.0; and support simultaneous scan and reconstruction.
(26) Software for the controller 106 may be loaded into the controller 106 using the computing device 107. Such software may implement a method(s) to process data acquired by the MRI apparatus 100, as described below. It is also possible the computing device 107 to operate such software. Yet further, the software implementing the method(s) of the disclosure may be distributed on removable media 114 so that the software can be read from the removable media 14 by the computing device 107 and be copied either into the controller 106 or operated on the computing device 107 itself.
(27) Generating GOC Sampling
(28) With reference to
(29) At 202, a first step of GOC is to create a uniformly random sampling pattern on a k-t Cartesian grid with Ns phase-encoding (PE) lines. For example, sampling on a smaller grid with N.sub.s PE lines
(30)
may be performed.
(31) At 204, the sampling pattern of the previous frame is circularly rotated. Starting from a randomly selected first PE index, p.sub.s (l), the index of the subsequent PE lines is sequentially advanced by gN.sub.s, with g=(1+√{square root over (5)})/2 (Eq. 1). Next, at 206, it is determined if all frames have been completed. If not, the process returns to 202. If completed, then at 208, a nonlinear stretching operation is applied to create a variable density sampling pattern with N PE lines.
(32) This strategy creates a sampling pattern that has both variable density and adjustable temporal resolution. In addition, the sampling pattern aids in the estimation of coil sensitivity maps when averaged over time. The time averaged CAVA pattern covers all of k-space with an envelope that favors the center of k-space. The samples p.sub.s(i) projected using a nonlinear stretching operation to a larger grid may yield:
(33)
where s controls the relative acceleration rate at the center of k-space compared to the overall acceleration rate, with s>1 ensuring that the sampling density is higher at the center of k-space, α>0 controls the transition from high-density central region to low-density outer region, [.] represents the rounding operation, and constant k is selected such samples ps=1 and ps=Ns are mapped to p=1 and p=N, respectively.
(34)
(35) Generating CAVA Sampling
(36) With reference to
(37) At 602, a first step is to generate a sequence of indices on a smaller grid of size N.sub.s=round(N/s). Starting from a randomly selected first PE index, p.sub.s(1), the PE indices of the subsequent readouts are sequentially advanced by gN.sub.s, yielding:
p.sub.s(i+1)=(p.sub.s(i)+gN.sub.s).sub.N.sub.
where p.sub.s(i) represents the PE index of the i.sup.th sample on the grid of size N.sub.s, g=(√{square root over (5)}−1)/2 is the golden ratio, and •
.sub.N.sub.
(38) Next, at 604, a nonlinear stretching operation is applied to create a variable density sampling pattern with N PE lines. This strategy creates a sampling pattern that has both variable density and adjustable temporal resolution. In addition, the sampling pattern aids in the estimation of coil sensitivity maps when averaged over time. The time averaged CAVA pattern covers all of k-space with an envelope that favors the center of k-space. The samples p.sub.s(i) projected using a nonlinear stretching operation to a larger grid may yield:
(39)
where p.sub.c (i) is the PE index of the i.sup.th sample on the grid with PE lines, s controls the relative acceleration rate at the center of k-space compared to the overall acceleration rate, with s>1 ensuring that the sampling density is higher at the center of k-space, α≥0 controls the transition from high-density central region to low-density outer region, [•] represents the rounding operation, and constant k≥0 is selected such samples p.sub.s=1 and p.sub.s=N are mapped to p=1 and p=N, respectively. In accordance with the disclosure, s=3 and α=3 may be used.
