LINEARITY CONTROL COMPOSITIONS AND METHODS OF USE
20230053128 · 2023-02-16
Inventors
- Rebecca Timmons (Framingham, MA, US)
- Chris Chace (East Freetown, MA, US)
- Louis Bonaparte, II (Plymouth, MA, US)
- Jaclyn Ruhl (Ballwil, CH)
- Lisbeth Blake (Guilford, CT, US)
Cpc classification
International classification
Abstract
The present disclosure provides methods and compositions for preparing linearity control slides to verify linearity of image-based hematology analyzers without the need to make such control slides over and over again for each analyzer each time the analyzer is verified.
Claims
1. A set of linearity control slides for verifying the analytical measuring interval (AMI) or reportable range, or both, of an image-based hematology analyzer comprising at least one, two, three, four, or all five, of: a first set of slides for combined red blood cell (RBC) hemoglobin (HGB) and hematocrit (HCT) components; a second set of slides for a white blood cell (WBC) component; a third set of slides for a platelet (PLT) component a fourth set of slides for a reticulocyte (RET) component; and a fifth set of slides for a nucleated red blood cell (NRBC) component; wherein each set of slides includes three or more slides, wherein each slide in the set contains one of a series of equidistant dilutions within a range of concentrations of the blood component of that set, and wherein each set of slides is fixed and stained, e.g., by a hematology analyzer, e.g., an image-based hematology analyzer.
2. The set of linearity control slides of claim 1, wherein after fixing and staining, the set is stable without refrigeration for at least six months.
3. The set of linearity control slides of claim 1, wherein staining comprises use of an eosin solution or a methylene blue solution.
4. (canceled)
5. The set of linearity control slides of claim 1, comprising at least the first (RBC), second (WBC), and third (PLT) sets of slides, wherein the first, second, and third sets of slides include the following high concentrations and low concentrations of the respective types of cells: TABLE-US-00022 High Low Concentration Concentration Parameter Linearity Samples Linearity Samples WBC ≤0.1 × 10.sup.3/μL ≥400 × 10.sup.3/μL >1.5 × 10.sup.6/μL RBC* PLT ≤10 × 10.sup.3/μL ≥5000 × 10.sup.3/μL >1.5 × 10.sup.6/μL RBC* RBC ≤0.4 × 10.sup.6/μL .sup. ≥8.0 × 10.sup.6/μL
6. A method of verifying linearity of enumeration of a blood component by an image-based hematology analyzer in an extended concentration range, the method comprising: obtaining a set of linearity control slides of claim 1 for a blood component; analyzing the set of linearity control slides with the image-based hematology analyzer using a standard protocol for the blood component; obtaining enumeration values for each blood component from the image-based hematology analyzer; determining linearity of the blood component enumeration using the enumeration values and a known target range for the blood component for the image-based hematology analyzer.
7. The method of claim 6, further comprising obtaining a set of linearity control slides for two or more blood components.
8. A method of producing a set of linearity control slides for verifying linearity on a hematology analyzer with respect to one or more blood components, the method comprising: obtaining a whole blood sample or a blood product sample comprising one or more blood components from a healthy human subject; preparing a high concentration sample from the whole blood sample or blood product sample for a first blood component; preparing a low concentration sample from the whole blood sample or blood product sample for the first blood component; preparing a plurality of equidistant dilutions for the first blood component, wherein the plurality of equidistant dilutions comprises the high concentration sample, the low concentration sample, a first dilution comprising one part of the high concentration sample and one part of the low concentration sample, a second dilution comprising one part of the first dilution and one part of the low concentration sample, and a third dilution comprising one part of the first dilution and one part of the high concentration sample; and preparing a slide for each of the plurality of equidistant dilutions by processing each of the equidistant dilutions for the first blood component in an image-based hematology analyzer to print, fix, and stain the equidistant dilution onto a slide.
9. The method of claim 8, wherein the high concentration sample and the low concentration sample originate from the same blood sample.
10. The method of claim 8, wherein the high concentration sample and the low concentration each comprises a predetermined concentration of the first component.
11. The method of claim 8, wherein the first blood component is selected from a group consisting of red blood cells (RBC), white blood cells (WBC), platelets (PLT), reticulocytes (RET), and nucleated red blood cells (NRBC).
12. The method of claim 8, further comprising conducting each of the recited steps for a second blood component to produce a second set of linearity control slides for the second blood component.
13. The method of claim 12, wherein the second blood component is selected from a group consisting of RBC, WBC, PLT, RET, and NRBC.
14. The method of claim 11, wherein the first blood component is RBCs and wherein the high concentration sample is ≥8.0×10.sup.6/μL of RBCs and the low concentration sample is ≤0.4×10.sup.6/μL of RBCs.
15. (canceled)
16. The method of claim 11, wherein the first blood component is WBCs, and wherein the high concentration sample is ≥400×10.sup.3/μL of WBCs and the low concentration sample is ≤0.1×10.sup.3/μL of WBCs.
17. (canceled)
18. The method of claim 16, wherein the low concentration blood sample of WBCs further comprises greater than 1.5×10.sup.6/μL RBCs.
19. The method of claim 11, wherein the first blood component is PLTs, and wherein the high concentration sample is ≥5000×10.sup.3/μL of PLTs and the low concentration sample is less than or equal to ≤10×10.sup.3/μL of PLTs.
20. (canceled)
21. The method of claim 19, wherein the low concentration sample of PLTs further comprises greater than 1.5×10.sup.6/μL RBCs.
22. The method of claim 8, wherein the plurality of equidistant dilutions is each processed on two, three, four, five, or six slides.
23. The method of claim 8, wherein the plurality of equidistant dilutions further comprises a fourth dilution comprising one part of the low concentration blood sample and one part of the second dilution, a fifth dilution comprising one part of the second dilution and one part the first dilution, a sixth dilution comprising one part of the first dilution and one part of the third dilution, and a seventh dilution comprising one part of the high concentration blood sample and one part of the third dilution.
Description
BRIEF DESCRIPTION OF THE DRAWINGS
[0034]
[0035]
[0036]
[0037]
[0038]
[0039]
[0040]
[0041]
[0042]
[0043]
[0044]
DETAILED DESCRIPTION
[0045] Verification of reportable ranges by linearity testing is often required for one or more measured parameters on image-based hematology analyzers, for example, such as those disclosed in U.S. Pat. Nos. 9,217,695; 10,094,764; and U.S. Pat. No. 10,764,538. To accomplish verification, the methods described herein can be used to prepare linearity control compositions and slides from specialized blood components and/or whole blood. The linearity control compositions include equidistant dilutions of blood components such as WBCs, RBCs, reticulocytes, PLTs, and RETs that are printed, fixed, and stained on a stabilized surface, e.g., a glass or plastic slide, e.g., a microscope slide, followed by digital image-based scanning to determine counts and concentrations for WBC, RBC (including HGB and HCT), PLT and RET.
[0046] A quantitative analytical method is said to be “linear” when measured results from a series of test samples are directly proportional to the known concentration of cellular or blood components in each of the test samples. Thus, a straight line can be used to characterize the relationship between measured test sample results and the known concentrations of the test samples for a predetermined range of concentrations resulting in the AMI or reportable range.
[0047] Before the development of the new methods and compositions described herein, verification protocols with specialized blood products were collected from donors each time the linearity verification protocol is to be executed. There was no commercially available linearity material specific for image-based hematology analyzers. The standard method for a verification protocol was to prepare slides of nine concentrations of a sample. Deviation from linearity was determined from the slides but was highly dependent on sample material and user experience. The slides were discarded once the verification protocol was verified.
