FIBROUS STERILIZABLE MATERIAL FOR PACKAGING OF MEDICAL DEVICES AND TRAY OBTAINED FROM THE MATERIAL

20240279877 ยท 2024-08-22

    Inventors

    Cpc classification

    International classification

    Abstract

    The present disclosure provides for a sterilizable fibrous material for the packaging of medical devices intended to be sterilized, characterized in that the sterilizable fibrous material is in the form of a cardboard having a basis weight of at least 180 g/m2; and the sterilizable fibrous material comprises a mixture of fibers containing at least 75% by weight of natural cellulose fibers, the length of which is less than 5 mm; and the sterilizable fibrous material has a permeability to air of at least 1.7 ?m/Pas at a pressure of 1.47 kPa, measured according to the ISO 5636-3 standard. The present disclosure also provides for a sterilizable tray obtained from the sterilizable fibrous material.

    Claims

    1. A sterilizable fibrous material for the packaging of medical devices intended to be sterilized, characterized in that: the sterilizable fibrous material is in the form of a cardboard having a basis weight of at least 180 g/m.sup.2, advantageously at least 200 g/m.sup.2; and the sterilizable fibrous material comprises a mixture of fibers containing at least 75% by weight of natural cellulose fibers, the length of which is less than 5 mm; and the sterilizable fibrous material has a permeability to air of at least 1.7, preferably at least 3.4, advantageously strictly greater than 3.4 ?m/Pa.Math.s at a pressure of 1.47 kPa, measured according to the ISO 5636-3 standard; and the sterilizable fibrous material has pores having a maximum pore diameter less than 50 ?m, measured according to appendix C of standard EN 868-3.

    2. The material according to claim 1, characterized in that the mixture of fibers contains: between 20% and 100% by weight of natural long cellulose fibers, the length of which is between 1 and 5 mm; and between 0% and 80% by weight of natural short cellulose fibers, the length of which is less than 1 mm.

    3. The material according to claim 2, characterized in that the long cellulose fibers have a length between 1 and 3 mm, and preferably between 1.4 and 2.5 mm, and the short cellulose fibers have a length between 0.2 and 1 mm, preferably between 0.3 and 1 mm.

    4. The material according to claim 2, characterized in that the ratio of long cellulose fibers to short cellulose fibers is between 2 and 30.

    5. The material according to claim 2, characterized in that the mixture of fibers further comprises between at least 1% by weight, preferably between 1 and 20% by weight, of chemical fibers having a titer between 0.3 and 10 dtex and a length between 2.5 and 20 mm.

    6. The material according to claim 5, characterized in that the chemical fibers are chosen from the group comprising artificial fibers such as Lyocell and rayon, and synthetic fibers such as biopolymers like polylactic acid, polyhydroxyalcanoate, polybutylene succinates, polybutylene succinate co-adipates, polycaprolactones, polybutyrate adipate terephthalate, poly(hydroxybutyrate-co-hydroxyvalerate) or their copolymers.

    7. The material according to claim 2, characterized in that the mixture of fibers is biosourced and/or recyclable and/or biodegradable.

    8. The material according to claim 7, characterized in that it is biodegradable to a degree of 90%, or even 95% and meets the requirements of biodegradation of the standard EN 13432.

    9. The material according to claim 1, characterized in that it has the pores have a mean pore diameter less than 35 ?m preferably between 10 ?m and 35 ?m, measured according to appendix C of standard EN 868-3.

    10. The material according to claim 1, characterized in that the sterilizable fibrous material it further comprises a wet strength agent, representing between 0.15 and 1% by dry weight with respect to the dry weight of cellulose.

    11. The material according to claim 10, characterized in that the wet strength agent is chosen from the group comprising polyamine epichlorhydrin (PAE), glyoxalated resins, such as glyoxalated polyamide (GPAM), formaldehyde-based resin.

    12. The material according to claim 1, characterized in that the sterilizable fibrous material it further comprises a sizing agent representing between 0.15 and 1% by dry weight in relation to the dry weight of cellulose.

    13. The material according to claim 12, characterized in that the sizing agent is chosen from the group comprising alkyl ketene dimer (AKD), alkenyl succinic anhydride (ASA), and rosin-based resin.

    14. The material according to claim 1, characterized in that the sterilizable fibrous material it has a COBB at 60 seconds, measured according to the standard ISO 535, of less than 20 g/m.sup.2.

