Feet positioning system for magnetic resonance imaging studies
10765341 · 2020-09-08
Assignee
Inventors
- Carlos Alberto Cabal Mirabal (Havana, CU)
- Evelio Rafael González Dalmau (Havana, CU)
- Luis Manuel Flores Díaz (Havana, CU)
- Jorge Amador Berlanga Acosta (Havana, CU)
- Luis Saturnino Herrera Martinez (Havana, CU)
Cpc classification
A61B5/055
HUMAN NECESSITIES
A61B5/706
HUMAN NECESSITIES
A61B90/39
HUMAN NECESSITIES
International classification
A61B5/055
HUMAN NECESSITIES
A61N5/10
HUMAN NECESSITIES
A61B5/00
HUMAN NECESSITIES
A61B90/00
HUMAN NECESSITIES
Abstract
The present invention is related to a system and method to ensure the reproducibility of the position of the feet and lower parts of the legs during Magnetic Resonance Imaging (MRI) studies, and to obtain robust quantitative information through the time. The system includes a device that is inserted into the radiofrequency coils of any MRI equipment. The device includes a foot support section, a leg support section and a base adapted to allow that the abovementioned sections be fixed in it. With this device and method, and through external and internal markers, quantitative studies of the evolution of pathophysiological phenomena that affect the anatomy and physiology of the feet and lower parts of the legs are performed.
Claims
1. A magnetic resonance imaging (MRI) system to control the orientation of a foot of an individual during a MRI scanning process, the system comprising a foot positioning device, wherein said foot positioning device comprises: (a) a foot support section that includes: 1. a foot support surface to situate said foot of said individual in a fixed position relative to the foot positioning device; wherein said foot support surface comprises at least two image marker elements that are visible when MRI images are recorded; 2. an arch for a heel adapted to be positioned behind the heel of the individual and sliding regarding said foot support surface; 3. means to fix the foot of the individual on said foot support surface during the scanning process; (b) a leg support section including: 1. a leg support to situate a leg of said individual in a fixed position regarding the foot positioning device; 2. means to fix said leg of the individual to said leg support during the scanning process; (c) a foot positioning device base adapted to allow that the foot support surface and the leg support section be fixed in said base and positioned on a scanning bed; wherein said foot positioning device is adapted to be inserted into a radiofrequency (RF) coil of the MRI system for scanning in a reproducible position regarding an axis of a static magnetic field of said MRI system.
2. The system according to claim 1, wherein said foot positioning device base is adapted to allow that said foot support section could be placed in a series of fixed positions indicated on a scale regarding said foot positioning device base.
3. The system according to claim 1, wherein the at least two image marker elements are located in parallel to said foot support surface and define a plane of a magnetic resonance image.
4. The system according to claim 1, wherein said foot positioning device comprises material not visible under visualization of magnetic resonance images.
5. The system according to claim 4, wherein said material is polyvinyl chloride.
6. The system according to claim 1, wherein said foot positioning device comprises an additional foot support surface, an additional arch for an additional heel, and an additional leg support, wherein said additional foot support surface, additional arch, and an additional leg support are positioned on the foot positioning device base.
7. A method to control the orientation of a foot of an individual regarding a magnetic resonance imaging system during a magnetic resonance imaging scanning process, the method comprising: (a) situating a foot positioning device in a fixed and reproducible position in a radiofrequency coil of the magnetic resonance imaging system; (b) locating the foot of the individual in the foot positioning device while the foot positioning device is in said fixed and reproducible position with regard to image marker elements; (c) recording first images of magnetic resonance imaging to check correct foot positioning; (d) optionally correcting the foot positioning regarding the image marker elements and the foot positioning device in the magnetic resonance imaging system; and optionally (e) recording second magnetic resonance images; wherein said foot positioning device is a device comprising: a foot support section that includes a foot support surface to situate said foot of said individual in a fixed position relative to the foot positioning device; wherein said foot support surface comprises said image marker elements, wherein said image marker elements are visible when magnetic resonance imaging images are recorded; an arch for a heel adapted to be positioned behind the heel of the individual and sliding regarding said foot support surface; means to fix the foot of the individual on said foot support surface during the scanning process; a leg support section including a leg support to situate a leg of said individual in a fixed position regarding the foot positioning device; means to fix said leg of the individual to said leg support during the scanning process; a foot positioning device base adapted to allow that the foot support surface and the leg support section be fixed in said base and positioned on a scanning bed; wherein said foot positioning device is adapted to be inserted into a radiofrequency (RF) coil of a magnetic resonance imaging system for scanning in a reproducible position regarding an axis of a static magnetic field of said magnetic resonance imaging system.
