FLUORESCENCE TEMPERATURE MEASUREMENT MATERIAL, PREPARATION METHOD THEREFOR, AND USE THEREOF

20240279545 ยท 2024-08-22

    Inventors

    Cpc classification

    International classification

    Abstract

    A fluorescence temperature measurement material, a preparation method therefore and use thereof are disclosed, which belong to the technical field of fluorescence temperature sensing. The fluorescence temperature measurement material has a chemical composition of Na.sub.1-xSr.sub.xTaO.sub.3:yPr.sup.3+, x=0.1-0.2 and y=0.4%-0.6%. The fluorescence temperature measurement material is prepared by a high-temperature solid-phase method and generates blue light at 492 nm (.sup.3P.sub.0.fwdarw..sup.3H.sub.4) and red light at 610 nm (.sup.1D.sub.2.fwdarw..sup.3H.sub.4) under the excitation of 290 nm ultraviolet light. The fluorescence intensity ratio (.sup.1D.sub.2.fwdarw..sup.3H.sub.4/.sup.3P.sub.0.fwdarw..sup.3H.sub.4) of two emission peaks has an exponential function relationship with temperature, so that the fluorescence temperature measurement material can calibrate temperature and has good temperature-sensitive performance. Moreover, the fluorescence temperature measurement material has a particle size of <1 ?m, a good spatial resolution and a significant CIE color coordinate change along with temperature.

    Claims

    1. A fluorescence temperature measurement material, having a chemical composition of Na.sub.1-xSr.sub.xTaO.sub.3:yPr.sup.3+, wherein x=0.1-0.2, and y=0.4%-0.6%.

    2. The fluorescence temperature measurement material according to claim 1, wherein x=0.15 and y=0.5%.

    3. A preparation method for the fluorescence temperature measurement material according to claim 1, comprising the following steps: weighing raw materials based on the chemical composition, uniformly mixing, adding a solvent, grinding, pre-sintering, regrinding, and calcining to obtain the fluorescence temperature measurement material.

    4. The preparation method for the fluorescence temperature measurement material according to claim 3, wherein the raw materials comprise Na.sub.2CO.sub.3, SrCO.sub.3, Ta.sub.2O.sub.5, and Pr.sub.6O.sub.11.

    5. The preparation method for the fluorescence temperature measurement material according to claim 3, wherein the grinding is performed for 20-40 min.

    6. The preparation method for the fluorescence temperature measurement material according to claim 3, wherein the pre-sintering is performed at a temperature of 300-500? C. for 1-3 h.

    7. The preparation method for the fluorescence temperature measurement material according to claim 3, wherein the regrinding is performed for 10-20 min.

    8. The preparation method for the fluorescence temperature measurement material according to claim 3, wherein the calcining is performed at a temperature of 900-1050? C. for 6-10 h.

    9-10. (canceled)

    11. A temperature sensor, comprising the fluorescence temperature measurement material according to claim 1.

    12. A method for calibrating temperature, wherein the method comprises the steps of exciting the fluorescence temperature measurement material according to claim 1 with an ultraviolet light with a wavelength of 290 nm, and measuring the ratio of an emission peak intensity of the fluorescence temperature measurement material at 492 nm to the emission peak intensity of the fluorescence temperature measurement material at 610 nm.

    Description

    BRIEF DESCRIPTION OF DRAWINGS

    [0022] FIG. 1 is an XRD pattern of fluorescence temperature measurement materials according to Examples 1-3;

    [0023] FIG. 2 is an emission spectrum (?.sub.ex=290 nm) at room temperature of the fluorescence temperature measurement materials according to Examples 1-3;

    [0024] FIG. 3 is an SEM image of the fluorescence temperature measurement materials according to Example 1;

    [0025] FIG. 4 is a temperature-dependent spectrum (303 K to 483 K) of the fluorescence temperature measurement material according to Example 1 with an excitation wavelength of 290 nm;

    [0026] FIG. 5A is an intensity (303 K to 483 K) of an emission peak at 492 nm (blue) of the fluorescence temperature measurement material according to Example 1, and FIG. 5B is an intensity (303 K to 483 K) of an emission peak at 610 nm (red) of the fluorescence temperature measurement material according to Example 1;

    [0027] FIG. 6 is a fitted plot of a fluorescence intensity ratio (.sup.1D.sub.2.fwdarw..sup.3H.sub.4/.sup.3P.sub.0.fwdarw..sup.3H.sub.4) of the fluorescence temperature measurement material according to Example 1;