[0048] Preparing linearity control compositions and slides using only whole blood does not provide results or measurements that span the manufacturer's linearity claims compared to linearity control compositions and slides prepared from specialized blood products and components as disclosed herein. Furthermore, performance of whole blood procedures requires a trained phlebotomist to draw patient or donor blood at the site where the image-based hematology analyzer is installed. The whole blood linearity range is also limited by available samples (patient population or donors) specific to the installation site. In addition, specialized onsite laboratory equipment and a trained technologist with proven technical competency are required to execute whole blood protocols. Successful linearity test outcomes are highly dependent on technical expertise (see
[0049] In contrast, linearity control slides prepared by the manufacturer from specialized blood products to create high, low, and intermediate concentration samples as described herein are cost-effective, efficient, and fulfill the manufacture's claims for linearity. The linearity control slides provide reproducible results, are stable up to 6 months, and can be used repeatedly on one or several image-based hematology systems.
General Methodology
[0050] The methods described herein provide for linearity control compositions, i.e., linearity control slides, which can be used to verify the reportable range of an image-based hematology analyzer. High and low concentration samples are prepared to specific concentrations for the specific blood component being tested. Equidistant intermediate dilutions are then prepared from the high and low concentration samples representing the AMI of the blood components. All samples are printed, fixed, and stained on slides to generate the linearity control slides, typically with multiple replicates, e.g., 6, 7, 8, or more replicates. The linearity slides can be used repeatedly on different image-based hematology analyzers resulting in reproducible linear range results for verifying the linearity.
[0051] Methods of Sample Preparation for Linearity Control Slides
[0052] Blood samples and specialized blood products are prepared to determine and/or verify linearity of the image-based hematology system for each blood component, e.g., WBC, RBC, PLT, and RET, and for calculated parameters, e.g., HGB, and HCT.
[0053] High concentration blood samples and low concentration blood samples are prepared generally as shown in
[0054] In some embodiments, if the result of the high concentration linearity sample is less than the target value, the sample can be further processed using the steps described above. Table 1 shows the target values for the low concentration linearity samples and the high concentration linearity samples for RBC, WBC, and PLT parameters.
TABLE-US-00002 TABLE 1 Target Values for Low and High Concentration Control Linearity Slides for Parameters: WBC, PLT, and RBC (*Required for Imaging) Low Concentration High Concentration Parameter Linearity Samples Linearity Samples WBC ≤0.1 × 10.sup.3/μL ≥400 × 10.sup.3/μL >1.5 × 10.sup.6/μL RBC* PLT ≤10 × 10.sup.3/μL ≥5000 × 10.sup.3/μL >1.5 × 10.sup.6/μL RBC* RBC ≤0.4 × 10.sup.6/μL .sup. ≥8.0 × 10.sup.6/μL
[0055] The high concentration and low concentration samples are used to prepare equidistant dilutions where there are three to seven intermediate dilutions between the high and low concentration samples resulting in five to nine total equidistant concentrations.
[0056] First, equal parts of the high and low concentration blood samples produce the middle intermediate concentration (Tube 5). Second, further dilutions between the low concentration and the middle concentration or the high concentration and the middle concentration result in intermediate concentrations (Tubes 3 and 7). Third, the intermediate concentrations can further be diluted with the middle concentration (Tubes 4 and 6), or with the high concentration (to produce Tube 8) or with the low concentration (to produce Tube 2). All dilutions are equal parts of the beginning concentrations.
[0057] To prepare the new linearity control slides (shown in
[0062] In further detail, each sample is printed in a monolayer on a slide, e.g., glass or plastic slide, (e.g., microscope slide), which are then fixed and stained. After staining, the sample slides are digitally imaged and the blood components are counted and compared to known values.
[0063] In some embodiments, the equidistant concentrations are processed either in open tube mode or closed tube mode (i.e., rack mode). Open tube mode refers to when the equidistant dilutions are processed using the open tube probe to aspirate the blood from the tube. Closed tube mode refers to when the equidistant dilutions are processed automatically by the image-based hematology analyzer automatically in closed tubes. For example, the equidistant dilutions can each be automatically retrieved from a tube rack and aspirated by the system using a sample probe that is designed or configured to pierce through a rubber stopper on each closed tube.
[0064] In some embodiments, the equidistant concentrations are processed onto multiple different slides to prepare replicates, e.g., two, three, four, five, six, or more replicates per concentration. In one example, each of the 9 prepared equidistant concentrations is printed, fixed, and stained on glass slides 6 times (6 replicates) generating a total of 54 linearity control slides for a specific blood component, e.g., RBC. The linearity control slides are analyzed on the image-based hematology analyzer and compared to the known lower and upper target values (range of AMI) as well as to the maximum allowable deviation from linearity as shown in Table 2. Results are analyzed according to the standards described in Clinical Laboratory Standards Institute: H26-A2 Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard—Second Edition, June 2010 and EP06-A Evaluation of the Linearity of Quantitative Measurement Procedures; A Statistical Approach Approved Guideline—June 2003.
TABLE-US-00003 TABLE 2 Acceptable Lower Limit and Upper Limit for WBC, PLT, RBC, HGB, and HCT and Maximum Allowable Deviation from Linearity (manufacturer's claims) using the specialized blood products as described in the Verification Protocols Acceptable Acceptable Lower Limit Upper Limit for for Analytical Analytical Maximum Allowable Parameter Measuring Measuring Deviation from [Units] Interval Interval Linearity WBC [10.sup.3/μL] ≤0.1 × 10.sup.3/μL ≥400 × 10.sup.3/μL ±0.50 10.sup.3/μL for WBC <3.33 × 10.sup.3/μL 15% for WBC ≥3.33 × 10.sup.3/μL PLT [10.sup.3/μL] ≤10 × 10.sup.3/μL ≥5,000 × 10.sup.3/μL ±20 PLT 10.sup.3/μL for PLT <133 × 10.sup.3/μL 15% for PLT ≥133 × 10.sup.3/μL RBC [10.sup.6/μL] ≤0.4 × 10.sup.6/μL ≥8.00 × 10.sup.6/μL ±0.20 10.sup.6/μL for RBC <2.00 × 10.sup.6/μL 10% for RBC ≥2.00 × 10.sup.6/μL HGB [g/dL] ≤1.2 g/dL ≥24 g/dL ±0.50 g/dL .sup. for HGB <5 g/dL 10% for HGB ≥5 g/dL HCT [%] ≤3.60% ≥72% 1% for HCT <10% HCT 10% for HCT ≥10% HCT
[0065] In some embodiments, the linearity control slides are stored without refrigeration at room temperature. Further, the linearity slides can be stored without refrigeration in the dark for up to six months.
[0066] Methods of Using the Linearity Control Slides
[0067] The linearity control slides, as described herein, are used to verify the reportable range on image-based hematology analyzers. The preserved glass slides from each linearity series is saved and can be re-used (re-imaged) on one or more additional Cobas® m 511 integrated hematology analyzers that have been temporarily re-configured to bypass printing and staining, thus allowing the Cobas® m 511 system to perform digital identification and counting of cells on the preserved glass slide. Recovered counts and concentrations of the blood component on each linearity control slide, obtained for example, on the Cobas® m 511 system, are compared to the known value on each respective linearity control slide. Linearity of the system is determined by the relationship of the recovered counts and concentrations and known values.
[0068] In some embodiments, the linearity control slides can be analyzed repeatedly on the same image-based hematology analyzer or on different image-based hematology analyzers.
[0069] Linearity control slides can be used to verify the reportable range on an image-based hematology analyzer when the system is newly installed, re-installed, after receiving maintenance, major repair, troubleshooting, or at any interval determined by the end-user.
[0070] Software Used in Cobas® m 511 System
[0071] The Cobas® m 511 system includes three major software components: viewing station software, analyzer software, and the image analysis application.
[0072] Viewing Station Software: The viewing station software provides the user interface to the Cobas® m 511 analyzer, including: managing system configuration and consumables; reporting analyzer and sample processing status; troubleshooting and error recover; quality control processing and results; reviewing sample results; classifying/reclassifying cells; and evaluating cell morphology.