(40)
(41)
(42)
(43) Methods
(44) Data Acquisition from a Pulsatile Flow Phantom
(45) A pulsatile flow phantom experiment was performed to acquire fully sampled PC-MRI data at high temporal resolution. The pulsatile flow was applied to a pipe, with inner diameter of ⅝ inch, placed in a U-shape arrangement inside the magnet bore. Two water bottles supported the pipe, as shown in
(46) The imaging plane was approximately perpendicular to the flow direction. All data were acquired using a 1.5 T (MAGNETOM, Avanto, Siemens Healthcare, Erlangen, Germany) scanner with 18-channel cardiac array. A total of eight fully sampled data sets were collected, each with a different pulsatile waveform set using a programmable pump (MR 5000 by Shelley Medical Imaging Technologies, Toronto, Ontario, Canada). In addition to the shape of the waveform, the duration of the pulsatile cycle was varied between 750 ms and 1,200 ms across the eight acquisitions. The data were collected with prospective triggering using a traditional GRE-based sequence, with flow-encoded and flow-compensated readouts interleaved. To maintain high temporal resolution, segment size was selected to be one, that is, one velocity encoded and one velocity-compensated readouts were recorded in each segment. With TR=5.13 ms, it led to a temporal resolution of 10.26 ms. The other relevant scan parameters were: TE=3.18 ms, flip angle=15 degrees, acquisition matrix=90×144, FOV=178×280 mm, scan time=68 to 108 s, VENC=150 cm/s, slice thickness=8 mm, acquisition bandwidth=503 Hz/pixel, and asymmetric echo off. The data were retrospectively under-sampled with two Cartesian sampling patterns: VISTA15 and CAVA. VISTA is a variable density pseudo-random sampling pattern that is based on Riesz energy minimization. Starting from a random initialization, VISTA iteratively updates the sampling pattern to maximize separation among the neighboring samples. VISTA allows variable density but does not permit retrospective adjustment of temporal resolution. We recently extended VISTA for PC-MRI. 16 Using VISTA and CAVA patterns, the under-sampling process was repeated to simulate six different readout lines per frame (LPF), that is, LPF (k)=4, 5, 6, 8, 10, and 15, leading to temporal resolutions (2×k×TR) of 41.0, 51.3, 61.6, 82.1, 102.6, and 153.9 ms, respectively. To mimic the real-time acquisition, only one readout was selected from each fully sampled frame. For example, the first CAVA index was picked from the first fully sampled frame, the second CAVA index was picked from the second frame, and so forth. For CAVA, only one sampling pattern was used, and different temporal resolutions were realized by binning k consecutive samples into one temporal frame. In contrast, for VISTA, a separate sampling pattern was generated for each temporal resolution. Since exact realizations of CAVA and VISTA depend on the starting point, ps(1), and initial distribution of samples, respectively, the retrospective under-sampling process was repeated for three different randomly selected values of ps(1) for CAVA and three random initializations for VISTA.
(47) Data Acquisition from Healthy Volunteers
(48) To further evaluate the performance of CAVA, PC-MRI data were collected from 10 healthy volunteers. All data were acquired using a 3 T (MAGNETOM, Prisma, Siemens Healthcare, Erlangen, Germany) scanner with 48-channel cardiac array. Each volunteer was imaged with both fully sampled BSI and FRI CAVA acquisition.
(49) The data were collected using a traditional GRE-based sequence, with flow-encoded and flow-compensated readouts interleaved. BSI data were collected with retrospective triggering, while the CAVA data were collected continuously for 10 s. The order of two acquisitions was randomized across volunteers. The imaging plane was prescribed perpendicular to the ascending aorta above the aortic valve. The other relevant scan parameters were: TR=4.21 ms, TE=2.12 ms, flip angle=15 degrees, acquisition matrix=84×128, FOV=250×300 mm, CAVA scan time=10 s, VENC=150 cm/s, slice thickness=6 mm, acquisition bandwidth=698 Hz/pixel, and asymmetric echo on. The data from each CAVA scan were retrospectively re-binned to create six data sets with 4, 5, 6, 8, 10, and 15 LPF (k) for temporal resolutions of 33.7, 42.1, 50.5, 67.4, 84.2, and 126.3 ms, respectively. For BSI, the relevant scan parameters were: TR=4.64 ms, TE=2.47 ms, acquisition matrix=(108−131)×(144−176), FOV=(230−282)×(280−340) mm, VENC=150 cm/s, slice thickness=6 mm, acquisition bandwidth=451 Hz/pixel, isotropic image resolution=1.7-1.9 mm, and temporal resolution=37.1 ms. To highlight the impact of respiration on FRI flow quantification, we additionally collected a CAVA data set from one of the healthy volunteers over a span of 27 s under deep breathing conditions.
