RELATIVE OPTIMIZED LINEARIZATION FOR RADIOCHROMIC FILM DOSIMETRY WITH NONUNIFORMITY CORRECTION
20260133327 ยท 2026-05-14
Assignee
Inventors
Cpc classification
A61N5/1075
HUMAN NECESSITIES
G01T1/2921
PHYSICS
A61N2005/1076
HUMAN NECESSITIES
A61N5/1071
HUMAN NECESSITIES
G01T1/06
PHYSICS
International classification
G01T1/06
PHYSICS
A61N5/10
HUMAN NECESSITIES
G01T1/29
PHYSICS
Abstract
Systems and methods for relative optimized linearization (ROL) for radiochromic film (RCF) dosimetry may eliminate the need for dose-response curve measurements while incorporating non-uniformity corrections. The ROL method may include, for each of multiple color channels, determining linearization parameters that minimize a cost function that evaluates an error between expected and calculated doses. The ROL method may use the linearization parameters to generate a variation map including dose-independent variation values and use the variation map to determine a corrected net optical density of the RCF. The ROL method may use the linearization parameters and the corrected net optical density to produce a dose image including corrected doses.
Claims
1. A method comprising: for each color channel c of multiple color channels, for each pixel location i of a digital image of a radiochromic film (RCF) that was exposed to radiation generated by a linear accelerator (LINAC), using a pixel value PV.sub.c,i at the pixel location i to calculate a dose D.sub.c,i of radiation in the color channel c absorbed by the RCF at a location corresponding to the pixel location i, wherein an optical density of the RCF at the location has changed in response to the dose of the radiation absorbed by the RCF at the location; for each color channel c of the multiple color channels, determining linearization parameters that minimize a cost function that evaluates an error between expected doses D.sub.i at pixel locations i and the calculated doses D.sub.c,i of the radiation in the color channel c absorbed by the RCF at the locations corresponding to the pixel locations i; generating a variation map including, for each pixel location i of the digital image of the RCF, a dose-independent variation value V.sub.i for the pixel location i, wherein generating the variation map comprises, for each pixel location i of the digital image of the RCF: for each color channel c of the multiple color channels, using the linearization parameters determined for the color channel c to calculate an optimized linearized dose D.sup.rol.sub.c,i of the radiation in the color channel c absorbed by the RCF at the location corresponding to the pixel location i, and determining the variation value V.sub.i for the pixel location i based on differences between the optimized linearized doses D.sup.rol.sub.c,i calculated for the multiple color channels c; for each color channel c of the multiple color channels, for each pixel location i of the digital image of the RCF, using the pixel value PV.sub.c,i for the pixel location i in the color channel c and the variation value V.sub.i for the pixel location i to determine a corrected net optical density netOD.sup.corrected.sub.c,i of the RCF for the color channel c at the pixel location i; and producing a dose image comprising a corrected dose for each color channel c of the multiple color channels at each pixel location i of the digital image of the RCF, wherein producing the dose image comprises, for each color channel c of the multiple color channels and each pixel location i of the digital image of the RCF, using the linearization parameters determined for the color channel c and the corrected net optical density netOD.sup.corrected.sub.c,i of the RCF for the color channel c at the pixel location i to calculate the corrected dose for the color channel c at the pixel location i.
2. The method of claim 1, wherein the multiple color channels include red, green, and blue channels.
3. The method of claim 1, further comprising using the LINAC to generate the radiation and expose the RCF to the radiation.
4. The method of claim 1, further comprising using a scanner to scan the RCF across the multiple color channels to generate the digital image of the RCF with responses of the RCF to the dose of absorbed radiation in the multiple color channels.
5. The method of claim 1, wherein, for each color channel c of the multiple color channels, for each pixel location i of the digital image of the RCF that was exposed to the radiation, using the pixel value PV.sub.c,i at the pixel location i to calculate the dose D.sub.c,i of radiation in the color channel c absorbed by the RCF at the location corresponding to the pixel location i comprises: using the pixel value PV.sub.c,i at the pixel location i to determine an uncorrected net optical density netOD.sub.c,i of the RCF in the color channel c at the location of the RCF corresponding to the pixel location i; and using the uncorrected net optical density netOD.sub.c,i of the RCF in the color channel c at the location of the RCF corresponding to the pixel location i to calculate the dose D.sub.c,i of radiation in the color channel c absorbed by the RCF at the location corresponding to the pixel location i.
6. The method of claim 5, wherein using the pixel value PV.sub.c,i at the pixel location i to determine the uncorrected net optical density netOD.sub.c,i of the RCF in the color channel c at the location of the RCF corresponding to the pixel location i comprises normalizing the pixel value PV.sub.c,i at the pixel location i by an averaged pixel value PV.sub.c,unexp of a digital image of an unexposed RCF in the color channel c.
7. The method of claim 6, further comprising using a scanner to scan the unexposed RCF across the multiple color channels to generate the digital image of the unexposed RCF.