    15. The material according to claim 1, characterized in that the sterilizable fibrous material complies with the standard DIN 58953-6 sections 3 and 4 in terms of a bacterial barrier.

    16. The material according to claim 1, characterized in that the sterilizable fibrous material complies with the standard ISO 11607-1.

    17. The material according to claim 1, characterized in that the sterilizable fibrous material has a thickness of at least 200, preferably 300 ?m.

    18. The material according to claim 1, characterized in that the sterilizable fibrous material further comprises a coating layer, the composition of which is able to render it heat-scalable.

    19. The material according to claim 1, characterized in that it the sterilizable fibrous material is calendered.

    20. The material according to claim 2, characterized in that the mixture of fibers further comprises between 1 and 40% by weight mercerized cellulose fibers.

    21. The material according to claim 1, characterized in that the sterilizable fibrous material has a resistance to bending of at least 1000 mN in a machine direction and a cross-machine direction, preferably at least 2000 mN, and even more preferably at least 3000 mN, according to ISO 2493-2, Paper and Board, Determination of resistance to bending, part 2: Taber Type tester Standard.

    22. The material according to any one of the preceding claims, characterized in that the sterilizable fibrous material has an overall water vapor transmission rate (WVTR) of at least 400 g/m.sup.2/day, preferably at least 800 g/m.sup.2/day, and more preferably at least 1000 g/m.sup.2/day at 38? C. and 90% relative humidity and as determined by the ISO 2528 standard.

    23. Tray obtained from the sterilizable fibrous material specified by claim 1.

    24. Use of the material specified by claim 1 for the fabrication of a tray.

    25. Packaging for medical sterilization, comprising: a tray according to claim 23; and a means of tight sterilizable closure of the tray.

    26. A container for the packaging of medical devices intended to be sterilized comprising: a tray comprising a sterilizable fibrous material, the tray extending from a first end to a second end, and extending from a first side to a second side, the first end including a first end wall and the second end including a second end wall, and the first side including a first side wall and the second side including a second side wall, with the first and second end walls and the first and second side walls defining a recess or cavity within internal boundaries of the first and second end walls and the first and second side walls, and with a bottom wall at a bottom of the recess or cavity; wherein the tray has a basis weight of at least 180 g/m.sup.2, advantageously at least 200 g/m.sup.2; and wherein the tray comprises a mixture of fibers containing at least 75% by weight of natural cellulose fibers, the length of which is less than 5 mm; and wherein the tray has a permeability to air of at least 1.7, preferably at least 3.4, advantageously strictly greater than 3.4 ?m/Pa.Math.s at a pressure of 1.47 kPa, measured according to the ISO 5636-3 standard; and wherein the tray has pores having a maximum pore diameter less than 50 ?m, measured according to appendix C of standard EN 868-3.

    27. A method for fabricating a container for the packaging of medical devices intended to be sterilized comprising: providing a sterilizable fibrous material; and forming the sterilizable fibrous material in to a tray, the tray extending from a first end to a second end, and extending from a first side to a second side, the first end including a first end wall and the second end including a second end wall, and the first side including a first side wall and the second side including a second side wall, with the first and second end walls and the first and second side walls defining a recess or cavity within internal boundaries of the first and second end walls and the first and second side walls, and with a bottom wall at a bottom of the recess or cavity; wherein the tray has a basis weight of at least 180 g/m.sup.2, advantageously at least 200 g/m.sup.2; and wherein the tray comprises a mixture of fibers containing at least 75% by weight of natural cellulose fibers, the length of which is less than 5 mm; and wherein the tray has a permeability to air of at least 1.7, preferably at least 3.4, advantageously strictly greater than 3.4 ?m/Pa.Math.s at a pressure of 1.47 kPa, measured according to the ISO 5636-3 standard; and wherein the tray has pores having a maximum pore diameter less than 50 ?m, measured according to appendix C of standard EN 868-3.

    28. The method of claim 27, wherein forming the sterilizable fibrous material in to the tray comprises thermoforming the sterilizable fibrous material or molding the sterilizable fibrous material.

    Description

    BRIEF DESCRIPTION OF THE DRAWINGS

    [0053] FIG. 1 is a picture of a tray of the present disclosure prior to sterilization, filled with a dye mixture.

    [0054] FIG. 2 is a picture of a tray of the present disclosure after sterilization, filled with a dye mixture.

    [0055] FIG. 3 is a picture of a tray of the present disclosure containing metallic parts after sterilization.