(50) Image Recovery and Analysis for Pulsatile Flow Phantom
(51) To reconstruct the accelerated CAVA data sets, the Reconstructing Velocity Encoded MRI with Approximate message passing aLgorithms (ReVEAL) technique was used. ReVEAL exploits wavelet sparsity as well as magnitude and phase similarities across velocity encodings to jointly reconstruct the velocity encoded and velocity compensated images. ReVEAL introduces three tuning parameters. The first parameter, λ controls the regularization strength for wavelet domain sparsity. The second parameter, ω, represents the variance of noise in k-space. The third parameter, σ, represents a measure of magnitude and phase similarity across velocity encodings. The k-space noise variance, ω, was estimated from the outer portions of k-space. The other two parameters were tuned by subjectively evaluating the image quality from a single CAVA data set with k=6. The parameter values were held constant for all pulsatile waveforms, sampling realizations, and temporal resolutions. ReVEAL was implemented in custom Matlab software (Mathworks, Natick, Mass.) and run offline on a Windows 7 PC equipped with a Tesla K40c GPU (Nvidia, Santa Clara, Calif.). The software utilizes GPU acceleration for a runtime of approximately 3 minutes. Open source software for ReVEAL is available online at GitHub (Open source software available at https://github.com/arg-min-x/ReVEAL). Coil sensitivity maps were estimated using ESPIRiT. 19 The reconstructed image series were converted to DICOM, and the cross section of the pipes was then segmented using Segment software. 20 The peak velocity (PV), stroke volume (SV), and recovery signal-to-noise ratio (rSNR) for each waveform and sampling realization were defined as follows:
(52)
(53) where {circumflex over (x)}.sub.i,j.sup.k represents images reconstructed using ReVEAL from under-sampled data, and x.sub.j.sup.k represents fully sampled images with temporal resolution matched to that of {circumflex over (x)}.sub.i,j.sup.k. To generate x.sub.j.sup.k images, k consecutive frames of the original high-resolution images were averaged. We refer to these fully sampled images with matched resolution as FS-MR. To assess variability of SV and PV in images recovered from CAVA and VISTA data sets, normalized values of SV and PV were calculated as a percentage of the reference values, yielding:
(54)
(55) where PVREF.sub.j and SVREF.sub.j represent peak velocity and stroke volume, respectively, calculated from the fully sampled high-resolution reference. For both CAVA and VISTA, t-tests (α=0.05) were performed to compare NSV.sub.i,j.sup.k and NPV.sub.i,j.sup.k to the reference value of 100. The values of rSNR.sub.i,j.sup.k from CAVA and VISTA were also compared to each other using t-test (α=0.05). For each k value, the t-test was performed by aggregating all entries for different i and j values. Due to observable difference in the flow profiles, the two cross sections (ROI-1 and ROI-2 shown in
(56) Image Recovery and Analysis for Healthy Volunteers
(57) The accelerated CAVA data sets were reconstructed using ReVEAL. The reconstructed image series were converted to DICOM, and the aortic cross section was then segmented using Segment software. The PV and SV values for each heartbeat were calculated and computed against the fully sampled BSI reference.
(58) For flow quantification, we define the following parameters:
SV.sub.i,j.sup.k=Stroke Volume: i.sup.th Hearbeat,j.sup.th Volunteer,LPF=k (10)
PV.sub.i,j.sup.k=Peak Velocity: i.sup.th Hearbeat,j.sup.th Volunteer,LPF=k (11)
(59) for i=1, 2, . . . , I j=1, 2, . . . , J k=4, 5, 6, 8, 10, 15
(60) To assess variability of SV and PV in images recovered from CAVA data sets, normalized values of SV and PV were calculated as a percentage of the reference values, yielding:
(61)
where PVREF.sub.j and SVREF.sub.j are the peak velocity and stroke volume for the jth volunteer calculated from the BSI reference. In addition, t-tests (α=0.05) were performed to compare NSV.sub.i,j.sup.k and NPV.sub.i,j.sup.k values from all heartbeats (i) and volunteers (j) to the reference value of 100. The t-test was separately performed for each value of k.