[0073] Analyzer Software: The analyzer software runs the analyzer computer and manages all hardware elements, sample processing, data management, and viewing station communications.
[0074] Image Analysis Application: The image analysis application runs on the analyzer computer and communicates with the cameras and motion-control hardware to acquire images at both the low- and high-magnification imaging stations. The program identifies and measures cells in the images, and it reports the cell counts and measurements to the analyzer software. It also saves images for display on the viewing station.
EXAMPLES
[0075] The invention is further described in the following examples, which do not limit the scope of the invention described in the claims.
[0076] The examples disclosed below describe methods of preparing high and low concentration samples, which are used to generate the intermediate equidistant dilutions. Each blood cell type has a different protocol to meet the linearity requirements of the hematology analyzer. The equidistant dilutions are used to prepare the linearity control slides. The linearity control slides can be used multiple times and provide reproducibly accurate readings.
Linearity Protocols Prepared from Specialized Blood Products
Example 1: White Blood Cell Analytical Measuring Interval
[0077] The purpose of Example 1 was to establish that the AMI for an image-based hematology analyzer for the parameter WBC meets the linearity requirements.
[0078] To study WBC AMI, a fresh leukapheresis sample and three corresponding whole blood samples were obtained from a single volunteer donor. Approximately 120-130 mL of the leukapheresis sample was collected in anticoagulant citrate dextrose solution A (ACD-A). Each of the three 10 mL whole blood samples were collected in K2-EDTA sample collection tubes. The samples were concentrated and diluted (as necessary) to create a high concentration sample, low concentration sample, and seven intermediate equidistant concentration samples resulting in a unique linearity series (as shown in
[0079] High Concentration WBC Sample
[0080] The high concentration WBC sample was prepared to meet the WBC target concentration of ≥400×10.sup.3/μL. The leukapheresis bag was mixed for approximately five minutes and the contents transferred to conical tubes, approximately 10 mL per tube. A baseline sample was retained, and processed on another SYSMEX® Hematology Analyzer to obtain the WBC concentration.
[0081] If the WBC baseline result was less than the target value of ≥400×10.sup.3/μL, then the remaining leukapheresis sample was centrifuged for twelve minutes at 1,000 rpms creating a WBC pellet. The supernatant was removed, leaving approximately 2 mL in each tube. The WBC pellet and supernatant were re-suspended and combined. If platelet aggregates were present, then the sample was discarded and the protocol was repeated with a new leukapheresis sample. A 1:10 dilution in STE buffer (Sodium Chloride-Tris-EDTA, 1× Solution, pH 8.0) of the concentrated sample was processed on another SYSMEX® Hematology Analyzer and the WBC concentration was obtained.
[0082] If the concentrated sample was less than the target value of ≥400×10.sup.3/μL, the sample was re-centrifuged and additional supernatant was removed. The sample was again resuspended in STE buffer and a 1:10 dilution was analyzed on another SYSMEX® Hematology Analyzer for the WBC concentration. When the target value was met, the sample was labeled as the “high concentration” sample.
[0083] Low Concentration WBC Sample
[0084] The low concentration WBC sample was prepared to meet two target values, ≤0.1×10.sup.3/μL for WBC and ≥1.5×10.sup.6/μL for RBC. RBCs were added to meet the minimum number of cells necessary for the Roche image-based hematology analyzer (Cobas® m 511 system) to initiate sample processing and facilitate imaging. The three whole blood donor samples were centrifuged for twelve minutes at 4,000 rpms. The buffy coat layer was removed and the remaining contents were transferred to one tube. The sample was mixed and processed on another SYSMEX® Hematology Analyzer to obtain a WBC and a RBC result.
[0085] If the low concentration WBC target value for the sample was not met, then an equal volume of STE buffer was added and the sample was centrifuged for twelve minutes at 4,000 rpms. An additional buffy coat layer was removed and the sample was resuspended and analyzed on another SYSMEX® Hematology Analyzer. When the WBC and RBC target values were met, the sample was labeled as the “low concentration” sample.
[0086] Linearity Series Preparation
[0087] An equidistant dilution series between the high concentration sample and the low concentration sample was prepared to create the linearity series (see
[0088] Tube 5 was prepared combining one part from Tube 1 and one part from Tube 9. Tube 3 was prepared combining one part from Tube 1 and one part from Tube 5. Tube 7 was prepared combining one part from Tube 5 and one part from Tube 9. Tube 2 was prepared combining one part from Tube 1 and one part from Tube 3. Tube 8 was prepared combining one part from Tube 7 and one part from Tube 9. Tube 4 was prepared combining one part from Tube 3 and one part from Tube 5. Tube 6 was prepared combining one part from Tube 5 and one part from Tube 7.
[0089] Sample Processing
[0090] Each dilution (concentration) was thoroughly mixed to prepare for processing by the Cobas® m 511 system using standard steps. Each sample was printed on a glass microscope slide in a monolayer and then automatically fixed and stained by the analyzer. The analyzer then automatically performed digital identification and counting of WBCs. Automated results for each dilution were compared to target values of each dilution to determine linearity of the samples.
[0091] Results for WBC Linearity Samples
[0092] Fifty-four WBC linearity slides (nine concentrations, six replicates per concentration), were processed on separate Cobas® m 511 systems and showed reproducible results. The nine equidistant data points spanning the complete WBC linear range (≤0.1-400.0×10.sup.3/μL) are shown in circles (
TABLE-US-00004 TABLE 3 Deviation from Linearity for WBC Linearity Slides in FIG. 4A Deviation.from. Target Linearity Deviation [10.sup.3/μL] [10.sup.3/μL] [%] Criteria Pass/Fail 6e−02 3.8e−03 6.3 0.5 [10.sup.3/μL] OK 57.62 −2.88 −5.0 15 [%] OK 115.17 −4.24 −3.7 15 [%] OK 172.72 −4.24 −2.5 15 [%] OK 230.28 −3.1 −1.3 15 [%] OK 287.83 −0.99 −0.3 15 [%] OK 345.39 1.9 0.6 15 [%] OK 402.94 5.38 1.3 15 [%] OK 460.5 9.25 2.0 15 [%] OK
TABLE-US-00005 TABLE 4 Deviation from Linearity for WBC Linearity Slides Analyzed on Separate cobas ® m 511 System Deviation.from. Target Linearity Deviation [10.sup.3/μL] [10.sup.3/μL] [%] Criteria Pass/Fail 4e−02 6.3e−03 15.8 0.5 [10.sup.3/μL] OK 70.17 −7.91 −11.3 15 [%] OK 140.31 −8.47 −6.0 15 [%] OK 210.44 −4.34 −2.1 15 [%] OK 280.57 1.85 0.7 15 [%] OK 350.7 7.44 2.1 15 [%] OK 420.83 9.77 2.3 15 [%] OK 490.97 6.2 1.3 15 [%] OK 561.1 −5.92 −1.1 15 [%] OK
[0093] Tables 3 and 4 illustrate the deviation from linearity for WBC linearity slides that were processed and analyzed on separate Cobas® m 511 systems. Preparation of the blood samples and slides was as described above. Deviation from linearity was within acceptable limits for both sets of slides and systems that were used.
Example 2: Red Blood Cell, Hemoglobin, and Hematocrit Analytical Measuring Interval
[0094] The purpose of Example 2 was to establish that the AMI for an image-based hematology analyzer for combined parameters RBC-HGB-HCT meet the linearity requirements.
