HIV VIRAL LOAD TESTING
20190018035 ยท 2019-01-17
Assignee
Inventors
Cpc classification
B01L3/0275
PERFORMING OPERATIONS; TRANSPORTING
B01L2200/16
PERFORMING OPERATIONS; TRANSPORTING
B01L2200/025
PERFORMING OPERATIONS; TRANSPORTING
B01L2300/0609
PERFORMING OPERATIONS; TRANSPORTING
G01N35/10
PHYSICS
G01N2035/1053
PHYSICS
B01L3/5635
PERFORMING OPERATIONS; TRANSPORTING
B01L2200/026
PERFORMING OPERATIONS; TRANSPORTING
B01L2200/10
PERFORMING OPERATIONS; TRANSPORTING
B01L2300/046
PERFORMING OPERATIONS; TRANSPORTING
B01L3/502
PERFORMING OPERATIONS; TRANSPORTING
G01N35/1011
PHYSICS
B01L2200/0668
PERFORMING OPERATIONS; TRANSPORTING
B01L2200/023
PERFORMING OPERATIONS; TRANSPORTING
G01N2035/00277
PHYSICS
C12Q1/6806
CHEMISTRY; METALLURGY
International classification
G01N35/10
PHYSICS
C12Q1/6806
CHEMISTRY; METALLURGY
B01L3/00
PERFORMING OPERATIONS; TRANSPORTING
Abstract
Methods of testing HIV viral load are described. The methods comprise detecting HIV viral RNA in a sample of leukocyte-depleted blood. Such methods can be carried out on low-volume samples obtained without the need for venipuncture or a centrifuge. The methods are particularly suited for HIV viral load testing in resource-limited settings. Methods for monitoring HIV infection are also described, as well as kits for carrying out the methods.
Claims
1-26. (canceled)
27. A kit for HIV viral load testing, which comprises a leukoreduction filter, and reagents required for reverse transcription of HIV viral RNA, and subsequent isothermal nucleic acid amplification of a product of the reverse transcription.
28. The kit according to claim 27, further comprising a means for applying a pressure differential across the leukoreduction filter.
29. The kit according to claim 27, further comprising an isotonic solution for diluting a sample of whole blood prior to leukodepletion of the sample.
30. The kit according to claim 27, further comprising a lancet for obtaining a sample of whole blood from a subject by finger prick or heel prick.
31. The kit according to claim 27, further comprising a blood collector for collecting a sample of blood from a subject.
32. (canceled)
33. The kit according to claim 27, further comprising a visually detectable label for labelling a product of the isothermal nucleic acid amplification and/or a chromatographic test strip for capturing and detecting a product of the isothermal nucleic acid amplification.
34. The kit according to claim 27, further comprising means for extracting nucleic acid from a sample of leukocyte depleted blood.
35. A method of testing HIV viral load, comprising: i) filtering a sample of whole blood through a leukoreduction filter of the kit of claim 27 to provide a sample of leukocyte depleted blood; ii) reverse transcribing HIV viral RNA of the sample of leukocyte depleted blood using reagents of the kit of claim 27; iii) amplifying, by isothermal nucleic acid amplification, a product of the reverse transcription using reagents of the kit of claim 27; and iv) detecting the product of the isothermal nucleic acid amplification; wherein the method is carried out without preparing a plasma sample.
Description
[0066] Embodiments of the invention are described in the following examples with reference to the accompanying drawings in which:
[0067]
[0068]
[0069]
[0070]
[0071]
[0072]
[0073]
[0074]
[0075]
EXAMPLES
Materials and Methods
[0076] Whole blood was spiked with cells of the 8E5 cell line containing one copy of proviral HIV DNA, or with culture supernatant of HIV-1 LAI (RNA, subtype B).