Description
BRIEF DESCRIPTION OF THE FIGURES
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
(11)
(12)
(13)
(14)
EXAMPLES
(15) The following examples are shown for illustrative purposes, and should not be considered as limiting the invention.
Example 1
Positioning of the Individual
(16) The first step to ensure reproducibility of quantitative measurements for the scan of MRI was to correctly position both feet and legs of the individual in the device shown in
(17) The feet and the lower parts of the legs were placed and fixed, to obtain simultaneously MRI (e.g., coronal and axial sections) and Magnetic Resonance Spectra (MRS) of both feet, without changing the position. This allowed us to compare, on an equal condition, both limbs along the studies, and so that a lower member serves as reference to the other one.
(18) Each foot carefully rested on the support surfaces 1 and heels were leaned on arches 3. At unison, the lower parts of the legs were supported on supports 2, which were conveniently fitted with pins 8, according to the dimensions of the feet. The positions of the support surfaces 1 and the heel arches 3 on the scale attached to the device (denoted by 5) were recorded.
(19) People examined bent the legs slightly, as shown in
Example 2
Checking and/or Correction of the Position of the Feet
(20) Once positioned the feet of the individual, he was placed at the isocenter of the magnet system, and proceeded to record the planning MRI in three sections: coronal, sagittal and axial. The MRI showed the external markers 7. The correct position of the feet was checked, so that the MRI of soles appeared fully supported on the support surfaces 1, determined by the pairs of external markers 7 for each foot. In case the positioning was not correct, it was corrected as in Example 1. If the positioning was correct, the final planning of study sections proceeded.
Example 3
Planning and Orientation of the Sections
(21) The first section to be recorded was oriented parallel to the plane determined by the external markers, although it could be any plane, according to a preset angle in reference to the one determined by such markers. Other necessary sections were determined in connection with this first one, according to the study to be performed.
(22) In
Example 4
Determination of Internal Markers: Position of Anatomical Structures
(23) In addition to external markers (denoted as 7), there were established internal controls that allowed determining the position, its reproducibility and evaluation of error in the serial MRI studies. This was essential, especially for those patients with inflammatory processes, since in these cases the determination of the sizes and relative positions of the anatomical parts and their evolution, either naturally or due to treatment regimens, is difficult.
(24) As an internal marker it was defined an internal anatomical structure of the foot, which was chosen in the way that it was not affected or it was far from the pathological processes affecting the foot, in particular the inflammatory ones. In this case it was taken as an internal marker the perpendicular distance L.sub.o from the center of the tolocalcaneum interosseum ligament to the segment joining the two external markers (see
(25) In Table 1 the values of L.sub.1 and L.sub.2 measured on images from 10 healthy volunteers are shown, recorded at two different times, in which the feet were always placed in identical positions. Surprisingly, as shown in Table 1, the mean changes L.sub.1 and L.sub.2 (variation between two successive positions in two different studies), are less than 1.0 mm (The maximum variation was 6.7%), which evidences how robust the device and the procedure are.
(26) TABLE-US-00001 TABLE 1 Criteria of the feet correct positioning. Measurements L.sub.1 and L.sub.2 in the sagittal MRI sections of 10 healthy volunteers studied in two separate occasions. Volunteer No 1 2 3 4 5 6 7 8 9 10 Length (mm) Study 1 (S.sub.1) L.sub.1 71 54 58 64 65 69 60 44 68 55 L.sub.2 13 5 7 7 8 20 16 31 13.5 16 Study 2 (S.sub.2) L.sub.1 70 54 57 62 65 69 58 42 66 54 L.sub.2 13 5 7 7 8 19 15 31 14 17 between L.sub.1 1 0 1 2 0 0 2 3 2 1 studies L.sub.2 0 0 0 0 0 1 1 0 1 1 (L.sub.x = L.sub.xS1 L.sub.xS2) x = 1 2 % Variation L.sub.1 1.4 0 1.8 3.2 0 0 3.4 4.5 3.0 1.9 (L.sub.x/L.sub.xS2*100) L.sub.2 0 0 0 0 0 5.3 6.7 0 3.6 5.9 L1 and L2 are variations of L.sub.1 and L.sub.2 from a single healthy volunteer from one study to another, at different times.
Example 5
Determining Internal Markers: Anatomical Structures Area
(27) Besides the external markers (indicated as 7), and the first internal marker described in Example 4, a second internal marker was defined as the area of a predetermined anatomical structure, according to MRI assessments that were required to perform.
