System and method of delicate membrane condensed phase membrane introduction mass spectrometry (CP-MIMS)
09583325 ยท 2017-02-28
Inventors
Cpc classification
International classification
Abstract
Systems and methods for analyzing a sample comprising an analyte selected from a volatile organic compound, a semi-volatile organic compound, a non-volatile organic compound, a polar organic compound and a halogenated non-volatile organic compound are provided. The systems comprises an ionization source, a flow cell or an immersion probe with a delicate membrane, the flow cell or immersion probe for accepting the sample, and the delicate membrane interface in fluid communication with the ionization source and a mass spectrometer. The flow cell system further comprises a simultaneously matched pumping in and out delivery (SMPIOD) system for delivering an acceptor phase comprising the analyte from the delicate membrane interface to the mass spectrometer at a constant acceptor flow pressure and a constant acceptor flow rate.
Claims
1. A system for analyzing a sample comprising an analyte, the system comprising: an acceptor phase supply comprising an acceptor phase; an ionization source; a mass spectrometer; and a membrane interface device, the device comprising a membrane interface in fluid communication with an acceptor phase carrier, the membrane interface configured for bathing in the sample, under ambient pressure, the acceptor phase carrier in fluid communication with the acceptor phase supply, the ionization source and the mass spectrometer.
2. The system of claim 1, wherein the membrane interface is a hollow fibre membrane (HFM) comprising polydimethylsiloxane, of no more than about 225 microns in thickness.
3. The system of claim 2, wherein the membrane is a thin membrane.
4. The system of claim 2, wherein the membrane is a supported liquid membrane.
5. The system of claim 2, wherein the probe is a J probe.
6. The system of claim 2, wherein the probe is a miniature coaxial probe.
7. The system of claim 6, further comprising an autosampler.
8. The membrane interface device of claim 6, wherein the miniature coaxial probe is an immersion probe, the probe comprising a membrane interface coaxial with-an acceptor phase delivery capillary and for fluid communication with the acceptor phase, the membrane interface configured for bathing in the sample, under ambient pressure, the acceptor phase delivery capillary in fluid communication with the acceptor phase supply, the ionization source and the mass spectrometer.
9. The system of claim 1, further comprising a mixer for mixing the sample.
10. An immersion probe for use with an ionization source and a mass spectrometer, the immersion probe comprising a membrane interface coaxial with an acceptor phase delivery capillary and for fluid communication with an acceptor phase, the membrane interface configured for bathing in the sample, under ambient pressure, the acceptor phase delivery capillary for fluid communication with acceptor phase supply, the ionization source and the mass spectrometer.
11. The probe of claim 10, wherein the membrane interface is a hollow fibre membrane (HFM) comprising polydimethylsiloxane, of about 0.5 microns to about 225 microns in thickness.
12. The probe of claim 11, wherein the HFM is a composite PDMS micro-porous polypropylene HFM, a thin PDMS HFM or a supported liquid membrane HFM.
13. A method of quantifying and measuring a trace level analyte in a sample, the sample being between about 1.0 L to about 1 mL, the method comprising: exposing a membrane interface device to a sample, such that the membrane interface device is bathed in the sample; moving the sample over the membrane interface device; delivering an acceptor phase to the membrane interface device via an acceptor phase carrier; delivering the analyte to an ionization source and to a mass spectrometer; and obtaining an output, thereby quantifying and measuring the trace level analyte.
14. The method of claim 13, wherein the measuring is direct.
15. The method of claim 14, wherein the sample is a biological sample or an environmental sample.
16. The method of claim 15, further comprising rapid prescreening the sample and providing the sample for further analyzing.
17. The method of claim 15, wherein the measuring and quantifying provides direct, in vivo or in situ monitoring of the biological or the environmental sample.
18. The method of claim 14, wherein the acceptor phase includes at least one of an internal standard, an acceptor phase ionization enhancer and an acceptor phase modifier.
19. The method of claim 18, further comprising varying the acceptor flow rate.
20. The method of claim 19, wherein the acceptor phase includes an acceptor phase modifier to provide a PDMS-PIM (Polymer Inclusion Membrane) HFM.
21. The method of claim 14, wherein the membrane interface device is an immersion probe.
Description
BRIEF DESCRIPTION OF THE DRAWINGS
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
(11)
(12)
(13)
(14)
(15)
(16)
(17)
(18)
(19)
(20)
(21)
(22)
(23)
DESCRIPTIONS OF THE PREFERRED EMBODIMENTS
(24) Except as otherwise expressly provided, the following rules of interpretation apply to this specification (written description, claims and drawings): (a) all words used herein shall be construed to be of such gender or number (singular or plural) as the circumstances require; (b) the singular terms a, an, and the, as used in the specification and the appended claims include plural references unless the context clearly dictates otherwise; (c) the antecedent term about applied to a recited range or value denotes an approximation within the deviation in the range or value known or expected in the art from the measurements method; (d) the words herein, hereby, hereof, hereto, hereinbefore, and hereinafter, and words of similar import, refer to this specification in its entirety and not to any particular paragraph, claim or other subdivision, unless otherwise specified; (e) descriptive headings are for convenience only and shall not control or affect the meaning or construction of any part of the specification; and (f) or and any are not exclusive and include and including are not limiting. Further, The terms comprising, having, including, and containing are to be construed as open-ended terms (i.e., meaning including, but not limited to,) unless otherwise noted.
(25) To the extent necessary to provide descriptive support, the subject matter and/or text of the appended claims is incorporated herein by reference in their entirety.
(26) Recitation of ranges of values herein are merely intended to serve as a shorthand method of referring individually to each separate value falling within the range, unless otherwise indicated herein, and each separate value is incorporated into the specification as if it were individually recited herein. Where a specific range of values is provided, it is understood that each intervening value, to the tenth of the unit of the lower limit unless the context clearly dictates otherwise, between the upper and lower limit of that range and any other stated or intervening value in that stated range, is included therein. All smaller sub ranges are also included. The upper and lower limits of these smaller ranges are also included therein, subject to any specifically excluded limit in the stated range.
(27) Unless defined otherwise, all technical and scientific terms used herein have the same meaning as commonly understood by one of ordinary skill in the relevant art. Although any methods and materials similar or equivalent to those described herein can also be used, the acceptable methods and materials are now described.
(28) Definitions
(29) Volatile organic compounds. Volatile compounds (VOC) are molecules characterized by a relatively high vapor pressure, typically greater than about 1,000 Pa.
(30) Semi-volatile compounds. Semi-volatile compounds (SVOC) are molecules with vapor pressures in the range of from about 0.1 Pa to about 1,000 Pa.