    [0056] FIGS. 4A to 4D are pictures of samples of the present disclosure, and comparative samples, in a home composter.

    DETAILED DESCRIPTION OF THE DISCLOSURE

    [0057] The exemplary embodiments disclosed herein are illustrative of advantageous sterilizable packages, and systems of the present disclosure and methods/techniques thereof. It should be understood, however, that the disclosed embodiments are merely exemplary of the present disclosure, which may be embodied in various forms. Therefore, details disclosed herein with reference to exemplary sterilizable packages and associated processes/techniques of assembly and use are not to be interpreted as limiting, but merely as the basis for teaching one skilled in the art how to make and use the advantageous sterilizable packages and/or alternative sterilizable packages of the present disclosure.

    EXAMPLE 1: Composition of the Samples

    [0058] a. Composition of the Samples of the Present Disclosure

    [0059] Pour samples were fabricated, the compositions of which are given in Table 1:

    TABLE-US-00001 TABLE 1 Sample Sample Sample Sample Sample Sample 1 2 3 4 5 6 Mixture Sodra 76% 72% 76% 72% 72% 32% of fibers Blue Z Navia 19% 18% 19% 18% 18% 16% Botnia 32% Nordic Buckeye 20% HPZ Danufil 5% 10% 1.7 dtex - 5 mm Danufil 10% 1.7 dtex - 3 mm Lyocell 5% 10% 1.7 dtex - 5 mm Wet Kymene 0.4 0.4 0.4 0.4 0.4 0.4 strength 20 Xcel agent (in dry weight of cellulose) Starch Hi Cat 0.5 0.5 0.5 0.5 0.5 0.5 1134A Sizing Fennosize 0.5 0.5 0.5 0.5 0.5 0.5 agent 860D Coating 4.3 with 5.4 with 4.5 with 5.2 with 7 with 27 with (g/m.sup.2)? 90% PVA, 90% PVA, 90% PVA, 90% PVA, 61% PVA, 63% PVA, 5% AKD 5% AKD 5% AKD 5% AKD 7% AKD, 4% AKD, and 5% and 5% and 5% and 5% 30.5 acrylic 32% acrylic Kymene Kymene Kymene Kymene latex, 1.5% latex, 1% zirconium zirconium salt salt
    b. Comparison Sample

    [0060] This was a spunbond composed of fibers of polypropylene and polyolefin with a basis weight of 90.5 g/m.sup.2 marketed by AMCOR 095134 under the brand name ULTRA?.

    EXAMPLE 2: Characteristics of the Samples

    [0061] The major characteristics of the different samples are given in Table 2.

    TABLE-US-00002 TABLE 2 EN Sample Sample Sample Sample Sample Sample Comp. Test Unit 868-3 1 2 3 4 5 6 sample Weight ISO g/m.sup.2 297 275 278 293 284 332 90.5 536 Thickness ISO ?m 466 474 450 485 441 476 174 534 Permeability ml/ >400 ml/ 655 1332 650 1104 2014 974 1696 to air ISO min min at 5636-3 1.47 kPa Cobb (60 s) g/m.sup.2 <20 g/m.sup.2 19.66 19.64 19.4 19.66 18.7 17 2.8 ISO 535 Pore diameter ?m <=35 12.6 15.9 14.1 15.2 14.6 15 26 EN 868-3 (max 50) (18) (22.6) (19.3) (21.5) (21) (19.5) (41) Appendix C

    [0062] Samples 1 to 5 were compliant with the standard ISO 11607-1.

    EXAMPLE 3: Characteristics Of The Sealing Of The Tray According To The Present Disclosure

    [0063] a. By a Brugger Test

    [0064] Sample 4 was calendered (30 kN/m at 50? C.) and subjected to a heat sealing test with a medical film using Brugger clamps at 180? C. with a force of 4 bars for 10 seconds. No tear-off of fibers was observed and the seal showed a value greater than 1.5 N/15 mm.

    b. By Injection of a Dye Per Standard ASTM F 1929
    i. Quality of the Seal in Absence of Sterilization