8. The method of claim 1, wherein the linearization parameters for each color channel c include a scaling value a.sub.c and a power value p.sub.c.
9. The method of claim 1, wherein the error evaluated by the cost function is a mean absolute error between the expected doses D.sub.i at the pixel locations i and the calculated doses D.sub.c,i of the radiation in the color channel c absorbed by the RCF at the locations corresponding to the pixel locations i.
10. The method of claim 1, wherein the corrected net optical density netOD.sup.corrected.sub.c,i of the RCF in the color channel c at the pixel location i is determined using an averaged pixel value PV.sub.c,unexp of a digital image of an unexposed RCF in the color channel c.
11. The method of claim 10, wherein determining the corrected net optical density netOD.sup.corrected.sub.c,i of the RCF for the color channel c at the pixel location i using the averaged pixel value PV.sub.c,unexp of the digital image of the unexposed RCF in the color channel c comprises: using the averaged pixel value PV.sub.c,unexp of the digital image of the unexposed RCF in the color channel c to calculate a scanned optical density scanOD.sub.c,unexp of the digital image of the unexposed RCF in the color channel c; and using the scanned optical density scanOD.sub.c,unexp of the digital image of the unexposed RCF in the color channel c to determine the corrected net optical density netOD.sup.corrected.sub.c,i of the RCF in the color channel c at the pixel location i.
12. The method of claim 1, wherein, for each color channel c of the multiple color channels, for each pixel location i of the digital image of the RCF, using the pixel value PV.sub.c,i at the pixel location i in the color channel c and the variation value V.sub.i for the pixel location i to determine a corrected net optical density netOD.sup.corrected.sub.c,i of the RCF in the color channel c at the location of the RCF corresponding to the pixel location i comprises: using the pixel value PV.sub.c,i at the pixel location i in the color channel c to calculate a scanned optical density scanODc,i for the pixel location i in the color channel c; and using the scanned optical density scanODc,i for the pixel location i in the color channel c and the variation value V.sub.i for the pixel location i to determine the corrected net optical density netOD.sup.corrected.sub.c,i of the RCF in the color channel c at the location of the RCF corresponding to the pixel location i.
13. The method of claim 1, further comprising using: comparing the corrected doses with the expected doses Di and determining that the LINAC meets acceptable performance criteria; and exposing a patient to treatment radiation generated by the LINAC only if the LINAC was determined to meet the acceptable performance criteria.
14. The method of claim 1, further comprising: using the dose image to recalibrate the LINAC; and using the recalibrated LINAC to generate treatment radiation that exposes a patient to a radiation dose distribution in accordance with an expected dose distribution.
15. The method of claim 14, wherein recalibrating the LINAC comprises: determining adjustments to the LINAC that would cause the LINAC to generate radiation that would expose the patient to a radiation dose distribution in accordance with the expected dose distributions; and making the determined adjustments to the LINAC.
16. The method of claim 14, wherein recalibrating the LINAC comprises adjusting the expected dose distribution to match the dose image.
17. An apparatus configured to: for each color channel c of multiple color channels, for each pixel location i of a digital image of a radiochromic film (RCF) that was exposed to radiation generated by a linear accelerator (LINAC), use a pixel value PV.sub.c,i at the pixel location i to calculate a dose D.sub.c,i of radiation in the color channel c absorbed by the RCF at a location corresponding to the pixel location i, wherein an optical density of the RCF at the location has changed in response to the dose of the radiation absorbed by the RCF at the location; for each color channel c of the multiple color channels, determine linearization parameters that minimize a cost function that evaluates an error between expected doses D.sub.i at pixel locations i and the calculated doses D.sub.c,i of the radiation in the color channel c absorbed by the RCF at the locations corresponding to the pixel locations i; generate a variation map including, for each pixel location i of the digital image of the RCF, a dose-independent variation value V.sub.i for the pixel location i, wherein generating the variation map comprises, for each pixel location i of the digital image of the RCF: for each color channel c of the multiple color channels, using the linearization parameters determined for the color channel c to calculate an optimized linearized dose D.sup.rol.sub.c,i, of the radiation in the color channel c absorbed by the RCF at the location corresponding to the pixel location i, and determining the variation value V.sub.i for the pixel location i based on differences between the optimized linearized doses D.sup.rol.sub.c,i calculated for the multiple color channels c; for each color channel c of the multiple color channels, for each pixel location i of the digital image of the RCF, use the pixel value PV.sub.c,i for the pixel location i in the color channel c and the variation value V.sub.i for the pixel location i to determine a corrected net optical density netOD.sup.corrected.sub.c,i of the RCF for the color channel c at the pixel location i; and produce a dose image comprising a corrected dose for each color channel c of the multiple color channels at each pixel location i of the digital image of the RCF, wherein producing the dose image comprises, for each color channel c of the multiple color channels and each pixel location i of the digital image of the RCF, using the linearization parameters determined for the color channel c and the corrected net optical density netOD.sup.corrected.sub.c,i of the RCF for the color channel c at the pixel location i to calculate the corrected dose for the color channel c at the pixel location i.