    [0028] FIG. 7 is a CIE color coordinate of the fluorescence temperature measurement material according to Example 1 in a range of 303 K to 483 K;

    [0029] FIG. 8 is a curve of an absolute sensitivity Sa and a relative sensitivity S.sub.r of the fluorescence temperature measurement material according to Example 1;

    [0030] FIG. 9 is a temperature-dependent spectrum (303 K to 483 K) of the fluorescence temperature measurement material according to Example 2 with an excitation wavelength of 290 nm;

    [0031] FIG. 10 is a temperature-dependent spectrum (303 K to 483 K) of the fluorescence temperature measurement material according to Example 3 with an excitation wavelength of 290 nm;

    [0032] FIG. 11A is an XRD pattern of the fluorescence temperature measurement material according to Comparative Example 1, and FIG. 11B is an emission spectrum (?.sub.ex=290 nm) at room temperature of the fluorescence temperature measurement material according to Comparative Example 1;

    [0033] FIG. 12 is a temperature-dependent spectrum (303 K to 483 K) of the fluorescence temperature measurement material according to Comparative Example 1 with an excitation wavelength of 290 nm;

    [0034] FIG. 13A is an XRD pattern of the fluorescence temperature measurement material according to Comparative Example 2, and FIG. 13B is an emission spectrum (?.sub.ex=290 nm) at room temperature of the fluorescence temperature measurement material according to Comparative Example 2; and

    [0035] FIG. 14 is a temperature-dependent spectrum (303 K to 483 K) of the fluorescence temperature measurement material according to Comparative Example 2 with an excitation wavelength of 290 nm.

    DETAILED DESCRIPTION

    [0036] To better illustrate the objectives, technical solutions, and advantages of the present disclosure, the present disclosure will be further described below with reference to specific examples. Those skilled in the art should understand that the specific examples described herein are merely illustrative of the present disclosure and do not limit the protection scope of the present disclosure.

    [0037] Unless otherwise stated, the test methods used in the following examples are conventional methods. The materials, reagents, and the like used in the following examples can be commercially available unless otherwise stated. The material of the present disclosure is used for non-contact temperature measurement.

    Example 1

    [0038] 0.3 mmol of SrCO.sub.3, 0.85 mmol of Na.sub.2CO.sub.3, 1 mmol of Ta.sub.2O.sub.5, and 0.00167 mmol of Pr.sub.6O.sub.11 were uniformly mixed, 5 mL of anhydrous ethanol was added, a resulting mixture was ground in an agate mortar for 30 min, and the grinded mixture was filled into a corundum crucible and then put into a muffle furnace for pre-sintering for 2 h at 400? C. After the sample was cooled, the sample was placed in a mortar to be ground for 15 min at a constant speed. The reground powder was loaded into a crucible and calcined in a muffle furnace at 1000? C. for 8 h, and finally the cooled sample was reground to uniform particles to obtain a Na.sub.0.85Sr.sub.0.15TaO.sub.3:0.5% Pr.sup.3+ material.

    Example 2

    [0039] 0.2 mmol of SrCO.sub.3, 0.9 mmol of Na.sub.2CO.sub.3, 1 mmol of Ta.sub.2O.sub.5, and 0.00167 mmol of Pr.sub.6O.sub.11 were uniformly mixed, 5 mL of anhydrous ethanol was added, a resulting mixture was ground in an agate mortar for 30 min, and the grinded mixture was filled into a corundum crucible and then put into a muffle furnace for pre-sintering for 2 h at 400? C. After the sample was cooled, the sample was placed in a mortar to be ground for 15 min at a constant speed. The reground powder was loaded into a crucible and calcined in a muffle furnace at 1000? C. for 8 h, and finally the cooled sample was reground to uniform particles to obtain a Na.sub.0.9Sr.sub.0.1TaO.sub.3:0.5% Pr.sup.3+ material.

    Example 3

    [0040] 0.4 mmol of SrCO.sub.3, 0.8 mmol of Na.sub.2CO.sub.3, 1 mmol of Ta.sub.2O.sub.5, and 0.00167 mmol of Pr.sub.6O.sub.11 were uniformly mixed, 5 mL of anhydrous ethanol was added, a resulting mixture was ground in an agate mortar for 30 min, and the grinded mixture was filled into a corundum crucible and then put into a muffle furnace for pre-sintering for 2 h at 400? C. After the sample was cooled, the sample was placed in a mortar to be ground for 15 min at a constant speed. The reground powder was loaded into a crucible and calcined in a muffle furnace at 1000? C. for 8 h, and finally the cooled sample was reground to uniform particles to obtain a Na.sub.0.8Sr.sub.0.2TaO.sub.3:0.5% Pr.sup.3+ material.