    [0065] Sample 5 was thermoformed to produce a tray. The tray was sealed with a medical film for sterilization by steam or ethylene oxide made of PP/PET. An aqueous solution with a blue dye and a surfactant was injected in the tray. The colored solution was moved to the area of the sealing points on each side and maintained for 5 seconds each time, for a total duration of 20 seconds per the standard ASTM F 1929. No leak was observed (see FIG. 1). Following the standard EN 868-5, calendered sample 6 was subjected to a heat sealing test as with a medical film using Brugger clamps at 190? C. with a force of 5 bars for respectively 2, 3, 4 and 5 seconds. All samples had a peel strength greater than 1.5 N/15 mm but remained less than 5 N/15 mm. This showed that a tray according to the present disclosure can be properly sealed with a medical film and be opened without significant effort and without fibers tearing. As shown in FIG. 1, the tray extends from a first end to a second end (from left to right), and extends from a first side to a second side (from the top of FIG. 1 to the bottom of FIG. 1). The first end includes a first end wall, and the second end includes a second end wall. The first side includes a first side wall, and the second side includes a second side wall. The first and second end walls and the first and second side walls define a recess or cavity within the internal boundaries of the first and second end walls and the first and second side walls (e.g., with the recess/cavity housing the dye mixture), with a bottom wall at the bottom of the recess/cavity.

    ii. Quality of the Seal after Sterilization

    [0066] Sample 5 was thermoformed to produce a tray. The tray was then sealed with a medical film for sterilization by steam or ethylene oxide made of PP/PET, and then sterilized with steam for 18 min at 134? C. An aqueous solution with a blue dye and a surfactant was then injected in the tray. The colored solution was moved to the area of the sealing points on each side and maintained for 5 seconds each time, for a total duration of 20 seconds per the standard ASTM F 1929. No leak was observed (see FIG. 2).

    EXAMPLE 4: Test for Evaluation of Microbiological Contamination of a Tray According to the Present Disclosure

    [0067] The purpose of this test was to show the ability of the material of the present disclosure to let the sterilizing steam pass through it, thus allowing a sterilization of the elements contained in it.

    [0068] To do this, a tray was fabricated from sample 5. On this tray was arranged 22 metallic pieces contaminated with microorganisms contained in saliva. The tray was then heat sealed with a film for sterilization by steam or ethylene oxide made of PP/PET. The tray was then sterilized with steam at 134? C. for 18 minutes. The tray was stored under normal conditions of pressure and temperature for 15 days prior to microbial analysis of 3 metallic pieces, chosen at random from the 22 pieces.

    [0069] The microbial contamination was evaluated by the method ISO 8784-1 and was based on a counting of bacteria and fungi. The medium used to culture the bacteria was Tryptic Soy Agar with an incubation of 48 h at 37? C. The medium used to culture the fungi was PDA with an incubation of 5 days at 29? C.

    [0070] The results are shown in Table 3:

    TABLE-US-00003 TABLE 3 Sample N of fungi N of bacteria Metallic piece 1 <10 CFU/part <10 CFU/part Metallic piece 2 <10 CFU/part <10 CFU/part Metallic piece 3 <10 CFU/part <10 CFU/part CFU: colony forming unit N: total number of microorganisms calculated in UFC per piece

    [0071] The results show that, after sterilization, the metallic pieces are no longer contaminated, which proves that the material of the present disclosure is permeable to the sterilizing steam and remains sterile after a storage at room temperature for 15 days. This test also showed the mechanical strength of the tray (see FIG. 3). In fact, it was possible to sterilize 22 metallic pieces weighing more than 1 kg and store them in sterile manner without any tearing or opening.

    EXAMPLE 5: Test for Bacterial Resistance of the Material of the Present Disclosure

    [0072] a. In Humid Conditions

    [0073] The test was performed on the 4 samples (samples 1-4) according to the present disclosure and one comparison sample under the conditions of the standard DIN 58953-6, section 3. Basically, the test involved sterilizing each sample with steam at 134? C. for 4 minutes. Each sample was then inoculated with 500 ?L of the microbe S. epidermidis on one surface. After drying, the opposite surface of the sample was placed in contact with a culture medium and incubated for 24 h at 37? C. The two surfaces of each sample were tested.

    [0074] The results are shown in Table 4.