18. The apparatus of claim 17, wherein the multiple color channels include red, green, and blue channels.
19. The apparatus of claim 17, wherein the apparatus is further configured to use the LINAC to generate the radiation and expose the RCF to the radiation.
20. The apparatus of claim 17, wherein, for each color channel c of the multiple color channels, for each pixel location i of the digital image of the RCF that was exposed to the radiation, using the pixel value PV.sub.c,i at the pixel location i to calculate the dose D.sub.c,i of radiation in the color channel c absorbed by the RCF at the location corresponding to the pixel location i comprises: using the pixel value PV.sub.c,i at the pixel location i to determine an uncorrected net optical density netOD.sub.c,i of the RCF in the color channel c at the location of the RCF corresponding to the pixel location i; and using the uncorrected net optical density netOD.sub.c,i of the RCF in the color channel c at the location of the RCF corresponding to the pixel location i to calculate the dose D.sub.c,i of radiation in the color channel c absorbed by the RCF at the location corresponding to the pixel location i.
21. The apparatus of claim 20, wherein using the pixel value PV.sub.c,i at the pixel location i to determine the uncorrected net optical density netOD.sub.c,i of the RCF in the color channel c at the location of the RCF corresponding to the pixel location i comprises normalizing the pixel value PV.sub.c,i at the pixel location i by an averaged pixel value PV.sub.c,unexp of a digital image of an unexposed RCF in the color channel c.
22. The apparatus of claim 17, wherein the linearization parameters for each color channel c include a scaling value a.sub.c and a power value p.sub.c.
23. The apparatus of claim 17, wherein the error evaluated by the cost function is a mean absolute error between the expected doses D.sub.i at the pixel locations i and the calculated doses D.sub.c,i of the radiation in the color channel c absorbed by the RCF at the locations corresponding to the pixel locations i.
24. The apparatus of claim 17, wherein the corrected net optical density netOD.sup.corrected.sub.c,i of the RCF in the color channel c at the pixel location i is determined using an averaged pixel value PV.sub.c,unexp of a digital image of an unexposed RCF in the color channel c.
25. The apparatus of claim 24, wherein determining the corrected net optical density netOD.sup.corrected.sub.c,i of the RCF for the color channel c at the pixel location i using the averaged pixel value PV.sub.c,unexp of the digital image of the unexposed RCF in the color channel c comprises: using the averaged pixel value PV.sub.c,unexp of the digital image of the unexposed RCF in the color channel c to calculate a scanned optical density scanOD.sub.c,unexp of the digital image of the unexposed RCF in the color channel c; and using the scanned optical density scanOD.sub.c,unexp of the digital image of the unexposed RCF in the color channel c to determine the corrected net optical density netOD.sup.corrected.sub.c,i of the RCF in the color channel c at the pixel location i.
26. The apparatus of claim 17, wherein, for each color channel c of the multiple color channels, for each pixel location i of the digital image of the RCF, using the pixel value PV.sub.c,i at the pixel location i in the color channel c and the variation value V.sub.i for the pixel location i to determine a corrected net optical density netOD.sup.corrected.sub.c,i of the RCF in the color channel c at the location of the RCF corresponding to the pixel location i comprises: using the pixel value PV.sub.c,i at the pixel location i in the color channel c to calculate a scanned optical density scanODc,i for the pixel location i in the color channel c; and using the scanned optical density scanODc,i for the pixel location i in the color channel c and the variation value V.sub.i for the pixel location i to determine the corrected net optical density netOD.sup.corrected.sub.c,i of the RCF in the color channel c at the location of the RCF corresponding to the pixel location i.
27. The apparatus of claim 17, wherein the apparatus is further configured to compare the corrected doses with the expected doses Di and determine whether the LINAC meets acceptable performance criteria, wherein a patient is exposed to treatment radiation generated by the LINAC only if the LINAC was determined to meet the acceptable performance criteria.
28. A system comprising: the apparatus of claim 17; the LINAC; and the scanner.
29. A method comprising: scanning a radiochromic film (RCF) across multiple color channels to generate a digital image, wherein an optical density of the RCF changes as a result of radiation exposure, the RCF contains an area exposed to a pattern of radiation doses, and the digital image includes responses of the RCF in multiple color channels; determining an optimal mathematical function that linearizes and scales a response of the RCF in each of the multiple color channels so that a transformed pixel response of each of the multiple color channels is optimally matched to an expected dose distribution for the RCF; applying the optimal mathematical function to determine a contribution of the dose-independent portion, wherein removing the dose-independent portion minimizes differences in radiation dose values across the multiple color channels; and removing the dose-independent portion from the color channels to produce a dose image that reflects only dose-dependent values.