[0095] To study RBC-HGB-HCT AMI, ten-10 mL whole blood samples, collected in K2-EDTA tubes, were obtained from a single volunteer donor. One of the ten tubes was processed on another SYSMEX® Hematology Analyzer to obtain the complete blood count (CBC). The samples were concentrated and diluted (as necessary) to create a high concentration sample, low concentration sample, and seven intermediate equidistant concentration samples resulting in a unique linearity series (as shown in
[0096] High and Low Concentration RBC Sample
[0097] The high concentration blood sample was prepared to meet the target concentration of RBC ≥8.00×10.sup.6/μL. One of the 10 mL K2-EDTA donor sample tubes was labeled as “baseline” and processed on another SYSMEX® Hematology Analyzer to obtain a baseline CBC result. This tube was then set aside for later use. The remaining nine donor sample tubes were centrifuged for ten minutes at 4,000 rpm. During this centrifugation step, two 50 mL conical tubes were obtained and labeled as “plasma” and “packed RBC's” respectively.
[0098] After centrifugation was complete, the plasma from each of the nine (9) donor sample tubes was transferred into a 50 mL conical tube labeled “plasma.” Then the packed RBCs were transferred into a fresh 50 mL conical tube labeled “packed RBC's” and thoroughly mixed. A 500 μL aliquot of “packed RBCs” was transferred into a red/grey-top tube. Into this same red/grey-top tube a 500 μL aliquot from the “plasma” tube was added, thus creating a 1:1 dilution. The 1:1 dilution tube was processed on another SYSMEX® Hematology Analyzer to obtain RBC, HGB, and HCT results. If the RBC result met the target value of ≥8.00×10.sup.6/μL, the tube was labeled as “high concentration” sample.
[0099] The low concentration blood sample was prepared by diluting the donor sample previously set aside, to meet the target concentration of ≤0.40×10.sup.6/μL. When the target value was met, the sample was labeled as the “low concentration” sample.
[0100] Linearity Series Preparation
[0101] Equidistant dilutions were prepared using the high and low concentration samples as described in Example 1 and as shown in
[0102] Sample Processing
[0103] Each dilution (concentration) was thoroughly mixed to prepare for processing by the Cobas® m 511 system. Each sample was printed on a glass microscope slide in a monolayer, and then automatically fixed and stained by the analyzer. The analyzer then automatically performed digital identification and counting of RBCs and evaluated RBCs—for cellular morphology and content (e.g., HGB and HCT). Automated results for each dilution were compared to target values of each dilution to determine linearity of the samples.
[0104] Results for RBC Linearity Samples
[0105] Fifty-four RBC linearity slides (nine concentrations, six replicates per concentration), were processed on three separate Cobas® m 511 systems and showed reproducible results. Nine equidistant data points spanned the linear ranges:
[0106] ≤0.4-≥8.00×10.sup.6/μL for RBC,
[0107] ≤1.2-≥24.0 g/dL for HGB, and
[0108] ≤3.6-≥72.0% for HCT.
TABLE-US-00006 TABLE 5 Deviation from Linearity for RBC Linearity Slides Deviation.from. Target Linearity Deviation [10.sup.6/μL] [10.sup.6/μL] [%] Criteria Pass/Fail 0.37 2.6e−02 7.2 0.2 [10.sup.6/μL] OK 1.36 −2.8e−02 −2.1 0.2 [10.sup.6/μL] OK 2.34 −3.5e−02 −1.5 10 [%] OK 3.33 −1.1e−02 −0.3 10 [%] OK 4.32 2.9e−02 0.7 10 [%] OK 5.3 6.9e−02 1.3 10 [%] OK 6.29 9.2e−02 1.5 10 [%] OK 7.27 8.3e−02 1.1 10 [%] OK 8.26 2.6e−02 0.3 10 [%] OK
TABLE-US-00007 TABLE 6 Deviation from Linearity for HGB Linearity Slides Deviation.from. Target Linearity Deviation [g/dL] [g/dL] [%] Criteria Pass/Fail 1.1 0.11 9.7 0.5 [g/dL] OK 3.99 −1.8e−02 −0.5 0.5 [g/dL] OK 6.88 −7.8e−02 −1.1 10 [%] OK 9.76 −8.4e−02 −0.9 10 [%] OK 12.65 −5.0e−02 −0.4 10 [%] OK 15.54 1.3e−02 0.1 10 [%] OK 18.43 9.1e−02 0.5 10 [%] OK 21.31 0.17 0.8 10 [%] OK 24.2 0.25 1.0 10 [%] OK
TABLE-US-00008 TABLE 7 Deviation from Linearity for HCT Linearity Slides Deviation.from. Deviation Target Linearity [%] Criteria Pass/Fail 3.2 5.6e−02 1.8 1 OK 11.82 −0.29 −2.4 10 [%] OK 20.45 −0.34 −1.7 10 [%] OK 29.07 −0.19 −0.7 10 [%] OK 37.7 5.3e−02 0.1 10 [%] OK 46.33 0.31 0.7 10 [%] OK 54.95 0.48 0.9 10 [%] OK 63.58 0.46 0.7 10 [%] OK 72.2 0.17 0.2 10 [%] OK
[0109] Slides were preparation as described above. Slides were measured on a Cobas® m 511 system for RBC, HGB, and HCT, using standard steps. The mean value of six replicates for each of the nine equidistant samples fell within acceptable limits for deviation from linearity for RBC (Table 5), HGB (Table 6), and HCT (Table 7).
Example 3: Platelet Analytical Measuring Interval
[0110] The purpose of Example 3 was to establish that the AMI for an image-based hematology analyzer for the parameter PLT meets the linearity requirements.
[0111] To study PLT AMI, a single-unit PLT apheresis sample was obtained from a single donor. Each single-unit of this component donor contained a minimum of 150 mL to a maximum of 400 mL product collected in ACD-A anticoagulant. Also required was 80-100 mL pooled ABO/Rh compatible whole blood that was inspected to ensure that clots and aggregates were not present. The samples were concentrated and diluted (as necessary) to create a high concentration sample, low concentration sample, and seven intermediate equidistant concentration samples resulting in a unique linearity series (as shown in
[0112] High and Low Concentration PLT Sample
[0113] The high concentration sample was prepared to meet the target value of ≥5,000×10.sup.3/μL for PLT. A conical tube with 4 mL STE buffer was prepared with 100 mg/mL K.sub.2-EDTA. The PLT apheresis sample was transferred into the conical tube and mixed with RBCs and buffer solutions. The conical tube was centrifuged for twelve minutes at 4,000 rpm. The supernatant was removed, the PLT pellet was re-suspended, and a dilution of the remaining solution for PLT and RBCs was analyzed on a SYSMEX® Hematology Analyzer to obtain results for the PLTs in the sample.
[0114] The low concentration blood sample was prepared to meet the target value of ≤10×10.sup.3/μL for PLT and ≥1.5×10.sup.6/μL for RBC. All ABO/Rh compatible whole blood samples were centrifuged for twelve minutes at 4,000 rpm. The buffy coat layer was removed. The remaining RBCs were then washed using STE buffer. The sample was then tested on a SYSMEX® Hematology Analyzer to assign target values for PLT and RBC.
[0115] Linearity Series Preparation
[0116] Equidistant dilutions were prepared using the high and low concentration samples as described in Example 1 and as shown in
[0117] Sample Processing
[0118] Each dilution (concentration) was thoroughly mixed to prepare for processing by the Cobas® m 511 system. Each sample was printed on a glass microscope slide in a monolayer, and then automatically fixed and stained by the analyzer. The analyzer then automatically performed digital identification and counting of PLTs. Automated results for each dilution were compared to target values of each dilution to determine linearity of the samples.
[0119] Results for PLT Linearity Samples
[0120] Fifty-four PLT linearity slides (nine concentrations, six replicates per concentration), were processed on separate Cobas® m 511 systems using standard steps and showed reproducible results. Nine equidistant data points spanning the complete PLT linear range ≤10×10.sup.3/μL and ≥5,000×10.sup.3/μL) were used.