    TABLE-US-00004 TABLE 4 Test for bacteria resistance in humid conditions per standard DIN 58953-6, section 3 Number of CFU/agar plate Sample Surface 1 2 3 4 5 ? Comparison sample A 0 0 0 0 0 0 (Amcor ultra white B 0 0 0 0 0 0 spunbond 90 g/m.sup.2 Sample 1 A 0 0 0 0 0 0 B 0 0 0 0 0 0 Sample 2 A 0 0 0 0 0 0 B 0 0 0 0 0 0 Sample 3 A 0 0 0 0 0 0 B 0 0 0 0 0 0 Sample 4 A 0 0 0 0 0 0 B 0 0 0 0 0 0 Sample 6 A 0 0 0 0 0 0 B 0 0 0 0 0 0 CFU = colony forming unit

    [0075] No colony was found on any of the samples of the present disclosure or the comparison sample. The test for bacteria resistance in humid conditions according to standard DIN 58953-6, section 3 was thus validated.

    b. In Dry Conditions
    i. Standard DIN 58953-6, Section 4

    [0076] The test was performed on 4 samples (samples 1-4) according to the present disclosure and one comparison sample under the conditions of the standard DIN 58953-6, section 4. Basically, the test involved sterilizing each sample with steam at 121? C. for 20 minutes. Each sample was then inoculated with 250 mg of sand contaminated with endospores of B. subtilis. Incubations were then done at 8? C. and 50? C. to create a flow of air through the contaminated sand and the sample as far as the agar plate. The samples were then incubated for 24 h at 37? C. The 2 surfaces of each sample were tested.

    TABLE-US-00005 TABLE 5 Test for bacteria resistance in dry conditions per standard DIN 58953-6, section 4 Number of CFU/agar plate Sample Surface 1 2 3 4 5 ? Comparison sample A 0 0 0 0 0 0 (Amcor ultra white B 0 0 0 0 0 0 spunbond 90 g/m.sup.2 Sample 1 A 0 0 0 0 0 0 B 0 0 0 0 0 0 Sample 2 A 0 0 0 0 0 0 B 0 0 0 0 0 0 Sample 3 A 0 0 0 0 0 0 B 0 0 0 0 0 0 Sample 4 A 0 0 0 0 0 0 B 0 0 0 0 0 0 Sample 6 A 0 0 0 0 0 0 B 0 0 0 0 0 0 CFU = colony forming unit

    [0077] As expected, no colony was found on any of the samples of the present disclosure or the comparison sample. This was expected, inasmuch as the test in humid environment according to standard DIN 58953-6 section 3 is challenging, water serving as the vector of contamination. The test for bacteria resistance under dry conditions according to standard DIN 58953-6, section 4 was thus validated.

    ii. Bacterial Filtration Efficiency (BFE)

    [0078] This test was performed according to the standard ASTM F2101, using an aerosol of Staphylococcus aureus. This test expresses the ratio of the number of bacteria arrested per sample, divided by the number of bacteria sprayed onto the sample tested.

    TABLE-US-00006 TABLE 6 Bacterial Bacterial Bacterial filtration Comparison filtration Comparison filtration Sam- efficiency sample, 1 efficiency sample, 2 efficiency ple 5 (BFE) % sheet (BFE) % sheets (BFE) % 1 99.9 1 76.8 1 93.6 2 99.9 2 71.0 2 91.0 3 99.9 3 71.5 3 92.0 4 99.9 4 73.4 4 94.3 5 99.9 5 63.5 5 94.9

    [0079] As shown by Table 6, sample 5 has a BFE of 99.9% as compared to the comparison sample, which has a BFE of 71.24% for a packaging of 1 sheet and 93.16% for a packaging of 2 sheets. The sample according to the present disclosure presents a more tortuous pathway for the microorganisms and thus minimizes the risks of contamination.

    [0080] Breathability is the ability of a material to allow water vapor to pass through it while preventing liquid water from passing through it, and is expressed interchangeably as water vapor transmission rate (WVTR), or moisture vapor transmission rate (MVTR). Accordingly, the material of the present disclosure has an overall water vapor transmission rate (WVTR) of at least 400 g/m.sup.2/day, preferably at least 800 g/m.sup.2/day, and more preferably at least 1000 g/m.sup.2/day at 38? C. and 90% relative humidity and as determined by the ISO 2528 standard.

    Example on MVTR and Bowie and Dick Test:

    [0081] The WVTR of sample 6 and comparative samples were measured according to the ISO 2528 standard, and the results are summarized in Table 7.

    TABLE-US-00007 TABLE 7 WVTR Samples (g/m.sup.2/day) Sample 6 (single layer) 2071 Sample 6 (double layer) 1698 Comparative sample 2146 (single layer) Comparative sample 1492 (single layer)

    [0082] The samples were measured following the ISO 2528 standard and at 38? C. and 90% relative humidity. The data shows that sample 6 has comparable WVTR to the comparative sample, while the basis weight of sample 6 is more than triple of the comparative sample. The material according to the present disclosure can thus be efficiently sterilized.