(31) Non-volatile organic compounds. Non-volatile organic compounds are molecules with vapor pressures between about 0.1 Pa to about 10.sup.6 Pa, or about 0.01 Pa to about 10.sup.4 Pa or about 10.sup.3 Pa and all ranges therebetween. Non-volatile organic compounds include for example, but not limited to naphthenic acids, organic acids, resin acids, fatty acids, natural organic matter, carboxylic acids, phenols, polyphenols, surfactants, substances of abuse, pharmaceutical compounds, metabolites, hormones, personal care products, flavorings, explosives and preservatives. They range in size from about 100 nl to about 1000 atomic mass units (amu), or about 100 to about 900 amu or about 300 to about 600 amu and all ranges therebetween.
(32) Halogenated non-volatile organic compounds. Halogenated non-volatile compounds are compounds that contain one or more halogen atom. Unlike the non-volatile compounds, they may have a size greater than 1000 amu, or be as large as 2000 amu, depending upon the halogen(s) (iodine, bromine, chlorine and/or fluorine) embodied in their molecular structure.
(33) Polar organic compounds. Polar compounds include for example, but not limited to naphthenic acids, organic acids, resin acids, fatty acids, natural organic matter, carboxylic acids, phenols, polyphenols, surfactants, substances of abuse, pharmaceutical compounds, metabolites, hormones, personal care products, flavorings, explosives and preservatives
(34) Charged compounds. Charged compounds are those compounds that contain ionizable functional groups and are present in ionic form in solution at ambient pH.
(35) On-line measurements. On-line measurements provide an analytical signal by passing the sample through a device without the need for subsequent sample handling.
(36) On-line, real-time analysis. On-line, real-time analysis refers to a method that allows for non-destructive analysis that, in turn, allows for repeated analysis in real-time. The samples can be re-circulated or continuously probed, and changes in the concentrations of measured molecules over time can be monitored.
(37) Analyte refers to a particular molecule or group of molecular species of interest.
(38) Permeate refers to a sample (analyte) after passage through a membrane introduction interface.
(39) Delicate membrane. Non-limiting examples of delicate membranes are thin PDMS, composite PDMS/micro-porous polypropylene HFM and supported liquid membranes. Advantageously, the supported liquid membranes can include carriers to improve or allow analysis of specific analytes. The supported liquid membranes may be about 10 microns to about 250 microns thick, and all ranges therebetween. A thin membrane would be, for example, about 0.5 microns to about 100 microns, or about 5 microns to about 90 microns, or about 25 microns to about 75 microns, or about 35 microns or about 50 microns, and all ranges therebetween. The membranes may be sheets, hollow fibre or other conformations as would be known to one skilled in the art.
(40) Rapid analysis. Rapid analysis in the context of the present technology is less than ten minutes and may be as fast as 0.1 minute, or 0.5 minute, or two minutes and all ranges therebetween.
(41) Consistent acceptor phase flow rate. Consistent flow rate means that the flow rate does not change during the analysis. The rate can be selected from a range of about 100 nl to about 1000 L/minute and all ranges therebetween, but is set at a given rate.
(42) Consistent acceptor phase flow pressure through the membrane. Consistent flow pressure means that the flow pressure does not change during the analysis. The pressure can be selected from a range of about 90 kPa to about 110 kPa, but is set at a given pressure.
(43) Direct measurement. Direct measurement means that the sample is measured without cleanup, pre-concentration and/or chromatographic separations prior to analysis.
(44) Trace level analytes. Trace level analytes in the context of the present technology refers to polar, non-volatile analytes in the concentration range of parts-per-trillion up to parts-per-million.
(45) Membrane interface device. In the context of the present technology, membrane interface device is any unit that holds a membrane that can be in contact with the sample on one side and an acceptor phase carrier on the other such as a flow cell, an immersion probe, a J probe, a miniature coaxial probe, each having a delicate membrane CP-MIMS.
(46) Acceptor phase carrier. In the context of the present technology, a carrier is any article that can deliver acceptor phase to the membrane interface device, for example, but not limited to a capillary, a needle, or a hypodermic tube.
DETAILED DESCRIPTION
(47) Apparatus and Materials
(48) The mass spectrometer used for the presented work was a triple quadrupole system (Micromass Quattro Ultima LC, Waters-Micromass, Altrincham, UK) with an ESI probe and Z-spray source. Nitrogen gas (UHP grade, 99.999% pure) was supplied from a liquid nitrogen Dewar and argon collision gas (UHP grade) was supplied from a compressed gas cylinder (Praxair Inc., Nanaimo, BC, Canada). The base pressure in the vacuum system was 910.sup.6 Torr. Negative ion mode experiments used an ESI capillary voltage of 3 kV. The desolvation gas flow rate was 750 L/hr at 300 C. and the cone curtain gas flow rate was set to 60 L/hr. For MS/MS experiments, the collision cell was maintained at a pressure of 3 mTorr, collision energies and MS parameters are given in Table 1.
(49) The CP-MIMS interface was based on a flow-over capillary hollow fiber membrane (HFM) design described previously in Rapid Communications in Mass Spectrometry, 2011, 25, pp 1141, incorporated herein in its entirety and shown in
(50) A schematic diagram of the experimental apparatus is given in
(51) The flow cell allows for maintenance of sample integrity, allowing for non-destructive analysis. Coupling of the flow cell with a recirculation system then allows for recirculation of the sample, which in turn allows for repeated analysis. Repeated analysis in turn allows for tracking of time dependent changes in a sample. Note that there is no sample injection loop or acceptor phase injection loop. This configuration allows for constantly flowing the acceptor phase with analytes that permeate through the membrane to the mass spectrometer.
(52) Reagents, solvents and target analytes were obtained from a variety of suppliers and were ACS grade or better unless otherwise noted. Abietic acid (Tech. grade, 70%), triclosan (5-chloro-2-(2,4-dichlorophenoxy)phenol), 2,4,6-trichlorophenol, nonylphenol, estrone (3-hydroxy-13-methyl-6,7,8,9,11,12,13,14,15,16-decahydrocyclphena[a]-phenanthren-17-one), gemfibrozil (5-(2,5-dimethylphenoxy)-2,2-dimethyl-pentanoic acid) were all obtained from Sigma Aldrich. Table 1 lists target analytes studied, relevant MS scan parameters, and pertinent physical properties. A commercially available naphthenic acids mixture (Merichem Company, Houston, Tex., USA) that has been characterized by others was used for these studies, in addition to naphthenic acid mixtures and pyrene butyric acid from Sigma-Aldrich Canada. Natural water samples were collected from a typical muskeg river drainage in the Alberta Oil Sands bitumen deposit area (AB, Canada), with 21 ppm dissolved organic carbon (DOC) and pH=6.3. Steam assisted gravity drainage (SAGD) heavy oil extraction process water samples (Alberta Oil Sands) had a measured pH of 11.3 and a specific conductivity of 110,0000/cm. Because of their highly contaminated nature, the process water samples were diluted and the pH was adjusted to pH 4.5 prior to measurements using 6M HCl (Fisher Scientific).
