BLOOD CULTURE SYSTEM WITH LOW MEDIA VOLUME
20190078043 ยท 2019-03-14
Assignee
Inventors
- Mei Yang-Woytowitz (Baltimore, MD, US)
- Song Shi (Reisterstown, MD)
- Xiaofei Chang (Timonium, MD, US)
- James Y. Zhou (Columbia, MD, US)
- Michael A. Brasch (Gaithersburg, MD)
Cpc classification
C12N2501/999
CHEMISTRY; METALLURGY
C12M41/46
CHEMISTRY; METALLURGY
International classification
Abstract
A container for culturing a blood sample. The container has a reservoir that is no larger than about 40 ml in volume with culture media therein varying in amount by volume 0.5 ml to about 20 ml. The container is adapted to receive a blood sample drawn from a patient, wherein the blood volume is about 1 ml to about 20 ml. In some embodiments the ratio of blood volume to culture media volume is about 2:1 to about 1:2 and the volume of blood does not exceed about 10 ml. In some embodiments, the media is lytic media. A method for using the container to culture a blood sample is also contemplated. In such method, the container is inoculated with the blood sample. In certain embodiments, the volume of the blood sample does not exceed 10 mls.
Claims
1. A container for culturing a blood sample, the container comprising: a reservoir that is no larger than about 40 ml in volume; culture media varying in amount by volume 0.5 ml to about 20 ml; wherein the container is adapted to receive a blood sample drawn from a patient, wherein the blood volume is about 1 ml to about 20 ml.
2. The container of claim 1 wherein the ratio of blood volume to culture media volume is about 2:1 to about 1:2 and the volume of blood does not exceed about 10 ml.
3. The container of claim 1 wherein the media is lytic media.
4. The container of claim 3 wherein the ratio of blood volume to culture media volume is about 2:1 to about 1:2.
5. The container of claim 4 wherein the blood volume is about 4 ml to about 10 ml and the culture media volume is about 2 ml to about 20 ml and the volume of blood does not exceed about 10 ml.
6. The container of claim 4 wherein the blood volume is about 1 ml to about 3 ml and the culture media volume is about 0.5 ml to about 6 ml.
7. The container of claim 1 further comprising a sensor, wherein the sensor detects a change in the culture media combined with blood sample indicative of microbial growth in the container.
8. A method for culturing a blood sample, the method comprising: providing a consumable with a reservoir having a volume that does not exceed about 40 ml, the reservoir having a predetermined volume of culture media disposed therein, the predetermined volume selected to provide a ratio of blood volume to culture volume in the range of about 2:1 to about 1:2.
9. The method of claim 8 wherein the blood volume is in the range of about 4 ml to about 10 ml.
10. The method of claim 8 wherein the blood volume is about 1 ml to about 3 ml.
11. The method of claim 8 wherein the culture media is lytic media.
12. The method of claim 8 wherein the lytic media is about 0.25 grams of lytic reagent for 100 ml of culture media.
13. The method of claim 12 wherein the lytic reagent is saponin.
14. The method of claim 8 wherein the consumable has a sensor disposed therein capable of detecting a change in the consumable that occurs as a result of microbial growth therein.
15. The method of claim 8 further comprising incubating the consumable after the consumable has been inoculated with the blood sample.
Description
BRIEF DESCRIPTION OF THE DRAWINGS
[0013]
[0014]
[0015]
[0016]
DETAILED DESCRIPTION
[0017] Described herein are methods for preparing and testing blood culture samples in higher concentration of blood in broth (providing lowered blood-to-broth ratios). Preferably, higher concentration of blood in broth refers to a broth (or media) volume that is equal to or less than the blood volume of a culture sample (e.g. 1:1, 2:1). Media and broth are used interchangeably herein. Samples with higher concentrations of blood in broth (i.e. providing lower blood-to-broth ratios) described herein showed similar or even improved time-to-detection (TTD) results as compared with current industry blood culture sample bottles that use a more diluted blood-to-broth ratio (e.g. 1:3, 1:4). As used herein, time-to-detection or TTD is defined as the time elapsed between the time point that blood microorganisms are inoculated into the containers and the time point that the system can detect microorganism growth in the containers.
[0018] In one embodiment, the blood cultures do not contain antibiotic inhibitors. In another embodiment, the blood cultures contain a predetermined amount of antibiotic inhibitor.
[0019] The methods described herein use blood cultures having lower blood-to-broth ratios, of 1:1 up to about 2:1. The volume and configuration of blood culture bottles in which the blood cultures are disposed may vary. For example,
[0020] Using a lower blood-to-broth ratio allows for a smaller amount of growth media and reagents, allowing smaller volume blood culture bottles to be used. Smaller volume blood culture bottles may have a reduced diameter, a shorter height, or both. The reduced diameter bottles also allows for a more accurate amount of blood draw by indicating volume more precisely than the larger 80 ml. bottle. The reduction in volume also reduces the cost of the consumable and allows a reduction in the instrument footprint by permitting more vials to be tested at once compared to the current standard 80 ml bottles.