    [0083] In order to verify the uniformity and efficiency of the vapor penetration, sample 6 was thermoformed into a tray and a Bowie and Dick test kit was placed in the tray. The tray was then sealed with a medical film and was subjected to a sterilization cycle (Bowie and Dick cycle: 3.5 minutes at 134? C.). After sterilization, the Bowie and Dick kit test showed efficient and uniform steam penetration.

    Biodegradability and Compostability

    [0084] Two of each sample 6 (second from top row, and bottom row of FIGS. 4A to 4D) and comparative samples (top row, and one above the bottom row of FIGS. 4A to 4D) were placed in frame and were introduced in a home composter. The samples were then visually inspected after one, two weeks and four weeks. As it can be seen in FIGS. 4A to 4D, sample 6 has almost completely biodegraded after four weeks, whereas the comparative samples remained intact.

    Stiffness and Rigidity

    [0085] In order to form a rigid tray, the material according to the present disclosure had a higher stiffness compared to other sterilization materials such as pouches or sterilization wraps. The rigidity or stiffness of a material may be measured by determining the resistance to bending of the material following ISO 2493-2-Paper and BoardDetermination of resistance to bendingpart 2: Taber Type tester Standard. The material according to the present disclosure may have a resistance to bending of at least 1000 mN in MD and CD directions, preferably at least 2000 mN and even more preferably at least 3000 mN.

    Example On Resistance To Bending

    [0086] The resistance to bending of Sample 6 and comparative sample were measured in the machine direction (MD) and the cross-machine direction (CD) (see Table 8, below) following the ISO 2493-2 standard.

    TABLE-US-00008 TABLE 8 Resistance in Resistance in MD direction CD direction (mN) (mN) Sample 6 6292 4938 Comparative 282 238 Sample

    [0087] The Comparative sample had a basis weight of 90 g/m.sup.2 and sample 6 had a basis weight of 332 g/m.sup.2 but the material according to the present disclosure had a much higher resistance to bending. Such resistance makes sample 6 suitable for making a rigid tray according to the present disclosure.

    [0088] This disclosure further encompasses the following aspects.

    [0089] Aspect 1. A sterilizable fibrous material for the packaging of medical devices intended to be sterilized, characterized in that the sterilizable fibrous material is in the form of a cardboard having a basis weight of at least 180 g/m.sup.2, advantageously at least 200 g/m.sup.2; and the sterilizable fibrous material comprises a mixture of fibers containing at least 75% by weight of natural cellulose fibers, the length of which is less than 5 mm; and the sterilizable fibrous material has a permeability to air of at least 1.7, preferably at least 3.4, advantageously strictly greater than 3.4 ?m/Pa.Math.s at a pressure of 1.47 kPa, measured according to the ISO 5636-3 standard; and the sterilizable fibrous material has pores having a maximum pore diameter less than 50 ?m, measured according to appendix C of standard EN 868-3.

    [0090] Aspect 2. The material according to Aspect 1, characterized in that the mixture of fibers contains: between 20% and 100% by weight of natural long cellulose fibers, the length of which is between 1 and 5 mm; and between 0% and 80% by weight of natural short cellulose fibers, the length of which is less than 1 mm.

    [0091] Aspect 3. The material according to any one of the preceding Aspects, characterized in that the long fibers have a length between 1 and 3 mm, and preferably between 1.4 and 2.5 mm, and the short fibers have a length between 0.2 and 1 mm, preferably between 0.3 and 1 mm.

    [0092] Aspect 4. The material according to any one of the preceding Aspects, characterized in that the ratio of long fibers to short fibers is between 2 and 30.

    [0093] Aspect 5. The material according to any one of the preceding Aspects, characterized in that the mixture of fibers further comprises between at least 1% by weight, preferably between 1 and 20% by weight, of chemical fibers having a titer between 0.3 and 10 dtex and a length between 2.5 and 20 mm.

    [0094] Aspect 6. The material according to any one of the preceding Aspects, characterized in that the chemical fibers are chosen from the group comprising artificial fibers such as Lyocell and rayon, and synthetic fibers such as biopolymers like polylactic acid, polyhydroxyalcanoate, polybutylene succinates, polybutylene succinate co-adipates, polycaprolactones, polybutyrate adipate terephthalate, poly(hydroxybutyrate-co-hydroxyvalerate) or their copolymers.