(28) In this example, to illustrate, this second internal marker was defined as the areas of multiple calcaneus coronal sections. Five different coronal sections were chosen. The record of multiple coronal sections ensures several evaluations from different areas, located at dissimilar distances of possible inflammatory processes or changes in other regions of the foot and/or lower parts of the leg.
(29) The ratio of the measured areas in different studies, on images of different structures, was an undeniable internal control of foot positioning and orientation of the sections. This internal marker is totally conclusive, complements and is consistent with the results presented and related to the first internal marker (Example 4).
(30) In serial studies performed under the conditions described, variations in the size of different parts of the foot were below 4.5%. Any variation greater than this value is attributable solely to the evolution of pathophysiological processes of the feet. In Table 2, the coefficient of variation of the calcaneus area is shown, measured in five coronal sections, at two different times, in 10 volunteers.
(31) TABLE-US-00002 TABLE 2 Demonstration of the reproducibility of the feet position calculated from the calcaneus area as a second internal marker. Different sections of the calcaneus I II III IV V Coefficient of 1 0.86 1.48 0.54 1.21 1.32 variation % 2 0.15 0.08 0.48 0.57 1.03 (10 volunteers) 3 1.81 0.10 0.10 0.36 0.03 4 1.56 4.35 2.67 0.52 2.11 5 2.65 0.27 0.11 0.78 1.65 6 0.66 0.56 0.16 1.06 0.08 7 2.78 0.23 0.45 0.59 0.62 8 0.72 2.08 0.90 2.90 4.40 9 1.45 1.61 0.43 1.68 0.95 10 2.67 2.40 0.16 1.03 0.93
Example 6
Determination of Evolution of the Dimensions of Diabetic Foot Ulcers (DFU) Under Treatment
(32) The guarantee of proper positioning and reproducibility allowed the quantitative evaluation of the DFU cicatrization kinetics by measuring the DFU area and volume changes during the treatment with EGF. The MRI of 25 DFU patients, taken in identical positions by the system and method of the invention, allowed the measurement of the lesion sizes, with amazing accuracy.
(33) In
Example 7
Determination of Evolution of the Edema Volume in the Feet of Patients with Inflammatory Processes
(34) The guarantee of the feet correct positioning, and its reproducibility with the system and method of the invention, allowed the quantitative evaluation of the kinetics of edema volume changes (swelling) due to the DFU; which is applicable to any other pathology associated with edema. For the 25 patients evaluated in Example 6, the values of the edema volumes were determined, throughout the treatment period. An example of the behavior of edema in DFU patients, treated with EGF, is shown in
(35) The rate of change of edema with respect to treatment time can be calculated from the values shown in
Example 8
Quantitative Evaluation by MRI of the Texture Evolution of Foot Lesions
(36) A reproducible position of the feet, as a result of the system and method of the invention, allowed recording the Diffusion MRI of patients with DFU, at different times after the beginning of treatment with EGF, and from them the Apparent Diffusion Coefficients (ADC) were calculated in the 25 patients of Example 6. ADC is a complex function of several properties, including the texture of the tissue where the measurement takes place. Only the guarantee of the position accuracy enabled to establish that the relationship of the ADC were a function depending solely on texture changes.
(37) In
(38) Quantification of changes in the molecular mobility in the lesions and their edges, through MRI, gives unexpectedly valuable information to assess the response to treatment and the onset of the granulation and epithelization processes in the DFU. In
Example 9
Quantitative MRI Evaluation of the Evolution of In Vivo Metabolic Activity in Diabetic Foot Ulcers (DFU)
(39) The guarantee of reproducibility of the position of the feet is an essential condition to perform quantitative follow-up studies, of the metabolic activity of foot diseases, from in vivo MRS studies whether mono voxel or multi-voxel. The guarantee of the exact location of the voxel is a necessary condition to compare the spectra over serial studies, and assess the response to therapy. Once both feet and lower parts of the legs are placed, it is possible to ensure that the voxels in size, position and orientation are the same through the longitudinal studies. Two lines are highlighted in the spectra, corresponding to the Lipids (Lip) and Creatine (Cr). The MRS amplitudes of healthy foot are at least twofold the ones of the foot affected by DFU. Also, the ratio of amplitudes Lip/Cr changes from the spectrum of the healthy foot to the foot affected by DFU, which is one of the biomarkers of the status of DFU.
Example 10
Evaluation of the Evolution of an Osteomyelitis Patient
(40) In