[0077] The efficacy of whole blood filtration for nucleated cells was assessed in three different ways:
i) White cell count pre- and post-filtration using a sensitive flow cytometric assay. This method counts white cells present in the whole blood as well as the spiked 8E5 cells;
ii) Relative CD45+ cell amounts were assessed using the Taqman Gene Expression PCR assay (Life Technologies, primer/probe set Hs04189704_m1). The reverse transcription (RT) step required for this assay was completed using the SuperScript III First Strand Synthesis System (Invitrogen) according to the manufacturer's specifications. The CD45 PCR was completed using the Stratagene Mx3000 instrument. This method detects white cells present in the whole blood as well as the spiked 8E5 cells;
iii) Quantification of proviral DNA against a plasmid-derived standard. This was performed by real-time PCR (qPCR) using the Stratagene Mx3000 as previously reported (Candotti et al., J Virol Methods 2004; 118: 39-47: Multiplex real-time quantitative RT-PCR assay for hepatitis B virus, hepatitis C virus and human immunodeficiency virus type 1); HIV-1 RNA was quantified using real-time RT-PCR as previously described (Candotti et al. 2004). The reference used was a secondary standard calibrated using the Artus HlVirus-1 RG RT-PCR kit in copies/ml.
[0078] All nucleic acid extractions for PCR and RT-PCR reactions were performed with the High Pure Viral Nucleic Acid kit (Roche), according to the manufacturer's instructions.
[0079] Where filtered samples were tested alongside plasma as a comparison to the leukoreduction levels normally employed for viral load testing, plasma was generated by centrifugation at 4,000 rcf (relative centrifugal force) for 15 minutes.
Example 1
Comparison of Different Leukoreduction Filters
[0080] The efficacy of removal of leukocytes from whole blood spiked with 8E5 cells using three different leukoreduction filters (a Macopharmal filter, containing five layers, a Macopharma2 filter, containing six layers, and a Pall filter) was compared. Leukocyte removal by each filter was assessed using the three different methods described above: (i) white blood cell count by flow cytometric analysis; (ii) white blood cell quantification by RT and CD45 qPCR; and (iii) qPCR for HIV-1 proviral DNA.
[0081] The results as assessed by white blood cell (WBC) count by flow cytometry are shown in Table 1 below. The results as assessed by RT and CD45 qPCR are shown in
TABLE-US-00001 TABLE 1 Leukocyte removal from whole blood as assessed by flow cytometry Log reduction % reduction compared to compared to WBC count Log WBC Diluted Diluted Sample (cells/l) Count Whole Blood Whole Blood Diluted whole 723.00 2.86 blood Macopharma1 97.30 1.99 0.87 86.54 filtered blood Macopharma2 0.80 0.10 2.95 99.89 filtered blood Pall filtered 4.61 0.66 2.20 99.36 blood Centrifuged 2.46 0.39 2.47 99.66 plasma
TABLE-US-00002 TABLE 2 Percentage leukocyte removal from whole blood as assessed by RT and CD45 qPCR, and HIV-1 qPCR Log reduction in % Leukocyte Log reduction in % Leukocyte leukocytes from removal from leukocytes from removal from diluted whole diluted whole diluted whole diluted whole blood as assessed blood as assessed blood as assessed blood as assessed Filter by RT and CD45 qPCR by RT and CD45 qPCR by HIV-1 qPCR by HIV-1 qPCR Macopharma1 1.27 94.53 1.12 92.38 Macopharma2 2.77 98.83 1.99 98.97 Pall 2.61 98.80 1.54 97.14
[0082] The results as assessed by flow cytometry show that 86.54% of leukocytes were removed using the Macopharmal filter, 99.89% of leukocytes were removed using the Macopharma2 filter, and 99.66% of were removed using the Pall filter. The results as assessed by CD45 qPCR show that 94.53% of leukocytes were removed using the Macopharmal filter, 99.83% of leukocytes were removed using the Macopharma2 filter, and 98.80% of leukocytes were removed using the Pall filter. The results as assessed by HIV-1 qPCR show that 92.38% of leukocytes were removed using the Macopharmal filter, 98.97% of leukocytes were removed using the Macopharma2 filter, and 97.14% of leukocytes were removed using the Pall filter.
