Dual Tube Hybrid Coriolis Mass Flow Sensor
20220074775 · 2022-03-10
Inventors
Cpc classification
G01F1/6847
PHYSICS
G01F1/8472
PHYSICS
International classification
Abstract
A sensor with both Coriolis tube and thermal tube is used to measure the mass flow rate of the fluid using both the Coriolis principle and the thermal method simultaneously. Above certain flow rate, the flow rate is measured by the Coriolis tube and below that flow rate, it is measured by the thermal tube. The Coriolis tube and the thermal tube are arranged parallelly with the common inlet and outlet. Two resistant coils are wound on the thermal tube to do the thermal measurement and a magnetic disk is attached to the Coriolis tube, work together with an excitation coil and two optical sensors to do the Coriolis flow measurement. It takes the advantages of both technologies and create a flow sensor which is super accurate, gas type insensitive, long-term stable and fast responsive without too much pressure drop.
Claims
1. A hybrid mass flow sensor comprising: a Coriolis tube; a thermal tube; a base plate in which the Coriolis tube and thermal tube are installed airtightly; a pair of resistant coils wound on the thermal tube; a magnetic disk attached to the Coriolis tube; an excitation coil installed close to the magnetic disk without contact; a pair of optical sensors surrounding portions of the Coriolis tube without contact, and a PCB mounted on the base plate and anchoring the optical sensors and the excitation coil.
2. The hybrid mass flow sensor of claim 1 which has both Coriolis tube and thermal tube.
3. The hybrid mass flow sensor of claim 1 wherein the Coriolis tube and thermal are installed parallelly and share the same inlet and outlet.
4. The hybrid mass flow sensor of claim 1 wherein the Coriolis tube measures the high-end mass flow rate of the fluid by the Coriolis principle.
5. The hybrid mass flow sensor of claim 1 wherein the thermal tube measures the low-end mass flow rate of the fluid by the thermal principle.
6. The hybrid mass flow sensor of claim 1 wherein the total flow rate is the summation of the flow rates of the Coriolis tube and the thermal tube.
7. The hybrid mass flow sensor of claim 1 wherein the resistant coils on the thermal tube are covered by covers to create a mini stable environment around the resistant coils.
8. The hybrid mass flow sensor of claim 1 wherein the resistant coils are optionally replaced by the flexible film resistant elements or thermal sensitive chips.
9. The hybrid mass flow sensor of claim 1 wherein the excitation coil will excite the Coriolis tube by applying magnetic force on the magnetic disk and make the flow tube doing swing motion.
10. The hybrid mass flow sensor of claim 1 wherein the optical sensors will monitor the twist motion of the Coriolis tube produced by the Coriolis force caused by the medium flowing inside the Coriolis tube.
11. The hybrid mass flow sensor of claim 1 wherein the sensor PCB, firmware and software will treat the signals acquired by the optical sensors and convert them to the mass flow rate.
12. The hybrid mass flow sensor of claim 1 wherein the thermal measurement will be calibrated by the Coriolis measurement.
Description
BRIEF DESCRIPTION OF THE DRAWINGS
[0016]
[0017]
[0018]
[0019]
[0020]
[0021]
[0022]
[0023]
[0024]
[0025]
[0026]
[0027]
DETAILED DESCRIPTION OF THE INVENTION
[0028]
[0029] Thermal tube 2 has a cover 11 in the middle of its top horizontal beam. Inside cover 11, there are two coils 12 and 13 wound on thermal tube 2 (
[0030]
[0031] To create a stable thermal environment, cover 11 made of copper sheet is installed.
[0032]
[0033] They are laser-welded to sensor base 3 at 21 and 22. For Coriolis tube 1, to strengthen the connection and provide a stable support, it will be brazed at 23.
[0034] In this embodiment, Coriolis tube 1 is made of 316L with an ID 2.286 [mm], OD 2.413 [mm] and 194 [mm] long. With a pressure drop 14.7 [psi], the full flow rate for water is close to 110 [kg/h].