Description
BRIEF DESCRIPTION OF THE DRAWINGS
[0044] The patent or application file contains at least one drawing executed in color. Copies of this patent or patent application publication with color drawing(s) will be provided by the Office upon request and payment of the necessary fee.
[0045] The accompanying drawings, which are incorporated herein and form part of the specification, illustrate various, non-limiting embodiments of the present invention. In the drawings, like reference numbers indicate identical or functionally similar elements.
[0046]
[0047]
[0048]
[0049]
[0050]
[0051]
[0052]
[0053]
[0054]
[0055]
[0056]
[0057]
[0058]
[0059]
[0060]
[0061]
[0062]
[0063]
[0064]
[0065]
[0066]
DETAILED DESCRIPTION OF PREFERRED EMBODIMENTS
[0067] In some aspects, as shown in
[0068]
[0069] In some aspects, the apparatus 908 may be configured to perform relative optimized linearization (ROL). In some aspects, the ROL may improve the approach of Devic by introducing a variable power function where the values of a and p in Equation 4 can be determined numerically via optimization using a cost function that evaluates the mean absolute error between the expected and the calculated doses for each color channel:
where D.sub.i is the expected dose at pixel i, obtained, for example, from a treatment planning system (TPS); netOD.sub.c,i is the measured net optical density at pixel i for color channel c, and is a minimum dose threshold, which may prevent the optimization from disproportionately prioritizing large low-dose regions outside the radiation field. In this form of the cost function, both the power value p.sub.c and the scaling value a.sub.c are optimized simultaneously for each color channel c, resulting in a linearization and normalization that aims to match the measurement to the expected dose across all pixels where the expected dose is above the threshold . With regard to the scaling value a, this approach differs from traditional normalization methods, such as single-point maximum value or central-axis normalization, which are less robust due to potential film non-uniformity at the normalization point.
[0070] The multichannel dosimetry (MCD) method may utilize all three color channels to account for non-uniformities that introduce dose-independent film responses. The effectiveness of the MCD approach may depend on the extent to which all three channels accurately represent these non-uniformities so they can be properly corrected in the final result. This dependence may explain why relative dosimetry has not been applied to the film non-uniformity problem. For example, Devic reported that the red channel may not linearize well at doses exceeding 1 Gy. Moreover, because only the coefficients of determination (R2) for the linear fit were provided, and not the resulting linearization errors, it is difficult to fully evaluate the effectiveness of the technique from Devic.
[0071] To assess the feasibility of replacing the MCD method with a relative approach that does not require acquiring dose-response curves, the accuracy of the linearization method proposed by Devic was evaluated through simulations based on EBT4 (Ashland Inc., Wayne, NJ, USA) film dose-response data. This evaluation, which first used the original fixed power value p of , aimed to verify that the linearization process was sufficiently robustmaintaining accuracy within 1% across all three color channels, to provide reliable inputs for optimizing non-uniformity correction. The simulations were then repeated with optimized (non-fixed) power value and the results were compared to evaluate the effectiveness of the proposed technique.
[0072] The evaluation was based on a simulated one-dimensional linear dose array, D.sub.exp, ranging from 10 cGy to 3 Gy in 3 cGy increments. This uniformly increasing set of dose values was first converted into red, green, and blue pixel values using measured EBT4 dose-response curves, using the following rational function:
where scanOD is the scanner optical density determined using Equation 2, D is the delivered dose, and a, b, and c are the fitting parameters. The fit parameters for this simulation were determined from films irradiated with a 6 MV flattened beam at 0, 25, 50, 75, 100, 150, 200, 300, 400, 500, 800, and 1000 cGy. Pixel values were obtained by averaging regions of interest (ROIs) approximately 22 cm in size from each film patch.
[0073] All three arrays of pixel values were then transformed into net optical density (netOD) and then linearized using the second term in Equation 4. Scaling factors a for each color channel were then applied to ensure that the maximum linearized netOD for each channel matched the maximum of the original dose profile, yielding the corrected dose profile D.sub.lin. The results were compared to the original dose values to assess the technique's ability to linearize the channel-specific non-linear dose-response of EBT4 film, using global percentage errors (.sub.i) computed across all three color channels as follows:
where D.sub.exp,i is the input dose value at location i, Dlin,i is the output linearized pixel response at location i, and D.sub.exp,max is the maximum value in the input dose profile. This simulation was also repeated for a dose profile with maximum dose of 10 Gy. These dose ranges for these two initial simulations were selected to represent a typical clinical QA plan and the maximum dose specified in the EBT4 film datasheet.
[0074] While the initial simulations assessed linearization accuracy at specific maximum doses of 3 and 10 Gy, additional simulations were conducted to evaluate robustness across a broader range of maximum dose values, from 2 to 10 Gy in 10 cGy increments. Because the slope of the dose-response curve varies across the applicable dose range, and the maximum dose is commonly used as the normalization reference, this value may have an impact on the overall accuracy of the linearization process (as also reported by Devic). For each profile with a different maximum dose, a linearization error profile was computed using Equation 8, and the average error across the entire profile was calculated. This analysis provides a more comprehensive understanding of the method's performance when applied to measured films with varying maximum dose levels.