[0083] The highest level of leukocyte removal was achieved using the Macopharma2 filter material. The lowest level of leukocyte removal was achieved using the Macopharmal filter material, and the Pall filter performed at an intermediate level. The Macopharma2 filter was used for subsequent experiments.
Example 2
Optimisation of Filter Thickness
[0084] The effect of the number of layers of filter material used for leukoreduction was determined. The Macopharma2 filter material used in Example 1 contained six layers. Since leukoreduction filters rely primarily on depth filtration to deplete the cell content, it was expected that increasing the number of layers of filter material would improve the efficacy of leukoreduction.
[0085] Whole blood samples spiked with 100,000 8E5 cells/ml were filtered using six, eight, or ten layers of the selected Macopharma filter. The efficacy of leukoreduction was determined by RT and CD45 qPCR, as described above. The results are shown in
TABLE-US-00003 TABLE 3 Percentage CD45+ cell removal from samples filtered using different numbers of layers of filter material Log reduction in Percentage leukocyte leukocytes from removal from diluted whole diluted whole Number of blood as assessed blood as assessed layers by RT and CD45 qPCR by RT and CD45 qPCR 6 2.55 99.72 8 3.04 99.91 10 3.02 99.90
[0086] The results show that a filter containing eight layers performs better than a filter with six layers. Greater than 3 log reduction (>99.9%) of white blood cells was obtained using eight layers of the Macopharma2 filter. However there was no significant difference in performance between filters of eight and ten layers.
[0087] The dead volume of each filter was determined by inserting the filter into a pre-weighed tube and centrifuging at 13,000 rpm for one minute. The tube was then re-weighed and the difference used to calculate the volume of liquid collected (given that the diluted blood has approximately the same density as water). The dead volume of the filter increases with the number of layers, as shown in Table 4.
TABLE-US-00004 TABLE 4 Dead volume of different filter thicknesses, measured by liquid weight Number of layers Average Dead volume (l) 6 39.2 8 50.8 10 77.2
[0088] Given the small starting volume of sample, a filter comprised of eight layers was selected because it achieves the desired reduction of 3 log white blood cells, with a dead volume of only 50 l.
Example 3
Optimisation of Sample Dilution
[0089] In view of the relatively small volumes of whole blood that can be collected without venipuncture, the effect of dilution of a small volume of a whole blood sample on the efficacy of leukoreduction was determined.
[0090] 100 ul of whole blood (un-spiked) was diluted 1:2, 1:3, or 1:4 with phosphate buffered saline (PBS). The diluted samples were filtered using the selected Macopharma2 leukoreduction filter, and the level of white blood cell removal was assessed by RT and CD45 qPCR, as described above. No significant difference in the effectiveness of leukoreduction was observed using different sample dilutions.
[0091] The eight-layer Macopharma2 filter configuration was assessed for optimal sample dilution to pass sample through the filter, so as to achieve the optimal level of white blood cell removal when using an input of 100 l of whole blood sample. A dilution of 100 l of whole blood in 200 l of phosphate buffered saline (PBS) (i.e. a dilution factor of 1:2) was found to be sufficient to allow blood to pass through the filter without diluting the sample excessively.
Example 4
Optimisation of Filtration Pressure
[0092] Samples of diluted whole blood were filtered through the selected Macopharma2 filter. The volume of air used was constant (1 ml), but the speed was varied. The level of white blood cell removal at the different pump speeds was assessed by RT and CD45 qPCR, as described above. The results are shown in Table 5.