(53) Photolysis experiments were carried out in a Rayonet photochemical reactor equipped with 8254 nm lamps (Model RPR 100, Southern New England UV Co., Branford, Conn.) and a 700 mL quartz reaction flask with a 10 C. immersion cold finger for sample cooling. Reactions were monitored by re-circulating 1 L of aqueous (DI) solution containing 0.7 mM H.sub.2O.sub.2 (Fisher Scientific) in a closed loop through the MIMS interface using O.D. Teflon tubing using a peristaltic pump (model 77200-62 Masterflex Easy-Load II with LS-25 Viton Tubing, Cole-Parmer, Vernon Hills, Ill., USA). The reaction mixture was flowed through the system at a fixed rate of 250 mL/min. The methanol acceptor phase was handled as described above.
(54) Aqueous samples and standards were maintained at 301 C. using a constant temperature water bath (model BM100; Yamato Scientific, Santa Clara, Calif., USA) and were re-circulated through the various CP-MIMS interfaces using a peristaltic pump (model 77200-62 Masterflex Easy-Load II with LS-25 Viton Tubing, Cole-Parmer, Vernon Hills, Ill., USA) at 250 mL/min. The 500 mL glass reservoir used was constructed in-house, and had Teflon lined septum port for injecting aliquots of standards and sample spikes. All standard solutions were prepared in HPLC grade methanol, with gas tight analytical syringes (Hamilton Company, Reno, Nev., USA). Aqueous samples were re-circulated in a closed loop system constructed from short lengths of 0.25 OD Teflon tubing (Cole-Parmer). For all experiments, the membrane was flushed with deionized water (DI, Model MQ Synthesis A10, Millipore Corp., Billerica, Mass., USA) between runs until stable baseline signal was observed. Signals were characterized based on their background-subtracted intensities at steady state for analytical calibration and by their 10-90% response times. Detection limits presented are based on signal-to-noise (S/N) ratio of three.
EXAMPLE 1
(55) Acceptor and Sample Flow Rate
(56) The influence of both acceptor and donor (sample) flow rates on the sensitivity and response for 50 ppb aqueous solutions of gemfibrozil with delicate membrane CP-MIMS using the composite PDMS membrane interface was determined. For all work presented the aqueous sample was maintained in a water bath at 30 C. The acceptor phase fluid handling system was optimized by using SMPIOD, where a twelve roller peristaltic pump (two pumping channels) was used to simultaneously drive the flow of liquid into and out of the membrane at the same pressure and flow rate. Initial studies using the in-line acceptor phase micro pump or a single channel 12 roller peristaltic pump showed that the delicate membranes were easily ruptured by the (slight) pulsing of the acceptor phase flow and/or pressure differences across the length of the HFM. In order to overcome this deficiency, SMPIOD was accomplished by reversing one of the pump tube configurations in the peristaltic pump head, and resulted in simultaneously matched pumping in and out of the delicate HFMs. This ensured a constant and even flow of acceptor phase, through these fragile HFM interfaces to the electrospray source. While a single pump need not be used, in the situation where two pumps are used, they must be closely matched for both flow rate and pressure.
(57)
(58) Quantitation, Signal Response Times and Detection Limits
(59) The analytical performance of the various delicate membrane CP-MIMS interfaces was evaluated for a variety of non-volatile, polar target analytes (see Table 1). Analytes for this study were selected for their environmental and/or biological significance, as well as for their relatively poor analytical performance with GP- MIMS. The low vapour pressures (ranging from 1 to 10.sup.8 Pa) and polar character of the target analytes in this study result in impractically long response times and poor analytical sensitivity with GP-MIMS operated at ambient temperatures.
(60) Detection limits and signal response times for each PDMS HFM interface were determined for each of the target analytes in aqueous solutions and are summarized in Table 2. The average steady state analytical signals used for this work were found by averaging >100 data points obtained during the measurement of a known aqueous standard. Response times and detection limits (S/N=3) for the target analytes are reported for replicate experiments (n=3 to 5). As illustrated in Table 2, the composite PDMS HFM had dramatically decreased response times that were in general 10 faster for the less volatile and more polar analytes studied. MIMS signal response times reflect the time it takes for the analyte signal to approach a steady state flux across the membrane, in response to a step function increase in upstream (sample) concentration. Without being bound to theory, these response times are largely governed by the rate of diffusion of the analyte through the membrane, which can be related to the analyte's molar volume. In general, smaller analytes will permeate polymer membranes faster than larger ones. Some of the target analytes do exhibit fairly long response times with the standard PDMS HFM (215 m thick PDMS), such as abietic acid (t.sub.10-90%=16.5 min). The response time for abietic acid is shortened by approximately a factor of ten when the composite PDMS HFM (0.5 m thick PDMS) is used (t.sub.10-90%=1.6 min). Similar response time improvements were observed for all target analytes measured with this interface (Table 2). This may have a significant advantage for the monitoring of relatively fast changes in chemical concentrations, or for the rapid measurement of a large number of samples. Without being bound to theory, to study dynamic chemical processes, the rate at which the analyte permeates the membrane must be faster than the rate of the change in concentration so that no information is lost. Applications could include following the progress of chemical reactions, chemical process control, the on-line environmental monitoring of (less volatile) contaminant plumes. In addition to changes in the bulk concentration, the technique allows for monitoring changes in the compositional distribution of compounds. This information is useful in monitoring the extent of chemical weathering, the effectiveness of certain processes (both natural and industrial) and can be applied to source identification. The 35 m thick PDMS HFM showed modest improvements in response time, but its' analytical performance (e.g. detection limits) was possibly diminished by its shorter length (5 cm versus 10 cm), limited by the membrane stock available at the time of the study.
(61)
EXAMPLE 2
(62) Stopped Acceptor Flow Mode
(63) As described previously, the delicate membrane CP-MIMS interface can be operated with a continuous flow of acceptor phase (e.g. continuous monitoring mode) or with a static acceptor phase in the membrane for a period of time to increase analytical sensitivity (stopped flow mode). The relative signal enhancements for stopped flow mode over continuous monitoring mode was determined using the composite 0.5 m thick PDMS/polypropylene interface for continuously re-circulated 1 ppb aqueous solutions of gemfibrozil.
EXAMPLE 3
(64) Direct, On-Line Measurement of Naphthenic Acids in Complex Samples
(65) To demonstrate the potential utility of delicate membrane CP-MIMS systems for trace level the continuous on-line monitoring of trace levels of low volatility analytes, the composite PDMS HFM based system was used to measure mixtures of naphthenic acids directly in deionized water and complex sample matrices. Naphthenic acids (NA) are natural components found in heavy crude oil, and are extracted and concentrated in the aqueous wastes generated during the various processes used to separate heavy crude oil from bitumen. NA mixtures can be found in total concentrations up to approximately 100 mg/L in bitumen extraction waste water tailings ponds and surface waters. These compounds are a highly complex mixture of alkyl substituted linear and cycloaliphatic carboxylic acids, which have a general molecular formula of C.sub.nH.sub.2n+ZO.sub.x, where n is the carbon number (typically 8-30), Z is the hydrogen deficiency and x is the number of oxygen's (typically 2-5).