(62) Results
(63) In the first study, PC-MRI data for a pulsatile flow phantom was collected and processed. To create a high-quality reference, eight fully sampled data sets were collected, each with a different flow waveform. For SV and PV quantification, the high temporal resolution images served as a reference, while, for rSNR quantification, FS-MR images served as a reference. The quantitative results are shown in
(64) TABLE 1 shows Flow and rSNR quantification for phantom data:
(65) TABLE-US-00001 TABLE 1 k 4 5 6 8 10 15 rSNR CAVA- 0.011 −0.078 −0.226 −0.355 −0.192 −0.650 VISTA t-test 0 1 1 1 1 1 NSV CAVA- 0.468 0.394 0.026 −0.631 0.860 −0.740 VISTA t-test 0 0 0 1 1 0 CAVA-Ref. −3.156 −1.975 −1.596 −2.268 −0.430 −1.545 t-test 1 1 1 1 1 1 VISTA-Ref. −3.625 −2.369 −1.622 −1.637 −1.290 −0.805 t-test 1 1 1 1 1 0 NPV CAVA- −0.500 −1.114 0.815 −0.941 −2.070 −5.114 VISTA t-test 0 1 0 0 1 1 CAVA-Ref. −6.32 −5.157 −6.002 −8.036 −8.587 −11.61 t-test 1 1 1 1 1 1 VISTA-Ref. −5.827 −4.043 −6.818 −7.095 −6.517 −6.493 t-test 1 1 1 1 1 1
(66) To demonstrate the application of CAVA to real-time PC-MRI, we compared SV and PV from FRI CAVA with that from BSI using data from healthy volunteers.
(67) TABLE 2 Flow quantification results for volunteer data. Each entry represents the mean of the difference between CAVA and BSI reference.
(68) TABLE-US-00002 TABLE 2 (Ref.) k 4 5 6 8 10 15 NSV CAVA- −2.639 −0.982 0.246 1.989 1.036 9.064 Ref. t-test 1 0 0 0 0 1 NPV CAVA- −2.420 −1.996 −2.295 −3.006 −3.919 −6.535 Ref. t-test 1 1 1 1 1 1
(69) The overall performances of CAVA and VISTA were comparable, with VISTA exhibiting slight advantage in terms of rSNR. The advantage of VISTA was within 0.5 dB for k≤10; only at k=15, VISTA had larger than 0.5 dB advantage. For NSV, the quantification from VISTA and CAVA were comparable. Although the NSV values from VISTA and CAVA were statistically different for k=8 and k=10, the largest difference (0.860 at k=10) between the mean NSV values from VISTA and CAVA was less than 1% of the reference. The NSV values from both methods were statistically different (except for VISTA at k=15) from the high-resolution fully sampled reference, but their mean values were within 4% of the reference. For NPV, the quantification of VISTA and CAVA were comparable, with CAVA exhibiting slight underestimation compared to VISTA. The mean difference between CAVA and VISTA was within 2% for k≤8 but grew to over 5% for k=15. Also, the NPV values for both methods were statistically different from the high-resolution fully sampled reference but, except for CAVA at k=15, were within 10% of the reference on average.
(70)
(71) Both magnitude and phase images are shown for six different temporal resolutions, indexed by LPF.
(72) Discussion
(73) For MRI, non-Cartesian trajectories can offer several advantages, including variable density and incoherent artifacts. A variable density pattern maximizes k-space signal-to-noise ratio for a given readout duration, and incoherent artifacts facilitate the application of CS. More recently, golden angle-based radial and spiral trajectories have been proposed for 2D real-time CMR. These acquisition schemes provide the added flexibility of adjusting temporal resolution retrospectively. In contrast, existing Cartesian trajectories are not equipped to combine retrospective adjustment of temporal resolution and variable density. We have proposed a new sampling method, called CAVA, that successfully combines artifact incoherence, variable density that can be parametrically controlled, and ability to vary temporal resolution after the acquisition.
(74) Data collection with CAVA, when combined with ReVEAL reconstruction, can allow highly accelerated real-time imaging. In this work, the combination of CAVA and ReVEAL enabled PC-MRI at acceleration rates as high as R=22.5 (LPF=4). Previously, real-time PC-MRI at such high acceleration rates had only been possible with non-Cartesian acquisitions. 22 Our validation study, although small, supports the feasibility of real-time PC-MRI with Cartesian sampling and retrospective adjustment of temporal resolution.