[0075] To extend the new ROL technique to correct for dose-independent non-uniformities, the optimization framework of Micke (Equation 3) was modified by replacing the dose computed using MCD (Equation 1) with the dose computed using ROL (Equation 4, using optimized values of a and p).
where D.sup.rol.sub.i,j is the optimized linearized dose formula modified to incorporate local dose-independent variations as follows:
where a.sub.c and p.sub.c are the optimized scale and power values for color channel c, and netOD.sub.c,i is the net optical density represented in terms of scanner optical density scanOD and the variation V.sub.i as at pixel location i and color channel c. In Equation 10a, the net optical density netOD.sub.c,i is represented in terms of scanner optical density scanOD.sub.c,i and the variation V.sub.i at pixel location i and color channel c as:
[0076] As a result, the net optical density netOD.sub.c,i incorporates the non-uniformity correction based on scanOD directly into the relative optimized linearized dose (based on netOD). To compute the net optical density netOD.sub.c,i within this framework, as shown in Equation 10 b, each raw pixel value may be converted to scanner optical density (scanOD) using Equation 2.
[0077] After the variation value V.sub.i is calculated using Equation 9 (and Equations 10a and 10b), as shown in Equation 11, the scanner optical density scanOD.sub.c,i may then be corrected for non-uniformities on a pixel-by-pixel basis by multiplying it by the variation value V.sub.i. As is also shown in Equation 11, the corrected scanner optical density (i.e., scanOD.sub.c,i.Math.V.sub.i) may then converted back to a pixel value using the inverse of Equation 2, and finally converted to a corrected net optical density (netOD) using Equation 5. This corrected netOD (shown in Equation 11) may account for the dose-independent variations consistently with the MCD method while being incorporated into the new ROL approach.
Equation 11 may be derived as follows:
[0078] The corrected netOD and the optimized scale and power values a.sub.c and p.sub.c may then be used to calculate corrected doses as shown in Equation 12:
[0079]
[0080] In some aspects, as shown in
[0081] In some aspects, as shown in
[0082] In some aspects, as shown in
[0083] In some aspects, as shown in
[0084] In some aspects, the process 1100 may include a step 1105 of computing a variation map using the scanOD and the linearization parameters determined in steps 1102 and 1103, respectively, via optimization with Equation 3. In some aspects, Equation 10 b may be applied within the optimization to convert between scanOD and netOD, because the linearization may be defined in terms of netOD while the non-uniformities may be corrected using scanOD. In some aspects, the variation map may include a variation V.sub.i at each pixel location i. In some aspects, the variation map may be calculated using Equations 9, 10a, and 10b. In some aspects, the apparatus 908 may compute the variation map.
[0085] In some aspects, the process 1100 may include a step 1106 of computing a corrected netOD using Equation 11 together with scanODs and the previously determined variation map. In some aspects, the apparatus 908 may compute the corrected netOD.
[0086] In some aspects, the process 1100 may include a step 1107 of computing the corrected dose for each channel c using Equation 12, the corrected netOD from step 1106, and the linearization parameters from step 1103. In some aspects, the apparatus 908 may compute the corrected dose for each channel c.
[0087] In some aspects, the process 1100 may include a step of applying the computed corrected doses. In some aspects, the application step may include using the corrected doses to control the LINAC 100 to generate treatment radiation that exposes a patient 910 to a radiation dose distribution in accordance with a patient-specific treatment plan.
[0088]
[0089] In some aspects, as shown in
[0090] In some aspects, as shown in
[0091] In some aspects, as shown in
[0092] In some aspects, as shown in
[0093] In some aspects, as shown in
[0094] In some aspects, the corrected net optical density netOD.sup.corrected.sub.c,i of the RCF 902 in the color channel c at the pixel location i may be determined using the averaged pixel value PV.sub.c,unexp of the digital image of the unexposed RCF 904 in the color channel c. In some aspects, determining the corrected net optical density netOD.sup.corrected.sub.c,i of the RCF 902 for the color channel c at the pixel location i using the averaged pixel value PV.sub.c,unexp of the digital image of the unexposed RCF 904 in the color channel c may include: (i) using the averaged pixel value PV.sub.c,unexp of the digital image of the unexposed RCF 904 in the color channel c to calculate a scanned optical density scanOD.sub.c,unexp of the digital image of the unexposed RCF 904 in the color channel c and (ii) using the scanned optical density scanOD.sub.c,unexp of the digital image of the unexposed RCF 904 in the color channel c to determine the corrected net optical density netOD.sup.corrected.sub.c,i of the RCF 902 in the color channel c at the pixel location i.