(66) To demonstrate the potential utility of the delicate membrane CP-MIMS interfaces for the measurement of NA real environmental samples, several experiments were conducted. Aqueous solutions of a previously characterized NA mixture (Merichem) at ppb total levels were measured using the 0.5 m thick PDMS composite HFM interface in both DI water and in river water at ppb concentrations. The Merichem spiked de-ionized water was pH 5, and when this sample was further acidified (by addition of 6M HCl) to pH 4, the signals for NA m/z values increased, suggesting that a larger fraction of these acids were protonated, allowing more of them to permeate the membrane and be detected.
EXAMPLE 4
(67) In the continuous flow mode, both the sample and acceptor phases are continuously flowing which permits following the concentration of a dynamic system, in which changes in the concentration of analyte can be monitored over time. To illustrate the potential for rapid screening, on-line applications of delicate PDMS interfaces, a photochemical destruction study was conducted in which a 900 ppb Merichem NA sample in DI water was irradiated with 254 nm UV photons in the presence of 0.7 mM H.sub.2O.sub.2. The sample was measured at the beginning and the end of the experiment using the composite 0.5 m PDMS HFM interface (
EXAMPLE 5
(68) The experiments of Example 1 will be repeated using a 10 cm long 35 m thick PDMS HFM. Detection limits and signal response times for each PDMS HFM interface will be determined for each of the target analytes in aqueous solutions. The average steady state analytical signals used for this work will be found by averaging >100 data points obtained during the measurement of a known aqueous standard. Response times and detection limits (S/N=3) for the target analytes will be reported for replicate experiments (n=3 to 5). The composite PDMS HFM has dramatically decreased response times that in general are 10 faster for the less volatile and more polar analytes studied, as compared to the standard membrane. Similarly, the thin membrane (35 m thick PDMS HFM) will show much decreased response times. For the composite PDMS HFM detection limits will range from 40 pptr for gemfibrozil and 2,4,6-trichlorophenol to 3 ppb for estrone. Similar, if not better ranges will be found for the 10 cm long 35 m thick PDMS HFM interface.
EXAMPLE 6
(69) The experiments of Example 2 will be repeated using a 10 cm long 35 m thick PDMS HFM and the composite 0.5 m thick PDMS/polypropylene interface. The results of the stopped flow mode operation with the composite membrane will show improvements upon previously reported findings with the 215 m thick PDMS CP-MIMS HFM interface for both membrane systems.
EXAMPLE 7
(70) Direct, on-line measurement of naphthenic acids in a complex samples will be conducted using a 10 cm long 35 m thick PDMS HFM and the composite 0.5 m thick PDMS/polypropylene interface. To demonstrate the potential utility of the delicate membrane CP-MIMS interfaces for the measurement of NA real environmental samples, several experiments will be conducted. Aqueous solutions of a previously characterized NA mixture (Merichem) at ppb total levels will be measured using the 0.5 m thick PDMS composite HFM interface and the 35 m thick PDMS HFM in both DI water and in river water at ppb concentrations. The estimated S/N=3 detection limits of ca. 0.5-10 ppb for any individual component will be reported. The signal rise times range will range from 0.7-1.4 minutes, suggesting at least for the species detected, the delicate membrane CP-MIMS response times will allow rapid screening methods for NA as well as the continuous real-time monitoring of changing NA concentrations, such as could be observed during cleanup, remediation or destruction/reclamation processes. For comparison purposes and as an authentic environmental NA sample, Athabasca oil sands NA will also be measured in a diluted heavy oil extraction process water sample. Signal rise times for the observed m/z values will be consistent with the data obtained for the Merichem NA sample presented in Table 3.
EXAMPLE 8
(71) To illustrate the potential for on line real time monitoring applications using the delicate membrane PDMS interfaces, a photochemical destruction study was conducted in which a 900 ppb Merichem NA sample in DI water was irradiated with 254 nm UV photons in the presence of 0.7 mM H.sub.2O.sub.2, using the composite PDMS HFM interface and the 35 m thick PDMS HFM interface. On inspection of the full scan mass spectra obtained before irradiation and after 100 minutes of UV/H.sub.2O.sub.2treatment, shifts in the profile will be measurable.
EXAMPLE 9
(72) Studies will be conducted to demonstrate that the 35 m thick PDMS HFM and the composite 0.5 m thick PDMS/polypropylene interface can be used in delicate membrane CP-MIMS with SMPIOD to study large molecules. It will be found that application of additional driving forces (e.g., thermal, electrical, chemical or pH gradients) will allow for analysis of larger molecules.
EXAMPLE 10
(73) For continuous on-line measurements for trace level analytes, such as encountered in, for example, but not limited to bioanalytical measurements in blood or urine samples, a miniature CP-MIMS probe was developed. This can be dipped or immersed or otherwise introduced in the sample to be analyzed. Preferably, the sample is adequately mixed by stirring or agitation (for example, but not limited to sonicating, bubbling, jiggling, squirting, spraying, flowing or rocking) during the measurement, the flow rate could be 0 for direct immersion of the interface probes in the sample.
(74) The acceptor phase flow characteristics of the miniature CP-MIMS coaxial membrane probe were designed to facilitate minimal dead volumes within the probe itself while simultaneously maximizing the linear velocity of the acceptor phase on the permeate side of the membrane.