    [0095] Aspect 7. The material according to any one of the preceding Aspects, characterized in that the mixture of fibers is biosourced and/or recyclable and/or biodegradable.

    [0096] Aspect 8. The material according to any one of the preceding Aspects, characterized in that it is biodegradable to a degree of 90%, or even 95% and meets the requirements of biodegradation of the standard EN 13432.

    [0097] Aspect 9. The material according to any one of the preceding Aspects, characterized in that it has pores having a mean pore diameter less than 35 ?m preferably between 10 ?m and 35 ?m, measured according to appendix C of standard EN 868-3.

    [0098] Aspect 10. The material according to any one of the preceding Aspects, characterized in that it further comprises a wet strength agent, representing between 0.15 and 1% by dry weight with respect to the dry weight of cellulose.

    [0099] Aspect 11. The material according to Aspect 10, characterized in that the wet strength agent is chosen from the group comprising polyamine epichlorhydrin (PAE), glyoxalated resins, such as glyoxalated polyamide (GPAM), formaldehyde-based resin.

    [0100] Aspect 12. The material according to any one of the preceding Aspects, characterized in that it further comprises a sizing agent representing between 0.15 and 1% by dry weight in relation to the dry weight of cellulose.

    [0101] Aspect 13. The material according to Aspect 12, characterized in that the sizing agent is chosen from the group comprising alkyl ketene dimer (AKD), alkenyl succinic anhydride (ASA), and rosin-based resin.

    [0102] Aspect 14. The material according to any one of the preceding Aspects, characterized in that it has a COBB at 60 seconds, measured according to the standard ISO 535, of less than 20 g/m.sup.2.

    [0103] Aspect 15. The material according to any one of the preceding Aspects, characterized in that it complies with the standard DIN 58953-6 sections 3 and 4 in terms of a bacterial barrier.

    [0104] Aspect 16. The material according to any one of the preceding Aspects, characterized in that it complies with the standard ISO 11607-1.

    [0105] Aspect 17. The material according to any one of the preceding Aspects, characterized in that it has a thickness of at least 200, preferably 300 ?m.

    [0106] Aspect 18. The material according to any one of the preceding Aspects, characterized in that it further comprises a coating layer, the composition of which is able to render it heat-sealable

    [0107] Aspect 19. The material according to any one of the preceding Aspects, characterized in that it is calendered.

    [0108] Aspect 20. The material according to any one of the preceding Aspects, characterized in that the mixture of fibers further comprises between 1 and 40% by weight mercerized cellulose fibers.

    [0109] Aspect 21. The material according to any one of the preceding Aspects, characterized in that the material has a resistance to bending of at least 1000 mN in a machine direction and a cross-machine direction, preferably at least 2000 mN, and even more preferably at least 3000 mN, according to ISO 2493-2, Paper and Board, Determination of resistance to bending, part 2: Taber Type tester Standard.

    [0110] Aspect 22. The material according to any one of the preceding Aspects, characterized in that the material has an overall water vapor transmission rate (WVTR) of at least 400 g/m.sup.2/day, preferably at least 800 g/m.sup.2/day, and more preferably at least 1000 g/m.sup.2/day at 38? C. and 90% relative humidity and as determined by the ISO 2528 standard.

    [0111] Aspect 23. Tray obtained from the material specified by any one of Aspects 1 to 22.

    [0112] Aspect 24. Use of the material specified by any one of Aspects 1 to 22 for the fabrication of a tray.

    [0113] Aspect 25. Packaging for medical sterilization, comprising: a tray according to Aspect 23; and a means of tight sterilizable closure of the tray.

    [0114] Aspect 26. A container for the packaging of medical devices intended to be sterilized comprising: a tray comprising a sterilizable fibrous material, the tray extending from a first end to a second end, and extending from a first side to a second side, the first end including a first end wall and the second end including a second end wall, and the first side including a first side wall and the second side including a second side wall, with the first and second end walls and the first and second side walls defining a recess or cavity within internal boundaries of the first and second end walls and the first and second side walls, and with a bottom wall at a bottom of the recess or cavity; wherein the tray has a basis weight of at least 180 g/m.sup.2, advantageously at least 200 g/m.sup.2; and wherein the tray comprises a mixture of fibers containing at least 75% by weight of natural cellulose fibers, the length of which is less than 5 mm; and wherein the tray has a permeability to air of at least 1.7, preferably at least 3.4, advantageously strictly greater than 3.4 ?m/Pa.Math.s at a pressure of 1.47 kPa, measured according to the ISO 5636-3 standard; and wherein the tray has pores having a maximum pore diameter less than 50 ?m, measured according to appendix C of standard EN 868-3.