TABLE-US-00009 TABLE 8 Deviation from Linearity for PLT Linearity Slides Deviation.from. Target Linearity Deviation [10.sup.3/μL] [10.sup.3/μL] [%] Criteria Pass/Fail 5 −6.9e−03 −0.1 20 [10.sup.3/μL] OK 631.88 12.57 2.0 15 [%] OK 1258.75 24.92 2.0 15 [%] OK 1885.62 31.68 1.7 15 [%] OK 2512.5 27.47 1.1 15 [%] OK 3139.38 6.9 0.2 15 [%] OK 3766.25 −35.39 −0.9 15 [%] OK 4393.12 −104.8 −2.4 15 [%] OK 5020 −206.7 −4.1 15 [%] OK
[0121] The mean value of six replicates for each of the nine equidistant samples fell within acceptable limits as shown in the deviation from linearity graph (Table 8).
[0122] Linearity Protocols Prepared from Whole Blood Table 9 shows the target values for the low concentration linearity samples and high concentration linearity samples for parameters WBC, PLT, and RBC/HGB using whole blood protocols.
TABLE-US-00010 TABLE 9 Target Values for Low and High Concentration Linearity Samples for Parameters: WBC, PLT, RBC and HGB using Whole Blood Samples and Protocols (*Required for Imaging) Parameter Low Concentration High Concentration [Units] Linearity Samples Linearity Samples WBC [10.sup.3/μL] ≤0.7 × 10.sup.3/μL ≥99 × 10.sup.3/μL 1.5 − 0.3 × 10.sup.6/μL RBC* PLT [10.sup.3/μL] ≥0 × 10.sup.3/μL ≥999 × 10.sup.3/μL 1.5 − 0.3 × 10.sup.6/μL RBC* RBC [10.sup.6/μL] ≤0.3 × 10.sup.6/μL ≥7.0 × 10.sup.6/μL HGB [g/dL] 2.0 g/dL 24 g/dL
[0123] The high concentration sample and the low concentration sample are used to prepare equidistant dilutions similar to as previously described, however, there are only three intermediate dilutions between the high and low concentration samples resulting in a total of five equidistant concentrations. Similar to the
[0124] Each sample is printed in a monolayer on a slide, e.g., glass or plastic slide, (e.g., microscope slide) and are fixed and stained. After staining, the sample slides are automatically imaged and the blood components are counted and compared to known values.
[0125] In some embodiments, the equidistant concentrations are processed either in open tube mode or closed tube mode (i.e., rack mode). In the following examples, each of the 5 prepared equidistant concentrations was printed, fixed, and stained on glass slides 6 times (6 replicates) generating a total of 30 linearity measurements for each blood parameter. The linearity measurements were analyzed on the image-based hematology analyzer according to the standards described in Clinical Laboratory Standards Institute, H26-A2 Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard—Second Edition, June 2010, and compared to the known lower and upper target values (truncated range of AMI) as well as to the maximum allowable deviation from linearity as shown in Tables 10 (China Food and Drug Administration (CFDA)) and 11 (United States Food and Drug Administration (USFDA)).
TABLE-US-00011 TABLE 10 Representative (CFDA/truncated) Acceptable Lower and Upper Limits for WBC, PLT, RBC, and HGB AMI and Maximum Allowable Deviation from Linearity Using Whole Blood Protocols Acceptable Lower Limit Acceptable for Upper Limit Analytical for Analytical Parameter Measuring Measuring Maximum Allowable [Units] Interval Interval Deviation from Linearity WBC 1.0 × 10.sup.3/μL 99 × 10.sup.3/μL ±0.50 10.sup.3/μL for WBC: 1.0 − 10 × 10.sup.3/μL [10.sup.3/μL] 5% for WBC: 10.1 − 99.9 × 10.sup.3/μL PLT 20 × 10.sup.3/μL 999 × 10.sup.3/μL ±10 PLT 10.sup.3/μL for PLT: 20 − 100 × 10.sup.3/μL [10.sup.3/μL] 10% for PLT: 101 − 999 × 10.sup.3/μL RBC 0.3 × 10.sup.6/μL 7.00 × 10.sup.6/μL ±0.05 10.sup.6/μL for RBC: 0.3 − 1.0 × 10.sup.6/μL [10.sup.6/μL] 5% for RBC: 1.01 − 7.0 × 10.sup.6/μL HGB [g/dL] 2.0 g/dL 24 g/dL ±0.20 g/dL for HGB: 2 − 7 g/dL 3% for HGB: 7.1 − 24.0 g/dL
TABLE-US-00012 TABLE 11 Representative (USFDA/truncated) Acceptable Lower and Upper Limits for WBC, PLT, RBC, and HGB AMI and Maximum Allowable Deviation from Linearity Using Whole Blood Protocols Acceptable Lower Acceptable Upper Limit for Limit for Maximum Allowable Analytical Analytical Deviation from Parameter Measuring Measuring Linearity per CLIA [Units] Interval Interval Requirements WBC [10.sup.3/μL] ≤0.07 × 10.sup.3/μL ≥99 × 10.sup.3/μL 15% PLT [10.sup.3/μL] ≥0 × 10.sup.3/μL ≥999 × 10.sup.3/μL 25% RBC [10.sup.6/μL] .sup. 0.3 × 10.sup.6/μL 7.00 × 10.sup.6/μL 6% HGB [g/dL] 2.0 g/dL 24 g/dL 7%
Example 4: Comparison of WBC Linearity Using the Whole Blood Protocol
[0126] Without specialized samples and sample preparation, as described in Example 1, the entire AMI for WBC cannot be achieved. The purpose of Example 4 was to demonstrate the reduced (truncated) range of AMI for an image-based hematology analyzer when using samples prepared with a whole blood implementation protocol for WBC.
[0127] To study whole blood WBC AMI, a minimum of fifteen (15), 10 mL K2/K3-EDTA tubes from a single donor were obtained via standard venipuncture and mixed thoroughly. Using a wooden applicator stick, all tubes were inspected for clots. If any sample was clotted, it was discarded and recollected. The donor samples were concentrated and diluted (as necessary) to create high and low concentration linearity samples.
[0128] Baseline WBC Sample
[0129] One of the 15 tubes of donor blood was labeled “Baseline.” The Baseline tube was processed on the Cobas® m 511 system using standard steps, and set aside for future use. It is recommended to collect a donor sample with a baseline WBC result of ≥8.0×10.sup.3/μL.
[0130] High Concentration WBC Sample
[0131] The high concentration WBC sample was prepared to meet the WBC target value of ≥99×10.sup.3/μL. To prepare the high concentration sample, the remaining donor tubes were centrifuged for ten (10) minutes at 5000 RPM. After centrifugation was complete, ˜⅔ of the plasma was removed and transferred from each donor tube into a single 15 mL conical tube. This conical tube was labeled as “Plasma” and retain for later use. If disruption of the buffy-coat layer or RBC layer occurred, centrifugation steps were repeated.
[0132] After removing plasma, the buffy coat layer from each donor tube was removed and transferred into a single 15 mL conical tube. It is acceptable if some RBCs were pipetted from the buffy coat layer. This conical tube was labeled “high concentration” sample. The donor tubes (containing mostly RBCs) were retained for use in preparing the low concentration WBC sample.
[0133] The high concentration sample was vortexed for five (5) seconds and processed on the Cobas® m 511 system using standard steps. Result were reviewed to verify WBC was ≥99.0×10.sup.3/μL. If unable to achieve the high concentration WBC target value, a new donor sample with a baseline WBC result ≥8.0×10.sup.3/μL was collected.
[0134] Low Concentration WBC Sample
[0135] The low concentration WBC sample was prepared to meet two target values, ≤1.0×10.sup.3/μL for WBC and ≥0.30×10.sup.6/μL for RBC. RBCs were added to meet the minimum number of cells necessary for the Cobas® m 511 system to initiate sample processing and facilitate imaging.