TABLE-US-00005 TABLE 5 CD45+ cell removal from samples filtered under different pressures. Percentage Log reduction leukocyte CD45 PCR in leukocytes reduction Quantification from diluted from diluted Pump speed (l/s) (cells/ml) whole blood whole blood Not filtered (diluted 1,657,652 blood sample) 25 1,129 3.17 99.93 50 1,198 3.14 99.93 100 2,082 2.90 99.87 200 3,150 2.72 99.81
[0093] As shown in Table 5, at a lower pump speed (and, therefore, at a lower pressure), there is a slightly higher level of removal of white blood cells. However, we have found that with decreasing pump speed there is an increase in filter dead volume. Thus, there may need to be a compromise between the efficacy of removal of white blood cells and the volume of sample lost to the filter.
[0094] The eight-layer Macopharma2 filter configuration was assessed for optimal pump speed to pass sample through the filter, so as to achieve the optimal level of white blood cell removal when using an input of 100 l of whole blood sample. This was assessed by RT and CD45 qPCR as detailed above. A pump speed of 50 l is was found to be optimal for 100 l whole blood sample diluted in 200 l PBS.
Example 5
Assessment of Leukodepletion of Spiked Whole Blood
[0095] The leukodepletion performance of the eight-layer Macopharma2 filter was assessed. Whole blood samples spiked with 100,000 8E5 cells/ml were filtered, and leukodepletion of the filtered samples was assessed using the three different methods described above: (i) white blood cell count by flow cytometric analysis; (ii) white blood cell quantification by RT and CD45 qPCR; and (iii) qPCR for HIV-1 proviral DNA. Leukodepletion of the filtered samples was compared to plasma samples (i.e. samples in which the cells have been removed from whole blood by centrifugation) since these are conventionally used for HIV viral load tests. The results obtained are shown in Table 6 below, and in
TABLE-US-00006 TABLE 6 Leukoreduction by centrifugation or filtration from whole blood samples spiked with 100,000 8E5 cells/ml, assessed by flow cytometry. Log Percentage WBC count Average Log Average difference from reduction from Sample (cells/l) WBC count WBC count whole blood whole blood Diluted whole 1448.00 1510.50 3.18 blood 1573.00 Diluted plasma 0.77 1.31 0.12 3.06 99.91 1.85 Macopharma2 1.08 0.50 0.30 3.48 99.97 (8 layers) 0.15 filtered blood 0.26
TABLE-US-00007 TABLE 7 Percentage leukocyte removal from whole blood as assessed by RT and CD45 qPCR, and HIV-1 qPCR Log reduction of % Leukocyte Log reduction of % Leukocyte leukocytes from removal from leukocytes from removal from diluted whole diluted whole diluted whole diluted whole blood as assessed blood as assessed blood as assessed blood as assessed Sample by RT and CD45 qPCR by RT and CD45 qPCR by HIV-1 qPCR by HIV-1 qPCR Plasma 2.47 99.66 1.70 98.02 Macopharma2 3.15 99.93 1.86 98.63 (8 layers) filtered blood
[0096] The results as assessed by flow cytometry, and by RT and CD45 qPCR, show that filtration using eight layers of the selected Macopharma2 filter achieved greater than 3 log leukoreduction. By all methods of assessment, filtration of the whole blood sample achieved a greater level of leukoreduction than centrifugation.
Example 6
Assessment of Retention of HIV Particles by Leukoreduction Filter
[0097] This example describes an assessment of the degree to which HIV particles are retained by a leukoreduction filter when spiked whole blood samples are passed through the filter.
[0098] It is important that the filter does not retain any of the free HIV particles in circulation so that the viral load of the sample is not underestimated. To determine the level of retention of viral particles by the Macopharma2 filter, blood was spiked with various concentrations of culture supernatant of HIV-1 LAI (subtype B). The amount of HIV in the diluted blood samples before and after filtration was assessed by HIV-1 RT-qPCR. The results are shown in
TABLE-US-00008 TABLE 8 Amount of HIV particles in blood samples before and after filtration HIV-1 input HIV (c/ml) in Macopharma2 HIV (c/ml) in (c/ml) (8 layers) filtered blood non-filtered blood 10,000 5,368 5,347 5,000 2,868 3,318 1,000 358 336 500 133 142
[0099] The results show that there is no significant retention of HIV particles on the filter.