[0035] When flowing Nitrogen, the flow rate under 14.7 [psi] pressure drop is about 1 [kg/h], or 57 [SLM], which is shown on
[0036] Depend on the gas and other factors, when the flow rate below certain percentage of the full flow rate (here is 50 [SLM]), the error will be unacceptable. We assume that for Nitrogen this percentage is 5%. For Hydrogen, Helium and other light gases, this percentage may be 10% or higher. As a demonstration, we will use 10% as a divider, below 10%, that is 5 [SLM], the thermal measurement will be used to measure the flow rate and above 10%, Coriolis measurement will be used. From the pressure drop calculation (
[0037] We now need to find out at 0 to 7 [sccm], what kind of output thermal sensor can provide. In the thermal measurement of the mass flow rate, coils 12 and 13 will be heated up. There are different schemes to do the heating, such as constant current, constant temperature or constant temperature drop. In this demonstration, we will use the constant current scheme as shown in
[0038] The coil temperature change will result in its resistance change:
R=R.sub.0[(1+α(T˜T.sub.0)], (1)
where: R and R.sub.0 are the current and the initial coil resistances, respectively;
[0039] α is the temperature coefficient (1/K), for the resistant wire used, this value is around 0.0045;
[0040] T and T.sub.0 are the current and the initial coil temperatures, respectively.
[0041] If we assume that a constant 12-mA current i is applied to both coils and we also assume that the initial resistances for both coils are 308 ohms. Based on these values and Equation (1), the coil resistance change is showing in
[0042] The voltage drops V across each coil can be calculated by
V=i.Math.R, (2)
They are plotted in
[0043] With an addition of thermal tube, the calibration is a little more complicated the one with only Coriolis tube. As shown in
Q.sub.T=Q.sub.c+Q.sub.t, (3)
where Q.sub.T, Q.sub.c, and Q.sub.t are the flow rates of total, Coriolis tube and thermal tube.
[0044] For the Coriolis flow, the flow rate Q.sub.e is a linear function of phase angle difference, that is
Q.sub.C=M.Math.φ, (4)
where φ is the phase angle difference between upstream leg and downstream leg of the sensor; and M is a constant.
[0045] For the thermal tube flow, we can use a two-order polynomial equation to fitting the curve:
Q.sub.t=a+b.Math.φ+c.Math.φ.sup.2, (5)
where a, b and c are fitting coefficients.
[0046] We can combine Equations (3), (4) and (5) together as
Q.sub.T=a+(M+b).Math.φ+c˜φ.sup.2=A+B.Math.φ+C.Math.φ.sup.2, (6)
where constants A, B and C can be decided by the calibration and saved in the PCB RAM for the later use in operation.
[0047] It can be seen from
[0048] Below 10% of the flow rate, we will totally rely on the thermal output. Depend on the curve linearity, different scheme can be used to interpolate the data. For the near-straight-line V-Q curve as shown in
[0049] The calibration will also decide the value φ.sub.10%, that is the phase difference angle when the flow rate is 10% of the full flow. These values will be saved and retrieved during measurement. The procedure will be: first check whether the phase angle is above φ.sub.10, if yes, use Equation (6) to get flow rate; if not, use Equation (7). If the thermal output is not very linear, then more sophisticated linearization and interpolation scheme should be used.
[0050] It is known that the thermal sensor is not very age-stable, that is one of the reasons that people are trying to switch to other measuring technologies or trying to recalibrate the thermal sensor in-line in recent years. With this invention, the thermal sensor output can be recalibrated easily. If it is a controller, the recalibration can be implemented per schedule, such as every 6 months, or even each power-up. For example, at each power-up or scheduled recalibration instant, the controller will control the flow rate flowing from zero up to pass the thermal-Coriolis division flow rate (10% in this demonstration), while passing the 10% flow rate Q.sub.10%, the V.sub.10% will be recorded down and saved in the RAM. If it is a meter, it can also be recalibrated in-line with a little help. For example, at each power up, by using manual control valve or system-controlled valve upstream of the unit to make the flow rate going up from zero to pass the Q.sub.10% and record down the V.sub.10%.
[0051] Thermal sensor usually has better than 1% sensitivity. For the case showing here, it means that the sensitivity is 50 [sccm], 1% of Q.sub.10%, which is 5 [SLM]. The total turn-down ratio will be: 50/50,000=1:1000, an astonishing number.
[0052] As the thermal measurement is under the control of Coriolis measurement, the hybrid sensor will keep the benefit of Coriolis sensor, such as fluid insensitivity, etc.
[0053] In other embodiments, the full flow rates of the Coriolis sensor tubes can be in 10 [kg/h], 1000 [kg/h] or high flow rate levels. The diameters of the Coriolis tubes can be different with the same size of the thermal tube. From the accuracy point of view, higher flow rate units benefit more, because the flow rate of the thermal tube flow will take smaller part of the total flow. For flow rate 1000 [g/h] or less, it may lose too much accuracy due to the error caused by the thermal tube. In such case, one tube doing both Coriolis measurement and thermal measurement functions may be more suitable (in another patent). For some lighter gases, such as Hydrogen and Helium, the thinner thermal tube, such as 0.008″ ID, may be needed. For the thermal tube, instead of heating coils, MEM film sensor may be used.