[0095] In some aspects, the averaged pixel value PV.sub.c,unexp and Equation 2 may be used to calculate the scanned optical density scanOD.sub.c,unexp. In some aspects, the pixel value PV.sub.c,i and Equation 2 may be used to calculate the scanned optical density scanOD.sub.c,i. In some aspects, the scanned optical density scanOD.sub.c,i, the scanned optical density scanOD.sub.c,unexp, the variation value V.sub.i, and Equation 11 may be used to determine the corrected net optical density netOD.sup.corrected.sub.c,i.
[0096] In some aspects, as shown in
[0097] In some aspects, as shown in
[0098] In some aspects, step 1103 of the process 1100 and step 1208 of the process 1200 may determine an optimal mathematical function that linearizes and scales a response of the RCF 902 in each of the multiple color channels so that a transformed pixel response of each of the multiple color channels is optimally matched to an expected dose distribution for the RCF 902. In some aspects, step 1105 of the process 1100 and step 1210 of the process 1200 may apply the optimal mathematical function to determine the contribution of the dose-independent portion, wherein removing the dose-independent portion minimizes differences in radiation dose values across the multiple color channels. In some aspects, the dose-independent portions may be in the form of the variation map. In some aspects, steps 1106 and 1107 of the process 1100 and steps 1212 and 1214 of the process 1200 may remove the dose-independent portion from the color channels to produce a dose image that reflects only dose-dependent values.
Examples
[0099] To assess the ability of the ROL method to replicate dose distributions produced by MCD, EBT4 film, which is a type of RCF, was used to measure dose distributions for three treatment plans: open and wedged 1010 cm fields, and a volumetric modulated arc therapy (VMAT) plan. The comparison was based on dose-difference maps between the traditional MCD method and the ROL method across all three color channels.
[0100] An open field was chosen as the simplest example to demonstrate the technique's effectiveness. A wedge field was included to evaluate performance in mid-dose ranges, which are typically obscured within the penumbra. The VMAT plan was selected as a real-world clinical example to assess the method's applicability.
[0101] All plans were created in Eclipse (Siemens Healthineers, Erlangen, Germany), and the expected planar doses were exported in Digital Imaging and Communications in Medicine (DICOM) format. During both plan delivery and calibration, the RCF was positioned at a depth of 5 cm within a polystyrene slab, with an additional 5 cm of backscatter material placed behind it. The VMAT test plan included two spherical targets positioned approximately 7 cm apart in the coronal plane at isocenter, with diameters of approximately 4 and 5.5 cm. Each target was prescribed a dose of 300 cGy.
[0102] Films were scanned using an Epson 11000 XL scanner in transparency mode with 16 bits per channel at a resolution of 150 dpi, with all color adjustments disabled. The cross-plane direction of the film was aligned with the scanner's carriage travel direction. A glass plate was placed over the films to ensure they remained flat during scanning.
[0103] The one-scan method proposed by Lewis was used to scale the original dose-response curves, and the triple-channel non-uniformity correction method of Micke was subsequently applied. As part of the one-scan method, two reference films were scanned alongside each quality assurance (QA) RCF film: one unexposed and the other exposed to the maximum dose expected in the QA RCF film. Films were physically marked with fiducials based on the room lasers, and these marks were used to register the scanned images. Small registration adjustments were also made in software to optimally align the measured and expected distributions spatially. Ideal spatial coincidence was desired for benchmarking the technique; however, controlled spatial errors were also introduced to assess the method's robustness. Additionally, different threshold values (e.g., z in Equation 6) of 1%, 5%, 10% and 15% were evaluated to assess the effect of threshold selection, and to support the selection of a final threshold value for comparing ROL and MCD dose across all test plans.
[0104] To evaluate the robustness of the ROL method against discrepancies between measured and expected dose distributions, the VMAT test case was repeated with two types of errors: (1) a 3 mm spatial positioning error applied in the cross-plane direction, referred to as VMAT-shifted, and (2) a dose delivery error in which the second of the two arcs was prematurely terminated by approximately 25%, referred to as VMAT-partial. This analysis is of particular interest because the expected dose is used in the ROL cost function to optimize the linearization parameters, and any mismatch between the measured film dose and the expected dose can lead to inaccurate or misleading results. Spatial positioning and dose delivery errors are common in clinical QA tests. The ROL technique would therefore be impractical if the resulting dose distribution were strongly affected by such sources of error. These two scenarios were selected as worst-case conditions to evaluate the ROL method's robustness.
[0105] To evaluate the ability of the ROL method to detect errors arising from incorrect treatment planning modeling parameters, an analysis was performed using a C-shape test plan based on the American Association of Physicists in Medicine (AAPM) Task Group 119 guidelines. In this case, errors in the multileaf collimator (MLC) modeling, specifically, the dosimetric leaf gap (DLG), were simulated by systematically adjusting the MLC offset in the treatment planning system. This test evaluates the method's sensitivity to beam model discrepancies, such as those introduced during LINAC commissioning or TPS configuration.