[0136] To prepare the low concentration sample, an additional layer of RBCs (approximately 500 μL) was removed from three of the donor tubes retained as described above and transferred into a new 15 mL conical.
[0137] Plasma was transferred from the Plasma tube into the 15 mL conical tube containing the RBCs. This tube was labeled “low concentration.” The low concentration sample was thoroughly mixed and processed on the Cobas® m 511 system using standard steps. Results were reviewed to verify that WBC was ≤1.0×10.sup.3/μL and RBC was ≥0.30×10.sup.6/μL target values were achieved.
[0138] Linearity Series Preparation
[0139] An equidistant linearity series was prepared between the high concentration sample and the low concentration sample. Tubes 1 through 5 were labeled, where Tube 1 contained the low concentration sample and Tube 5 contained the high concentration sample. Tube 3 was prepared combining one part from Tube 1 and one part from Tube 5. Tube 2 was prepared combining one part from Tube 1 and one part from Tube 3. Tube 4 was prepared combining one part from Tube 3 and one part from Tube 5.
[0140] Sample Processing
[0141] Each tube of the linearity series was prepared for processing by a Cobas® m 511 system. Each sample was printed on a glass microscope slide in a monolayer, and then automatically fixed and stained by the analyzer. The analyzer also automatically performed digital identification and counting of WBCs. Automated WBC counts for each dilution were compared to target values of each dilution to determine linearity of the samples.
[0142] Results for WBC Linearity Samples
[0143] Thirty WBC linearity results (five concentrations, six replicates per concentration), were obtained on the Cobas® m 511 system. The five equidistant data points spanning a truncated WBC linear range (≤1.0-99.0×10.sup.3/μL) are shown in circles (
TABLE-US-00013 TABLE 12 Deviation from Linearity for WBC Using Whole Blood WBC Protocol in FIG. 5A Target Mean SD CV [10.sup.3/μL] N [10.sup.3/μL] [10.sup.3/μL] [%] Var Weight 0.37 6.00 0.38 2.4e-02 6.3 5.9e−04 7.3 36.28 6.00 40.34 0.66 1.6 0.43 7.6e−04 72.19 6.00 70.76 0.77 1.1 0.59 1.9e−04 108.09 6.00 109.5 1.16 1.1 1.35 8.6e−05 144.00 6.00 144.1 1.30 0.9 1.70 4.8e−05
[0144] Comparison of
Example 5: Comparison of RBC Linearity Using the Whole Blood Protocol
[0145] Without specialized samples and sample preparation, as described in Example 2, the entire AMI for RBC-HGB-HCT cannot be achieved. The purpose of Example 5 was to demonstrate the reduced (truncated) range of AMI for an image-based hematology analyzer when using samples prepared with a whole blood implementation protocol for RBC. Separate whole blood protocols can be developed for the analysis of HGB and HCT linearity.
[0146] To study whole blood RBC AMI, three (3), 10 mL K2/K3-EDTA tubes from a single donor were obtained via standard venipuncture and mixed thoroughly. Using a wooden applicator stick, all tubes were inspected for clots. If any sample was clotted, it was discarded and recollected. The donor samples were concentrated and diluted (as necessary) to create high and low concentration linearity samples.
[0147] Baseline RBC Sample
[0148] One of the 3 tubes of donor blood was labeled “Baseline.” The Baseline tube was processed on the Cobas® m 511 system using standard steps and set aside for future use. It is recommended to collect a donor sample with a baseline RBC result of ≥4.5×10.sup.6/μL.
[0149] High Concentration RBC Sample
[0150] The high concentration RBC sample was prepared to meet the RBC target value of ≥7.0×10.sup.6/μL. To prepare the high concentration sample, the remaining donor tubes were centrifuged for ten (10) minutes at 5000 RPM. After centrifugation was complete, ˜⅔ of the plasma was removed and transferred from each donor tube into a single 15 mL conical tube. This conical tube was labeled as “Plasma” and retained for later use. If disruption of the buffy-coat layer or RBC layer occurred, centrifugation steps were repeated.
[0151] Low Concentration RBC Sample
[0152] The low concentration RBC sample was prepared to meet a target value of ≤0.30×10.sup.6/μL. Aliquots of the Baseline and Plasma tubes previously set aside were combined and labeled “low concentration.” The low concentration sample was thoroughly mixed and processed on the Cobas® m 511 system using standard steps. Results were reviewed to verify that RBC target value of ≤0.30×10.sup.6/μL was achieved.
[0153] Linearity Series Preparation The linearity series was prepared by an equidistant dilution series between the high concentration sample and the low concentration sample as described in Example 4.
[0154] Sample Processing
[0155] Each dilution (concentration) was thoroughly mixed to prepare for processing by the Cobas® m 511 system. Each sample was printed on a glass microscope slide in a monolayer, and then automatically fixed and stained by the analyzer. The analyzer then automatically performed digital identification and counting of RBCs and evaluated RBCs—for cellular morphology and content (e.g. HGB and HCT). Automated results for each dilution were compared to target values of each dilution to determine linearity of the samples.
[0156] Results for RBC Linearity Samples Thirty RBC linearity results (five concentrations, six replicates per concentration), were obtained on the Cobas® m 511 system. The five equidistant data points spanning a truncated RBC linear range (≤0.3-7.0×10.sup.6/μL). The mean value of six replicates for each of the five equidistant samples fell within acceptable limits as shown in Table 13.
TABLE-US-00014 TABLE 13 Deviation from Linearity for RBC Using Whole Blood RBC Protocol in FIG. 10A Deviation.from. Target Linearity [10.sup.6/μL] [10.sup.6/μL] Deviation [%] Criteria Pass/Fail 0.27 −9.0e−05 −0.0 0.05 [10.sup.6/μL] OK 2 −6.8e−03 −0.3 5 [%] OK 3.73 5.7e−02 1.5 5 [%] OK 5.46 5.2e−02 0.9 5 [%] OK 7.19 −0.15 −2.1 5 [%] OK
[0157] Comparison of Tables 5 and 13 emphasizes the enhanced linear range of results, accuracy, and precision using samples prepared from the high, low, and seven intermediate concentration samples versus a linearity series prepared from high, low, and three intermediate concentration samples. Deviation using whole blood RBC protocol was higher than when using specialized blood products protocol as described herein.
Example 6: Comparison of PLT Linearity Using the Whole Blood Protocol
[0158] Without specialized samples and sample preparation, described in Example 3, the entire AMI for PLT cannot be achieved. The purpose of Example 6 was to demonstrate the reduced (truncated) range of AMI for an image-based hematology analyzer when using samples prepared with a whole blood implementation protocol for PLT.
[0159] To study whole blood PLT AMI, a minimum of five (5), 10 mL K2/K3-EDTA tubes from a single donor were obtained via standard venipuncture and mixed thoroughly. Using a wooden applicator stick, all tubes were inspected for clots. If any sample was clotted, it was discarded and recollected. The donor samples were concentrated and diluted (as necessary) to create high and low concentration linearity samples.
[0160] Baseline PLT Sample
[0161] One of the 5 tubes of donor blood was labeled as “Baseline.” The Baseline tube was processed on the Cobas® m 511 system using standard steps, and set aside for future use. It is recommended to collect a donor sample with a baseline PLT count of ≥300×10.sup.3/μL.
[0162] Platelet Rich Plasma Sample Preparation
[0163] To create the high concentration PLT sample to meet the PLT target concentration of ≥999×10.sup.3/μL, a Platelet Rich Plasma (PRP) fraction was first prepared by centrifuging the 4 remaining donor tubes for ten (10) minutes at 2000 RPM. While centrifugation occurred, 0.3 g of EDTA was weighed and transferred into each of two EDTA 15 mL conical tubes. After centrifugation was complete, the platelet rich plasma from each donor tube was removed and equally distributed between the two 15 mL conical tubes containing EDTA. It is acceptable if some RBCs were pipetted into the EDTA tubes during transfer of plasma due to unclear separation of RBCs and plasma layers. If separation of RBCs and plasma is not achieved based on a visual assessment, the donor tubes may be re-centrifuged for an additional five (5) minutes at 2000 RPM. The “PRP” tubes were mixed thoroughly to ensure the EDTA was dissolved into solution.