(75) The working PDMS HFM dimensions used were 0.94 mm o.d., 0.51 mm i.d. and 20 mm long (Silastic brand; Dow Corning, Midland, Mich., USA). The membrane was 215 m thick. The outer body of the probe stem was constructed from 22 gauge stainless steel hypodermic stock (Vita Needle Co., Needham, Mass., USA). The coaxially arranged acceptor phase delivery capillary was made from a short length of deactivated Siltek capillary column (0.25 mm ID, 0.37 mm OD, Restek Corp., Bellefonte, Pa., USA). The membrane was mounted over the hypodermic probe stem and a stainless steel end plug using hexane (ACS grade, Fisher Scientific, Ottawa, Ontario, Canada) as a membrane-swelling agent. A few turns of 30 gauge copper wire (GC Electronics Ltd, Rockford, Ill., USA) over the membrane at the end plug and on the probe stem were used to ensure that the membrane was not dislodged during preliminary testing and subsequent use. The probe was assembled using a 0.75 mm bore stainless steel tee union for 1/16 diameter tubing (Valco Instruments Co. Inc., Brockville, ON, Canada), which also acted as a connection point for acceptor phase delivery and its subsequent transport to the mass spectrometer. A schematic diagram of the apparatus is given in
(76) For this work, HPLC grade methanol (Fisher Scientific) was used as supplied, and was degassed using helium sparging (UHP Grade, Praxair Inc., Nanaimo, BC, Canada) when employed as an acceptor phase. Acceptor phase delivery to the membrane probe assembly was accomplished using 1/16 PEEK tubing (0.75 mm ID, Chromatographic Specialties, Brockville, ON, Canada), while acceptor phase transfer to the mass spectrometer used a smaller ID (0.25 mm) piece of PEEK tubing from the same supplier. The diameter of the tube used between the probe and the ESI source was intentionally smaller than the acceptor phase delivery tube to minimize any unintended dilution or signal broadening effects. Preliminary work showed that minor changes in the length of the tubing from the probe to the mass spectrometer did not adversely affect the signal rise time or maximum intensities observed. A triple quadrupole tandem mass spectrometer equipped with a low dead volume in-line micro-pump and ESI was used, as described above.
(77) Target analytes included phenol, 2-chlorophenol, 2,4-dichlorophenol, 2,4,6-trichlorophenol, triclosan (2,4,4-trichloro-2-hydroxydiphenyl ether), gemfibrozil (5-(2,5-dimethylphenoxy)-2,2-dimethylpentanoic acid) and nonylphenol (Sigma Aldrich, Oakville, ON, Canada). The chlorophenols and nonylphenol are representative drinking water contaminants, triclosan is an antifungal/bacterial agent shown to have endocrine disruptive capabilities, and gemfibrozil is a fibrate drug used to lower lipid levels. Stock analyte solutions were prepared at ppm levels in methanol, followed by dilution in deionized (DI) water (Model MQ Synthesis A10, Millipore Corp., Billerica, Mass., USA) or in other indicated matrices to low ppb levels for the presented work.
(78) For the chlorination experiment, a 20,000 ppm (as Cl.sub.2) working stock of sodium hypochlorite (Sigma Aldrich) was prepared in DI water, and a 0.5 M phosphate buffer (pH 7.0) was used for pH adjustment. Artificial urine, water from a closed system decorative Koi fish pond, lager beer with 5% alcohol by volume and primary sewage treatment effluent from a small municipal treatment plant were used without dilution or filtration for complex sample matrices in this study. The samples were contained in either 40 mL clear glass sample vials (Scientific Specialties Inc., Hanover, Md., USA), 4 mL HDPE autosampler cups (Pulse Instrumentation Inc., Milwaukee, Wis., USA), 1.8 mL clear glass chromatography vials (Agilent Technologies, Mississauga, ON, Canada) or custom made 0.5 mL vials fashioned by cutting standard NMR tubes (Norell, Landisville, N.J., USA) to shorter lengths, as indicated. All were washed three times with HPLC grade methanol and dried in air before use. Samples were either unstirred, stirred with a miniature Teflon stir bar and magnetic stirrer (LAB DISC S56, Fisher Scientific, Vancouver, BC, Canada) or mixed by helium bubbled through the sample via a short length of 22 gauge stainless steel hypodermic tube. All measurements were made at ambient temperature (25 C.) and pressure (101 kPa). For the automation experiments, a commercial autosampler from a flow injection analysis system (Model 301, Alchem Corp., Clackamas, Oreg., USA) was used, modified by mounting the miniature CP-MIMS probe assembly in the sampling turret.
(79) The optimum acceptor phase flow was determined by measuring the signal-to-noise (S/N) ratio obtained for the steady-state signals of ppb-level aqueous solutions of 2,4,6-trichlorophenol and triclosan at methanol acceptor phase flow rates ranging from 50 to 300 L/min. These measurements were made in 40 mL aqueous samples, using a magnetic stirrer and Teflon stir bar for mixing. After each measurement, the miniature probe was washed clean by immersing it in 40 mL of stirred DI water until the signal returned to baseline levels. The results are given in
(80) To wash this probe after a measurement or on-line monitoring experiment, it can simply be immersed in DI water until the signals for target analytes return to baseline levels. As noted above, the miniature CP-MIMS probe (unlike flow cell type interfaces) can be cleaned between measurements by simply dipping it in a small quantity of any suitable wash solvent, such as DI water, or any other (membrane-compatible) solvent, like methanol.
(81) Changing the wash solvent provides faster membrane cleaning times between samples (than possible with water), decreasing the wait time needed between samples. Although our previously discussed flow cell interface could also be cleaned with organic solvent, it required much greater quantities of solvent, increasing both the cost and waste produced by each measurement.
(82) An experiment was conducted to compare the analyte wash-out time after monitoring a 78 ppb aqueous triclosan solution. A 40 mL aliquot of stirred DI water was spiked with triclosan and monitored via direct insertion of the mini CP-MIMS probe. After the signal reached steady state, the probe was transferred to 40 mL of a stirred wash solvent (DI water or methanol). Depletion of the residual triclosan analyte was monitored as it was washed out of the PDMS membrane.
(83) A comparison study was conducted in which aqueous solutions (40 mL) containing 2,4,6-trichlorophenol (28 ppb), triclosan (77 ppb), gemfibrozil (78 ppb) and nonylphenol (53 ppb) were prepared by spiking with the appropriate quantity of stock solutions, followed by measurement with the miniature coaxial probe CP-MIMS system. One solution was not mixed, one mixed with the stir bar/stirrer, and the remaining solution was mixed by bubbling helium through the sample during the measurement. Although helium bubbling provides an approach to mixing samples in small volumes that may not be amenable to stir bar mixing, the potential for loss of volatile analytes exists. However, it should be noted that CP-MIMS in general is well suited for the analysis of low volatility analytes.
(84) The data collected were analyzed for the signal response times (t.sub.10-90%) and 3S/N ratio detection limits. The unmixed solution yielded the longest response times and poorest detection limits, whereas the helium bubbler and stir bar mixing gave similar results. Without being bound to theory, the increased analyte diffusion path length created by a depleted boundary layer developed at the membrane surface in unmixed samples adds mass transport resistance, reducing the permeability and increasing the transport time as predicted by Fick's law. As would be known to one skilled in the art, other means of mixing are contemplated, for example, but not limited to mechanical vibration.
(85) The results of this study yielded t.sub.10-90% response times of 2-10 min for the target analytes tested, and detection limits from pptr to low ppb levels.
(86) The depletion of analyte concentrations in small sample volumes was also studied to test if sample pre-screening using the miniature CP-MIMS probe could be followed by a second measurement (e.g., HPLC/MS), using the same sample for both. Using previously published acceptor phase calibration methodologies, when 40 mL of 70 ppb aqueous gemfibrozil was interrogated with the miniature CP-MIMS probe, a gemfibrozil concentration of 9 ppb was observed in the methanol acceptor phase. At an acceptor phase flow rate of 200 L/min, this corresponds to a total gemfibrozil mass transfer of ca. 2 ng/min from the sample under steady state conditions.