    Comparative Example 1

    [0041] 0.1 mmol of SrCO.sub.3, 0.95 mmol of Na.sub.2CO.sub.3, 1 mmol of Ta.sub.2O.sub.5, and 0.00167 mmol of Pr.sub.6O.sub.11 were uniformly mixed, 5 mL of anhydrous ethanol was added, a resulting mixture was ground in an agate mortar for 30 min, and the grinded mixture was filled into a corundum crucible and then put into a muffle furnace for pre-sintering for 2 h at 400? C. After the sample was cooled, the sample was placed in a mortar to be ground for 15 min at a constant speed. The reground powder was loaded into a crucible and calcined in a muffle furnace at 1000? C. for 8 h, and finally the cooled sample was reground to uniform particles to obtain a Na.sub.0.95Sr.sub.0.05TaO.sub.3:0.5% Pr.sup.3+ material.

    Comparative Example 2

    [0042] 0.6 mmol of SrCO.sub.3, 0.7 mmol of Na.sub.2CO.sub.3, 1 mmol of Ta.sub.2O.sub.5, and 0.00167 mmol of Pr.sub.6O.sub.11 were uniformly mixed, 5 mL of anhydrous ethanol was added, a resulting mixture was ground in an agate mortar for 30 min, and the grinded mixture was filled into a corundum crucible and then put into a muffle furnace for pre-sintering for 2 h at 400? C. After the sample was cooled, the sample was placed in a mortar to be ground for 15 min at a constant speed. The reground powder was loaded into a crucible and calcined in a muffle furnace at 1000? C. for 8 h, and finally the cooled sample was reground to uniform particles to obtain a Na.sub.0.7Sr.sub.0.3TaO.sub.3:0.5% Pr.sup.3+ material.

    Example of Use

    [0043] The XRD patterns of the fluorescence temperature measurement materials according to Examples 1-3 were measured by an X-ray diffractometer, the results were shown in FIG. 1, and it can be seen that the crystal structure of orthorhombic perovskite was not affected by the doping of 0.5% Pr.sup.3+.

    [0044] The emission spectra (?.sub.ex=290 nm) at room temperature of the fluorescence temperature measurement materials according to Examples 1-3 were measured by a fluorescence spectrometer, and the results were shown in FIG. 2, As can be seen from FIG. 2, under the excitation of ultraviolet light at 290 nm, the Na.sub.0.85Sr.sub.0.15TaO.sub.3:0.5% Pr.sup.3+ material of Example 1, the Na.sub.0.9Sr.sub.0.1TaO.sub.3:0.5% Pr.sup.3+ material al of Example 2, and the Na.sub.0.8Sr.sub.0.2TaO.sub.3:0.5% Pr.sup.3+ material of Example 3 all exhibited two strong main emission peaks at 492 nm and 610 nm, respectively, which were corresponding to .sup.3P.sub.0.fwdarw..sup.3H.sub.4 and .sup.1D.sub.2.fwdarw..sup.3H.sub.4 transitions.

    [0045] The SEM image of the fluorescence temperature measurement material according to Example 1 was detected by a scanning electron microscope, and the result was shown in FIG. 3. As can be seen from FIG. 3, the Na.sub.0.85Sr.sub.0.15TaO.sub.3:0.5% Pr.sup.3+ fluorescence temperature measurement material prepared in Example 1 had a particle size of less than 1 ?m, which had a good spatial resolution.

    [0046] The temperature-dependent spectrum test was performed on the fluorescence temperature measurement material according to Example 1 by using an FLS980 fluorescence spectrometer, FIG. 4 was a temperature change spectrum (303 K to 483 K) of the material according to Example 1, and it can be seen from FIG. 4 that the Na.sub.0.85Sr.sub.0.15TaO.sub.3:0.5% Pr.sup.3+ material in Example 1 had two strong main emission peaks at 492 nm and 610 nm in the range of 303 K to 483 K, which indicated that the material in Example 1 had better temperature sensing performance in the range of 303 K to 483 K; FIG. 5A was an intensity (303 K to 483 K) of an emission peak at 492 nm (blue) of the fluorescence temperature measurement material according to Example 1, FIG. 5B was an intensity (303 K to 483 K) of an emission peak at 610 nm (red) of the fluorescence temperature measurement material according to Example 1, as shown in FIG. 5A, the emission peak intensity I.sub.492 (integrated intensity between 480 nm and 510 nm) at 492 nm (.sup.3P.sub.0.fwdarw..sup.3H.sub.4) decreased significantly as the temperature increases from 303 K to 483 K, while the emission peak intensity I.sub.610 (integrated intensity between 585 nm and 638 nm) at 610 nm (.sup.1D.sub.2.fwdarw..sup.3H.sub.4) increased first and then decreased (as shown in FIG. 5B), and the fluorescence intensity ratio I.sub.610/I.sub.492 (.sup.1D.sub.2.fwdarw..sup.3H.sub.4/.sup.3P.sub.0.fwdarw..sup.3H.sub.4) had a certain exponential function relationship with the temperature; and FIG. 6 was a fitted plot of the fluorescence intensity ratio (.sup.1D.sub.2+3H.sub.4/.sup.3P.sub.0.fwdarw..sup.3H.sub.4) of the fluorescence temperature measurement material according to Example 1. The temperature of a to-be-measured object can be obtained by calculating the ratio I.sub.610/I.sub.492 of the emission peak intensity at 492 nm to the emission peak intensity at 610 nm and then comparing this ratio in an exponential function graph.