[0164] The following tubes were centrifuged for (10) minutes at 5000 RPM: [0165] a. The two (2) 15 mL conical PRP tubes [0166] b. The two (2) 10 mL EDTA donor tubes with PRP removed
[0167] High Concentration PLT Sample
[0168] After centrifugation was completed and without disturbing the PLT pellet, the platelet poor plasma (PPP) was removed from the PRP tube and transferred into a new 15 mL conical tube labeled as “PPP” and retained for later use.
[0169] The contents of the two (2) “PRP” tubes were thoroughly mixed until no visible PLT pellet remained. The contents of the two “PRP” tubes were combined and 500 μL of “Baseline” sample was added. This conical tube was labeled as “high concentration.”
[0170] The high concentration tube was mixed thoroughly and processed on the Cobas® m 511 system using standard steps. Results were reviewed to verify that PLT was ≥999×10.sup.3/μL. If unable to achieve high concentration PLT requirement, a new donor sample with a higher baseline PLT result ≥300×10.sup.3/μL was collected.
[0171] Low Concentration PLT Sample
[0172] The low concentration PLT sample was prepared to meet two target concentrations, ≤20.0×10.sup.3/μL for PLT and ≥0.30×10.sup.6/μL for RBC. RBCs were added to meet the minimum number of cells necessary for the Roche image-based hematology analyzer (Cobas® m 511 system) to initiate sample processing and facilitate imaging. To prepare the low concentration sample, the two (2) 10 mL EDTA donor tubes containing PRP were retrieved. Any remaining plasma and buffy coat was removed and discarded. The packed RBCs were transferred and combined into a new 15 mL conical tube labeled “low concentration into which 3.0-4.0 mL of PPP was transferred.
[0173] Contents of the low concentration tube were mixed thoroughly and processed on the Cobas® m 511 system using standard steps. Results were reviewed to verify that PLT was ≤20.0×10.sup.3/μL and RBC was ≥0.30×10.sup.6/μL.
[0174] Linearity Series Preparation
[0175] The linearity series was prepared by an equidistant dilution series between the high concentration sample and the low concentration sample as described in Example 4.
[0176] Sample Processing
[0177] Each dilution (concentration) was thoroughly mixed to prepare for processing by the Cobas® m 511 system. Each sample was printed on a glass microscope slide in a monolayer, and then automatically fixed and stained by the analyzer. The analyzer then automatically performed digital identification and counting of PLTs. Automated results for each dilution were compared to target values of each dilution to determine linearity of the samples.
[0178] Results for PLT Linearity Samples
[0179] Thirty PLT linearity results (five concentrations, six replicates per concentration), were obtained on the Cobas® m 511 system using standard steps. Five equidistant data points spanning the truncated PLT linear range (8-1379×10.sup.3/μL) were used. The mean value of six replicates for each of the five equidistant samples fell within acceptable limits as shown in Table 14.
TABLE-US-00015 TABLE 14 Deviation from Linearity for PLT Using Whole Blood PLT Protocol in FIG. 11A Target Mean SD CV [10.sup.3/μL] N [10.sup.3/μL] [10.sup.3/μL] [%] Var Weight 8 6 8.17 0.41 5.0 0.17 1.6e−02 350.75 6 356 2.9 0.8 8.4 8.1e−06 693.5 6 692.5 13.49 1.9 181.9 2.1e−06 1036.25 6 1052 7.15 0.7 51.07 9.3e−07 1379 6 1379 28.74 2.1 826.27 5.3e−07
[0180] Comparison of Tables 8 and 14 emphasizes the enhanced linear range of results, accuracy, and precision using specialized linearity samples prepared from the high, low, and seven intermediate concentration samples versus a linearity series prepared from whole blood and high, low, and three intermediate concentration samples. Using whole blood protocol yielded a higher deviation than when using special blood products protocol as described herein.
[0181] Linearity Control Slides
[0182] Data has been generated via methods using specialized blood products in combination with the gold standard verification protocols (Examples 1-3), and via methods using whole blood samples in combination with whole blood implementation protocols (Examples 4-6). Regardless of method chosen and AMI outcome, all 6 examples require investment of a significant amount of time, money, and technical expertise. Eliminating those obstacles is highly advantageous for customers.
Example 7: Comparison of Linearity Control Slides for RBC-HGB-HCT
[0183] The purpose of Example 7 was to compare the AMI of two Cobas® m 511 systems using linearity control slides. Parameter(s) RBC-HGB-HCT were selected to demonstrate feasibility of proof-of-principle.
[0184] The linearity dilutions were prepared and processed on Cobas® m 511 system C4 using (gold standard) methods previously described in Example 2. The counts for RBC-HGB-HCT were obtained. The linearity control slides resulting from the gold standard RBC method were retained. Printing and staining functions of the same Cobas® m 511 system C4 were then temporarily reconfigured.
[0185] The linearity control slides were then processed on two Cobas® m 511 systems: first on the same Cobas® m 511 system (C4), and then on a second Cobas® m 511 system (P37): [0186] 1) the linear range for RBC (from the linearity control slides) processed on the first Cobas® m 511 system is 0.35-8.66×10.sup.6/μL and processed on the second Cobas® m 511 system is 0.34-8.30×10.sup.6/μL as shown in
TABLE-US-00016 TABLE 15 Linear Range for RBC (0% Deviation) Target Mean SD % CV Deviation % Deviation Test 0.35 0.35 0.011 3.13 0.000 0.00 PASS 1.35 1.35 0.023 1.74 0.000 0.00 PASS 2.30 2.30 0.028 1.20 0.000 0.00 PASS 3.30 3.30 0.010 0.30 0.000 0.00 PASS 4.33 4.33 0.020 0.45 0.000 0.00 PASS 5.38 5.38 0.031 0.58 0.000 0.00 PASS 6.43 6.43 0.054 0.84 0.000 0.00 PASS 7.71 7.71 0.050 0.65 0.000 0.00 PASS 8.66 8.66 0.060 0.70 0.000 0.00 PASS [0187] 2) the acceptable limits and RBC deviation from linearity (using the linearity control slides) was calculated on the first Cobas® m 511 system as 0% deviation for all concentrations, and ranged from −0.68% to 0.07% deviation on the second Cobas® m 511 system as shown in
TABLE-US-00017 TABLE 16 Linear Range for RBC (<1% Deviation) Target Mean SD % CV Deviation % Deviation Test 0.35 0.34 0.010 0.00 −0.002 −0.68 PASS 1.35 1.30 0.023 1.80 0.000 0.01 PASS 2.30 2.21 0.028 1.27 0.001 0.07 PASS 3.30 3.17 0.010 0.33 0.002 0.06 PASS 4.33 4.17 0.019 0.45 0.001 0.03 PASS 5.38 5.17 0.033 0.64 −0.000 −0.00 PASS 6.43 6.17 0.051 0.82 −0.003 −0.05 PASS 7.71 7.40 0.043 0.58 −0.008 −0.10 PASS 8.66 8.30 0.054 0.65 −0.013 −0.15 PASS [0188] 3) the linear range for HGB (from the linearity control slides) processed on the first Cobas® m 511 system is 1.08-26.02 g/dL and processed on the second Cobas® m 511 system, is 1.00-25.57 g/dL as shown in Table 17, meeting the target values of ≤1.2-≥24.0×g/dL.