Example 1: Comparison of Time to Detection (TTD) for Various Microorganisms Using Standard Ratio/Standard Volume and Low Ratio/Low Volume
[0021] Time to Detection was compared for the standard ratio of about 1:5 blood to media volume (i.e. 8 ml of blood and 40 ml of culture media) with 1:1 blood to media volume. Both media with no lysis reagent (non-lytic media) and media with a lysis regent (lytic media) were used. The non-lytic media was the BACTEC? Aerobic Media. For the 1:4 ratio samples, 40 ml of media was used. The standard BACTEC Aerobic bottles contain 40 ml of media. So for the 1:1 ratio examples, 32 ml of media were removed from the bottle before the sample was introduced into the bottle. To make the media lytic, about 0.25% of saponin was added to the media.
[0022] As noted above, four sets of samples were prepared. Two sets had the 1:4 standard blood-to-broth ratio and two sets had the 1:1 ratio. One of the two sets for each ratio had non lytic media and the other set had lytic media. To create the samples with the 1:1 blood-to-broth ratio, 32 mls of the 40 mls of media in the standard BACTEC Aerobic bottle was withdrawn, leaving only 8 ml broth remaining. To make the broth lytic, a lytic reagent, saponin, was added to the remaining broth (40 mls for the 1:4 ratio and 8 mls for the 1:1 ratio) in an amount to yield a concentration of 0.25% in the respective standard ratio and low ratio samples. About 8 ml of blood and 10-100 cfu of each of the microorganism listed in Table 1 below were inoculated into each of the bottles. The BACTEC Standard Aerobic bottles with the original 40 ml media was used for those samples having the 1:4 blood to broth ratio. All inoculated bottles were placed into the BD BACTEC? FX instrument and cultured for 5 days at 35? C. The result showed that the time to detection for the eight microorganisms identified in Table 1 below were comparable with each other. Therefore, the 1:1 blood to media ratio supports microorganism growth in blood culture even though the blood dilution factor for the low 1:1 ratio was significantly less than the conventional 1:4.
TABLE-US-00001 TABLE 1 TTD for Various Organisms Using Standard Ratios and Volumes and Low Ratios and Volumes TTD (hrs) 1:4 BMR 1:4 BMR 1:1 BMR 1:1 BMR Organisms non lytic lytic non lytic Lytic Acinetobacter 15.0 15.2 14.6 21.1 lwoffii Candida albicans 25.8 26.6 27.0 28.3 Candida glabrata 64.0 50.0 70.0 68.0 Cryptococcus 75.0 80.0 80.0 65.0 neoformans Escherichia coli 10.9 10.7 10.7 10.4 Pseudomonas 15.1 14.9 16.7 15.5 aeruginosa Staphylococcus 12.5 11.8 13.6 11.4 aureus Streptococcus 14.5 19.1 14.1 14.7 pneumoniae
[0023] Surprisingly, the samples with the 1:1 ratio (low ratio/low volume) performed similarly to the samples with the 1:4 ratio (standard ratio/standard volume). Also, the above table shows that, for some microorganisms, the TTD was lower for the bottles with the lytic media than for the bottles with the non-lytic media. However, for some of the microorganisms tested, an increase in TTD was observed for the bottles having the lytic media compared to the bottles with the non-lytic media. Since the sample size was small, these observations are not necessarily indicative of a substantial difference to TTD for bottles with non-lytic media compared with bottles with lytic media.
[0024]
[0025] The liquid media is a modified media based on BACTEC Myco/F Lytic media with additional saponin. The samples are each loaded (i.e. spiked) with 10-100 CFU's of microorganisms 211. The microorganisms tested were Alcaligenes faecalis, Candida albicans, Candida glabrata, Enterococcus faecalis, Escherichia coli, Neisseria meningitides, Pseudomonas aeruginosa, Staphylococcus aureus (2 different strains), Staphylococcus epidermidis, Streptococcus pneumonia, and Streptococcus pyogenes. The sensors 212 used on the bottom of the tubes were made with the similar formulation as in BACTEC Myco/F Lytic bottles.
[0026] For each of the microorganisms, the blood culture bottles 203, 205, 208 were inverted 5 times immediately after the addition of the microorganisms 211 and then loaded into a BACTEC FX instrument. An adaptor was used to hold the tubes, so that the tubes stayed in the stations firmly and were located at the proper positions for detection. The tubes and bottles were incubated for 5 days and the signals from the culture bottle sensors 212 were detected for TTD analysis. Triplicate cultures were performed for each microorganism.