[0106] The C-shape plan consisted of a PTV surrounding a central cylindrical core. The PTV was an arc-shaped structure with an inner radius of 1.5 cm, an outer radius of 3.7 cm, and a length of 8 cm. The central core, which was separated from the PTV by a 0.5 cm margin, was a 1 cm-radius cylinder with a length of 10 cm. RCF was irradiated with a 6 MV flattened beam using a modular film phantom composed of 2 cm-thick slabs of 1515 cm Acrylonitrile Butadiene Styrene (ABS) material (density: 1.04 g/cm3), with laser-cut EBT4 film precisely registered within the phantom using a three-pin registration system.
[0107] The plan was based on a synthetic computed tomography (CT) dataset, with the origin aligned to the physical center of the film. This ensured accurate origin definition during treatment planning. Irradiations were performed on a Truebeam LINAC (Siemens Healthineers, Erlangen, Germany) with the film positioned in the axial orientation. Setup was performed via CBCT registration and verified using in-room lasers. No additional shifts were introducedall analysis was based solely on the geometric accuracy of the Cone-Beam Computed Tomography (CBCT) alignment and film registration system.
[0108] Dose distributions were calculated using Eclipse v18.0.1.261 (Siemens Healthineers, Erlangen, Germany), with MLC offset values varied from 0.0 to 0.4 mm in 0.1 mm increments. Film registration, including rotation and translation corrections, was executed using custom software based on fiducial markers and phantom pin alignment. Gamma analysis was conducted using a global 2 mm, 3% criterion, with pass rates computed over the region receiving more than 10% of the maximum dose.
[0109] While the principal aim was to evaluate the ROL method's sensitivity to small variations in the planning model, context for the results was provided by conducting a parallel evaluation of gamma pass rates between doses predicted by the TPS and those measured via the MCD technique on the same film. The MCD dose was derived using the same procedure described above which included dose response curves, one-scan film patches, and computation of variation maps. Gamma pass rates for each MLC offset were compared between the two methods.
ExamplesResults of Relative Linearization Without Optimization
[0110] Simulation results using measured EBT4 dose-response curves and the original (non-optimized) linearization method of Devic, applied to dose profiles with maximum doses of 3 and 10 Gy, are shown as dashed lines in
[0111] Repeating this type of simulation with datasets featuring different maximum doses ranging from 1 to 10 Gy yielded the average linearization errors shown as the dashed lines in
ExamplesResults of Relative Linearization With Optimization
[0112] The optimal power and scale values (a and p in Equation 4) for both 3 and 10 Gy simulations for all three color channels are shown in rows 1 and 2 of Table 1 below. The solid lines in
TABLE-US-00001 TABLE 1 p a Red Green Blue Red Green Blue Sim, 3 Gy 0.690 0.669 0.727 1065.5 1920.8 5940.0 Sim, 10 Gy 0.657 0.531 0.753 1020.8 1498.9 6366.9 Open 0.703 0.678 0.723 1157.0 2096.2 6614.7 Wedge 0.675 0.639 0.691 1128.4 1970.7 6233.6 VMAT 0.762 0.772 1.027 1204.8 2375.9 13976.3 VMAT-shifted 0.754 0.769 1.025 1185.5 2352.3 13813.8 VMAT-partial 0.730 0.713 0.823 1307.4 2438.0 9175.9
[0113] Measured response curves for EBT4 film were used with the MCD method to compute the planar dose across all three color channels for each film, serving as references to evaluate the effectiveness of the new ROL approach. The optimal power (p) and scaling values (a) determined for all test fields are shown in Table 1. The computed variation maps for the wedge field, generated using both MCD and ROL, are presented in
[0114] The average absolute error, and standard deviation of the raw error, evaluated within the low-, middle-, and high-dose regions, are shown in Table 2 below.
TABLE-US-00002 TABLE 2 Test Plan Channel Low Middle High Open Red 0.28 0.09 0.15 0.17 0.42 0.26 Green 0.50 0.22 0.37 0.31 0.63 0.33 Blue 0.23 0.27 0.54 0.36 0.36 0.43 Wedge Red 0.51 0.14 0.47 0.35 0.70 0.31 Green 0.87 0.28 0.51 0.37 0.37 0.43 Blue 0.43 0.26 0.71 0.42 0.61 0.42 VMAT Red 0.16 0.18 0.20 0.32 0.18 0.22 Green 0.31 0.33 0.76 0.33 0.19 0.24 Blue 0.32 0.31 0.66 0.37 0.38 0.36 VMAT shifted Red 0.16 0.16 0.13 0.28 0.39 0.24 Green 0.32 0.33 0.79 0.35 0.39 0.23 Blue 0.28 0.30 0.65 0.37 0.70 0.36 VMAT partial Red 0.92 0.40 3.43 1.73 10.19 0.59 Green 1.20 0.41 3.20 1.51 10.13 0.71 Blue 0.72 0.46 2.98 1.65 9.81 0.72
[0115] Dose profiles for the VMAT field, computed using both MCD and ROL, are shown in
[0116] Gamma analysis results from the MLC offset sensitivity tests are shown in
TABLE-US-00003 TABLE 3 Gamma pass rate (%) MLC Offset (mm) MCD ROL 0.0 99.69 99.34 0.1 99.80 99.35 0.2 99.79 99.39 0.4 99.62 99.35
[0117] The ROL method for radiochromic film dosimetry eliminates the need for dose-response curve measurements. The ROL method uses an improved linearization formula for pixel-to-dose conversion, demonstrating accuracy across the full EBT4 dose range up to 10 Gy. The ROL method incorporates non-uniformity corrections, which achieve high-accuracy film dosimetry.