    [0115] Aspect 27. A method for fabricating a container for the packaging of medical devices intended to be sterilized comprising: providing a sterilizable fibrous material; and forming the sterilizable fibrous material in to a tray, the tray extending from a first end to a second end, and extending from a first side to a second side, the first end including a first end wall and the second end including a second end wall, and the first side including a first side wall and the second side including a second side wall, with the first and second end walls and the first and second side walls defining a recess or cavity within internal boundaries of the first and second end walls and the first and second side walls, and with a bottom wall at a bottom of the recess or cavity; wherein the tray has a basis weight of at least 180 g/m.sup.2, advantageously at least 200 g/m.sup.2; and wherein the tray comprises a mixture of fibers containing at least 75% by weight of natural cellulose fibers, the length of which is less than 5 mm; and wherein the tray has a permeability to air of at least 1.7, preferably at least 3.4, advantageously strictly greater than 3.4 ?m/Pa.Math.s at a pressure of 1.47 kPa, measured according to the ISO 5636-3 standard; and wherein the tray has pores having a maximum pore diameter less than 50 ?m, measured according to appendix C of standard EN 868-3.

    [0116] Aspect 28. The method of Aspect 27, wherein forming the sterilizable fibrous material in to the tray comprises thermoforming the sterilizable fibrous material or molding the sterilizable fibrous material.

    [0117] While particular embodiments have been described, alternatives, modifications, variations, improvements, and substantial equivalents that are or may be presently unforeseen may arise to applicants or others skilled in the art. Accordingly, the appended claims as filed and as they may be amended are intended to embrace all such alternatives, modifications variations, improvements, and substantial equivalents.

    [0118] All ranges disclosed herein are inclusive of the endpoints, and the endpoints are independently combinable with each other (e.g., ranges of up to 25 wt. %, or, more specifically, 5 wt. % to 20 wt. %, is inclusive of the endpoints and all intermediate values of the ranges of 5 wt. % to 25 wt. %, etc.). Combinations is inclusive of blends, mixtures, alloys, reaction products, and the like. The terms first, second, and the like, do not denote any order, quantity, or importance, but rather are used to distinguish one element from another. The terms a and an and the do not denote a limitation of quantity and are to be construed to cover both the singular and the plural, unless otherwise indicated herein or clearly contradicted by context. Or means and/or unless clearly stated otherwise. Reference throughout the specification to some embodiments, an embodiment, and so forth, means that a particular element described in connection with the embodiment is included in at least one embodiment described herein, and may or may not be present in other embodiments. In addition, it is to be understood that the described elements may be combined in any suitable manner in the various embodiments. A combination thereof is open and includes any combination comprising at least one of the listed components or properties optionally together with a like or equivalent component or property not listed.

    [0119] Unless defined otherwise, technical and scientific terms used herein have the same meaning as is commonly understood by one of skill in the art to which this application belongs. All cited patents, patent applications, and other references are incorporated herein by reference in their entirety. However, if a term in the present application contradicts or conflicts with a term in the incorporated reference, the term from the present application takes precedence over the conflicting term from the incorporated reference.

    [0120] Unless specified to the contrary herein, all test standards are the most recent standard in effect as of the filing date of this application, or, if priority is claimed, the filing date of the earliest priority application in which the test standard appears.

    [0121] Although the materials, systems and methods of the present disclosure have been described with reference to exemplary embodiments thereof, the present disclosure is not limited to such exemplary embodiments and/or implementations. Rather, the materials, systems and methods of the present disclosure are susceptible to many implementations and applications, as will be readily apparent to persons skilled in the art from the disclosure hereof. The present disclosure expressly encompasses such modifications, enhancements and/or variations of the disclosed embodiments. Since many changes could be made in the above construction and many widely different embodiments of this disclosure could be made without departing from the scope thereof, it is intended that all matter contained in the drawings and specification shall be interpreted as illustrative and not in a limiting sense. Additional modifications, changes, and substitutions are intended in the foregoing disclosure. Accordingly, it is appropriate that the appended claims be construed broadly and in a manner consistent with the scope of the disclosure.