(87) To examine potential analyte depletion effects in small samples, 70 ppb aqueous gemfibrozil samples (1.8 mL and 400 L) were continuously monitored with the miniature CP-MIMS probe for a 1-h period. The results of this study are shown in
(88) To illustrate the use of the miniature CP-MIMS probe for quantitative measurements, a series of combined aqueous standards in the ppb range was interrogated with the mini CP-MIMS probe, using a magnetic stir bar for sample mixing. The background subtracted steady-state signals for each of the target analytes were subsequently used to generate calibration curves (
(89) The use of a miniature CP-MIMS probe allows the same potential applications for on-line monitoring as its flow cell MIMS counterparts, including its use in flowing streams, but it also can be employed in confined spaces and in smaller volumes of sample than would be practical with flow cell type interfaces. As an illustrative (and comparative) example, the chlorination reaction of aqueous phenol was monitored with the probe system in a 40-mL vial. A sample of DI water, buffered at pH 7, was stirred and spiked to a final concentration of 250 ppb phenol. After the phenol signal reached steady state, an aliquot of sodium hypochlorite was added to achieve an active chlorine concentration of 10 ppm (as Cl.sub.2), and the reaction allowed to proceed at 25 C in an uncapped vial.
(90) The resulting signals for reactants and chlorinated phenols are illustrated in
(91) The results suggest that the miniature CP-MIMS probe could be used for the monitoring of a wide range of chemical systems, environmental testing scenarios and industrial processes. The miniature CP-MIMS probe can readily be implemented in small volume samples or be directly inserted into a continuously flowing sample stream (e.g., a pipeline in an industrial scenario) and offers a simple alternative for in situ reaction monitoring. The miniature probe can be used in sample volumes that are as little as about 100 L or about 200 L or about 400 L. In comparison, the minimum flow cell sample volume is in the range of about 5 mL to about 10 mL, although it would be impractical to flow the sample with such a small volume. The flow cell is more robust than the miniature probe and is well suited to online continuous monitoring where sample volume is not a consideration. It is also more suited to automated flow dilution(s) and for process monitoring.
EXAMPLE 11
(92) To demonstrate automated use, a rotary tray autosampler system from a flow injection analyzer was adapted to use the miniature CP-MIMS probe. The PEEK transfer lines employed were flexible enough that the movement of the probe in the autosampler system was unhindered.
(93) For this study, the autosampler was equipped with 4-mL sample cups and used a maximum programmable sampling time of 99 s/sample. Trichlorophenol (50 ppb) and nonylphenol (100 ppb) were analyzed in a wide variety of sample matrices. Sample matrices, including DI water, Koi pond water, beer, artificial urine and primary sewage
(94) waste water effluent, were studied. The samples were well mixed before the autosampler cups were filled, and measured as part of a continuous sample sequence by a 99-s miniature CP-MIMS probe immersion in each sample. The probe was rinsed by automated immersion in clean methanol for 99 s between replicate samples and for 197 s between sample types. All samples were directly analyzed without dilution, pretreatment or filtration.
(95) In evaluating the use of this simple autosampler system, it should be noted that samples were not mixed during the probe immersion. Although the exposure time of the membrane probe to the sample was very reproducible (because of the automation), it was not long enough to allow for steady-state signal development. Analysis of the data showed similar relative standard deviations for both peak height and peak area measurements at 10% for three replicates within the same sample matrix. Fitting the autosampler with a mechanism to agitate the probe or to mix the samples would improve this precision. The signals observed for trichlorophenol (50 ppb) were much stronger than those for nonylphenol (100 ppb), but were slightly less reproducible. This
(96) analyte-dependent sensitivity/precision is readily explained by the fact that CP-MIMS is 5 more sensitive to trichlorophenol than to nonylphenol. Furthermore, the signal response time for nonylphenol to reach steady state is 4 times longer than that for trichlorophenol. Taken together, these factors contribute to the greater sensitivity observed for trichlorophenol over nonylphenol (
(97) Although some signal suppression effects for the complex samples were observed, with analyte signal levels ranging from 13 to 120% compared with those obtained for DI water samples, the effect appears to be both analyte- and matrix-dependent (
(98) The observations in
(99) These experiments demonstrate the potential of miniature probe CP-MIMS as a rapid, automated pre-screening technique to identify positive samples for subsequent quantitation by conventional methods (e.g., HPLC/MS). It is contemplated that this system, when coupled to a modern autosampler system, will provide the possibility of a logic driven automated pre-screen prior to more time-consuming cleanup and chromatographic analyses is foreseeable.
EXAMPLE 12
(100) Direct, in vivo/in situ monitoring was demonstrated with the mini probe as described above. A large, freshly cut celery plant stalk (Apium graveolens) was carefully pierced 1 cm from its base with a small twist drill bit to allow direct, horizontal insertion of the miniature CP-MIMS probe, such that the active membrane surface was completely inside the stem. The stalk with embedded probe was mounted vertically with 1 mm of its base immersed in 50 mL of 130 ppm aqueous gemfibrozil solution. The system was left undisturbed, and the subsequent osmotic transport of gemfibrozil up the celery stalk recorded using the miniature CP-MIMS probe system for nearly an hour (
EXAMPLE 15
(101) To illustrate the potential for rapid screening, on-line monitoring applications of delicate PDMS CP-MIMS interfaces, a photochemical destruction study was conducted in which a 900 ppb Merichem NA sample in DI water was irradiated with 254 nm UV photons in the presence of 0.7 mM H.sub.2O.sub.2. The sample was continuously monitored over the course of the experiment by both full scan and selected SIM scans using a 2.0 cm length of composite 0.5 m PDMS HFM mounted in a J-Probe immersion CP-MIMS interface (
EXAMPLE 16
(102) By adjusting the upstream sample pH, a sequence of CP-MIMS measurements can be carried out on a NA sample to provide additional compositional information. The pH is adjusted such that NAs are ionized (any pH>6) and the analyte is measured and quantified. The sample is then acidified (any pH<4) with a strong acid, and the sample is re-analyzed. The first experiment measures neutral compounds (those that are un-ionized at pH>6). The second experiment measures those which are un-ionized at pH 4 (this includes neutrals+NAs). The difference between the two, will be those compounds that were ionized at pH>6 and un-ionized at pH 4 (i.e., carboxylic acids such as NAs). Similarly, by adjusting the concentration of other cations (e.g., Ca.sup.2+, Pb.sup.2+, Cd.sup.2+) at a constant pH, one can use the resulting CP-MIMS signals for the determination of the corresponding metalligand formation constants. This information can be useful when one is dealing with complex mixtures and multiple formation constants can be determined for isomer classes in cases where the pure components are not available or known.