[0118] The results from the simulations performed provide insight into the average linearization errors that can be expected from ROL as a function of the maximum dose delivered to the film. For example, the results shown in
[0119] This behavior stands in contrast to the non-optimized linearization, where the accuracy is more sensitive to the maximum dose, and this sensitivity can vary by film type. For instance, Devic reported that linearization for EBT2 and EBT3 films resulted in inaccuracies in the red channel when the maximum dose exceeded 1 Gy. To confirm this, we ran additional simulations using published EBT3 dose-response curves and observed similar trends: the red channel showed the largest errors, with maximum deviations around 4% at 2 Gy, increasing to 7% at 10 Gy. In contrast, the EBT4 non-optimized simulation results presented here in
[0120] The ROL technique was evaluated by comparing it with established multichannel dosimetry across all color channels for various test cases, including open and wedge fields, as well as a VMAT plan. The results demonstrated that the ROL method successfully solved for the linearization parameters, yielding optimized power values ranging from 0.531 to 1.027. While these values deviate from the fixed exponent of , they remain within a similar range, suggesting that a variable power offers improved flexibility and accuracy across different dose distributions.
[0121] The ROL technique demonstrated the ability to produce variation maps comparable to those generated by MCD (see
[0122] As shown in
[0123] In some aspects, the linearization optimization cost function (e.g., Equation 6) may include a minimum dose threshold. Without this threshold, the size of the measured film may influence the results, as larger films contained a greater number of out-of-field pixels. The evaluation of different threshold values, shown in
[0124]
[0125] The sensitivity tests demonstrated that ROL performs well in the presence of spatial positioning errors between the expected and measured dose distributions. This is evident from the similarity of the gamma maps shown in
[0126] Similar trends were observed when small changes were introduced into the treatment planning model. The ROL method was able to resolve differences in DLG values based on variations in MLC offset, indicating that its output reflects sensitivity to the underlying model parameters rather than simply conforming to the expected dose distribution used during optimization. Notably, the MCD results suggested an optimal MLC offset of 0.1 mm, whereas the parallel analysis using ROL methods identified 0.2 mm as optimal, which matched the value previously determined and used clinically by the local physics team in their treatment planning system.
[0127] A suitable clinical use case for ROL is routine QA of patient-specific planar dose distributions, where measured film is compared to the expected dose from the TPS, provided that baseline pass rates have been established using MCD during the TPS commissioning and validation process. Another application is the evaluation of subtle changes in dose distribution resulting from modifications to the treatment planning dose model, where ROL can help isolate the dosimetric impact of specific parameter adjustments.
[0128] From a clinical implementation standpoint, ROL offers practical advantages, especially in busy clinics with limited resources. Traditional multichannel dosimetry workflows require time-intensive preparation: generating and analyzing dose-response curves for each film batch, performing calibration irradiations, scanning multiple films, and ensuring consistent scanner placement and orientation. Errors introduced by misaligned film placement or inconsistent scanning procedures have been reported, which emphasizes the complexity of current workflows.
[0129] The ROL method eliminates the need for batch-specific dose-response measurements and one-scan calibration films. Instead, the ROL method may be carried out using an unexposed RCF, which may be from the same batch and may be scanned and analyzed alongside the QA RCF. The ROL approach simplifies the process, reduces the potential for operator error, and streamlines QA film analysis, making it particularly attractive for high-throughput clinical environments. This, the ROL method is an accurate and efficient approach for converting measured RCF optical density to dose using relative techniques, simplifying the film dosimetry process by eliminating the need for time-consuming calibration curves.
[0130] While various embodiments are described herein, it should be understood that they have been presented by way of example only, and not limitation. Thus, the breadth and scope of this disclosure should not be limited by any of the above-described exemplary embodiments. Moreover, any combination of the above-described elements in all possible variations thereof is encompassed by the disclosure unless otherwise indicated herein or otherwise clearly contradicted by context.
[0131] Additionally, while the processes described above and illustrated in the drawings are shown as a sequence of steps, this was done solely for the sake of illustration. Accordingly, it is contemplated that some steps may be added, some steps may be omitted, the order of the steps may be re-arranged, and some steps may be performed in parallel. For example, steps 1102 and 1104 of the process 1100 of