Example 7
Viral Load Testing of Spiked Blood Samples for Use in Resource-Limited Settings
[0100] In resource-limited settings, it is necessary to determine whether HIV viral load is below or above 1000 copies/ml. This Example describes a method for determining whether the HIV viral load of a sample is above or below 1000 copies/ml.
[0101] Blood samples were spiked with different concentrations of the World Health Organisation (WHO) 3rd International Standard for HIV-1 RNA (subtype B) (National Institute for Biological Standards and Control, NIBSC). The NIBSC viral stock is quantified in International Units (IU). There are approximately 1.7 copies of HIV per 1 IU.
[0102] 100 l whole blood samples were diluted in 200 l phosphate buffered saline (PBS), and filtered through eight layers of the Macopharma2 filter material at a pump speed of 50 l air per second.
[0103] Viral RNA was extracted, amplified by isothermal nucleic acid amplification, and the amplification products were detected by rapid visual detection with a dipstick, using a simple amplification-based assay (SAMBA) method similar to the method described in Lee et al., Journal of Infectious Diseases 2010; 201(S1):S65-S71.
[0104] The results are shown in
Example 8
Viral Load Testing of Spiked Blood Samples for Use in Resource-Limited Settings
[0105] Blood samples were spiked with plasma samples from four patients collected in Namibia (HIV-1 subtype C) (each of which has a very high viral load; >30,000 copies/ml), and diluted in five doubling dilutions, with final viral load ranging between 120 and 6100 copies/ml.
[0106] 120 l whole blood samples were diluted in 240 l phosphate buffered saline (PBS), and filtered through eight layers of the Macopharma2 filter material at a pump speed of 50 l air per second.
[0107] Viral RNA was extracted, amplified by isothermal nucleic acid amplification, and the amplification products were detected by rapid visual detection with a dipstick, using a SAMBA method similar to the method described in Lee et al., Journal of Infectious Diseases 2010; 201(S1):S65-S71.
[0108] The results are shown in
Example 9
Viral Load Testing of Clinical Samples
[0109] 207 whole blood samples from infected individuals, most of whom were undergoing treatment, were leukodepleted by filtration through a leukodepletion filter. Viral RNA was extracted, amplified by isothermal nucleic acid amplification, and the amplification products were detected by rapid visual detection with a dipstick, using a SAMBA method similar to the method described in Lee et al., Journal of Infectious Diseases 2010; 201(S1):S65-S71. The assay was designed to monitor HIV viral load with a cut-off of 1,000 copies/ml.
[0110] In parallel, the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 test was used to quantify the HIV viral copy in plasma, obtained by centrifugation at 2,200 g for 5 minutes. Testing by the SAMBA and Roche tests was completed on the same day for each patient sample. Plasma samples were stored frozen at 80 C. for discordant analysis using the Abbott RealTime HIV-1 test. Discordant samples for which there was insufficient sample to complete repeat testing were removed from the final analysis.
[0111] Table 9 below shows the distribution of the 198 samples remaining for the final analysis. Given that quantitation by the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 test has an accuracy of 0.3 log, the comparative data were stratified into 4 categories: undetectable (no HIV target present); between 500 and 2,000 copies (i.e. within the 30.3 log 10 accuracy of the Roche test); <500 and >2,000 copies/ml.
TABLE-US-00009 TABLE 9 Distribution of HIV viral load measurement by SAMBA test compared to Roche COBAS AmpliPrep/COBAS TaqMan test Viral load measurement by Roche CAP/CTM (copies/ml) undetectable <500 500-2,000 >2,000 SAMBA >1,000 0 8 6 24 (copies/ml) <1,000 114 43 3 0 [0112] Overall concordance: 96%
[0113] It can be seen in this patient population that the leukodepletion filter was effective in removing the CD4+ cells, since the overall concordance between the Roche test, using plasma, and the SAMBA test, using leukodepleted whole blood, was 96% (190/198).