EXAMPLE 17
(103) Individual molecules within the class of NAs have different sizes and the response time with CP-MIMS is size dependent (as described above). Therefore, CP-MIMS can be used to provide information about molecular sizes (sometimes referred to as hydrodynamic volume). This information can be useful when one is dealing with complex mixtures and the exact chemical structure of individual compounds is not known and/or the pure components of the mixture are not available for individual study. Molecular size determination can be a important predictor of other transport phenomena and can be used to inform cellular uptake rates and toxicity studies.
EXAMPLE 18
(104) The method allows for monitoring changes in the compositional distribution of compounds. The distribution of individual components in a complex mixture can provide useful information when monitoring the extent of chemical weathering, the effectiveness of certain processes (both natural and industrial) and can be applied to source identification.
EXAMPLE 19
(105) A study was conducted in which the effect of pH upon the observed CP-MIMS signals for naphthenic acids (NA). Two spike levels of NA (0.4 and 1.4 ppm total Merichem NA) were added to a 40 mL stirred (magnetic stir bar) deionized water sample, and the CP-MIMS signal responses were recorded, as well as the pH of the solution. For this experiment, the J-Probe type CP-MIMS interface was used with a 2.0 cm length of PDMS HFM that was 170 m thick.
EXAMPLE 20
(106) To demonstrate the capabilities of CP-MIMS for the detection and measurement of NAs in a wide range of typical environmental samples, 40 mL EPA type water sample vials were filled with a variety of different water samples obtained from northern Alberta, including oil sands process water (OSPW-1,2), ground water (GW-1,2) and natural surface waters (SW-1,2,3). The samples were acidified to pH 4.0 with 2M HCl, then stirred (magnetic stir bar) and sequentially measured with the CP-MIMS system using the J-Probe type interface equipped with a 2.0 cm length of PDMS HFM that was 170 m thick. The sample probe was removed from each sample when steady signals were obtained, and the probe washed with methanol between each sample until baseline signals were achieved.
EXAMPLE 21
(107) To demonstrate the capabilities of CP-MIMS for the determination of intrinsic molecular properties such as diffusion coefficients and hydrodynamic volumes, the response times of various isomer classes to a step-function increase in upstream NA concentration was determined. The chromatograms displayed in
EXAMPLE 22
(108) Dilute aqueous solutions of Merichem naphthenic acids (50-1000 ppb) were analyzed using an immersion J-probe membrane interface coupled to CP-MIMS with a methanol acceptor phase. Because naphthenic acids are a complex mixture of structurally related isomer classes, typically ranging in molecular weight from 150-400 atomic mass units (amu), membrane performance (response time and sensitivity) was characterized at five different mass/charge ratios spanning 213-305 amu.
(109) Response times were measured as the time required for the signal intensity to rise from 10 to 90% of its full steady state value as a result of switching from a clean water sample to one that contains a Merichem naphthenic acid mixture. In general, response times are independent of chemical concentration. As can be seen in Table 4, the thin PDMS composite membrane demonstrates a 10 to 20-fold reduction in response time over the thicker 35 m PDMS membrane. These faster response times are advantages when making rapid measurements and significantly reduce the duty cycle. The 170 m PDMS interface, gives rise to similar or slightly longer response times than the 35 m membrane. It should be noted here that for identical materials, the response times are predicted to increase with the square of the thickness, however PDMS from different sources can and will have different perm-selectivities and viscosities due to differing polymer chain length, cross-linking and additives.
(110) The sensitivity of these membrane interfaces to various isomer classes was assessed by measuring the signal to noise ratios at various m/z values resulting from a 50 ppb sample of Merichem. The detection limit was estimated based on the concentration of Merichem naphthenic acid solution giving rise to a peak at 3 times the signal/noise ratio. In general, the thicker membrane with the larger outside diameter was the most sensitive to naphthenic acids with detection limits ranging from 5 to 10 ppb. The detection limits for the thinner membranes is somewhat higher, ranging from 10 to 40 ppb for the isomer classes monitored here.
EXAMPLE 23
(111) CP-MIMS for direct measurement of polyaromatic hydrocarbons (PAHs) in dilute aqueous solution, as shown in
(112) Water samples containing representative PAHs were analyzed by CP-MIMS using an immersion J-probe capillary PDMS interface. A methanol acceptor phase was flowed through the membrane capillary at 200 uL min.sup.1 and transported to an Atmospheric Pressure Photoionization source whereupon the molecular ions for both fluorene and pyrene ([M+]=m/z 166 and 202, respectively), were observed within 5 minutes by Fourier Transform Ion Cyclotron Resonance Mass Spectrometry (FT-ICR-MS).
EXAMPLE 24
(113) CP-MIMS for monitoring changes to the mass profile of complex naphthenic acid mixtures treated with adsorbent, as shown in
(114) Water samples containing a complex mixture of naphthenic acids at pH 4 and pH 7 were analyzed by CP-MIMS using a methanol acceptor phase and electrospray ionization (ESI) in negative ion mode and a triple quadrupole mass spectrometer.
(115) Full scan data between m/z=100-400 shows the mass profile of naphthenic acid isomer class families. At pH 4 both naphthenic acids are protonated and readily cross the membrane interface along with other neutral hydrophobic molecules, such as alcohols (
(116) The addition of activated charcoal as an adsorbent was followed by CP-MIMS in real-time in full scan mode showing a dramatic decrease in the naphthenic acid concentration after about 20 mins of stirring. Mass spectra for the pH 4 and pH 7 solutions after exposure to activated charcoal is shown in
(117) Panel F illustrates the ability to follow the kinetics of removal for specific isomer classes (mass/charge windows) in real-time over the course of the treatment process. This data demonstrates the ability to use CP-MIMS follow treatment processes of complex mixtures of naphthenic acids in heterogeneous aqueous solutions.
EXAMPLE 25
(118) Duncan et al. 2015 Journal of Mass Spectrometry, 50: 437-443, the contents of which are incorporated herein in their entirety by reference, discloses implementation of a continuously infused internal standard in the membrane acceptor phase, and modulation of the acceptor phase flow rate to mitigate suppressed analyte signals.
(119) Quantitation of analytes can be improved by varying the acceptor flow rate. At higher acceptor flow rates, for example, but not limited to 500 l min versus 200 l min, the sensitivity decreased but ionization suppression was also reduced. This extended the Linear Dynamic Range (LDR) to higher sample concentrations. This was accomplished iteratively. Flow rates can be as low as 100 nL/minute.
EXAMPLE 26
(120) An on-the-fly acceptor phase flow rate adjustment with an automatic calibration feedback system will be used. For example, if an observed analyte signal is too high for the measured LDR, the acceptor flow will be increased, providing an online dilution to effectively extend the method's LDR.