    [0047] FIG. 7 was a CIE color coordinate of the fluorescence temperature measurement material according to Example 1 in a range of 303 K to 483 K, in which the CIE color coordinates changed significantly with the increase of temperature, which indicated that the temperature can be determined by the change of the luminescence color of the sample. FIG. 8 was a curve of an absolute sensitivity Sa and a relative sensitivity S.sub.r of the fluorescence temperature measurement material according to Example 1, which indicated that the material had an ultrahigh temperature measurement sensitivity.

    [0048] FIG. 9 was a temperature-dependent spectrum of the fluorescence temperature measurement material according to Example 2, and FIG. 10 was a temperature-dependent spectrum of the fluorescence temperature measurement material according to Example 3, with an excitation wavelength of 290 nm. It can be seen from FIGS. 9 and 10, the Na.sub.0.9Sr.sub.0.1TaO.sub.3:0.5% Pr.sup.3+ material in Example 2 and the Na.sub.0.8Sr.sub.0.2TaO.sub.3:0.5% Pr.sup.3+ material in Example 3 both had two strong emission main peaks at 492 nm and 610 nm in the range of 303 K to 483 K, and the fluorescence intensity ratio I.sub.610/I.sub.492 (.sup.1D.sub.2.fwdarw..sup.3H.sub.4/.sup.3P.sub.0.fwdarw..sup.3H.sub.4) had a certain exponential function relationship with temperature, and therefore, the above materials can be used as temperature measurement materials.

    [0049] FIG. 11A was an XRD pattern of the fluorescence temperature measurement material according to Comparative Example 1, and FIG. 11B was an emission spectrum (?.sub.ex=290 nm) at room temperature of the fluorescence temperature measurement material according to Comparative Example 1; FIG. 12 was a temperature-dependent spectrum (303 K to 483 K) of the fluorescence temperature measurement material according to Comparative Example 1 with an excitation wavelength of 290 nm; it can be seen from FIG. 12 that the emission peak intensity at 492 nm and the emission peak intensity at 610 nm of the fluorescence temperature measurement material according to Comparative Example 1 both significantly decreased as the temperature increases, in this case, the 610 nm emission peak cannot be used as a reference signal, indicating that the Na.sub.0.95Sr.sub.0.05TaO.sub.3:0.5% Pr.sup.3+ material in Comparative Example 1 was not advantageous for use as a proportional temperature sensing material.

    [0050] FIG. 13A was an XRD pattern of the fluorescence temperature measurement material according to Comparative Example 2, and FIG. 13B was an emission spectrum (?.sub.ex=290 nm) at room temperature of the fluorescence temperature measurement material according to Comparative Example 2; FIG. 14 was a temperature-dependent spectrum (303 K to 483 K) of the fluorescence temperature measurement material according to Comparative Example 2 with an excitation wavelength of 290 nm; it can be seen from FIG. 14 that the Na.sub.0.7Sr.sub.0.3TaO.sub.3:0.5% Pr.sup.3+ material in Comparative Example 2 showed a slow decrease in emission peak intensity at 492 nm as the temperature increases, which was not conducive to monitoring signals as a temperature probe.

    [0051] Finally, it should be noted that the foregoing examples are merely intended for illustrating the technical solutions of the present disclosure and do not limit the protection scope of the present disclosure. Although the present disclosure is described in detail with reference to the preferred examples, those of ordinary skill in the art understand that the technical solutions of the present disclosure can be modified or equivalently substituted without departing from the essence and scope of the technical solutions of the present disclosure.