(75) The rSNR results from the phantom data highlight that performance of CAVA, in terms of overall image quality, is comparable to that of VISTA. At the very low temporal resolution of 153.9 ms, however, rSNR of CAVA is lower by more than 0.5 dB. For flow quantification, the performances of VISTA and CAVA are comparable for k<8, with CAVA exhibiting slightly higher underestimation of NPV for k≥8. The slightly inferior performance of CAVA, especially at lower temporal resolutions, can be attributed to the sampling distribution that is only approximately uniform and can lead to samples that are clustered together as seen in
(76) The quantitative results from volunteer imaging follow the results from phantom imaging. The NSV and NPV values for CAVA closely match the BSI values for k≤10. As is the case for phantom imaging, a slight underestimation is observed in NSV at k=4, which can be attributed to regularization-induced blurring at a high acceleration rate. CAVA also underestimates PV with respect to the high-resolution reference. The underestimation of in vivo data is smaller than the one observed for the phantom data. This can be attributed to the difference in the temporal resolution of the reference. For phantom, the reference has a temporal resolution of 10.26 ms, while it is 37.1 ms for in vivo. For phantom imaging, when the CAVA NPV values are compared to temporal resolution matched reference, FS-MR, the underestimation is reduced to less than 4% for all k. Another disparity between the two studies is that NSV at k=15 is overestimated for the in vivo study, while no such overestimation is observed for the phantom study. We attribute this overestimation to image artifacts (
(77) To highlight the dependence of cardiac output on respiration, additional data from one healthy volunteer was analyzed. The data were collected continuously for 27 s using CAVA sampling. To amplify the effect of inspiration, the volunteer was instructed to breathe deeply during the scan. The images were reconstructed at a resolution of 50.5 ms.
(78)
(79) Experimental Results
(80)
(81) Conclusion
(82) Thus, the present disclosure describes a novel sampling method, CAVA, that exhibits variable density and adjustable temporal resolution. CAVA provides a Cartesian alternative to the golden angle radial sampling and may benefit a wide range of dynamic MRI applications, including real-time PC-MRI.
(83) Although the subject matter has been described in language specific to structural features and/or methodological acts, it is to be understood that the subject matter defined in the appended claims is not necessarily limited to the specific features or acts described above. Rather, the specific features and acts described above are disclosed as example forms of implementing the claims. The present disclosure is capable of other implementations and of being practiced or carried out in various ways.
(84) It must also be noted that, as used in the specification and the appended claims, the singular forms “a,” “an” and “the” include plural referents unless the context clearly dictates otherwise. Ranges may be expressed herein as from “about” or “approximately” one particular value and/or to “about” or “approximately” another particular value. When such a range is expressed, other exemplary implementations include from the one particular value and/or to the other particular value.
(85) By “comprising” or “containing” or “including” is meant that at least the named compound, element, particle, or method step is present in the composition or article or method, but does not exclude the presence of other compounds, materials, particles, method steps, even if the other such compounds, material, particles, method steps have the same function as what is named.
(86) In describing example implementations, terminology will be resorted to for the sake of clarity. It is intended that each term contemplates its broadest meaning as understood by those skilled in the art and includes all technical equivalents that operate in a similar manner to accomplish a similar purpose. It is also to be understood that the mention of one or more steps of a method does not preclude the presence of additional method steps or intervening method steps between those steps expressly identified. Steps of a method may be performed in a different order than those described herein without departing from the scope of the present disclosure. Similarly, it is also to be understood that the mention of one or more components in a device or system does not preclude the presence of additional components or intervening components between those components expressly identified.
(87) As discussed herein, a “subject” (or “patient”) may be any applicable human, animal, or other organism, living or dead, or other biological or molecular structure or chemical environment, and may relate to particular components of the subject, for instance specific organs, tissues, or fluids of a subject, may be in a particular location of the subject, referred to herein as an “area of interest” or a “region of interest” (ROI).