EXAMPLE 27
(121) A second approach for extending the LDR and mitigating ionization suppression employed a continuously infused internal standard by the addition of internal standard to the acceptor phase. By infusing the internal standard into the acceptor phase, the concentration of analyte in the sample was determined with minimal sample modification. This approach maintained sample integrity for future or further analysis and had the added benefit of allowing for direct observation of and correction for any deviations in ionization efficiency. To demonstrate this strategy, 5 ppb of aniline-d5 was added directly to the acceptor phase, and aniline was successively spiked into 1 l of stirred de-ionized (DI) water to generate calibration data. The measured aniline signals were normalized with the internal standard signal and plotted against aniline concentration. Excellent calibration linearity was observed for the entire concentration range examined (spanning six orders of magnitude). High concentrations of aniline in the sample resulted in the aniline-d5 signal being suppressed by as much as 90%. However, when the sample was replaced with DI water, the aniline-d5 signal quickly returned to its original level. Therefore, adding an appropriate internal standard to the acceptor phase can enable the correction of ionization suppression effects and provide an excellent LDR when using ESI. Either labeled or non-labeled internal standards are viable alternatives when isotopically labeled molecules are not readily available. The internal standard is preferably the same compound as the analyte, but it may alternatively be a similar compound with similar characteristics.
EXAMPLE 28
(122) Duncan et al. 2015 Journal of Mass Spectrometry: A semi-quantitative approach for the rapid screening and mass profiling of naphthenic acids directly in contaminated aqueous samples, in press, the contents of which are incorporated herein in their entirety by reference, discloses approaches to increasing CP-MIMS sensitivity for Naphthenic acids including decreasing the acceptor phase flow rate, heating the membrane interface, increasing the surface area of the membrane interface, and the use of base (for example, but not limited to sodium acetate) within the acceptor phase to enhance ionization. Acceptor phase ionization enhancers include acids, bases, anions and cations, in addition to simple alkanes, simple aromatics, simple alkanes functionalized with amines or carboxylates, and simple aromatics functionalized with amines or carboxylates.
EXAMPLE 29
(123) Positive ion mode with CP-MIMS can use protonated amines or cationized acids.
EXAMPLE 30
(124) As shown in
(125) The foregoing is an example of the present technology. As would be known to one skilled in the art, variations that do not alter the scope of the technology are contemplated.
(126) TABLE-US-00001 TABLE 1 Physical data and MS scan parameters for the target analytes studied. Chemical Molar Vapour Molar MS MS Scan Target Abstract mass pressure Volume Entrance Parameters Analyte Service (g/mol) (Pa).sup.a Log K.sub.ow.sup.a (cm.sup.3/mol).sup.b Cone (V) (m/z).sup.c Abietic 514-10-3 302.45 4.3 10.sup.5 6.46 270 50 301 Acid Estrone 53-16-7 270.37 3 10.sup.8 3.25 230 50 269 .fwdarw. 145 Gemfibrozil 25812-30-0 250.33 8.1 10.sup.5 4.30 240 20 249 .fwdarw. 120 Nonyl- 25154-52-3 220.35 1.15 5.76 236 40 219 .fwdarw. 133 phenol 2,4,6- 88-06-2 197.45 2.3 3.67 124 40 197 Trichlorophenol Triclosan 3380-34-5 289.54 8.6 10.sup.5 4.76 195 20 289 .sup.aSRC Physical Properties Database .sup.bCalculated using ACD software .sup.cSIM or SRM (.fwdarw.) as indicated, negative ion mode for all
(127) TABLE-US-00002 TABLE 2 Detection limits and response times determined for a variety of analytes measured by CP-MIMS employing three different PDMS HFM interfaces 215 m thick 35 m thick Composite PDMS Lowest PDMS HFM PDMS HFM HFM (0.5 m thick) Measured t.sub.10-90%/ Detection t.sub.10-90%/ Detection t.sub.10-90%/ Detection Analyte Conc./ppb min limit/ppb.sup.a min limit/ppb min limit/ppb Abietic 29 16.5 1 12.5 2 1.6 0.6 Acid Estrone 66 7.5 4 1.8 12 0.6 3 Gemfibrozil 9 4.7 0.2 3.7 0.5 0.8 0.04 Nonyl- 5 9.5 0.1 3.5 2 1.0 0.5 phenol 2,4,6- 4 2.5 0.03 0.7 0.2 0.3 0.04 Trichlorophenol Triclosan 8 5.8 0.1 2.0 0.5 0.6 0.1 .sup.aDetection limits based on S/N = 3
(128) TABLE-US-00003 TABLE 3 Analysis of signal-to-noise ratios (S/N) and average signal rise times for selected NA obtained for Merichem NA in DI and River Water spiked at 440 ppb total NA. S/N Ratios Molecular Average Formula m/z DI Water River Water t.sub.10-90%/min. C.sub.9H.sub.18O.sub.2 157 25.5 22.1 0.7 C.sub.13H.sub.22O.sub.2 209 123 17.6 1.0 C.sub.13H.sub.24O.sub.2 211 81.1 12.5 1.0 C.sub.13H.sub.26O.sub.2 213 64.3 17.5 1.3 C.sub.15H.sub.26O.sub.2 237 180 22.5 1.2 C.sub.16H.sub.28O.sub.2 251 420 32.8 1.2 C.sub.17H.sub.30O.sub.2 265 70.4 44.0 1.5 C.sub.18H.sub.32O.sub.2 279 75.8 22.3 1.4
(129) TABLE-US-00004 TABLE 4 Performance characteristics for selected naphthenic acid isomer families (C.sub.nH.sub.2n+zO.sub.2) for Merichem mixture using J-probe immersion type interface with 2 cm membrane length. Membrane Response Times Sensitivity PDMS t.sub.10-90 (mins) Est. DLs (ppb) thickness m/z 213 223 237 251 305 213 223 237 251 305 Isomer 13, 0 14, 2 15, 2 16, 2 20, 3 13, 0 14, 2 15, 2 16, 2 20, 3 class n, z 14, 6 15, 8 16, 8 17, 8 21, 9 14, 6 15, 8 16, 8 17, 8 21, 9 0.5 m 0.6 0.6 0.6 0.7 3.0 20 20 40 30 30 composite.sup.1 35 m.sup.2 10 15 15 20 20 10 5 20 20 20 170 m.sup.3 10 15 25 25 30 10 5 5 5 10 .sup.1plasma deposited thin film PDMS on microporous polypropylene, neoMecs Inc. (O.D. = 0.264; I.D. = 0.263) .sup.2MedArray Inc (O.D. = 0.24 mm; I.D. = 0.17 mm) .sup.3Silastic Dow-Corning (O.D. = 0.64 mm; I.D. = 0.30 mm)