[0027] The results for the procedures are depicted in Table 2 below. As shown in Table 2, the blood culture bottles 203, 205 (
TABLE-US-00002 TABLE 2 Comparison of low volume/low ratio TTD with TTD for standard 1:4 ratio Blood to Blood to StdAer Organisms Broth 1:1 Broth 2:1 (Control) Alcaligenes faecalis 14.9 17.1 19.6 Candida albicans 20.3 22.7 22.6 Enterococcus faecalis 10.2 10.0 11.1 Escherichia coli 7.9 9.6 10.9 Neisseria meningitidis 12.7 18.2 19.1 Pseudomonas aeruginosa 11.4 12.4 15.4 Staphylococcus aureus 8.6 8.8 12.2 (strain #1) Staphylococcus aureus 8.5 9.0 11.9 (strain #2) Streptococcus epdidermidis 13.4 13.2 17.9 Streptococcus pneumoniae 12.2 12.8 14.0 Streptococcuspyogenes 8.7 9.0 10.8 (Strp. group A) Average 11.7 13.0 15.0 (Standard Deviation) (3.6) (4.6) (4.2)
[0028] The average TTD of the 11 tested microorganisms in the preparations having the 1:1 blood-to-broth ratio 207 was 11.7 hours. The BACTEC Standard Aerobic bottles 208 with a more diluted blood-to-broth ratio of at least 1:3 210 had an average TTD of 15.0 hours, which was longer than the 1:1 sample 207. Similarly, the average TTD for the 16 ml bottle 210 with a 2:1 blood-to-broth ratio 204 was 13.0 hours. The shorter TTD for the 1:1 and 2:1 blood-to-broth ratios 207, 204 is an unexpected result and improvement over current more diluted blood-to-broth ratios like the BACTEC Standard Aerobic bottles with at least a 1:3 blood-to-broth ratio 210.
Example 2: Growth with Antibiotics in the Culture Media Inhibitor
[0029]
[0030] The performance of 1:1 blood-to-broth ratio in the presence of an inhibitory antibiotic was tested in comparison to a standard diluted blood-to-broth ratio at least 1:3. Two different sample preparations for the 1:1 blood-to-broth ratio are described below.
[0031]
[0032]
[0033] The control test sample preparation is depicted in
[0034] Each sample preparation was tested with 64 antibiotic/microorganism combinations. The antibiotics tested were Gentamicin, Amoxicillin-Clavulanate, Aztreonam, Ceftriaxone, Cefepime, Imipenem, Meropenem, Piperacillin-Tazobactam, Vancomycin, Tigecycline, Ciprofloxacin, Tetracycline, Fluconazole. The drug quantity inoculated into each test bottle was equivalent to the drug amount at its peak serum level in 7 ml of blood. These antibiotics were combined with certain microorganisms that they generally treat. For example, Amoxicillin-Clavulanate was combined with S. aureus, E. coli, and E. faecalis. Each sample was tested for both TTD and recovery of the microorganism in the presence of drugs.
[0035] The media was inoculated with blood sample, antibiotics and saponin a few minutes apart from each other.
[0036] In the TTD tests, the Low Volume/Low Ratio sample preparation had a similar performance to the control sample preparation, (the 1:4 PlusAer), for Ciprofloxicin, Piperacillin/Tazobactam, and Tigecycline. The Low Volume/Low Ratio preparation had a faster TTD as compared to the PlusAer preparation when tested with Vancomycin, Cefepime, Amoxicillin-Clavulanic Acid, Ceftriaxone, and Meropenem. See Table 3 below. The Low Volume/Low Ratio preparation had a slower TTD as compared to the PlusAer preparation for only one antibiotic, Gentamicin. The performance is surprising given the conventional view in the art that acceptable TTD cannot be achieved with higher amounts of blood relative to broth (i.e. lower ratios) (e.g. volume blood to volume media of 1:1 and 2:1) at low volumes (i.e. 20 ml or less total).
TABLE-US-00003 TABLE 3 Comparison of TTD/positivity for different combinations of Antibiotics and Microorganisms Low Volume/low ratio TTD (hours) (approximate Plus Plus Antibiotic Microorganism average) Lytic Aerobic Vancomycin S. aureus 23 15 18 280/?g S. aureus .sup.17.5 17.25 17 E. faecalis 16 15.75 17 S. aureus 18 15.25 18.75 E. faecalis 20 18 19 Amoxicillin/ S. aureus 65 25 none clavulanic E. coli 10 10 10 acid 81.2 ?g E. faecalis none 40 none E. coli 10 10 10 S. aureus 50 .sup.37.5 none Ceftriaxone P. aeruginosa none none none 1050 ?g E. faecalis 21 14 18 S. aureus none .sup.42.5 none P. aeruginosa 19 15 18 S. aureus 18 14 19 Meropenem P. aeruginosa 20 21 20 343 ?g S. aureus none none none E. faecalis 51 34 49 P. aeruginosa 35 39 none E. faecalis 16 14 12 P. aeruginosa none none none Cefepime S. aureus none none none 1351 ?g E. faecalis 23 21 29.5 S. aureus none none none E. faecalis 10 10 10 E. Coli 11 none none E. faecalis 13 12 13
[0037] With regard to drug absorption efficiency, the low volume low ratio preparation proved to have a faster absorption rate. The overall antibiotic absorption efficiency in a lower volume preparation is similar to that of the PlusAer preparation.
[0038]
[0039]
[0040]
[0041]
[0042] The recovery rate comparisons of
[0043] Although the invention herein has been described with reference to particular embodiments, it is to be understood that these embodiments are merely illustrative of the principles and applications of the present invention. It is therefore to be understood that numerous modifications may be made to the illustrative embodiments and that other arrangements may be devised without departing from the spirit and scope of the present invention as defined by the appended claims.