TABLE-US-00018 TABLE 17 Linear Range for HGB (0% Deviation) Target Mean SD % CV Deviation % Deviation Test 1.08 1.08 0.041 3.77 0.000 0.00 PASS 3.97 3.97 0.082 2.06 0.000 0.00 PASS 6.75 6.75 0.105 1.55 0.000 0.00 PASS 9.67 9.67 0.052 0.53 0.000 0.00 PASS 12.72 12.72 0.075 0.59 0.000 0.00 PASS 15.83 15.83 0.163 1.03 0.000 0.00 PASS 18.95 18.95 0.138 0.73 0.000 0.00 PASS 22.88 22.88 0.133 0.58 0.000 0.00 PASS 26.02 26.02 0.286 1.10 0.000 0.00 PASS [0189] 4) the acceptable limits and HGB deviation from linearity (using the linearity control slides) was calculated on the first Cobas® m 511 system as 0% deviation for all concentrations, and ranged from −0.35% to 2.71% deviation on the second Cobas® m 511 system as shown in Table 18.
TABLE-US-00019 TABLE 18 Linear Range for HGB (<3% Deviation) Target Mean SD % CV Deviation % Deviation Test 1.08 1.00 0.052 0 0.029 2.71 PASS 3.97 3.82 0.075 1.97 0.022 0.55 PASS 6.75 6.48 0.098 1.52 0.001 0.01 PASS 9.67 9.33 0.052 0.55 −0.025 −0.25 PASS 12.72 12.28 0.075 0.61 −0.044 −0.35 PASS 15.83 15.28 0.133 0.87 −0.045 −0.28 PASS 18.95 18.45 0.138 0.75 −0.013 −0.07 PASS 22.88 22.33 0.137 0.61 0.094 0.41 PASS 26.02 25.57 0.250 0.98 0.247 0.95 PASS [0190] 5) the linear range for HCT (from the linearity control slides) processed on the first Cobas® m 511 system is 3-79% and processed on the second Cobas® m 511 system is 3-78% as shown in Table 19, respectively, meeting the target values of ≤3.6-≥72.0%.
TABLE-US-00020 TABLE 19 Linear Range for HCT (0% Deviation) Target Mean SD % CV Deviation % Deviation Test 3.00 3.00 0.001 2.34 0.000 0.00 PASS 12.00 12.00 0.002 1.99 0.000 0.00 PASS 21.00 21.00 0.002 0.99 0.000 0.00 PASS 30.00 30.00 0.001 0.40 0.000 0.00 PASS 40.00 40.00 0.002 0.42 0.000 0.00 PASS 49.00 49.00 0.004 0.78 0.000 0.00 PASS 59.00 59.00 0.006 0.99 0.000 0.00 PASS 71.00 71.00 0.007 1.01 0.000 0.00 PASS 79.00 79.00 0.006 0.80 0.000 0.00 PASS [0191] 6) the acceptable limits and HCT deviation from linearity (using the linearity control slides) was calculated on the first Cobas® m 511 system as 0% deviation for all concentrations, and ranged from −0.10% to 0.95% deviation on the second Cobas® m 511 system as shown in Table 20.
TABLE-US-00021 TABLE 20 Linear Range for HCT (<1% Deviation) Target Mean SD % CV Deviation % Deviation Test 3.00 3.00 0.001 1.60 0.000 0.28 PASS 12.00 12.00 0.002 2.00 0.000 0.09 PASS 21.00 20.00 0.003 1.39 −0.000 −0.10 PASS 30.00 29.00 0.002 0.60 −0.001 −0.20 PASS 40.00 39.00 0.002 0.48 −0.001 −0.20 PASS 49.00 48.00 0.004 0.79 −0.000 −0.09 PASS 59.00 57.00 0.005 0.92 0.001 0.13 PASS 71.00 69.00 0.006 0.87 0.004 0.54 PASS 79.00 78.00 0.009 1.18 0.008 0.95 PASS
[0192] The linearity control slides, prepared from specialized high, low, and intermediate concentration samples as described herein provide comparable correlation to freshly prepared linearity samples from specialized high, low, and intermediate concentration samples.
[0193] Additionally, linearity control slides prepared, as described, provide decreased variability, increased reproducibility and achieve the entire AMI compared to linearity samples prepared from whole blood samples. Comparison of Tables 13 and 16 emphasizes the enhanced accuracy, precision, and instrument-to-instrument standardization from the linearity control slides based on the linearity samples prepared from the high and low concentration blood samples. Deviation while using specialized blood products was consistent when measured on two Cobas® m 511 systems.
[0194] Lastly, since the linearity control slides are created/generated by the manufacturer, the time and cost savings to the customer are significant.
Example 8: Stability of Linearity Control Slides
[0195] This study was conducted to demonstrate that Cobas® m 511 system slides stored for 30 days produce similar results compared to original (baseline) results, with respect to manual microscopic reviews for the WBC differential and morphology. The results from all evaluations performed during this study were successful and met predefined acceptance criteria. This demonstrates that slides generated by the Cobas® m 511 system are stable for manual microscopic review for a minimum of 30 days up to 180 days.
[0196] To verify Cobas® m 511 slide stability performance, the protocol was performed using at least twenty (20) samples collected in K2 EDTA collection tubes, and processed on a SYSMEX® Automated Hematology Analyzer. The samples consisted of: [0197] six (6) random residual whole blood samples [0198] five (5) normal healthy donors [0199] nine (9) targeted residual whole blood samples; three (3) each from each of the following categories: [0200] Anemia HGB 6-10 g/dL [0201] Leukopenia WBC ≤2×10.sup.3/μL [0202] Thrombocytopenia PLT ≤50×10.sup.3/μL
[0203] Each sample was processed in triplicate in rack mode on one (1) Cobas® m 511 system within eight (8) hours of sample collection. The Cobas® m 511 system slides created by the Cobas® m 511 system were evaluated by a qualified Medical Technologist (MT) at baseline, after a minimum of thirty (30) days, and again after a minimum of one-hundred eighty (180) days.
[0204] The evaluation consisted of five (5) separate analyses. None of the individual evaluations is required to meet a specific PSD requirement; rather, the combined set of results will be used to determine slide stability. [0205] 1—Red Blood Cell Inclusions: Present/Not Present: report % agreement [0206] 2—WBC, RBC, and PLT Morphology: Grading 1-4+: report % agreement [0207] 3—WBC Diff Parameters Bias Assessment: report % bias [0208] 4—WBC Diff Parameters Regression Analysis: report slope and intercept [0209] 5—Macroscopic Slide Characteristics: Peeling, Flaking, Print, Color Fading: Grading 1-3+: report % agreement
[0210] The slide assessment of the first replicate (baseline) for each sample processed on the Cobas® m 511 system occurred within eight (8) hours of sample processing. The remaining slides for each sample were stored in Fisher Scientific Slide storage boxes for the remainder of the study at room temperature (15°-25° C.). After one-hundred eighty (180) days, a qualified Medical Technologist (MT) assessed one (1) of the remaining two (2) slides. Following assessment, the Cobas® m 511 system slides were placed back into the slide storage boxes.
[0211] Cell characteristics (RBC Inclusions and WBC, RBC, and PLT morphology) exhibit no difference between zero (0) and thirty (30) days nor zero (0) and one-hundred eighty (180) days. Percent bias and correlation coefficients for the 5-part WBC diff are acceptable and lastly, macroscopic assessments of Cobas® m 511 slides indicate no observations of sample peeling, sample flaking or change in print uniformity, and only a slight change in color (fading) when comparing baseline results to results obtained after at least thirty (30) days in storage and one hundred-eighty (180) days in storage.
[0212] This protocol demonstrated that slides generated by the Cobas® m 511 system are stable for a minimum of one-hundred eighty (180) days.
Other Embodiments
[0213] It is to be understood that while the invention has been described in conjunction with the detailed description thereof, the foregoing description is intended to illustrate and not limit the scope of the invention, which is defined by the scope of the appended claims. Other aspects, advantages, and modifications are within the scope of the following claims.